Reading a pump's log.
+ The questionWhat can a pump's event log and library reports show after an incident, and what can they not?
The pump's own account
A pump's own record settled within a day what had happened in the Alberta fluorouracil infusion of Chapter 16, and it could not settle why. The day after the four-day infusion ran out in about four hours, the unit manager and a nurse checked the pump's data history and found the rate it had been given, 28.8 mL/h. ISMP Canada's 2007 root cause analysis, a structured search for every cause behind an incident, reports that check and its result.
The record answered the first question any investigation asks, what the pump had been told. It could not say why 28.8 was entered when the order meant 1.2, and nothing in it showed what reached the patient beyond the pump's own count. This chapter reads what the fleet of Chapter 24 writes down, and takes its frame from a 2010 summit of AAMI and FDA whose first theme asked for standard ways to report, pool and analyze incidents like this one. What every such record can hold follows from the rule set out in "Before we start": a pump's records are written by the pump, so they hold what it was told, what its software decided, what its mechanism counted and what its sensors flagged, and nothing beyond its outlet.
What an event log holds
An event log, also called a history log, is the pump's time-stamped list of what happened to it, and its maker decides what goes in; as Chapter 17 showed, it keeps the number the pump accepted, not the keys pressed to reach it. The SIGMA Spectrum service manual of Chapter 24 describes downloading the log over an infrared link or a memory card as a file that can run to more than 300 pages of text. A 2012 study sponsored by CareFusion, now part of BD, used syringe-pump event logs from a Berlin intensive care unit to sort 133,601 infusion starts by how each was programmed (Chapter 21).
| Record | What it shows | What it cannot show |
|---|---|---|
| Pump event log | Values accepted and when; starts, stops, alarms and alerts; the volume the pump counted | What the order meant; what the bag held; what reached the vein |
| Library report | Alerts, overrides and corrected entries; drugs that alert most; share of infusions through the library; pumps on the current version | Errors inside the limits; infusions programmed around the library; why an alert was overridden |
| Incident report | What a person saw, judged worth reporting and wrote down | Incidents nobody noticed or reported; close calls that need not be reported |
The volume column of a log needs the most care, because it looks like a measurement and is not one. A large-volume pump computes volume infused from the strokes of its mechanism, and a syringe pump from the travel of its plunger (Chapters 4 and 5), so the figure in the log is exactly as right as the conversion behind it. A log that shows a bag fully delivered shows that the mechanism moved as far as that volume should take it.
Worn tubing, a syringe the pump took for another size, a line blocked for an hour before the alarm (Chapter 12) or a catheter leaking into tissue can each leave a log showing the programmed volume delivered. Fluid that ran by gravity with the set out of the pump (Chapter 14) never appears at all. Even the Ivenix pump, which measures the volume in its chamber from air pressure (Chapter 4), records the chamber and not the vein.
Reading a library report
The library keeps a second record. When an entry crosses a limit, the pump records the alert and what the nurse did next: override it and run the entry as programmed, change the entry, or cancel. Pharmacists download these records and read them as continuous quality improvement (CQI) reports, whose contents the AAMI Foundation's 2017 guide lists and the table above summarizes. An entry beyond a limit that is then corrected is the good catch of Chapter 21.
The 2019 analysis of Chapter 21, whose override rates averaged about three in four, also reports the override-to-reprogram ratio, the number of overrides for each corrected entry: 9.92 for high-alert drugs and 7.30 for the rest. The AAMI guide shows how far the ratio can climb: at an override rate of 98 %, with corrections making up the other 2 %, there are 49 overrides for every correction.
For illustration, the 2019 means are applied to a report of 2,000 alerts on high-alert drugs. Overrides, at 75.8 % (Chapter 21): 2,000 × 0.758 = 1,516. Corrected entries, the good catches: 1,516 ÷ 9.92 = about 153. The other 2,000 − 1,516 − 153 = 331 ended in neither, a cancelled infusion for example; the published summary does not break them down. About 1 alert in 13 (2,000 ÷ 153 = 13.1) led to a corrected entry, and those 153 are the report's whole record of errors caught. A mean of ratios across hospitals is not any one hospital's ratio.
What the 1,516 overrides mean, the report cannot say. An override is a keystroke: it records that a nurse ran an entry the library questioned, not whether the limit was too tight for the unit, the order was unusual and right, or the alert was misread. The 153 corrections are a floor on errors caught, not a count of errors made, since an entry inside the limits raises no alert, and an infusion programmed in plain mL/h, around the library, raises none at all. A speaker from Intermountain, a 22-hospital system, told the 2010 summit that 137 alerts had saved patients from harm over almost two years; how many alerts there were in all was not given.
Theme 1: incidents that are not counted
The summit, held at FDA's headquarters on 5 and 6 October 2010, drew 330 attendees. Its report, Infusing Patients Safely, set out five themes, and the first was to "standardize systems and processes for reporting, aggregating, and analyzing infusion device incidents." Four priority issues sat beneath it. Systems that pooled adverse-event reports, FDA's among them, were incomplete and lacked a standard form. Reported incidents understated the true number, partly because close calls, errors caught before they reached a patient, did not have to be reported. Makers often could not find root causes because they could not see hospitals' incident and quality data. And there was no safe, shared process for analyzing failures together.
The report's examples show the difficulty from inside hospitals. A speaker from Duke University Medical Center described pump quality data as a spreadsheet about 30 columns wide with 85,000 rows, which staff searched for hours and which "never really answers the question, why?" That comes to some 2.55 million cells. Of 978 root cause analysis reports involving devices at the Veterans Health Administration since 1999, 129 involved infusion pumps, about 13 %, and 69 of those concerned patient-controlled analgesia.
Some of what the summit asked for now exists. A research group led from Purdue University pooled alert records from 49 hospitals for the 2018 study of library delays, and pooled logs revealed what no single report could: an old library still raising alerts 661 days after its replacement (Chapter 24). Pooling cannot add what the pumps never recorded, and Chapter 31 sets out the limits of FDA's own database of device reports.
How the pump knows, and why it can still be wrong
The records make the answer to this guide's question visible, because they hold exactly what the pump knows: the numbers it was told and the motion of its own mechanism. A large-volume pump knows how many strokes its mechanism has made and how much each should move in tubing of its design bore (Chapter 4); a syringe pump knows how far its screw has turned and which syringe it was told it holds (Chapter 5). Neither meets the drug. The Ivenix pump measures the volume in its chamber from air pressure, and DEKA's RemunityPRO+ monitors its delivery by acoustic volume sensing (Chapters 4 and 6), but both stop at their own outlet. Everything past the outlet is assumed.
Sensors are the pump's only check on those assumptions, and each answers narrowly or late. Pressure climbs slowly against the give of syringe and line, so, by its maker's figures, one syringe pump at 1 mL/h with a 50 mL syringe may take up to about an hour to alarm at its middle setting (Chapter 12). Air detectors see bubbles above a threshold (Chapter 13), and nothing in the pump detects a catheter leaking into tissue or, as Mount Sinai's paper noted, a line that has come apart. Under its maker's test conditions, the volume a perfect pump counts matches the volume leaving its outlet within the ±5 % makers commonly state (Chapter 8). Outside those conditions, and everywhere past the outlet, the count rests on assumptions that a sensor may contradict only after a delay.
The wrong dose begins where that knowledge stops. The dose is the pump's volume multiplied by numbers people supply, carried through a line nobody measures, so a pump can work as built and still deliver a dose nobody ordered. An entry can carry a tenfold slip, a key read twice or a wrong unit (Chapters 16 and 17), and a 2025 recall of Ivenix software told users to avoid programming two leading zeros. A bag can hold a concentration other than the one selected. Mount Sinai's added tubing, about 20 mL by its paper's figure, takes over an hour to clear at a 10 mL/h carrier (Chapter 3), and a change in another infusion on the catheter pushes drug already in the shared line into the vein (Chapter 11).
Software and people complete the list. A library catches only entries that cross its limits, and only on infusions programmed through it (Chapter 21); the software that does the arithmetic can itself fail, as the Ivenix versions of Chapter 20 and BD's 2020 Alaris recall showed. People bridge every gap the machine leaves: they choose the bag, prime the line, answer the alarm and judge the alert (Part V). A pump's records therefore describe its half of the chain completely and say nothing of the other half, where a perfect pump's wrong doses begin.
After an incident, keep the pump, its set and the bag, and download the event log and library records with the serial number and library version. Read the log for what it can prove: entries and their times, alerts and what followed them, alarms. Then gather what it cannot show: the order, the bag's label and contents, the line as set up, other infusions on the catheter, and the accounts of the people involved. In library reports, read every override rate beside the limits that produced it, and treat corrections as a floor on errors caught.
- A pump's event log and library records are written by the pump, so they show what it was told, decided, counted and flagged, and nothing past its outlet.
- In a library report an override is a keystroke and a correction is a floor on errors caught: at the 2019 means, about 1 high-alert alert in 13 ended in a corrected entry.
- The first theme of the 2010 AAMI and FDA summit asked for standard ways to report, pool and analyze incidents, and pooled pump data now show what single reports cannot, but not what pumps never recorded.
- A pump knows its mechanism's motion and tests its assumptions only through sensors that answer late, so a perfect pump gives a wrong dose whenever the entry, bag, line, another infusion, software or people carry the error.