07 — Elastomeric

A pump with no motor.

+ The questionHow can a stretched balloon and a narrow flow restrictor deliver a drug for days, and what changes their rate?

Rubber pushes, a restrictor resists

An elastomeric pump fits in one hand: a clear shell around a balloon of stretched rubber filled with drug, a length of tubing, and, near the patient's end, a narrow flow restrictor. There is no motor, battery, screen or alarm. The pumps of Chapters 4 to 6 count the motion of a mechanism and infer a volume from it; an elastomeric pump has nothing to count and infers nothing. FDA describes it in one line: the fluid sits in a stretchable balloon, and pressure from the balloon's elastic walls drives delivery. The rate is whatever that pressure can push through the restrictor, so it is set entirely by the physics of Chapter 2, and every condition that changes that physics changes the rate.

Chapter 2 showed that the flow through a narrow tube is the driving pressure divided by the tube's resistance, and that the resistance rises in proportion to the liquid's viscosity and the tube's length and falls with the fourth power of its radius. In an elastomeric pump the flow restrictor is by design the narrowest part of the path, so it sets the resistance. Makers choose the restrictor's bore and length to give a labeled rate: Avanos's catalogue lists an ON-Q pump of 400 mL at 4 mL/h, which would take about 100 hours to empty. Some non-electric infusers use a spring instead of a balloon, and EU classification guidance revised in April 2026 counts both kinds among active devices, the EU's term for devices that depend on a source of energy other than gravity or the energy the body generates for that purpose (Chapter 29).

17 — Rubber pushes; a restrictor sets the rate
RATE = BALLOON PRESSURE ÷ RESTRICTOR RESISTANCE BALLOON PRESSURE stretched rubber balloon of drug in a shell SKIN Infusor range calibrated with the restrictor at the skin, 31.1 or 33.3 °C TO PATIENT FLOW RESTRICTOR sets the rate; its resistance follows viscosity NO MOTOR NO BATTERY NO SENSOR NO ALARM schematic; calibration figures from Baxter's 2021 brochure (maker's figures)
Fluid and drugPressure and forceFocal detail
Plate 17 — The balloon supplies the pressure and the restrictor the resistance, so the rate is set by physics alone. Nothing in the device counts, measures or alarms.

Nothing in the device responds when the rate drifts. An electronic pump that meets more resistance raises its pumping pressure to hold the set rate, up to its occlusion limit (Chapter 12). A balloon has only the pressure its rubber gives, so anything that adds resistance, lowers the pressure or thickens the liquid lowers the rate, and nothing records or reports the change. The only readout is the balloon itself, shrinking over hours or days, set against the expected end time on the label. A pump that finishes hours early or late may be working exactly as its physics dictates, under conditions other than those it was set for.

Viscosity follows temperature

The restrictor's resistance is proportional to the viscosity of the liquid, and the viscosity of water falls steeply as it warms. The international formulation published in 2009 gives 1.0016 mPa·s at 20 °C, 0.93213 at 23 °C, 0.76441 at 32 °C and 0.74881 at 33 °C. With the pressure unchanged, flow rises in the same proportion as viscosity falls, by about 2.1 % for each degree near skin temperature. Where the restrictor sits therefore matters: taped to the skin it runs warm, while lying on a bedside table it runs at room temperature. A pump calibrated for one of those places runs at a different rate in the other.

Makers state where their figures hold. Baxter's 2021 Canadian brochure gives an accuracy of ±10 % for its Infusor range, calibrated with 5 % dextrose at 31.1 or 33.3 °C with the restrictor at the skin, and ±15 % for its Intermate range, calibrated with saline at 21.1 °C. A pharmacy checklist from BC Cancer, reviewed in March 2023, relays Baxter's guidance that flow changes by about 2.3 % per degree, slightly more than viscosity alone predicts; the source of the difference is not documented. The ±10 or ±15 % applies only at the stated temperature, fluid and fill.

18 — Warmer liquid, faster flow
±10 % (INFUSOR CLAIM) 0.70 0.80 0.90 1.00 1.10 20 25 30 35 37 FLOW RELATIVE TO A 33 °C CALIBRATION TEMPERATURE (°C) INTERMATE CALIBRATED (SALINE) INFUSOR CALIBRATED (5 % DEXTROSE, RESTRICTOR AT SKIN) 1.00 on a table at 23 °C: 0.803, about 20 % slow computed from water viscosity (IAPWS 2009 formulation, via NIST); balloon pressure assumed constant
ReadoutFocal detail
Plate 18 — With the pressure unchanged, flow rises as viscosity falls, about 2.1 percent a degree near skin temperature. A restrictor calibrated at 33 °C that lies on a table at 23 °C runs at 0.803 of its label, outside a ±10 percent claim.
Solved example: a skin-calibrated pump at room temperature

A pump calibrated at 33 °C, close to Baxter's 33.3 °C, delivers its labeled rate with the restrictor at skin temperature. If the restrictor lies on a table at 23 °C instead, flow goes as the inverse of viscosity, so the rate becomes 0.74881 ÷ 0.93213 = 0.803 of the label, about 20 % slow. An illustrative 250 mL pump labeled 5 mL/h should empty in 250 ÷ 5 = 50 hours; it runs at about 5 × 0.803 = 4.0 mL/h and takes about 250 ÷ 4.0 = 62 hours. Baxter's rule of about 2.3 % per degree gives 10 × 2.3 = 23 % slow for the same 10 degrees. The reverse case is a restrictor calibrated at 20 °C and worn against the skin at 32 °C: 1.0016 ÷ 0.76441 = 1.31, so it runs about 31 % fast.

Bench studies show the same effect. In a 2025 Japanese study of two elastomeric and two electronic pumps at 4 mL/h, with devices supplied by their makers, warming from 25 to 32 °C took one elastomeric pump from 98.2 to 117.5 % of its set rate and the other from 78.9 to 99.3 %. Viscosity alone predicts a rise of 0.89002 ÷ 0.76441, about 16 %, smaller than the rises of about 20 and 26 % that the two pumps showed. Time within ±15 % of the set rate fell from 65.5 to 43.3 % for the first and rose from 31.0 to 89.1 % for the second, while the electronic pumps barely changed. A 2003 study, with pumps donated by their makers, found flow within ±15 % for 29 to 100 % of the infusion across its five pumps, one of them elastomeric.

Fluid, fill and height

Temperature is not the only condition on the label. The fluid matters: the BC Cancer checklist states that an Infusor filled with saline runs about 10 % faster than with the 5 % dextrose it was calibrated with, and Baxter names the calibration fluid for each range. The fill matters, because a balloon's pressure depends on how far its rubber is stretched; the checklist warns that a device filled to less than 81 % of its nominal volume runs faster. Height matters because the pump's height above the catheter adds the pressure of a column of liquid. The checklist gives about 0.5 % per inch, faster with the pump above the outlet and slower below, so a pump carried 40 cm (15.7 inches) above the catheter runs about 8 % fast.

19 — Conditions that move an elastomeric pump's rate
±15 % INTERMATE ±10 % INFUSOR RESTRICTOR AT 23 °C, CALIBRATED AT 33 °C CALIBRATED AT 20 °C, WORN AT 32 °C SALINE IN A 5 % DEXTROSE DEVICE PUMP 40 cm ABOVE THE CATHETER FILLED BELOW 81 % OF NOMINAL FASTER (NO FIGURE GIVEN) −20 % +31 % about +10 % about +8 % the diluent alone can use the whole ±10 % band −30 −20 −10 0 +10 +20 +30 CHANGE IN RATE (%) temperature rows: the guide's arithmetic from water viscosity; fluid, height and fill: BC Cancer checklist relaying Baxter (maker's figures)
ReadoutFocal detail
Plate 19 — Each condition on the label moves the rate: saline in a dextrose-calibrated device adds about 10 percent, a pump 40 cm above the catheter about 8 percent, and a 10-degree mismatch in restrictor temperature 20 to 31 percent, enough to leave either accuracy band.

Resistance downstream adds to the restrictor's. The checklist notes that catheters narrower than 22 G may slow the flow, and a 2019 bench study of four elastomeric pumps and one electronic pump measured the effect of everyday conditions directly. Under ideal conditions the pumps delivered 88.5 to 99 % of their total volume. Raising or lowering them by 20 or 40 cm, or applying back pressures of 10 to 30 mmHg of the kind met in home infusion, changed the flow and the delivered volume of the elastomeric pumps, while the electronic pump was barely affected. Every one of these effects is a term in the same division of pressure by resistance, and the pump has no means of correcting any of them.

The labeled rate holds at one temperature, one fluid and one fill

An elastomeric pump's rate and its accuracy band are measured under conditions printed beside them. Filled with another fluid, filled short, carried high or low, or kept at another temperature, it runs at another rate with no sign of the change, and the ±10 or ±15 % no longer describes it. Two devices can be compared only under the same stated conditions.

Nothing to alarm and nothing to record

Removing the motor removes whole classes of failure: there is no software to update, no battery to run down and no key to bounce. It adds one. The device cannot tell anyone that it is wrong. It has no occlusion alarm, so a kinked line simply stops it; it has no air detector, and it keeps no record of what it delivered. Baxter presents its range for pain management, antibiotics and oncology, and its accuracy claims of ±10 and ±15 % are wider than the ±5 % many large-volume pumps claim (Chapter 8). What the device delivers is fixed by the restrictor, the balloon and the conditions of use, so the label carries the accuracy claim as much as the device does.

The standards also separate pumps with a motor from devices without one. IEC 60601-2-24, the international standard for electrically driven infusion pumps, excludes the devices covered by ISO 28620. ISO 28620 covers non-electrically driven portable infusion devices that deliver continuously or as boluses around the spinal cord (neuraxial), into blood vessels or under the skin, and excludes devices driven by gravity alone. Its second edition was published on 16 January 2020 and confirmed by ISO in August 2026. In the EU, the revised guidance places elastomeric pumps in risk class IIa, one class below infusion pumps, and India's 2017 risk classification lists "infusion pump or elastomeric infusion device" in Class C. Chapters 27, 29 and 30 set those rules beside the mechanisms of this part.

Fill and label for the conditions of the claim

For hospital pharmacists who fill elastomeric pumps: use the diluent the maker calibrated the device with and fill to the nominal volume, because saline in a dextrose-calibrated device runs about 10 % fast and an underfilled one runs faster too. Write the expected end time with its tolerance; for a 250 mL device at 5 mL/h, ±10 % means about 45 to 56 hours. Tell the patient where the restrictor must sit, taped to the skin for a skin-calibrated device, and to keep the pump at about the level of the catheter. Record the lot, fill and diluent on the dispensing record, because the device keeps no record of its own.

+ What this chapter established
  • An elastomeric pump has no motor or sensor: the balloon's pressure pushes the drug through a narrow restrictor, and the rate is that pressure divided by the restrictor's resistance.
  • The restrictor's resistance follows the viscosity of the liquid, which falls about 2 % per degree near skin temperature, so a pump calibrated at 33 °C runs about 20 % slow at 23 °C.
  • Makers state accuracy of about ±10 or ±15 % at a named temperature, fluid and fill, and the fluid, the fill, the height and the back pressure each move the rate away from it.
  • With nothing to alarm and nothing to record, an elastomeric pump's accuracy depends on its conditions of use, so the conditions on its label are part of the accuracy claim.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.