29 — European Union

A pump under the MDR.

+ The questionWho decides whether a pump may be sold in the European Union, and against which rules?

A certificate instead of a clearance

On 17 June 2026 the Official Journal of the European Union carried a Commission decision of 11 June that added the general safety standard for medical electrical equipment, EN 60601-1, to the list of standards that support the EU's medical device law. EN 60601-1 is the European adoption of IEC 60601-1, and each EN standard in this chapter is the European adoption of the IEC or ISO standard with the same number. The decision noted that no reference to EN 60601-1 had been published under that law before. The particular standard for infusion pumps and controllers, EN 60601-2-24, was not added, and as of October 2026 it is still missing. The decision shows where the European system differs from the American one of Chapter 28. In the United States, FDA itself reviews a pump before clearing it. In the EU, much depends on which published standards a manufacturer can point to.

The law is Regulation (EU) 2017/745, the Medical Device Regulation, called here the MDR. In the United States the same letters stand for the medical device reports of Chapter 31. No EU agency reviews a pump before it is sold. The manufacturer compiles technical documentation showing that the pump meets the Regulation's general safety and performance requirements. For all but the lowest-risk devices, that documentation is assessed by a notified body, an independent organization designated by a national authority, which also audits the maker's quality management system. When the notified body is satisfied it issues a certificate. The maker then declares conformity and affixes the CE mark, and the pump may be sold in every member state.

A harmonized standard is a European standard whose reference the Commission has published in the Official Journal for this purpose. A device that meets it is presumed to meet the requirements the standard covers, so the notified body checks the test results rather than an argument that they were the right tests. The MDR's list began with an implementing decision of 16 July 2021 and had been amended ten times by 11 June 2026. Which standards are on it decides how much of a pump's evidence the maker must defend clause by clause.

Rule 12 puts a pump in class IIb

The MDR has four classes, I, IIa, IIb and III, assigned by the rules of its Annex VIII. Rule 12 covers active devices intended to administer or remove medicinal products, body liquids or other substances. Such devices are class IIa "unless this is done in a manner that is potentially hazardous", in which case they are class IIb. The hazard is judged by the substance, the part of the body and the mode of application. In April 2026 the Medical Device Coordination Group, or MDCG, where the member states' authorities meet, revised its classification guidance as MDCG 2021-24 Rev.1. Its examples settle the question: infusion pumps are class IIb, beside ventilators, dialysis equipment and blood pumps for heart-lung machines.

The same guidance puts feeding pumps and "elastomeric or balloon pumps for infusion" in class IIa. An elastomeric pump has no motor (Chapter 7), but the guidance treats as active any device that delivers a drug from stored energy, whether in a stretched balloon or a spring. The split matches the mechanisms of the earlier chapters. A powered pump pushes a drug into a vein at whatever rate it is told (Chapter 16). It keeps pushing against rising pressure until its occlusion alarm sounds (Chapter 12). An elastomeric pump runs at a fixed rate, so no wrong rate can be keyed into it, and a feeding pump delivers into the gut rather than into a vein. In its notes on the same phrase under other rules, the guidance says that "potentially hazardous" refers to the device's characteristics, not to its user's competence.

The disposable parts have classes of their own. Under Rule 2, tubing that serves as a channel in an active drug delivery system is class IIa, which covers a pump's administration set. Syringes made for syringe pumps are class IIa too, while a gravity set is class I. A pump set is therefore a device with its own documentation and, when it fails, its own corrective action. Such actions have reached tens of millions of sets at a time (Chapter 26).

74 — How MDR Rule 12 classes a pump and its set
DEVICE THAT DELIVERS A MEDICINAL PRODUCT ACTIVE DEVICE? includes stored energy: balloon or spring RULE 12 POTENTIALLY HAZARDOUS MANNER? substance · part of body · mode of application device characteristics, not user competence CLASS IIb infusion pumps · ventilators dialysis equipment · blood pumps for heart-lung machines YES YES NO CLASS IIa feeding pumps elastomeric or balloon pumps for infusion THE QUESTION THAT MAKES A PUMP IIb THE SET IS A DEVICE OF ITS OWN TUBING OR SYRINGE FOR A PUMP RULE 2 CLASS IIa GRAVITY ADMINISTRATION SET CLASS I MDCG 2021-24 Rev.1, April 2026
Fluid and drugReadoutFocal detail
Plate 74 — Under the MDR an infusion pump is class IIb because Rule 12 treats its delivery as potentially hazardous; elastomeric and feeding pumps stay in class IIa, and the pump's set is a class IIa device of its own.
Solved example: classifying a pump and its set, and the time left on a legacy certificate
  1. A large-volume pump with a motor, for intravenous drugs, administers medicinal products, so Rule 12 applies, and the guidance lists infusion pumps in class IIb.
  2. Its administration set is tubing in an active drug delivery system: Rule 2, class IIa. A gravity set for the same bag is class I.
  3. In this example the pump holds a certificate under the Medical Devices Directive that preceded the MDR. Its maker had a quality management system and an application with a notified body by 26 May 2024, and a written agreement by 26 September 2024. As a class IIb device that is not implanted, it may then be placed on the market until 31 December 2028.
  4. From 11 October 2026 to 11 December 2028 is 26 months, and the last 20 days add about 0.7 of a month: about 26.7 months.

From October 2026, the maker of this class IIb pump therefore has about two years and three months to complete its MDR certification before the legacy route closes.

A second opinion on the clinical evidence

For a pump, class IIb brings a step that most class IIb devices escape. Recital 56 of the MDR says that notified bodies should generally ask an expert panel to review their assessment of the clinical evaluation for two groups: class III implants, and class IIb active devices intended to administer or remove a medicinal product. An infusion pump fits the second group exactly. Article 54 sets out this clinical evaluation consultation procedure. Before it certifies a new pump, the notified body asks a panel of experts appointed under the Regulation for its view on the clinical evaluation assessment report, the document in which the notified body judges the maker's clinical evidence.

Article 54(2) exempts three cases, and the second is interpreted in the MDCG's guidance of 2020, MDCG 2019-3 rev.1. The first is the renewal of a certificate already issued under the MDR. The second is a modification of a device that the same maker already markets for the same purpose, where the change does not worsen the balance of benefit and risk. In the MDCG's reading, devices sold under the old directives count as already marketed. The third is a device whose clinical evaluation is covered by a common specification. A maker updating its own established pump may therefore avoid the panel, while a new design, or a pump from a new maker, goes through it.

Standards that carry no presumption

A pump's technical documentation rests on tests to the standards of Chapter 27. The Commission's summary list of harmonized standards, generated on 17 June 2026, has 71 entries. Two of them bear directly on a pump: EN 60601-1 with its 2024 amendment A13, cited since June 2026, and the risk management standard EN ISO 14971:2019 with its 2021 amendment, cited since 2022. EN 60601-2-24 is absent. So are the collateral standards EN 60601-1-2 for electromagnetic compatibility and EN 60601-1-8 for alarm systems, the software life-cycle standard EN 62304, the usability standard EN 62366-1 and EN ISO 28620 for pumps without electric drive.

The difference is one of argument rather than of testing. Nothing prevents a maker from testing to EN 60601-2-24, whose accuracy and occlusion tests remain the obvious evidence for a pump (Chapters 9 and 12). For alarms, software, usability and the particular pump tests, however, no presumption applies. The maker must show that the standard it used reflects the state of the art and covers the relevant requirements, and the notified body judges that case. In standards usage the state of the art means generally acknowledged current practice, not necessarily the newest technology. FDA does not recognize IEC 60601-2-24 either, so a US submission cannot cite it as a recognized standard (Chapter 28). India, as Chapter 30 shows, names its own identical version of the particular standard first.

75 — Pump standards: cited in the EU, recognized in the US
STANDARD EU · CITED UNDER THE MDR summary list, 17 June 2026 US · FDA RECOGNIZED database, 11 Oct 2026 60601-1 general safety EN 60601-1:2006 + A13:2024 · since June 2026 19-49 (Ed. 3.2) 60601-1-2 EMC not cited 19-36 (partial) 60601-1-8 alarm systems not cited 5-131 60601-2-24 infusion pumps and controllers not cited no record 62304 software life cycle not cited 13-79 62366-1 usability not cited 5-129 ISO 14971 risk management EN ISO 14971:2019/A11:2021 · since 2022 5-125 ISO 28620 pumps without electric drive not cited no record AAMI TIR101 flow testing US document 6-482 THE PUMP STANDARD: NOT CITED, NOT RECOGNIZED India: IS 13450 (Part 2/Sec 24):2019 is identical to IEC 60601-2-24:2012 (Chapter 30). EU summary list is informative; the Official Journal governs. US: FDA Recognized Consensus Standards database.
ReadoutFocal detail
Plate 75 — Of the standards a pump is built to, only EN 60601-1 and EN ISO 14971 are cited under the MDR, and the particular pump standard, IEC 60601-2-24, is neither cited in the EU nor recognized by FDA as of October 2026.

EUDAMED, the 2028 deadline and a proposal

EUDAMED is the EU's database of devices, makers, certificates and, in time, incidents. The Commission declared its first four modules functional on 26 November 2025, in a decision published the next day. They cover actor registration, the registration of devices with their unique device identifiers, notified bodies and certificates, and market surveillance. Regulation (EU) 2024/1860 set a gradual roll-out in which each obligation applies six months after a notice of functionality is published in the Official Journal. Registration in those modules became mandatory on 28 May 2026, with later deadlines for devices and certificates already on the market. The vigilance module is still in development as of October 2026. Until it runs, field safety notices are published nationally (Chapter 31). Since 10 January 2025 the same 2024 regulation has also obliged a maker to give at least six months' notice of an expected end of supply that could cause serious harm. That notice matters to a hospital whose fleet depends on one pump (Chapter 24).

On 16 December 2025 the Commission adopted a proposal to simplify the MDR and its companion regulation for diagnostics, COM(2025) 1023. It would limit the expert panel consultation to class III implants, taking infusion pumps out of it. It would replace the five-year maximum validity of certificates with periodic reviews scaled to risk. It would also extend the deadline for reporting a serious incident from 15 to 30 days, except for deaths, serious deteriorations of health and public health threats. The Parliament's public health committee received a draft report on 30 June 2026, and the Parliament's indicative date for a plenary vote at first reading is 14 December 2026.

The expert panel step still applies to pumps

Until COM(2025) 1023, or a revised version of it, is adopted and applies, a new class IIb pump still goes through the consultation procedure unless an exemption fits, certificates still run for at most five years, and the 15-day reporting deadline still holds.

Pumps certified under the old directive have a deadline of their own. Regulation (EU) 2023/607, published on 20 March 2023, extended directive certificates to 31 December 2027 for class III and implantable class IIb devices, and to 31 December 2028 for other class IIb devices, class IIa and some class I devices. The extension holds only while the device still complies with the directive, presents no unacceptable risk and has no significant change in design or intended purpose, and only if the maker had a quality management system and an application with a notified body by 26 May 2024, and a written agreement with it by 26 September 2024. A substantial redesign of a legacy pump therefore needs an MDR certificate.

76 — EU dates that govern a pump, 2023 to 2028
legacy class IIb route: directive certificate, conditions met 2023 2024 2025 2026 2027 2028 2029 20 Mar 2023 Reg (EU) 2023/607 published 26 May 2024 QMS and application to a notified body 26 Sep 2024 written agreement with the notified body 10 Jan 2025 6 months' notice of supply interruption (Art. 10a) 26 Nov 2025 EUDAMED notice: 4 modules functional 16 Dec 2025 COM(2025) 1023 proposed 28 May 2026 EUDAMED modules mandatory 17 Jun 2026 EN 60601-1 cited (Decision 2026/1231) 14 Dec 2026 plenary vote indicated · proposal 31 Dec 2027 legacy class III and implantable IIb end OCT 2026: THIS GUIDE 31 Dec 2028 LEGACY INFUSION PUMPS: LAST DAY about 26.7 months after 11 Oct 2026 Reg (EU) 2023/607 · Reg (EU) 2024/1860 · Decision (EU) 2025/2371 · Decision (EU) 2026/1231 · COM(2025) 1023 · EP procedure 2025/0404(COD)
Software and dataReadoutFocal detail
Plate 76 — EUDAMED registration became mandatory in May 2026 and the MDR revision is still a proposal, while a legacy infusion pump's directive certificate carries it only to 31 December 2028.
Write the EU file for standards that give no presumption

Map each general safety and performance requirement to the standard used. Where that standard is not harmonized, as EN 60601-2-24 and EN 62304 are not, state why it represents the state of the art for the pump. Plan a new pump's certification around the expert panel consultation, and record the reasoning when a modified pump claims the exemption. Register the company and the pump's unique device identifier data in EUDAMED, and keep the set's class IIa file in step with the pump's. For a legacy pump, count back from 31 December 2028 and treat any significant design change as the end of the extension.

+ What this chapter established
  • In the EU a notified body, not a regulator, assesses a pump against the MDR and issues the certificate behind its CE mark.
  • Rule 12 puts an infusion pump in class IIb, while elastomeric and feeding pumps are class IIa and pump sets are class IIa devices in their own right.
  • A new pump goes through expert panel consultation, and its particular standard, EN 60601-2-24, carries no presumption of conformity as of October 2026.
  • EUDAMED's first four modules have been mandatory since 28 May 2026, legacy class IIb certificates run to 31 December 2028, and the 2025 revision of the MDR is still a proposal.