Bubbles the pump must see.
+ The questionHow does a pump detect air, and how much air matters?
A detector that tells liquid from gas
An air-in-line detector is a sensor clamped around the tubing that tells liquid from gas as the line passes through it. A blockage can only be inferred from rising pressure (Chapter 12), but air can be watched directly, because it changes how the tube carries a beam of sound or light. MHRA's 2013 guidance, since withdrawn, described the detectors on volumetric pumps as ultrasonic or optical. In an ultrasonic detector a small transmitter on one side of the tube sends sound to a receiver on the other. Liquid carries the sound across, while a bubble carries it far less well, so the receiver's signal drops for as long as the bubble takes to pass.
What the detector reports is therefore a gap with a duration, and turning that gap into a size takes two more facts. At a known flow through a tube of known bore, the fluid moves at a known speed, so the duration gives the length of the bubble, and the bore turns the length into a volume. Baxter's 510(k) summary for its SIGMA Spectrum pump, cleared in July 2025, states its threshold in exactly these terms: an alarm for bubbles longer than 2.5 cm, or 1 inch, which the maker puts at about 140 µL in its own sets. That volume holds only for the bore of those sets. Air detection, like the volume on the display, rests on an assumption about the tubing.
Some designs also catch air before it reaches the detector. According to its operating manual, the cassette of ICU Medical's Plum 360 holds up to 1 mL of air in an air trap before the pump raises a cassette alarm, and the pump carries air sensors on both its inlet, or proximal, side and its outlet, or distal, side. Syringe pumps are a different case: a primed syringe has no container to run dry, and the same MHRA guidance rated the risk of air with syringe pumps as minimal. For volumetric pumps it recommended that every pump have some means, by design or by detector, of preventing air embolism, the entry of air into the bloodstream that the next section describes. It also recommended that a detector offered as an option always be bought.
Single bubbles and accumulated air
Pumps apply two limits, because air can arrive as one bubble large enough to matter or as a stream of small ones, each below the single limit, that add up. Baxter's summary gives both: the 2.5 cm single bubble, and more than 1 mL of air accumulated within 15 minutes at room temperature, or more than 1.5 mL at 15.5 °C, with bubbles smaller than 10 µL left out of the count. ICU Medical's specification sheet for the Plum 360, as reproduced by a reseller, gives a distal limit of 0.1 mL for a single bubble and 0.25 mL accumulated in 4.9 mL of fluid, with proximal limits of 0.5 mL and 1.0 mL. The withdrawn MHRA guidance said most volumetric pump detectors stop single bubbles of about 100 µL.
Sensitivity has a price. MHRA noted that more sensitive detectors pick up what it called champagne bubbles, fine bubbles that can cause nuisance alarms, alarms that sound when nothing needs to be done. BD's tip sheet on air-in-line alarms for its Alaris system, which gives no thresholds, names a common cause of nuisance alarms: tubing not fully inserted in the detector. Air alarms are frequent without being the main burden. A 2019 paper by two BD staff analyzed BD Alaris pump logs from European critical care units. Of 467,437 alarms of all kinds, 17,100 were single-bubble alarms and 6,975 air-accumulation alarms, or 3.7 and 1.5 % (Chapter 15).
How much air harms
Air that enters a vein travels with the blood toward the heart and lungs, and enough of it at once, a venous air embolism, can be fatal. A 2016 review gives the lethal venous dose for adults as "between 200 and 300 cc, or 3–5 mL/kg", figures drawn from adult cases and animal models that give no safe volume for a newborn. The same review reports that small amounts of venous air are seen on CT scans in up to 23 % of patients with peripheral intravenous lines, usually without symptoms. Air that reaches the arteries, directly or by crossing from the venous side, is another matter: the review says it can harm "even when the volume of air is small", and gives no threshold.
Set against those figures, a single bubble at a pump's limit is small. At Baxter's 140 µL, it would take about 1,400 such bubbles to reach the lower adult figure of 200 mL. The detector's limits sit far below the adult venous range, and the harmful volumes cited here arrived by routes no pump detector watched. Open lines are one. The 2016 review cites a report in which a large 15 French peel-away sheath admitted 300 mL of air, the top of the adult lethal range, in half a second, through an opening no pump was watching.
Devices with no detector are another. A New South Wales safety newsletter of 2003 relayed reports to Australia's regulator, the TGA, of deaths and serious injuries from air delivered by high-flow infusers and fluid warmers, mostly when partly emptied bags were reconnected. ECRI's list of top health technology hazards for 2023 included inflatable pressure infusers, which squeeze a bag to drive fluid in fast and can deliver fatal air emboli from it. In each case the air reached the patient along a path with no detector on it, from equipment built to move fluid fast.
Air that enters below the detector, at a port or a loose connection near the patient, passes unseen, and so does air from any device that has no detector, such as a pressure infuser. A pump that has not alarmed for air has checked only its own segment of the line.
Length is what the detector sees
The same volume of air looks very different in different tubing. After the pump, air may pass into an extension of much narrower bore, where a given volume stretches into a long column. BD lists a microbore extension set with a 1 mm bore, 250 cm long, holding 2.1 mL when primed. At that bore a bubble of 140 µL fills about 18 cm of tube, so a long white gap in microbore tubing can hold less air than a short one in a standard administration set. The same rule works against the patient inside the pump: if a set of wider bore than the maker specified were loaded, the same 2.5 cm limit would let a larger volume through before alarming.
- Baxter sets its single-bubble limit at 2.5 cm, about 140 µL, in its own sets (the maker's figure). A cylinder's volume is its cross-section times its length, so the cross-section is 140 mm³ ÷ 25 mm = 5.6 mm². A circle of that area has a radius of 1.34 mm (3.14 × 1.34 × 1.34 = 5.6), so the set's bore is about 2.7 mm.
- In BD's 1 mm microbore extension the cross-section is 3.14 × 0.5 × 0.5 = 0.785 mm². The same 140 µL there is 140 ÷ 0.785 = 178 mm long, about 18 cm, and a 2.5 cm bubble holds only 0.785 × 25 = 19.6 µL, about a seventh as much.
- Against the review's lower adult figure of 200 mL, or 200,000 µL, it takes 200,000 ÷ 140 = about 1,400 bubbles at the limit.
- The detector's limit is a length, which the bore turns into a volume, and the same length can mean seven times more air in one tube than in another.
The arithmetic explains why makers state air limits for named sets and why a set from another maker, even one that fits, changes what the detector protects against. It also explains why the accumulated-air limit matters as much as the single-bubble one. Eight bubbles, each just under Baxter's 140 µL single limit, add up to more than 1 mL without tripping it, and only a count kept over time, like Baxter's 15-minute window, will catch them.
In the ward and the workshop, check that the tubing is fully seated in the detector before suspecting the pump, and use only the sets the pump's maker lists, because the air limit was set for their bore. Never clear repeated air alarms by lifting the tubing out of the detector. For a maker, in India or elsewhere: state each air limit as a volume for each compatible set, give the bore it assumes, and test both the single-bubble and the accumulated limit with fine bubbles as well as large ones.
- An air-in-line detector, ultrasonic or optical, sees a gap in its signal as a bubble passes, and the pump turns the gap into a length and, through the bore of its set, a volume.
- Pumps set two limits, one for a single bubble (2.5 cm, about 140 µL, in one maker's sets) and one for air accumulated over time, such as 1 mL in 15 minutes.
- A bubble at such a limit is about 1 part in 1,400 of 200 mL, the lower end of the 200 to 300 mL that a review gives as the lethal venous dose for adults, though arterial air can harm in small amounts and no figure covers newborns.
- The harmful volumes of air cited in this chapter came through routes no pump detector watches, such as pressure infusers, reconnected bags and open lines.