The order goes straight to the pump.
+ The questionWhat changes when the health record programs the pump, and what new risks come with the connection?
Fifteen keystrokes become two
At one 286-bed hospital, according to a 2018 report, programming an infusion took a mean of 15 keystrokes before its pumps were connected to the electronic health record and 2 afterward, a fall of 86 % in the first eight months. The keystrokes that disappeared are the ones in which Chapters 16 and 17 found the tenfold errors, and Chapter 21 showed that a drug library catches only some of those. Interoperability, the exchange of data between the pump and the hospital's information systems, removes most of them at the source. The same report states that every prepopulated infusion and dose change ran under the pump's dose limits.
The usual arrangement has two directions. The electronic health record is the hospital's electronic chart of the patient and of the orders for that patient. In auto-programming, an order written in it travels as data to the pump. The nurse scans a barcode on the patient's wristband, one on the bag and one on the pump channel, and the record matches the bag to an active order. An integration server then sends a programming request with the drug, concentration, dose and rate to that channel. The pump shows the values, checks them against its drug library and waits for the nurse to confirm. In auto-documentation, the running pump sends its rate, the volume infused and its state back to the record, which writes them into the chart.
The change is larger than it looks. A keypad entry is built by a person from an order; an auto-programmed entry is copied by software from an order. The nurse's work shifts from composing numbers to checking numbers composed elsewhere, and the pump, which in Chapter 20 trusted its keypad, now trusts a message from a server. The error that remains is whatever was already in the order, or whatever the copying gets wrong.
Removing the keypad removes the typing slip and keeps every error made upstream. A wrong dose in the order, a weight recorded wrongly in the chart or a bag matched to the wrong order arrives at the pump already formed, and it looks as authoritative as a correct one. The pump's library still checks the values against its limits, which is why the dose-limit check matters as much after the connection as before it.
What the connection removed
A 2024 before-and-after study at Norton Children's Hospital, which has 274 beds, measured the effect in a pediatric setting with BD Alaris pumps; its authors declared no conflicts of interest. They compared January to June 2020, before interoperability went live in September 2020, with January to June 2022. Infusions rose from 143,997 to 165,343. Library alerts fell from 32,422 to 8,335, errors caught before they reached the patient from 197 to 20, and events the system recorded as harm averted from 1,731 to 295. Library compliance at the main hospital rose from 94 % to 97.9 %, above the 95 % target of Chapter 21.
These falls need careful reading. Fewer alerts and fewer caught errors could mean that fewer errors were typed, because there was less typing, or that the same errors were no longer caught. A comparison of two periods cannot separate those. It does show how the remaining alerts were treated: the override rate barely moved, from 73 to 70.6 %, and the share of overrides made in under two seconds changed little (61.5, then 58.5 %). What changed was the kind of alert overridden: overrides the study classed as high-risk fell from 5,851, or 24.6 % of all overrides, to 207, or 3.5 %.
Figures from the 2024 pediatric study, first half of 2020 against first half of 2022; keystrokes from the 2018 report at another hospital. - Keystrokes: 15 − 2 = 13 fewer per program, or 130,000 fewer key presses per 10,000 programs. - Library alerts: 32,422 ÷ 143,997 × 10,000 = about 2,252 before, and 8,335 ÷ 165,343 × 10,000 = about 504 after, a fall of about 78 % per infusion. - Errors caught before administration: 197 ÷ 143,997 × 10,000 = about 13.7 before, and 20 ÷ 165,343 × 10,000 = about 1.2 after. - Harm-averted events: 1,731 ÷ 143,997 × 10,000 = about 120 before, and 295 ÷ 165,343 × 10,000 = about 18 after, a fall of about 85 %. - Paused infusions not restarted: none before, and 6 ÷ 165,343 × 10,000 = about 0.36 after.
Per 10,000 infusions the connection removed about 1,750 alerts and about 12 caught errors, and added about one third of a failure of a kind the earlier period had not recorded.
New ways to fail
A connection adds parts, and each can fail in ways a standalone pump could not. The pediatric study found one at once: reports of a paused infusion that was never restarted rose from none to six, a change its authors found statistically significant. A 2024 descriptive report from an 11-hospital system using ICU Medical Plum 360 pumps recorded programming edits falling from 1.27 to 0.18 per 100 programs; its authors include recipients of research grants from ICU Medical. It also recorded one episode in which auto-programming failures, which had run at 80 to 90 cases over two months, rose to more than 900 in two weeks.
The servers between record and pump are devices with their own recalls. In February 2025 BD began a correction because its Alaris Systems Manager could send an outdated auto-programming request to a pump; FDA placed it in Class I. Network details become safety details too. In February 2026 BD began correcting Alaris labeling that did not say which version of TLS, the encryption protocol of the secure wireless link, had to be enabled. FDA placed that recall, covering about one million units, in Class II. Each of these failures lies outside the pump's mechanism and outside its keypad, in the data path that replaced the keypad.
A pump on the network can be attacked
A connection also makes the pump reachable. ICS-CERT is the US government team that issues advisories on industrial control systems. In 2015 it reported, first on its secure portal on 23 June, that the Hospira Symbiq infusion system, version 3.13 and earlier, had a vulnerability "which can be exploited to remotely control the device". It scored 7.1 on version 2 of the Common Vulnerability Scoring System, whose scale runs to 10. On 31 July 2015 FDA strongly encouraged hospitals to move to other infusion systems, according to trade-press reports of its safety communication. News reports called it the first time FDA had told users to stop using a device over a cybersecurity risk. FDA knew of no case of remote access, and Hospira had stopped making Symbiq in 2013.
Later findings show where the openings lie. A security researcher who examined Baxter's SIGMA Spectrum pumps reported in 2022 that their Wi-Fi credentials were stored in the battery's memory, so a battery sold on after a pump left service could carry a hospital's network password, and that an unauthenticated message could change the pump's gateway address. The flaws were reported to Baxter on 20 April 2022, and the researcher and Baxter both published on 8 September 2022, 141 days later. That interval, in which the maker confirms a flaw and prepares a fix before it becomes public, is the core of coordinated vulnerability disclosure.
Since 29 March 2023, under section 524B of the Federal Food, Drug, and Cosmetic Act, a US submission for a networked pump has had to include a plan for such vulnerabilities after sale, including coordinated disclosure, and processes to patch them (Chapter 28). A patch is a new software version, with the risks Chapter 20 described. The Ivenix correction of August 2024 fixed a cybersecurity vulnerability together with other software anomalies in one release, the kind of update the section expects a maker to be able to issue. The device-software guide in this series treats device cybersecurity in depth. For a pump, the plan matters for as long as it stays on a hospital network, because every update and every new connection is a route by which its instructions can change.
Before go-live, and after every server, record or pump software update, test what happens when the network drops during an infusion, when an order changes while a request waits, and when a paused infusion's order is discontinued. Keep keypad programming procedures and library limits ready for downtime. Ask each maker for its software bill of materials (Chapter 28) and its vulnerability-disclosure contact, record the TLS and wireless settings its labeling requires, and erase network credentials from pumps and batteries before they leave the hospital.
- Interoperability sends the order from the health record to the pump and the pump's data back to the chart, cutting programming at one hospital from 15 keystrokes to 2.
- In a 2024 pediatric study, library alerts fell from about 2,252 to 504 per 10,000 infusions and caught errors from about 14 to 1, while the override rate barely changed.
- The connection adds failures of its own: paused infusions not restarted, bursts of auto-programming failures and an outdated programming request that FDA placed in Class I.
- A networked pump can be attacked, as the 2015 Symbiq advisory showed, and since 2023 section 524B has required makers submitting such devices to FDA to plan for vulnerabilities and to patch them.
Let's Talk