Pumps that leave the bedside.
+ The questionHow do ambulatory and patient-controlled analgesia pumps limit what a patient can give, and where do insulin pumps fit?
A button, a lockout and a limit
A patient-controlled analgesia pump, used for pain relief, gives a patient in pain a dose of a strong opioid, a morphine-like painkiller, whenever the patient presses a button. The arrangement is safe only as long as nobody else presses it. The pump supplies half of that safety and the patient the other half. The large-volume and syringe pumps of Chapters 4 and 5 do only what a clinician has programmed. A patient-controlled analgesia (PCA) pump also takes requests from the patient, and its software decides which to honor. Each press of the handset delivers a fixed demand dose; a press during the lockout interval that follows a dose is refused; and a limit over one or four hours caps the total. Some orders add a continuous basal rate beneath the demand doses.
One health system's standard adult order set, the Alberta Health Services form of October 2020, shows typical numbers for morphine at 5 mg/mL: demand doses of 0.5, 1, 1.5 or 2 mg, lockouts of 10, 8 or 6 minutes, a suggested start of 1 mg every 6 minutes, a 4-hour limit of 30 mg, and no basal rate. Its hydromorphone and fentanyl orders follow the same pattern at 1 mg/mL and 25 mcg/mL. A 2017 review in US Pharmacist gives similar starting doses for adults new to opioids, morphine 1 mg, hydromorphone 0.2 mg and fentanyl 20 mcg, but a longer lockout for morphine, 10 to 15 minutes. Protocols differ on the lockout; the arithmetic of what any lockout allows is the same.
The Alberta order's suggested start is morphine at 5 mg/mL, 1 mg a press, a 6-minute lockout and a 4-hour limit of 30 mg. One dose is 1 ÷ 5 = 0.2 mL. The lockout allows at most 60 ÷ 6 = 10 doses an hour, so 10 mg, or 2 mL, an hour. Over 4 hours the lockout alone would allow 40 doses, 40 mg; the limit stops delivery at 30 mg, so the 31st press is refused even though its lockout has passed. A patient pressing at every opportunity reaches the limit after 30 doses, 3 hours of pressing, and further presses in that 4-hour period deliver nothing. If a visitor presses while the patient sleeps, every 6 minutes for 2 hours, the pump gives 20 doses, 20 mg, every one of them allowed by the lockout and inside the limit.
The patient is the sensor
The pump's limits are a backstop, because PCA relies on a safeguard outside the pump. An opioid makes a patient drowsy before it slows breathing dangerously; the US Pharmacist review puts it as "sedation always precedes respiratory depression". A patient who has had too much falls asleep and stops pressing, so the loop through the patient's own hand closes before the limits are reached. A basal rate keeps running whether the patient is awake or not, so it never passes through that loop, and the same review advises that a basal rate supply no more than about half of the daily requirement. The software counts presses and times; the patient's level of consciousness is the sensor that decides whether there are any presses to count.
The design fails when someone else presses. In December 2004 the Joint Commission, which accredits US hospitals, issued Sentinel Event Alert 33 on "PCA by proxy": serious harm when family members, caregivers or clinicians activate the pump for the patient without authorization. A 2005 trade report on the alert, a secondary source, gave its context: one such event in the Joint Commission's own database, and 6,069 PCA error reports collected by the US Pharmacopeia, 460 of them with harm, including 5 deaths. The alert recommended criteria for selecting patients, monitoring, teaching patients and families, and warning tags on the handset; AHRQ's patient-safety network records it as retired in June 2016.
The pump judges the timing and the total of presses, never the hand that made them. A press by a relative, a visitor or a clinician who believes the patient is in pain is honored exactly like the patient's own, and the safeguard that rests on drowsiness stopping the presses is gone. Monitoring sedation and breathing, which the 2004 alert recommended, is what can catch it; the pump cannot.
The alert did not deal with people who are allowed to press. A 2007 position statement from the American Society for Pain Management Nursing drew that line: dosing by someone the care team has authorized, which it called authorized agent-controlled analgesia, is distinct from unauthorized dosing, which is PCA by proxy. Nothing in a press identifies the hand that made it. A pump's records can show when doses were given, but not who pressed, so an investigation of an opioid overdose on PCA has to read them against the nursing notes and the accounts of the people present. Chapter 25 shows what pump logs can and cannot establish.
Pumps carried by the patient
An ambulatory pump is built to be carried, so that an infusion can continue at home or at work. The CADD-Solis, first sold by Smiths Medical and now by ICU Medical, weighs 595 g with four AA batteries. A 2025 study describes it as pumping by sequential squeezing of a tube. Smiths Medical's 2010 manual for the VIP model lists cassette reservoirs of 50 to 250 mL or bags of up to 3 L, clinician boluses of up to 50 mL, a limit on doses per hour, and rates up to 500 mL/h. ICU Medical's 2026 brochure states a delivery accuracy of ±6 % (nominal), without conditions. On a 100 mL reservoir at 4 mL/h, ±6 % means a nominal 25-hour infusion may end from 23.6 to 26.6 hours.
The pump does not see the reservoir on the patient's belt. A reservoir volume, programmable from 0 to 9,999 mL, is entered when the reservoir is attached, and the pump tracks what remains by subtracting its own count. That reckoning is only as good as the count. In the 2025 bench study, whose pumps were supplied by their makers, a CADD-Solis at 4 mL/h delivered 98.1 % of its set rate at 25 °C and 102.4 % in motion. Through a 20 G epidural catheter 100 cm long, however, it delivered 93.1 %, and a 3 mL bolus pushed through that catheter delivered 2.7 mL. A count of the mechanism's motion cannot see any of these differences.
At the smallest scale the measurement question returns. DEKA's RemunityPRO+, cleared by FDA on 24 July 2026, is a small pump for subcutaneous treprostinil that delivers 8 to 225 µL/h in steps of 1 µL/h from a 3 mL cassette used for 72 hours, and it uses acoustic volume sensing as feedback (Chapter 4). DEKA's 510(k) summary claims ±6 % and states that a blockage alarms within 12 minutes at 100 µL/h or more but may take up to 8 hours below that. At the 8 µL/h minimum, 8 hours is 64 µL not delivered. What changed from its predicate, the legally sold device its 510(k) compared it with, was its connection, to the cloud and to the patient's own device, which Chapter 28 takes up.
Insulin pumps, in one section
Insulin pumps belong to the same family and differ in scale. A cartridge in Tandem's t:slim X2 holds up to 3 mL, or 300 units of U-100 insulin, so one unit is 10 µL. Tandem's 2015 specification for the original t:slim gives basal rates up to 15 units an hour in steps of 0.001 unit, which is 10 nanoliters, and boluses up to 25 units. A basal rate of 0.5 units an hour is 5 µL/h, a twentieth of the newborn's syringe infusion in Chapter 5 and a 25,000th of the 125 mL/h of Chapter 4. A full cartridge holds 7.5 days' supply at 40 units a day.
At that scale too the pump counts what its mechanism moved, not what entered the tissue. In 2026 FDA classified two recalls of Insulet's Omnipod 5 Pods, a patch pump worn on the skin, as Class I, for cannula damage that could cause under-delivery. Insulin pumps also differ in control. Some now take their rate from a glucose sensor through an algorithm, which FDA regulates as a device of its own called a controller. FDA files interoperable insulin pumps that work with such controllers under a separate product code, FDA's three-letter code for a type of device. A De Novo grant, FDA's route for a new low- to moderate-risk type of device with no predicate, created the code in February 2019, and FDA had recorded 28 clearances under it by September 2026. That loop, its algorithms and its evidence are the subject of a planned guide in this series on automated insulin delivery. This guide treats insulin pumps only as pumps, which count steps like the others.
For pharmacists and pain teams: check every PCA order with the arithmetic of the solved example, working out what the lockout allows in one hour and in four and how long a patient pressing at every chance takes to reach the limit, and set the limit from the doses the order intends. Build each protocol into the drug library so the pump rejects a concentration or lockout outside it (Chapter 21). Tell families in writing, and on a tag on the handset, that only the patient presses unless the team has authorized someone, and pair every PCA with sedation and breathing checks at set intervals.
- A PCA pump delivers a fixed demand dose per press, refuses presses during the lockout and caps the total over one or four hours; at one health system's suggested morphine start, the lockout allows at most 10 mg an hour and the limit 30 mg in 4 hours.
- The design relies on the patient as the sensor, because sedation comes before dangerous slowing of breathing and a drowsy patient stops pressing; a press by anyone else defeats it, which the Joint Commission called PCA by proxy in 2004.
- Ambulatory pumps such as the CADD-Solis track the reservoir by their own count, and what they deliver depends on the catheter and on motion; DEKA's RemunityPRO+ takes acoustic feedback on microliter volumes.
- Insulin pumps move steps as small as 10 nanoliters and, like the other pumps, count what the mechanism moved rather than what reached the tissue.