A key that counted twice.
On Friday 23 April 2010, Jeffrey Shuren, director of the Center for Devices and Radiological Health at the US Food and Drug Administration, used a meeting of the Association of Health Care Journalists in Chicago to announce what his agency called the Infusion Pump Improvement Initiative. The white paper behind it, dated April 2010, counted about 56,000 reports of adverse events with infusion pumps from 2005 to 2009, and 87 recalls, 14 of them in Class I, the category for problems that could cause serious harm or death. Citing market research, FDA put the number of external infusion pumps in US hospitals and other settings in 2006 at over two million.
The problems, the paper said, had been seen "across multiple manufacturers, pump types, and use environments". It named the most common types as software defects, user-interface issues, and mechanical or electrical failures, and gave examples of each. In one, a software fault called a key bounce made a pump occasionally register a single press of the zero key as two, so that "0" became "00". In others, a screen did not make clear whether a patient's weight was to be entered in pounds or in kilograms; alarms failed to sound when something went wrong, or sounded when nothing had; housings broke in routine use, batteries failed early, and pumps sparked or caught fire. FDA's reading was that many of these problems appeared "to be related to deficiencies in device design and engineering."
A week later, on 30 April 2010, FDA wrote to Baxter ordering it to recall and destroy the Colleague volumetric pumps then in use in the United States. Baxter had stopped shipping new Colleague pumps in July 2005, after recall notices over batteries and over the pump's user interface and failure codes, both later classed Class I. In October 2005 the government seized several thousand of its pumps in Northern Illinois, and in June 2006 Baxter signed a consent decree, a court-enforced agreement with the government. On 13 July 2010 a final order covered the Colleague pumps in American hospitals, about 200,000 by Baxter's count.
Few of the examples in the white paper described a pump that could not push fluid. They described a key that counted twice, a screen that left a unit unclear, an alarm that stayed silent and a battery that gave out. A pump can move fluid with great precision and still deliver a dose nobody ordered, because what it delivers depends on what it was told, on the line between it and the vein, and on everything inside it that is not the pumping mechanism.
Let's Talk