Infusions that share a vein.
+ The questionWhat happens to each drug when several pumps feed one catheter?
Three pumps, one line
In a 2015 laboratory study in Utrecht, three pumps fed one infusion line: a carrier of plain fluid at 8 mL/h, the sedative midazolam at 5 mL/h, and norepinephrine at 0.1 mg/mL and 2 mL/h, joined in a set whose shared part held 1.8 mL. When the norepinephrine was started, it took a mean of 3,797 seconds, more than an hour, to reach 95 % of its intended delivery. When the rate of another pump on the line was raised, the midazolam delivered rose by an average of 27.5 % for about 8 minutes, and the norepinephrine by 23.5 %. The study traced these deviations to the physical properties of the infusion system, not to a faulty pump.
Chapter 3 showed that drug waits in the volume of a line before it reaches the vein, and Chapter 10 showed that short-acting drugs follow their delivery within minutes. A line shared by several pumps combines both facts. Each pump decides how much of its own drug enters the line, but none of them decides when that drug leaves it, because every fluid in the shared part moves at the total flow of all the pumps together. So when one pump changes its rate, it changes the timing of every other drug in the shared part.
A shared line holds a mixture
Beyond the junction where the lines meet, the shared tubing and catheter hold a mixture, and its strength is set by the ratio of the flows. Norepinephrine at 0.1 mg/mL and 2 mL/h enters the junction at 200 mcg/h, and a carrier at 8 mL/h dilutes it into 10 mL/h, so the mixture holds 20 mcg per mL. In steady running the patient receives 200 mcg/h whatever the carrier's rate, because the drug's own pump alone decides how much drug enters. The shared volume, however, holds a stock of drug, 36 mcg in 1.8 mL, and the carrier decides how fast that stock moves.
Any change in total flow changes the speed at which the stock leaves. When the carrier's rate is raised, the stored mixture is pushed out faster, so the patient briefly receives more drug though no one touched its pump. When the carrier's rate is lowered or the carrier is stopped, the stock leaves slowly, so the patient receives less until fresh drug fills the line. A 2005 study that traced a blue dye through administration sets showed these effects: the lag after a change grew with the shared volume and shrank with the total flow, and a carrier stopped and restarted cut delivery for a long time and then released a bolus. The simplest way to size them is to treat the line as plug flow (Chapter 3).
- Inputs, from the Utrecht study with its third pump left out: norepinephrine 0.1 mg/mL at 2 mL/h (200 mcg/h), a carrier at 8 mL/h, and 1.8 mL of shared volume.
- The mixture holds 200 ÷ 10 = 20 mcg/mL, so the shared volume stores 20 × 1.8 = 36 mcg.
- Carrier doubled to 16 mL/h: the stored mixture now leaves at 18 mL/h, carrying 20 × 18 = 360 mcg/h, 80 % more than ordered, for 1.8 ÷ 18 = 0.1 h, or 6 minutes. In those 6 minutes 36 mcg reach the vein instead of 20 mcg, an extra 16 mcg.
- Carrier stopped: the mixture leaves at 2 mL/h, carrying 40 mcg/h, 80 % less than ordered, for 1.8 ÷ 2 = 0.9 h, or 54 minutes. The patient misses 144 mcg.
- Carrier restarted at 8 mL/h after those 54 minutes: the line now holds undiluted drug at 100 mcg/mL, which leaves at 10 mL/h, carrying 1,000 mcg/h for 10.8 minutes. The extra is again 144 mcg: the drug withheld returns as a bolus.
- In words, on a plug-flow model, a carrier change produces a step in delivery whose height is set by the change in total flow and whose length is set by the shared volume.
Real lines smear the step. Fluid near the wall of a tube moves more slowly than fluid at its center, so a sharp front spreads as it travels, and the compliance of each syringe softens every change in rate (Chapter 5). Applied to the rig's second step, which raised the carrier from 16 to 25 mL/h with all three pumps running, the plug model gives about +39 % for 3.4 minutes; the Utrecht rig measured deviations of about a quarter, lasting about 8 minutes. Its authors found that shared volume and syringe compliance push in opposite directions, the first delaying a change and the second smoothing it. A 2017 model from the same group predicted measured delays with a standard deviation of about 16 seconds, and showed in an illustrative case that a fivefold increase in one pump can briefly triple another drug's output.
A carrier is plain fluid, and changing its rate looks like changing nothing. On a shared line it changes how fast every drug already in the line reaches the vein, so a flush, a bolus of fluid, a carrier bag running out or a carrier slowed to save fluid is, for a few minutes, a change in the dose of each drug on that line.
Late starts and lines that run backward
The same physics explains the slow start. A drug started into a shared line must take up its syringe's slack, fill its own tubing and then replace the shared volume, which moves only at the total flow. A rule of thumb from studies of this shared volume, also called dead volume, puts the time to steady delivery at one to three times the shared volume divided by total flow. A standard Y-set, a Y-shaped piece that joins two lines into one, holds 1.5 mL, so at a total of 10 mL/h the rule gives 9 to 27 minutes. Yet a 2015 bench study, with norepinephrine at 0.5 mg/mL, measured a median of 40 minutes with such a set, with an overshoot lasting about 35 minutes, and 12.6 minutes with a set whose shared volume was 0.046 mL. The maker of that low-volume set supplied the devices, and four authors had travel paid by the company.
Lines also run backward. When the shared line is blocked, or one pump stops while others keep running, fluid from the running pumps can travel up the quiet line instead of on to the vein. This backflow stores drug where no one expects it and returns it later as a bolus; at the syringe changeovers of Chapter 10, a running carrier made both backflow and forward surges larger. An anti-reflux valve, a one-way valve fitted to each line above the junction, prevents backflow, and one low-crack model opened at an average of 5.3 mmHg in a 2014 study. An anti-siphon valve stops a syringe emptying by gravity when the pump hangs above the patient; it opens only above about 100 to 150 mmHg, a range the 2018 bench study below cites from earlier work.
Valves have costs. A 2018 bench study measured the pressure a syringe pump needed to push through three-way devices fitted with anti-siphon valves: 178.7 mmHg at 10 mL/h, rising to 243.3 mmHg at 400 mL/h, about five times the 37 to 44 mmHg needed without them. The authors, who ran the pump at a 500 mmHg limit, inferred that at its lowest setting, 75 mmHg, it would alarm rather than run through the anti-siphon devices; that setting was not tested. Joined lines also pool their compliance. In the 2014 study, with the occlusion alarm set at 1,000 mmHg, a 50 mL/h line alone alarmed 81 seconds after a block downstream; joined to a 10 mL/h line behind an anti-reflux valve it took 133 seconds, 64 % longer, and adding an anti-siphon valve brought the time back to 74 seconds (Chapter 12).
Two drugs through one pump
Some pumps combine lines inside the machine. ICU Medical's Plum 360 can deliver two compatible drugs at independent rates through one set, because its secondary line joins its cassette directly (Chapter 4). ICU Medical's specification gives a minimum of 0.5 mL/h on each line and a maximum of 500 mL/h combined. In pumps that run secondary infusions, software makes the switch between primary and secondary infusions, and a 2025 correction of Fresenius Kabi's Ivenix pump concerned exactly that moment (Chapter 20).
Long lines magnify everything in this chapter. Mount Sinai's 2020 report gives a volume of 20 mL for its lines, a figure Chapter 3 found hard to reconcile with the tubing the report names. Through 20 mL, the plug arithmetic above would stretch from minutes to more than an hour. For makers, FDA's 2014 guidance asks for performance testing in every configuration of the system, or in a justified representative set of them.
For the nurses, pharmacists and biomedical engineers who set up critical care lines: record which drugs share each lumen, or channel, of the catheter, where they join and how much volume lies between the junction and the vein. Give a short-acting, high-alert drug its own lumen where one is free, or join it as close to the catheter as possible, and fit an anti-reflux valve to every line above a junction. Check that valves and the pump's occlusion setting are compatible, and treat a change in carrier rate as a change in the dose of every drug on that line.
- Beyond a junction the line holds a mixture whose strength is set by the ratio of flows, so the shared volume stores drug that every pump on the line moves.
- On a plug-flow model, doubling an 8 mL/h carrier raises a 2 mL/h drug's delivery by 80 % for 6 minutes, and stopping it cuts delivery by 80 % for 54 minutes, with the shortfall returning as a bolus on restart.
- Real lines spread and soften these steps, and bench studies measured start-ups longer than an hour and deviations of about a quarter lasting several minutes.
- Anti-reflux and anti-siphon valves prevent backflow and siphoning, but they add opening pressure and change how quickly a blockage is detected.