15 — Alarms

Alarms someone acts on.

+ The questionWhy do so many pump alarms go unanswered, and what makes an alarm worth hearing?

Two and a half alarms for every infusion

Pump logs from hospitals in Europe recorded about 2.2 million alarms during about 1.6 million infusions; 467,437 of those alarms came from critical care units, during 187,441 infusions. The logs came from 1,183 BD Alaris pumps, 566 large-volume and 617 syringe pumps. Two BD employees analyzed these logs in a 2019 paper, and BD funded the study. Chapters 12 to 14 showed what a pump can sense: pressure in a blocked line, air passing its detector, a door or set out of place. Each of those ends in an alarm, and the alarm is the only way the pump's knowledge reaches a person. A sensor whose alarm goes unanswered protects nobody.

Two words need their field sense here. An alarm is a signal from the pump that a condition needs attention, from a blocked line to an infusion about to finish. An alert is a prompt on the screen that asks the user to confirm or change an entry, such as a drug library's warning that a dose exceeds a soft limit (Chapter 21). IEC 60601-1-8, the international standard for alarm systems in medical electrical equipment, sorts alarm conditions into priorities by degree of urgency, so that a pump can tell a person how soon to come as well as that something is wrong. It is in force as the 2006 edition with amendments of 2012 and 2020, and FDA recognizes it.

What pumps alarm for

The 2019 paper breaks the critical care alarms down by type. Downstream occlusion led with 136,148, or 29.1 %, the blockages of Chapter 12. Close behind came 134,754 callback alarms, 28.8 %; a callback is a reminder that a pump has been left unattended, not a fault. Near end of infusion accounted for 107,236, or 22.9 %, and end of infusion for 20,762. Single-bubble and air-accumulation alarms, the subject of Chapter 13, totaled 24,075, about 5 %. Syringe disengaged (15,400), upstream occlusion (14,602), drive engage failure (10,850) and end of syringe (3,396) made up nearly all the rest.

40 — What critical care pumps alarmed for
DOWNSTREAM OCCLUSION 136,148 · 29.1 % SINGLE BUBBLE 17,100 · 3.7 % SYRINGE DISENGAGED 15,400 · 3.3 % UPSTREAM OCCLUSION 14,602 · 3.1 % DRIVE ENGAGE FAILURE 10,850 · 2.3 % AIR ACCUMULATION 6,975 · 1.5 % CALLBACK 134,754 · 28.8 % NEAR END OF INFUSION 107,236 · 22.9 % END OF INFUSION 20,762 · 4.4 % END OF SYRINGE 3,396 · 0.7 % FAULTS 201,075 43.0 % 56.9 % EXPECTED OR REMINDER 0 50,000 100,000 150,000 ALARMS Waterson and Bedner 2019 (authors BD staff; BD-funded). Grouping is the guide's. 214 other alarms not shown.
Pressure and forceAirReadoutFocal detail
Plate 40 — In one maker's critical care logs, downstream occlusion was the largest single alarm type, but more than half of all alarms were callbacks and near-end or end warnings: reminders and expected events, not faults.

The near-end and end alarms count down the volume to be infused, the amount the user programmed the pump to deliver. At the end a pump can fall back to a keep-vein-open rate, a trickle that keeps the catheter open until someone comes. These alarms are predictable by design: the pump knows from the start when they will sound. Sorted by what each type asks of a nurse, the counts fall into two groups. Callbacks, near-end and end-of-infusion alarms and end-of-syringe alarms announce an expected event or a pump left waiting. Occlusions, air, a syringe out of its holder and a drive that failed to engage report a condition that has stopped or threatens delivery.

Solved example: how often an alarm reports a fault, from one published log
  • All critical care alarms, 467,437, divided by all infusions, 187,441, give 2.49 alarms per infusion. The paper reports 4.50, the plain average of four unit ratios: (3.71 + 4.36 + 1.33 + 8.61) ÷ 4 = 4.50.
  • Expected events and reminders: callbacks 134,754 + near end 107,236 + end of infusion 20,762 + end of syringe 3,396 = 266,148, or 56.9 % of all alarms.
  • Faults that stop or threaten delivery: downstream occlusion 136,148 + upstream occlusion 14,602 + single bubble 17,100 + air accumulation 6,975 + syringe disengaged 15,400 + drive engage failure 10,850 = 201,075, or 43.0 %. The other 214 fall in neither list: 14 door-open alarms and 200 that the paper's counts by type do not account for.
  • Per infusion that is 201,075 ÷ 187,441 = 1.07 fault alarms and 266,148 ÷ 187,441 = 1.42 others.
  • In this log fewer than half of all alarms reported a fault, and because the log cannot tell a blocked catheter from a kink cleared in seconds, 43 % is an upper bound on alarms that reported a fault in the pump or line.

The paper does not say how many alarms were actionable, so the sorting above is this guide's, not the paper's. It is generous to the fault group, since a downstream occlusion alarm may come from a catheter that has clotted or from an arm bent for a minute. Even so, more than half of what a critical care nurse hears from these pumps is a reminder or the announcement of something expected. The two averages tell their own story: the pooled figure is lower because most infusion starts were in units with fewer alarms per infusion, while the pediatric ICU, at 8.61, contributed only 904 of the 187,441 starts.

An alarm rate depends on how it is averaged

Dividing total alarms by total infusions and averaging each unit's ratio give different answers from the same log, here 2.49 and 4.50, because a small unit counts as much as a large one in the second method. A comparison of alarm burden between hospitals, units or products means little until the method is stated.

When alarms stop being heard

One reason alarms go unanswered is their number. The Joint Commission's 2013 report on alarm system safety noted that alarm signals of all kinds can reach several hundred per patient per day on some units. Each one asks for a judgment, and when most need nothing urgent, staff learn to treat them as background, a response hospitals call alarm fatigue. In its Sentinel Event Alert of 8 April 2013 the Joint Commission counted 98 alarm-related events reported to it from January 2009 to June 2012, of which 80 ended in death, 13 in permanent loss of function and 5 in extra care or a longer stay; infusion pumps were among the devices named. FDA's database held 566 alarm-related deaths from January 2005 to June 2010, across all devices.

The response took the form of a goal. The Joint Commission's National Patient Safety Goal NPSG.06.01.01, "Improve the safety of clinical alarm systems", came in two phases for hospitals. Phase I, effective 1 January 2014, asked leaders to make alarm safety a priority by 1 July 2014 and hospitals to identify their most important alarm signals during 2014. Phase II, from 1 January 2016, required policies and procedures and staff education. Its scope named "IV machines". On 1 January 2026 the Joint Commission replaced these goals with National Performance Goals, and alarm safety became NPG.01.05.01, which asks hospitals to identify their most important alarm signals and to document policies and procedures for them. A 2012 survey cited in the report drew about 1,600 responses: 90 % called alarm management a safety issue, and fewer than half had an organization-wide process. Trade reports of ECRI's hazard lists put alarm hazards first for 2012 and 2014, with infusion pump medication errors third in 2012 and second in 2014.

Alarms also fail in the other direction. FDA's 2010 white paper reported that "some pumps fail to activate pre-programmed alarms when problems occur", while "others activate an alarm in the absence of a problem." Both kinds have recurred. BD's Class I recall of its Alaris system in February 2020 covered, among five issues, a low-battery alarm that could fail and the priority given to keep-vein-open and end-of-infusion alarms. Fresenius Kabi's Ivenix software recall begun on 14 November 2025 covered a misreported battery capacity that could shut the pump down without a low-battery alarm. In 2025 ICU Medical's Smiths Medical unit recalled CADD-Solis VIP ambulatory pumps for false upstream occlusion alarms, a cause FDA records as labeling design.

Seconds against half-lives

Some answers can be timed. The same BD-authored paper measured, for infusions of the short-acting drugs epinephrine, norepinephrine, dobutamine and dopamine, the time from an interruption to the infusion's restart. Median times were short: 17.5 seconds in the neonatal ICU, 12.75 in the pediatric ICU, 7.5 in the general ICU and 8.38 in the cardiac ICU. The neonatal mean was 58.73 seconds, and the longest neonatal restart in the paper's table was 444.25 seconds, although its text mentions some responses beyond 13 minutes. The paper set these times against half-lives of 180 seconds for epinephrine, 120 for dobutamine, 60 to 120 for dopamine and 120 to 180 for norepinephrine (Chapter 10).

41 — Restart times against drug half-lives
DOPAMINE 60–120 s DOBUTAMINE 120 s NOREPINEPHRINE 120–180 s EPINEPHRINE 180 s median mean maximum GENERAL ICU 7.5 s CARDIAC ICU 8.4 s PEDIATRIC ICU 12.8 s NEONATAL ICU MEAN 58.7 s MAX 444 s 17.5 s ≈ 2.5 EPINEPHRINE HALF-LIVES: ABOUT 18 % OF EFFECT LEFT (SIMPLE MODEL) 0 60 120 180 240 300 360 420 480 SECONDS Waterson and Bedner 2019 (authors BD staff; BD-funded): interruption to restart, short-acting drug infusions; half-lives as the paper gives them. 18 % is the guide's calculation.
Fluid and drugReadoutFocal detail
Plate 41 — Median restarts took seconds, a small fraction of a half-life, but the longest neonatal restart, 444 seconds, spans about 2.5 half-lives of epinephrine.

Against a half-life of 180 seconds, a 17.5 second gap costs little, about a tenth of a half-life. The longest restart in the paper's table is another matter. At 444 seconds, about 2.5 half-lives of epinephrine, a simple model in which the drug clears at a constant fractional rate leaves about 18 % of its steady level: two half-lives leave a quarter, and the further 0.47 of a half-life reduces that to about 18 %. The model ignores how the body redistributes a drug. The medians describe units that answer quickly, and the patient at risk is the one in the tail.

An alarm is worth hearing when it is rare enough to be believed, reports a condition that needs a person, and says how urgently. The earlier chapters supply the levers: an occlusion limit matched to the line cuts false occlusion alarms (Chapter 12), seated tubing and the right set cut nuisance air alarms (Chapter 13), and priorities by urgency, as IEC 60601-1-8 defines them, let the urgent alarm stand out from the reminders. The Joint Commission's goal asks hospitals to do the same sorting for every device that alarms, starting from the signals that matter most, and to back the result with documented policies and procedures.

Count the alarms before changing the settings

For the biomedical engineer or pharmacist who looks after a fleet: pull alarm data from the pumps' own logs (Chapter 25), count them by type and by unit, and divide totals by totals. Start with the largest category that needs no action, such as callbacks, and the largest fault category, in this log downstream occlusion, and trace each to its cause: a limit too low for the line, a set that kinks, a reminder nobody needs. Write the resulting settings into policy, as the Joint Commission's goal requires, and check that every change leaves the urgent alarms loud.

+ What this chapter established
  • In one maker's logs from European critical care units, pumps raised 467,437 alarms in 187,441 infusions, 2.49 per infusion when totals are divided, though the paper's average of unit ratios gives 4.50.
  • Downstream occlusion (29.1 %), callbacks (28.8 %) and near-end warnings (22.9 %) led, and 56.9 % of all alarms announced an expected event or a reminder rather than a fault, so the urgent alarm has to stand out from the routine ones.
  • The Joint Commission counted 80 deaths among alarm-related events from 2009 to 2012 and made alarm safety a goal from 2014, and FDA's white paper and recalls of 2020 and 2025 describe pump alarms that failed to sound or sounded falsely.
  • Median restart times for short-acting drug infusions were 7.5 to 17.5 seconds against half-lives of minutes, but the longest in the paper's table, 444 seconds, would leave about 18 % of the drug's level in a simple model.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.