A connector that cannot fit.
+ The questionHow did feeding tubes come to be connected to veins, and what has a new connector standard changed for feeding pumps and their sets?
Feeds that reached a vein
In 2014 the Joint Commission, which accredits US hospitals, cited 116 published cases, in 34 papers, in which an enteral feed had been delivered into an intravenous line, and 21 of those patients died. Enteral feeding delivers liquid formula into the stomach or small intestine through a tube, and an enteral feeding pump moves it at a set rate. Formula is made for the gut, not the bloodstream. A feeding line joined to a catheter in a vein delivers it into the blood, and the pump, which infers volume from its mechanism and senses at most pressure and air in the line, has no means of knowing where its line ends. The rate can be exactly the one ordered and the patient still harmed, because the fluid went into the wrong tube.
Chapter 18 showed how a maker tests whether people can program a pump correctly. For this error, a clearer screen or more training does not reach the cause. The Joint Commission's earlier alert, of 3 April 2006, described nine cases reported to it, in seven adults and two infants, eight of them fatal, according to a trade publication's summary at the time. Four were enteral feeds into intravenous catheters and two were blood-pressure cuff tubes joined to intravenous catheters; the others put barium into a central line, a feed into a peritoneal dialysis catheter and intravenous fluid into a tracheostomy cuff. The common factor was not a drug or a dose but a connection.
The problem was wider than feeding. In a 2010 review, the Pennsylvania Patient Safety Authority counted 36 reports of wrongly joined lines in 21 months, 35 of them between two lines that both carried liquid. They included eight secondary intravenous sets attached to the wrong port, five reversed hemodialysis lines, three reversed gastric and jejunal feeding ports, two swapped epidural and patient-controlled analgesia lines, and two feeding sets connected to central venous catheters. A misconnection, a line joined to a port meant for another, can happen between almost any two tubes at a bedside that share a connector.
One connector that fit everything
That shared connector was the Luer connector, the small tapered fitting on intravenous lines and syringes. Feeding tubes, epidural lines and cuff tubes also used it, or a fitting close enough to mate with it. A connector that fits everything turns each connection into a choice that only the user's eyes can check. Lines lie close together at a bedside and can look alike, and training and labels ask the user to be right on every occasion. A connector that physically cannot join the wrong line removes the choice. This is a forcing function, a design that makes the wrong action impossible rather than unlikely, and unlike a label or a training session it does not depend on attention.
FDA wrote to makers of enteral feeding tubes on 9 July 2010 that misconnections were "associated with death and serious injury" and probably under-reported, and pointed to a connector standard then in preparation, ISO 80369-1. FDA's own account dates its recognition of the ISO 80369 series to 2011. The series divides small-bore connectors into families by application, each made so that it does not mate with the others. The Joint Commission's alert of 20 August 2014, which updated the 2006 alert, named the families: intravascular, which keeps the Luer; limb cuff; enteral; neuraxial, for epidural and spinal lines; and breathing and driving gases.
A standard that makes connections fail on purpose
Part 1, the general requirements, sets out how a maker assesses what the standard calls "non-interconnectable characteristics". Its third edition was published on 10 October 2025, replacing that of 2018, and FDA recognized it in December 2025. Part 7, for intravascular and hypodermic connectors, is the Luer itself, in a second edition of 4 May 2021. Part 3, for enteral connectors, was published in July 2016 and amended in 2019, and ISO confirmed it in 2023. It covers feeding sets, enteral syringes, drainage sets and access ports. Many connectors made to it are sold under the trade name ENFit. Because the intravascular line keeps the old fitting, every other family has to change, and the change runs through every enteral item a hospital stocks: tubes, pump sets, syringes and the ports between them.
FDA did not mandate the new connectors. It issued guidance on small-bore connectors for enteral use on 11 February 2015, and its letter of 7 September 2018 recommended enteral devices with ISO 80369-1 or ISO 80369-3 connectors, "many sold under the name ENFit." It asked makers to plan the removal of older devices and hospitals to keep transition adapters, which join a new connector to an old one, to a minimum. The letter counted 2 deaths, 24 serious injuries and 32 malfunctions reported for enteral misconnections since 2011, and noted that FDA's testing found gravity flow through ENFit "slightly slower" with 24 Fr tubes, which it did not consider a safety concern.
An adapter lets an ENFit set join an older feeding tube, and so keeps in use the older fitting, the one able to mate with lines of other families. A ward that still holds legacy enteral stock and adapters holds part of the old hazard with them, and FDA asked in 2018 that adapters be kept to a minimum.
The transition is not finished. GEDSA, the makers' trade association, announced in July 2021 that its members had stopped making the older feeding tubes and cross-application connectors on 1 July 2021 and would stop stand-alone transition sets and adapters on 1 January 2022. It said that about 80 % of the top 25 US hospitals were using or converting to ENFit. In January 2026 ECRI placed "Tubing Misconnections Remain a Threat Amid Slow ENFit and NRFit Adoption" fifth on its list of health technology hazards for the year; NRFit is the counterpart connector for neuraxial lines.
Enteral feeding pumps, and doses too small for the connector
In the United States an enteral feeding pump is regulated as an infusion pump. Its FDA product code is LZH, "pump, infusion, enteral" (Chapter 6). LZH falls under 21 CFR 880.5725, the same regulation as other infusion pumps, in device Class II, and pumps under it are cleared through 510(k). IEC 60601-2-24, the particular standard for infusion pumps, includes enteral nutrition pumps in its scope, and the EU's classification guidance gives feeding pumps as an example of class IIa (Chapter 29). FDA's database holds 34 clearance records under the code, the most recent for the Luminoah FLOW Enteral Nutrition System, cleared on 20 April 2026. Like most pumps of Part II, a feeding pump infers its volume from its mechanism, and like them it delivers whatever enters its set into whatever the set is joined to.
A feeding pump's specification reflects its job. Cardinal Health's figures for its Kangaroo OMNI pump, as a reseller reproduces them, give rates of 1 to 400 mL/h in steps of 1 mL/h and an automatic water flush at 875 mL/h every 1 to 24 hours. The stated accuracy is ±5 % or 0.5 mL/h for standard formula, and ±10 % for thick formula. At 60 mL/h for 20 hours the pump delivers 1,200 mL, and ±5 % allows 1,140 to 1,260 mL. At 5 mL/h the 0.5 mL/h figure is itself 10 %, and the specification does not say whether the larger of the two applies.
The new connector created a problem of its own at the smallest doses, given by enteral syringe. The ENFit syringe tip has a small recess between the tip and its collar, which industry testing calls the moat. Liquid that collects there can be added to the dose or lost from it, which matters most when the whole dose is a fraction of a milliliter. A low-dose tip design, used with a tap to clear the moat, addresses the problem. A poster from GEDSA, hosted by a maker and dated about 2020, reports tests of 1 mL syringes delivering low volumes. Against a range of ±10 % that a clinician consensus judged acceptable, the low-dose tip ranged from −2.90 % to +10.47 %, the standard ENFit tip from −24.06 % to +28.15 % and oral syringes from −7.37 % to +9.69 %. These are the industry's own figures, with no affiliations or funding stated.
The target is 0.2 mL from a 1 mL syringe, so the acceptable range of ±10 % is 0.18 to 0.22 mL. The bounds are those on the GEDSA poster. - Standard ENFit, −24.06 % to +28.15 %: 0.2 × 0.7594 = 0.152 mL to 0.2 × 1.2815 = 0.256 mL. - Low-dose tip, −2.90 % to +10.47 %: 0.2 × 0.9710 = 0.194 mL to 0.2 × 1.1047 = 0.221 mL. - Oral syringe, −7.37 % to +9.69 %: 0.2 × 0.9263 = 0.185 mL to 0.2 × 1.0969 = 0.219 mL. In these tests a standard ENFit syringe gave from about three quarters to about one and a quarter of a 0.2 mL dose. The oral syringe stayed inside the clinicians' range, and the low-dose tip stayed inside it except for an upper bound of 0.221 mL, about one thousandth of a milliliter over.
For biomedical engineers, pharmacists and purchasing teams, list every enteral item in stock, including feeding pump sets, tubes, syringes, extension sets and medication ports, and confirm that each carries an ISO 80369-3 connector. Treat each remaining transition adapter as a hazard to be removed, and record where any remains and why. Use low-dose tip syringes for small enteral doses, and train staff in the tap that clears the moat. For makers of feeding pumps, design the set with ISO 80369-3 connectors and assess non-interconnectability by ISO 80369-1:2025.
- Feeds, air and other fluids reached veins because one connector, the Luer, fitted lines with very different destinations, and a pump cannot sense where its line ends.
- The ISO 80369 series separates small-bore connectors into families that do not mate, keeping the Luer for intravascular lines and giving enteral lines ENFit (ISO 80369-3, 2016, amended 2019), under a general part in its third edition of 2025.
- FDA recommended rather than mandated the new connectors, and GEDSA's members said they ended legacy enteral connectors in 2021 and would end adapters in 2022, but ECRI still ranked slow adoption among the hazards of 2026.
- Enteral feeding pumps are regulated as infusion pumps, and in industry testing a standard ENFit syringe gave about 0.15 to 0.26 mL for a 0.2 mL dose, where a low-dose tip stayed close to the acceptable range.