14 — Free flow

When the pump lets go.

+ The questionWhat happens when a set comes out of a pump with its clamp open, and how did designs stop it?

A pump is also a valve

In 1991 an aide switched off an infusion pump delivering oxytocin, a drug given to start or strengthen labor contractions, without closing the manual clamp on its set, and the mother received nearly 35 times the prescribed dose, as ECRI later recounted. The pump stopped when told to; the open clamp then left gravity in control of the line. Chapters 12 and 13 described what a pump senses while it holds the line. This chapter is about the moment it stops holding it, and about the long effort to make sure that something else holds it then.

A large-volume pump controls flow by closing the tube. The fingers of a peristaltic pump press it shut in sequence, and the valves of a cassette pump close off its chamber (Chapter 4), so that at every moment some part of the mechanism blocks the path from bag to patient. That closure makes the pump the line's valve as well as its motor. Once the set is out of the pump, or the door is open, only two things remain between the bag and the vein: the roller clamp of Chapter 1, if someone has closed it, and the resistance of the set and catheter. ECRI defines what follows: "Free-flow, or unrestricted gravity flow, refers to the uncontrolled delivery of an infusion to a patient."

38 — A pump as the line's valve
SET IN THE PUMP GRAVITY PUSH PUMP FINGERS VEIN 20 G HEART LEVEL 1 m ABOVE THE HEART 73.6 mmHg 30 mL/h, AS PROGRAMMED SET OUT, CLAMP OPEN PUMP DOOR OPEN ANTI-FREE-FLOW CLAMP: CLOSES ON REMOVAL, IF FITTED VEIN 20 G HEART LEVEL 1 m 61 mL/min = 3,660 mL/h maker's figure, 20 G, test conditions not stated THE ONLY VALVE LEFT Schematic. Head computed: 1,000 kg/m³ × 9.81 m/s² × 1 m. Flow: BD Insyte Autoguard product page.
Fluid and drugPressure and forceReadoutFocal detail
Plate 38 — While the set is in the pump, the mechanism is the line's valve. Out of the pump with the roller clamp open, only the resistance of set and catheter limits a 74 mmHg gravity push, and the flow is about 120 times the programmed rate.

Gravity's push is easy to estimate. A bag hanging 1 m above the heart drives the fluid with the weight of a 1 m column of water, about 74 mmHg, and MHRA's withdrawn guidance gave about 110 mmHg for a typical hanging bag (Chapter 1). A pump cannot measure flow it no longer controls. In a 2017 column summarizing its work on the hazard, ECRI noted that a pump's alarm can show only that its safety mechanism is not engaged, not that fluid is running, and that some pumps do not alarm at all. Makers, the column added, name the roller clamp as the primary safeguard.

Syringe pumps have their own version. A syringe held above the patient is a small hanging container, and if its plunger is not held by the pump's driver it can empty by gravity, which is called siphonage. The same MHRA guidance recommended anti-siphon valves for syringe users: valves that open only above a set pressure (Chapter 11), so that gravity alone cannot draw the syringe down. Neither an anti-siphon valve nor an anti-free-flow set, MHRA added, replaces closing the roller clamp, which it called "the primary means of occlusion".

Stopping a pump can leave its line running

The stop key halts the mechanism. If the set is out of the pump, or the pump does not hold the tube shut, the line keeps running at whatever gravity and resistance allow. A 2003 New South Wales safety newsletter advised that in an emergency the roller clamp must be closed, because stopping the pump is not enough.

How fast an open set runs

An open set runs far faster than any drug rate, and its speed is set by the narrowest parts of the path. By Poiseuille's law, the 20 G catheter of Chapter 2 needs only 0.61 mmHg to pass 100 mL/h of water at 20 °C. On the same law it would pass about 200 mL/min at the push of a bag 1 m up, but at that speed the flow is no longer smooth and the law overstates it; the flows makers measure are far lower. BD's product pages give gravity flow rates for its Insyte Autoguard catheters of 20 mL/min at 24 G, 61 mL/min at 20 G and 95 mL/min at 18 G, under test conditions the pages do not state.

39 — Programmed rate against gravity flow
PROGRAMMED, NOREPINEPHRINE 16 mcg/mL 30 mL/h · 250 mL BAG LASTS 8.3 h GRAVITY, 24 G × 40 · BAG EMPTY IN 12.5 min GRAVITY, 20 G × 122 · 4.1 min GRAVITY, 18 G × 190 · 2.6 min 122 TIMES THE ORDERED RATE 10 100 1,000 10,000 100,000 FLOW (mL/h, TENFOLD STEPS) Gravity flows: BD Insyte Autoguard product pages (maker's figures; test conditions not stated). Bag times for 250 mL.
Fluid and drugSoftware and dataFocal detail
Plate 39 — On one maker's gravity flows, an open set runs about 40 to 190 times faster than a 30 mL/h drug infusion, depending on the catheter, and empties a 250 mL bag in 2.6 to 12.5 minutes.
Solved example: gravity flow through an open set from a bag 1 m above the heart
  • Push of a 1 m column of water: 1,000 kg/m³ × 9.81 m/s² × 1 m = 9,810 Pa. One mmHg is 133.3 Pa, so the push is 9,810 ÷ 133.3 = 73.6 mmHg. The vein's own pressure would subtract a little; it is left out here.
  • Why Poiseuille's law fails here: scaled from 0.61 mmHg per 100 mL/h, 73.6 mmHg would pass 100 × 73.6 ÷ 0.61 = about 12,000 mL/h, or 200 mL/min. Through a 0.80 mm bore, whose cross-section is 0.50 mm², that is 3,333 mm³/s ÷ 0.50 mm² = 6.6 m/s. The Reynolds number, density × speed × bore ÷ viscosity, is then 1,000 × 6.6 × 0.0008 ÷ 0.001 = about 5,300, far above the 2,000 or so below which flow in a tube stays smooth, so 200 mL/min is not a real ceiling and the example uses measured flows.
  • Catheter as BD rates it (test conditions not stated): 20 mL/min at 24 G to 95 mL/min at 18 G, or 1,200 to 5,700 mL/h; at 20 G, 61 mL/min is 3,660 mL/h.
  • Against an order: a 250 mL bag of norepinephrine at 16 mcg/mL running at 30 mL/h gives 8 mcg/min, the label's lowest starting dose, and lasts 250 ÷ 30 = 8.3 h. Open through a 20 G catheter it runs 3,660 ÷ 30 = 122 times faster, giving about 976 mcg/min, and empties in 250 ÷ 61 = 4.1 minutes.
  • On these figures an open set runs about 40 to 190 times faster than this drug rate, depending on the catheter, and empties the bag in minutes.

The ratio, more than the absolute flow, is what makes free flow dangerous. The pump's rate set the dose rate. Unrestrained, gravity multiplies the dose rate by the same factor as the flow, so a drug whose dose was worked out to the microgram arrives at about a hundred times its ordered rate. The oxytocin case, at nearly 35 times the prescribed dose, is of the same order. Short-acting drugs of the kind Chapter 10 described act within minutes, so the bag can empty before anyone has looked at the line.

Making the set close itself

Free flow was recognized as a design problem long before it was solved. The 2017 ECRI column records that ECRI's 1982 evaluation of infusion controllers raised the need for safeguards, that in 1992 it recommended phasing out devices without set-based free-flow protection, and that in 1997 it rated devices lacking it "Unacceptable". A set-based anti-free-flow device is a clamp built into the administration set that closes the tube automatically whenever the set is taken out of the pump. The protection then travels with the set instead of depending on someone remembering the roller clamp.

Pressure also came from the Joint Commission, the main accreditor of US hospitals. On 30 November 2000 it issued a Sentinel Event Alert, its bulletin on a kind of serious harm, on infusion pumps. A trade newsletter reported the alert together with recommendations from the US Pharmacopeia, the body that sets US drug standards: identify pumps that can free-flow, quarantine or phase them out, and ask FDA and makers to stop such pumps reaching the market. Among the alert's cases, as the newsletter reported, was 60 mEq of potassium delivered in 20 minutes, three times the intended amount. From 1 January 2003 the National Patient Safety Goals of the Joint Commission, then called JCAHO, required "free-flow protection on all general-use and PCA (patient controlled analgesia) intravenous infusion pumps". Those are the goal's words as a trade newsletter quoted them. A New South Wales newsletter reported the start date that year.

Makers built the answer into the set. In May 2003 FDA cleared Deltec's CADD high-volume administration set with FlowStop, a feature that closes the tube if the set is misloaded or detached from the pump. FlowStop replaced a separate anti-siphon valve. Baxter's 2025 summary for its SIGMA Spectrum pump lists free-flow protection by a slide clamp in the set. No FDA rule is known to have required such protection; the push came from ECRI's ratings and the Joint Commission's goals.

Protection that depends on loading and service

A set that closes itself protects only if the pump, its maintenance and its instructions let it. MHRA listed incidents of free flow after a door-contact check was missed during maintenance, and after the wrong sets were used. In January 2025 BD began a Class II recall because the manual for its Alaris Pump Module 8100 told users to leave the safety clamp open when reloading a set, which risked over-infusion. FDA's record lists about 242,000 systems. In October 2025 BD began a Class I recall because the module's bezel assembly may be damaged if dropped or jarred, with unregulated flow among the possible results. Free-flow protection is a chain of parts and steps, and a missed check at service or a wrong instruction can break it.

Check free-flow protection at every service and set change

In the workshop, include the door and set-detection checks in every preventive maintenance, and confirm after any repair that the pump stops the flow or alarms when its door is opened with a set loaded. On the ward, close the roller clamp before opening the door or removing a set, and take out of service any pump that lets fluid run with the door open. For a maker, in India or elsewhere: build the protection into the set so that it closes on removal, list only sets that carry it, and read the instructions for any step that leaves a clamp open.

+ What this chapter established
  • A large-volume pump is also the line's valve, so once a set leaves it with the roller clamp open, gravity alone drives the flow, and the pump cannot measure it.
  • A bag 1 m above the heart pushes with about 74 mmHg, and one maker gives gravity flows of 20 to 95 mL/min through common catheters, about 40 to 190 times a 30 mL/h infusion.
  • ECRI raised the hazard in 1982 and rated pumps without set-based protection unacceptable in 1997, and the Joint Commission's goals required free-flow protection from 2003.
  • Protection travels with the set but depends on loading, service and instructions, as BD's 2025 recall over a manual that said to leave a safety clamp open showed.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.