Screens that invite errors.
+ The questionWhich ways of entering a number lead to wrong entries, and what has research measured?
A pump that stored a different number
In a 2010 study of number entry, Thimbleby and Cairns keyed 10,000 mg/h into a Graseby 3400 syringe pump, and the pump recorded 100.00 mg/h without a warning. Every key had been pressed as intended, yet the number in the pump was a hundredth of the number typed. The same paper reported that the Graseby 500 discarded every digit after the first one past the decimal point, and that an unfinished entry on the 3400 reset itself after about 4 seconds. These were pumps and software of that era, not current products, but they answer the question Chapter 16 left open. Arithmetic that is right on paper still has to pass through a keypad, a program and a screen before it becomes a rate.
Pumps take numbers in two main ways. In serial entry, digits are typed in order on a numeric keypad, as on a calculator. In incremental entry, arrow keys or on-screen chevrons step a displayed number up or down until it reads correctly. Each fails in its own way. A keypad accepts any sequence of keys, including sequences that are not well-formed numbers, and the pump's software must decide what to make of them. Arrow keys can produce only numbers the pump is able to hold, but they are slow, as the measurements later in this chapter show. The interface therefore does more than pass a number from a nurse to a motor: it decides which mistakes are possible, which are caught and which pass silently into the rate.
Malformed numbers, and the zero that doubled
A malformed number is a keystroke sequence that is not a valid number for its field: two decimal points, digits beyond the field's precision, or superfluous zeros at the front. Thimbleby and Cairns argued that a pump should block such an entry rather than repair it silently as the Graseby pumps did. They tested the idea in a simulation of 5,000 runs of random keying errors against target numbers from 0.01 to 99.99, with a wrong keystroke at most 1 time in 10, and probably 1 in 100 to 1 in 20. Blocking malformed numbers removed between 32.7 % and 55.2 % of the errors that left a number out by ten, wrong by a factor of ten or more, depending on the assumptions. The figures come from a model, not from wards.
The key bounce of FDA's 2010 white paper is the opposite case: not a person's wrong keystroke but a digit that nobody typed. The fault lay in the program that reads the keys, and Chapter 20 shows the timing that prevents it. Where such a fault strikes matters. A 2013 analysis of real pump logs found that the digit 0 is used far more than any other, and that the values 1,000, 100 and 50 appear in nearly half of all infusions, so the key pressed most often is the very key that the white paper's example doubled.
Malformed numbers were still the subject of a recall in 2025. On 14 November 2025 Fresenius Kabi began a correction of software 5.10.1 and earlier on its Ivenix large-volume pump, which FDA classified Class I. One of its two faults froze the screen when a rate entered with two leading zeros exceeded the drug library's limit, leaving the pump in what FDA's record calls a fail-stop state (Chapter 20). FDA's record lists 30 units, with two serious injuries and no deaths reported as of 18 November 2025. The fix was pump software 5.10.2 with version 5.2.2 of the management software. Until it arrived, users were advised to "avoid programming infusates with two leading zeros when possible." Fifteen years after the white paper, zeros at the front of a number were again a hazard, and Chapter 20 follows what became of the fix.
What laboratory studies have measured
Laboratory studies have compared the two styles directly. In a 2015 study, 22 participants entered 90 numbers on each of a keypad and an incremental interface. Their error rates were 1.36 % with the keypad and 1.41 % with arrow keys, a difference too small to be significant, but each number took 4.95 seconds on the keypad and 25.45 seconds with arrows, about five times as long. In 74.95 % of trials the participants did not look back at the display before confirming. They were students working at a laboratory task, not nurses on a ward, so the figures compare the two styles rather than measure error rates in hospitals.
According to later papers that summarize them, two earlier studies point to the checking step as the place where entry errors are caught or missed. A 2011 study found that incremental interfaces led participants to notice more of the errors deliberately inserted into their entries, and to look at the display for longer. A checksum is a short number worked out from the digits of an entry, so that a wrong digit almost always changes it. A 2013 study of checking found that participants caught every error when they had to enter a checksum themselves, but only 36 % when they were asked to verify one shown to them. A check that makes a person produce a number catches more than a check that asks them to look at one, a finding that bears on how a second nurse checks a pump before a high-alert infusion starts.
The 2015 laboratory rates are 1.36 % per number on the keypad and 1.41 % with arrow keys. A program of three numbers, a weight, a rate and a volume to be infused, is entered without a slip with probability 0.9864³ = 0.960 on the keypad and 0.9859³ = 0.958 with arrows. - Chance of at least one slip: 1 − 0.960 = 4.0 % on the keypad, and 1 − 0.958 = 4.2 % with arrows. - Time for the three numbers: 3 × 4.95 = 14.9 s on the keypad, and 3 × 25.45 = 76.4 s with arrows. - Over an illustrative 20 programs in a shift: 20 × 0.040 = 0.8 programs with a slip on the keypad, against 20 × 0.042 = 0.84 with arrows. At these rates about 1 program in 25 carries a slip whichever style is used, and the arrows cost about a minute more per program without making slips rarer.
A screen that shows the number the pump will use
FDA's white paper filed its other examples under user-interface issues. It cited "numerous reports of confusing or unclear on-screen user instructions," and it described the screens of the opening story that left a patient's weight unclear between pounds and kilograms, "leading to inappropriate dosing." A weight in the wrong unit is an error the size of the conversion factor, the same arithmetic as in Chapter 16, and nothing in a keypad prevents it. Only the screen can, by showing the unit beside the number and the consequence beside the rate: the dose as the pump has understood it, and the time the infusion will take. In the Alberta case of Chapter 16, the pump did not show the duration.
A report in FDA's adverse event database, quoted in the 2010 paper by Thimbleby and Cairns, describes a pump programmed at 68 mL/h instead of 6.8 mL/h: "The patient expired the next day." The authors point out that the record cannot say how the error arose, because a pump keeps the number it accepted, not the keys pressed to reach it. A missed decimal key, a key that registered twice and a misread order all leave the same 68 behind.
An event log shows the rate a pump ran, when it was entered and, on pumps with a drug library, the alerts it raised. It does not show whether the user pressed the decimal key, whether a key registered twice or what the order said. An investigation that finds 68 where 6.8 was meant has found the outcome of an error, not its mechanism, and Chapter 25 sets out what logs can and cannot establish.
An interface can also enter numbers that nobody chose. In May 2026 Fresenius Kabi began a further recall of Ivenix large-volume pumps because a drop or a severe jolt could loosen the touchscreen's input cable so that "unintended touch inputs may register at arbitrary screen locations." FDA classified the recall Class I; its record lists 18,444 units. A key that bounced in the 2000s and a touchscreen that registers phantom touches in 2026 are one hazard at two ends of the period this guide covers: input the user did not give, arriving in a field that the pump will read as part of a dose.
For pump makers, including those designing for Indian hospitals, reject malformed numbers, such as two decimal points, digits beyond the field's precision or leading zeros, and ask for the entry again instead of trimming it. Show every accepted number with its unit, and beside the rate show the dose it gives and the time to the end of the volume to be infused. Test number entry with the values real logs are full of, 0, 50, 100 and 1,000, and with a zero that registers twice. Record keystrokes as well as accepted values in the event log, so that an investigation can tell a user's slip from a software fault.
- An interface can change a correct number on its way into a pump: two Graseby pumps tested in 2010 altered or truncated entries without a warning, and one reset unfinished entries.
- A simulation found that blocking malformed numbers removes about a third to a half of out-by-ten errors, and an entry with two leading zeros still froze a large-volume pump in a Class I recall begun in November 2025.
- In a laboratory study, keypads and arrow keys gave similar error rates of about 1.4 % per number, but the arrows took about five times as long and most participants never looked back at the display.
- A pump's log keeps the accepted number and not the keystrokes, so the screen, showing unit, dose and duration, is where an interface can expose a wrong entry before it runs.
Let's Talk