03 — Dead volume

Drug the patient has not yet received.

+ The questionWhy does a change at the pump take minutes, sometimes hours, to reach the patient?

Pumps outside the room, lines through the door

In 2020 the intensive care units of Mount Sinai Morningside in New York moved the infusion pumps of patients isolated with COVID-19 out of their rooms and ran the lines to them through the doors. A brief report published on 30 July 2020 set out how. Each line gained two 210 cm "microbore transducer extensions", in the paper's words, passed through the gap between double doors rather than under a door. A second identification band was taped inside the glass of the door so that the patient's barcode could still be scanned from outside. Chapter 2 showed what that length of narrow tubing does to the pressure a pump must make, and why the team raised its pumps' occlusion limits.

The paper also stated the cost in volume: "Transducer extension tubing adds approximately 20 mL of circuitry." Before connection, the lines were flushed with a priming volume that the paper estimated at about 30 mL. To keep drugs moving through that circuit, the lines ran a carrier, a plain fluid infused continuously into the same line to push the drugs along, started at 50 mL/h and later individualized down to a minimum of 10 mL/h. The report listed the limits of the arrangement as well: rates above 300 mL/h could set off false occlusion alarms, rates below 5 mL/h could delay real ones, and the pumps could detect a blockage but not a disconnection. It named no pump and measured no patient outcomes.

06 — Mount Sinai Morningside, 2020: pumps outside the room
DOUBLE DOORS SECOND ID BAND INSIDE THE GLASS FOR SCANNING CARRIER 50 mL/h, LATER AT LEAST 10 mL/h DRUG PUMP PUMP OCCLUSION LIMITS 300 → 500 mmHg 2 × 210 cm MICROBORE EXTENSIONS PER LINE MANIFOLD BED ABOUT 20 mL OF CIRCUITRY (THE PAPER'S FIGURE) Schematic. Shah et al., Crit Care Explor 2020. The paper names no pump.
Fluid and drugPressure and forceSoftware and dataFocal detail
Plate 06 — Moving the pumps outside the room put a long, narrow line between each pump and its patient. By the paper's own figure the circuit held about 20 mL, which every change of rate had to cross.

The drug moves only as fast as the fluid behind it

Every infusion line holds fluid between the point where a drug enters it and the tip of the catheter. That fluid is the line's dead volume for that drug, and the drug cannot reach the vein until the fluid ahead of it has been pushed out. To a first approximation, liquid in a narrow tube at infusion rates moves in plug flow: what enters first leaves first, with some smearing at the front because, as Chapter 2 showed, liquid near the axis moves faster than liquid near the wall. The time a change takes to cross the line is then the dead volume divided by the total flow through it, the transit time. Nothing the pump does alters that arithmetic except the flow itself. A drug running at 2 mL/h on its own through 20 mL of line takes 10 hours to arrive.

The patient therefore receives whatever is in the line, at the rate the line delivers it, not what the pump is set to. When a nurse titrates a drug, adjusting its rate step by step to hold a measured effect such as blood pressure, each new rate reaches the vein one transit time later, and until then the old one continues. A carrier shortens the delay because it adds to the total flow: the drug joins a faster stream. It also ties the drug's delivery to the carrier, since drug already mixed into the line arrives as fast as the carrier pushes it, and Chapter 11 shows the surges and gaps that follow when a carrier's rate is changed.

07 — A rate change reaches the vein one transit time later
0 20 40 60 80 100 120 140 160 180 TIME (min) RATE SET AT THE PUMP OLD RATE NEW RATE NURSE CONFIRMS NEW RATE RATE ARRIVING AT THE VEIN REAL LINES: LATER AND ROUNDED 1 h 40 min: 20 mL AT 12 mL/h Schematic. Plug flow; 10 mL/h carrier + 2 mL/h drug through the paper's 20 mL.
Fluid and drugHuman input and use errorFocal detail
Plate 07 — The pump's screen shows the new rate at once; the vein receives it only when the fluid ahead of the drug has been pushed out. Real lines round the step and add delay.
Solved example: how long a new rate takes to cross 20 mL of line

Transit time is the dead volume divided by the total flow. The dead volume is the paper's 20 mL; the drug rate of 2 mL/h is chosen for illustration. With the carrier at 50 mL/h, the total flow is 52 mL/h, and 20 divided by 52 is 0.385 h, about 23 minutes. With the carrier at its 10 mL/h minimum, the total is 12 mL/h, and 20 divided by 12 is 1.67 h, or 1 hour 40 minutes. With no carrier, the drug alone at 2 mL/h takes 20 divided by 2, or 10 hours. If the line held 1.4 mL instead, the volume BD gives for two of its 213 cm microbore extensions, the same three cases would take about 1.6, 7 and 42 minutes. At the minimum carrier rate, a new drug rate set outside the door would reach the patient 1 hour 40 minutes later, if the paper's 20 mL is right.

A pump's display changes before the vein does

The number on the screen changes the moment the nurse confirms it. The vein sees the change one transit time later, and meanwhile the drug in the line keeps arriving at the old rate. The same holds when a drug's pump is stopped while a carrier keeps running: the drug already in the shared line goes on reaching the patient until the carrier has flushed it through.

A volume that does not fit the tubing

The paper's 20 mL is hard to reconcile with the tubing it names. BD, one maker of such tubing, gives its 213 cm microbore extension a priming volume of 0.7 mL, so two would hold about 1.4 mL. Holding 20 mL in 420 cm takes a bore of about 2.5 mm, and BD classes tubing as microbore only below 2 mm. The paper is not consistent on length either, describing the tubing elsewhere as "six feet", or 183 cm, against two lengths of 210 cm. Either its 20 mL covers more of the circuit than the two extensions, or the tubing was wider than microbore. The 20 mL stands in the chapters that follow as the paper's own figure, with the smaller volume beside it.

The two figures differ by a factor of about 14, and transit time scales with whichever is right. At the 10 mL/h minimum carrier and a 2 mL/h drug, 1.4 mL is crossed in about 7 minutes and 20 mL in 1 hour 40 minutes. Norepinephrine, a drug infused to raise blood pressure, has a half-life of about 2.4 minutes according to its FDA-approved label: in that time the body clears half of it from the blood. The label adds that its effect on blood pressure stops within 1 to 2 minutes after the infusion stops. Both delays are long beside these times. Chapter 10 returns to drugs this short-lived. ECRI's alert of 1 April 2020 advised that pumps for vasopressors and other drugs titrated to effect stay in the room (Chapter 12). Vasopressors are drugs, such as norepinephrine, that raise blood pressure.

08 — Transit time falls as total flow rises
1 10 100 1,000 0 10 20 30 40 50 60 TRANSIT TIME (min) TOTAL FLOW THROUGH THE LINE (mL/h) 10 mL/h CARRIER + 2 mL/h DRUG 50 mL/h CARRIER + 2 mL/h DRUG 20 mL, THE PAPER'S FIGURE 1.4 mL, TWO 213 cm MICROBORE (MAKER'S VOLUMES) 10 h 1 h 40 min 23 min 42 min 7 min 1.6 min MOUNT SINAI'S MINIMUM CARRIER Transit time = dead volume / total flow, plug flow; the guide's arithmetic. Drug rate 2 mL/h chosen for illustration.
Fluid and drugReadoutFocal detail
Plate 08 — Transit time is the dead volume divided by the total flow, so the 14-fold gap between the paper's 20 mL and the microbore geometry carries straight into the delay. At Mount Sinai's minimum carrier the paper's figure means 1 hour 40 minutes.

Carrier flow trades delay for fluid

A faster carrier is the obvious remedy, and Mount Sinai's two settings show its price. A carrier at 50 mL/h delivers 1,200 mL of fluid a day; at 10 mL/h it delivers 240 mL. Lowering the carrier from the first rate to the second cuts that fluid to a fifth and, by the arithmetic of the solved example, multiplies the transit time through 20 mL by more than four, from about 23 minutes to 1 hour 40 minutes. No carrier rate removes the trade, because the delay falls only as the fluid rises. The team also flushed 20 mL of saline through after drugs that contained dextrose, so even the flushes added volume that the patient received.

The carrier also changes what each drug does in the line. Once a drug joins a carrier, it is diluted in the line by the ratio of its own rate to the total flow, and the patient receives that mixture at whatever speed the total flow sets. A change in the carrier is then a change in the dose of every drug already mixed into it, for one transit time, even though no drug pump has been touched. Chapter 11 measures those effects in lines that share a catheter. Here the point is narrower: a long line makes the delay longer, and a carrier that shortens it makes the drug's delivery depend on a second pump.

Real lines are slower than the arithmetic

Plug-flow arithmetic gives the shortest possible delay, not a forecast. A 2020 bench study modeled a neonatal epinephrine infusion: the drug at 0.2 mL/h from a syringe pump joined a 5 mL/h carrier and crossed an extension with a priming volume of 0.4 mL. The formula gives 0.4 mL divided by 5.2 mL/h, about 4.6 minutes. The measured mean times for the drug to reach its target concentration were 30, 21 and 46 minutes with 3, 10 and 60 mL syringes, with wide spreads, and the drug column stalled in 3 of 7 runs with the 10 mL syringe and 3 of 4 with the 60 mL one. The authors also saw fluid flow backward at the connector.

The extra delay belongs to the pump and the line together: a syringe pump's slow start and the give in the syringe and tubing, their compliance, add to the time the plug needs (Chapters 5 and 10). The abstract of a 2022 bench study of lag times concludes that peristaltic and syringe pumps are "similarly susceptible to dead volume effects", so the delay is a property of the line more than of one kind of pump. Mount Sinai's report measured no delays at all. Its remote pumps return in Chapter 12, where the same long line slows the detection of a blockage, and in Chapter 24, which follows what the arrangement asked of the hospital's fleet.

Count the dead volume before choosing where a drug joins the line

For a pharmacist or an intensive care team planning infusion lines, list every component between the point where each drug enters and the catheter tip, add their priming volumes from the makers' instructions, and divide by the total flow to get the transit time at the rates in use. Compare it with how quickly the drug's effect must follow a change, and favor joining titrated drugs close to the catheter, where the volume downstream of the junction is smallest. Treat the result as the shortest delay possible, since pump start-up and compliance add to it. Where a remote pump makes the delay too long for a drug, keep that pump at the bedside, as ECRI advised in 2020.

+ What this chapter established
  • A drug reaches the vein only after the fluid ahead of it has been pushed out, so a change at the pump arrives one transit time later, the dead volume divided by the total flow.
  • Mount Sinai's remote pumps added two 210 cm microbore extensions and, by the paper's figure, about 20 mL of circuitry, which at a 10 mL/h carrier and a 2 mL/h drug means 1 hour 40 minutes for a new rate to arrive.
  • The paper's 20 mL does not fit microbore geometry, in which two such extensions would hold about 1.4 mL by BD's figures, and transit time scales directly with whichever volume is right.
  • A carrier shortens the delay at the cost of fluid and of tying the drug to a second pump, and real lines are slower than the plug-flow arithmetic because pump start-up and compliance add to it.