18 — Usability

Testing with the people who will use it.

+ The questionHow does a maker show that nurses can program a pump safely before it is sold?

Infusion pumps as FDA's example

On 29 May 2026 FDA finalized its guidance on the human factors information that a medical device marketing submission should contain. When it needed an example of a device type that "historically has been associated with known use error issues," it named infusion pumps. The arithmetic of Chapter 16 and the keypads of Chapter 17 are where a working pump meets the person who programs it. A laboratory study of students cannot show how one particular pump, with its own screens, sets and users, will fare. A maker has to show that for its own design before the pump is sold. The discipline that does so is called usability engineering, or human factors engineering, and FDA's guidance on the subject puts both names in its title.

Its central idea is the use error: an action, or a failure to act, by which a user brings about a result that the maker did not intend or the user did not expect. A rate entered in the wrong unit is a use error, and so is a set loaded with its clamp left open (Chapter 14) or an alarm silenced and forgotten (Chapter 15). The term deliberately avoids blame. It treats the error as the outcome of a meeting between a person and a design, which the designer can change, rather than as a fault of the person. That makes the maker's question concrete: which use errors does this pump allow, and how likely and how harmful is each?

A process that starts from the tasks

IEC 62366-1 sets out the work as a process to analyze, specify, develop and evaluate a device's usability as it bears on safety. Its first edition was published on 25 February 2015, replacing IEC 62366 of 2007, and an amendment of 2020 was merged with it into a consolidated edition 1.1 on 17 June 2020. ISO records the standard as confirmed in June 2021 and current, with no successor in development as of October 2026. It covers normal use, including use errors, and leaves out what it calls abnormal use. The process begins with tasks rather than screens: what the users must do, in what conditions, and what could go wrong at each step.

For a pump the task list is long: hang a bag or load a syringe, seat the set, prime it, select a drug, enter a concentration, a weight, a dose and a volume to be infused, start, answer an alarm, change a bag. Each task is examined for the use errors it allows and the harm each could cause. The result picks out the critical tasks, which FDA's guidance defines as user tasks that, if performed incorrectly or not performed at all, "would or could cause serious harm." A task is critical because of what its failure would do, not because it is difficult.

FDA's infusion pump guidance of 2 December 2014 asks that the hazard analysis cover "drug loading, priming, programming, infusion." Its table of hazards includes a key struck twice, a wrong key, programming sequences "beyond user expectations or capabilities," an incorrect concentration and a wrongly loaded set. The table reads as a summary of Chapters 14, 16 and 17, and it turns the errors those chapters trace into hazards for a maker to analyze and test before a pump is sold.

Design then proceeds in rounds. Formative evaluations, smaller studies with users during development, find where people stumble while the design can still change, and each round feeds the next version of the screens, the set and the instructions. The last step is a human factors validation test, also called a summative test. Representative users perform the critical tasks on the final design, and every use error, close call and difficulty is recorded and traced to its cause. The 2014 guidance says that a pump submission "should include a Human Factors Validation study report." A use error on a critical task found at this stage has to be explained, and a cause in the design can send it back to an earlier round.

46 — From the tasks to a validation test
TASKS load, prime, program, start, answer alarms USE-RELATED RISK ANALYSIS use errors and harms, task by task CRITICAL TASKS failure would or could cause serious harm FORMATIVE EVALUATIONS smaller studies with users in development VALIDATION TEST ≥ 15 PER DISTINCT USER GROUP FDA 2014 hazards: key struck twice, wrong key, wrong concentration, set loaded wrongly DESIGN CHANGE rounds FDA SUBMISSION, FROM 1 AUG 2026 CATEGORY 1 CATEGORY 2 CATEGORY 3 INFUSION PUMPS: FDA'S EXAMPLE AT DECISION POINT D Schematic. Process after IEC 62366-1:2015+A1:2020; categories from FDA final guidance of 29 May 2026; minimum of 15 from FDA guidance revised 3 Aug 2026.
Human input and use errorReadoutFocal detail
Plate 46 — Usability engineering starts from the tasks and ends in a validation test with representative users, and FDA's 2026 guidance points an infusion pump toward the full report of Category 3.

Two FDA documents of 2026

In 2026 FDA finalized one of its human factors documents and revised the other. The guidance of 29 May 2026 covers the content of human factors information in marketing submissions. It finalizes a revised draft of 9 December 2022, which had itself replaced a 2016 draft list of the devices of highest priority for human factors review. FDA generally does not expect the new content in submissions received before 1 August 2026. The guidance sorts submissions into three categories, chosen through four decision points, A to D. At D, a device type with a history of known use errors is one for which validation data are likely to be needed. Infusion pumps are the example, which points a new pump toward the third category.

CategoryWhat the submission contains
1A conclusion that human factors are unaffected
2A rationale: there are no critical tasks, or no validation data are needed
3A full human factors and usability engineering report, including validation testing

The second document, Applying Human Factors and Usability Engineering to Medical Devices, was first issued on 3 February 2016 and revised on 3 August 2026. The revision keeps the rule of the 2016 text for validation tests: "the minimum number of participants should be 15" for each distinct user group. A pump programmed by intensive care nurses, configured by pharmacists who build its drug library and serviced by biomedical technicians may have three user groups, and so, on that rule, at least 45 participants. Both documents are final guidance, not drafts, and Chapter 28 places them in the clearance process for a pump.

Why fifteen participants

Fifteen is not a figure for finding every problem but for finding the common ones. The revised guidance's appendix on sample size cites a 2003 study by Faulkner. The model behind such figures is simple. If every participant has the same chance of meeting a given problem, independently of the others, the chance that at least one of them meets it is one minus the chance that all of them miss it. A common use error is then almost certain to appear in a test of fifteen, and a rare one is likely to be missed, as the example below shows.

Solved example: the chance that 15 participants see a use error

A use error is made by 1 user in 10 who attempt the task. Each of 15 participants misses it with probability 0.9, so all 15 miss it with probability 0.9¹⁵. - 0.9¹⁵ = 0.206, so at least one participant meets the error with probability 1 − 0.206 = 0.794, about 79 %. - For an error made by 1 user in 20: 0.95¹⁵ = 0.463, and 1 − 0.463 = 54 %. - For an error made by 1 user in 100: 0.99¹⁵ = 0.860, and 1 − 0.860 = 14 %. Fifteen participants will usually reveal a use error that one user in ten makes, have about even odds of revealing one that one in twenty makes, and will usually miss one that one in a hundred makes.

47 — What fifteen participants can see
0 % 25 % 50 % 75 % 100 % 1 5 10 15 20 25 30 CHANCE AT LEAST ONE MEETS THE ERROR PARTICIPANTS FDA MINIMUM PER USER GROUP: 15 1 USER IN 10 1 IN 20 1 IN 100 79 % 54 % 14 % 1 IN 100: SEEN IN ONLY 14 % OF TESTS Guide's arithmetic: each participant meets the error independently with probability p; the model behind FDA's sample-size appendix (Faulkner 2003).
ReadoutFocal detail
Plate 47 — Fifteen participants per user group usually reveal a use error made by 1 user in 10, but a test of that size sees one made by 1 user in 100 only 14 % of the time.
A passed validation test can leave a one-in-a-hundred use error unseen

A pump that is programmed many times a day across a hospital will meet use errors that fifteen participants per group are unlikely to show. A clean validation test means the known critical tasks were performed without errors that matter; it does not show that rare errors are absent. Those are reduced by design, such as limits in the drug library or connectors that cannot be joined wrongly, and found after release.

What a simulated ward cannot show

A validation test measures tasks, not working days. The 2017 study of ten US hospitals in Chapter 16 found infusions run without the smart pump's drug library among its most frequent errors. Two of its authors were BD staff, and the CareFusion Foundation was among its funders. Running an infusion around the library is a decision taken on a ward, under that ward's pressures. A test that asks participants to program a dose through the library, under observation, is not built to measure how often they would bypass it over a full shift.

Human factors work therefore continues after clearance. Chapter 21 shows how often drug libraries are bypassed in practice, Chapter 25 what pump logs reveal about it, and Chapter 31 how use errors reach FDA as reports once pumps are in service. The validation test establishes that the final design, used by the people it was designed for, does not invite the errors its makers could foresee. It cannot establish how a hospital will use it, and the connectors of Chapter 19 show the other route the field has taken: a design that makes one class of error impossible, whatever the user does.

Plan the validation test from the risk analysis

For a pump maker, including an Indian maker preparing a first US submission, write the use-related risk analysis first and derive the critical tasks from it: concentration and weight entry, dose and volume to be infused, set loading, alarm response. Recruit at least 15 participants from each distinct user group, drawn from the people who will use the pump in the market it is sold into, and test the final design with its real instructions and training. Plan for Category 3, the full report, under FDA's May 2026 guidance for any submission received from 1 August 2026. Keep the IEC 62366-1 usability engineering file as the record for every market, noting that EN 62366-1 was not on the EU's list of harmonized standards for the MDR in June 2026 (Chapter 29).

+ What this chapter established
  • Usability engineering, set out in IEC 62366-1 (2015, consolidated with its 2020 amendment), starts from the tasks users perform and picks out the critical tasks whose failure could cause serious harm.
  • FDA's 2014 pump guidance asks for a human factors validation report and names a key struck twice, an incorrect concentration and a misloaded set among the hazards to analyze.
  • In 2026 FDA finalized guidance on human factors content in submissions, with infusion pumps as its example of a device likely to need validation data, and revised its 2016 guidance while keeping a minimum of 15 participants per user group.
  • Fifteen participants will usually reveal a use error made by 1 user in 10 but will usually miss one made by 1 in 100, so validation confirms a design rather than proving rare errors absent.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.