+ Part V · Told what to do
Told what to do.
FDA's white paper of April 2010 recorded that "a software problem called a 'key bounce' caused an infusion pump to occasionally register one keystroke (e.g., a single zero, '0') as multiple keystrokes (e.g., a double zero, '00')." The paper named no pump and gave no dose. The paper filed the fault under software defects, yet the number the fault corrupted was one that a person had typed, and every infusion passes through such numbers before any mechanism moves. Chapter 16 turns a dose into a rate and shows where tenfold errors enter, Chapter 17 sets out which keypads and screens invite wrong entries, and Chapter 18 explains how a maker shows that a pump can be programmed safely. Chapter 19 turns to the connector at the end of the line, which can take a correct dose to the wrong place, and to the feeding pumps behind a new connector standard.
From an order to a keypad.
+ The questionHow does a dose in micrograms per kilogram per minute become a rate in milliliters an hour, and where does a tenfold error come from?
Three numbers stand between a dose and a rate
A dose rate is the amount of a drug that a patient is to receive in a unit of time, and for many drugs given by infusion it is also scaled to body weight. An order for norepinephrine, a drug that raises blood pressure, might read 0.1 micrograms per kilogram per minute, written 0.1 mcg/kg/min on charts and pump screens. The pump has no use for that figure. It moves a volume at a rate in milliliters an hour, and the sensors of Part IV watch only the line that the volume enters: the pressure behind a blockage, a bubble crossing an ultrasonic beam, a door or set out of place. None of them can tell whether the rate itself was the one the order meant.
Three numbers turn the dose rate into a pump rate, each in a separate step. The patient's weight in kilograms turns the dose per kilogram into a dose per minute. Sixty turns minutes into hours. The drug's concentration, the mass of drug in each milliliter of the bag, divides the dose per hour into milliliters per hour. Of the three, only sixty is fixed. The weight comes from a scale, a chart or an estimate, the concentration from a bag label or a drug library (Chapter 21). A pump in dose-calculation mode does the arithmetic once the weight, concentration and dose are entered; in plain milliliter mode it receives the answer worked out by a nurse. Either way, a person supplies the numbers that the arithmetic rests on.
A program carries more than a rate. The nurse also enters the volume to be infused, the amount after which the pump ends the infusion, and many pumps then drop to the keep-vein-open rate of Chapter 4. A secondary infusion (Chapter 4) hands back to the primary infusion when it is done. Each of these is another number typed or confirmed. The volume to be infused, divided by the rate, also gives the time the infusion should take, the simplest check on a rate that a screen can show, and the worst case in this chapter turns on its absence.
The bag sets the concentration
The concentration is the number most easily taken for granted, because it is printed on the bag. The FDA-approved prescribing information for Baxter's premixed norepinephrine, revised in February 2026, lists three strengths in 250 mL bags: 4 mg, 8 mg and 16 mg, which are 16, 32 and 64 mcg/mL. Those strengths are stated as norepinephrine base, the drug molecule alone. What is dissolved is a salt, norepinephrine bitartrate monohydrate, which weighs about twice as much for the same amount of drug, and the same label gives it as 31.9, 63.8 and 127.6 mcg/mL. A library entry or a hand calculation built on the salt figure for the first bag would deliver about half the intended dose, since 16 divided by 31.9 is 0.50.
The strength also decides the number that a nurse types. On the arithmetic of the solved example below, a dose that runs at 30 mL/h from the 16 mcg/mL bag runs at 7.5 mL/h from the 64 mcg/mL bag. A rate is therefore meaningless without its concentration, and one carried over from a handover note or a previous bag of another strength is wrong by the ratio of the strengths, here 2 or 4. The label itself doses in micrograms a minute, not per kilogram: it gives 8 to 12 mcg/min to start and 2 to 4 mcg/min to maintain; at 16 mcg/mL the maintenance dose is 7.5 to 15 mL/h and the starting dose 30 to 45 mL/h. Many intensive care units order the same drug per kilogram instead, so two conventions for one drug can circulate in one building.
The order is 0.1 mcg/kg/min for a patient of 80 kg, an illustrative weight, from the 4 mg in 250 mL bag on the FDA label. - Concentration: 4 mg is 4,000 mcg, and 4,000 ÷ 250 = 16 mcg/mL. - Dose per minute: 0.1 × 80 = 8 mcg/min, the lowest starting dose on the label. - Dose per hour: 8 × 60 = 480 mcg/h. - Rate: 480 ÷ 16 = 30 mL/h. - Decimal point moved one place to the right, 1.0 mcg/kg/min: 80 mcg/min, 4,800 mcg/h, 300 mL/h. - Moved one place to the left, 0.01 mcg/kg/min: 0.8 mcg/min, 48 mcg/h, 3 mL/h. The intended rate is 30 mL/h. One misplaced decimal point gives ten times or one tenth of the dose, and the pump receives 300 or 3, numbers it has no reason to refuse.
Norepinephrine is also titrated: its rate is raised or lowered in small steps against the patient's measured blood pressure, so one infusion may be reprogrammed many times, and each change is a fresh entry. The concentration entry can fail even in software built to check it. In February 2020 BD began a recall of its Alaris system for five issues, one of them errors in programming custom concentrations. FDA classified the recall Class I, and Chapter 26 follows it. ISMP, the Institute for Safe Medication Practices, places intravenous adrenergic agonists such as norepinephrine on its list of high-alert medications, drugs that "bear a heightened risk of causing significant patient harm when they are used in error."
How a decimal point multiplies a dose by ten
The factor of ten in a tenfold error is no coincidence. In written numbers each position is worth ten times the one to its right, so a digit added, a digit lost or a decimal point moved by one place multiplies or divides the whole number by exactly ten. Such tenfold errors arise wherever a number is written, read or keyed: in the order, in the pharmacy, at the pump. The white paper's key bounce produces the same factor mechanically. If the zero of a 30 mL/h rate registered twice, the pump would run at 300 mL/h.
A tenfold error is also large without being absurd, and that is what makes it hard to catch. One maker's 510(k) summary states the accuracy of its large-volume pump from 0.1 to 999 mL/h, and both 3 and 300 mL/h lie well inside that span, so a pump that checks only whether a rate is within its mechanical range cannot tell 300 from 30. Other conversion factors produce errors of their own size: 1,000 between milligrams and micrograms, 60 between minutes and hours, and the patient's weight between a dose per minute and a dose per kilogram per minute.
An entry of 8 meant as mcg/min but read as mcg/kg/min, for the same 80 kg patient, asks for 640 mcg/min. From the 16 mcg/mL bag that is 2,400 mL/h, beyond the range, and the pump refuses it; from the 64 mcg/mL bag it is 600 mL/h, and the pump accepts it. Whether a slip is caught therefore depends on the other numbers in the arithmetic, not on the size of the slip. A drug library narrows the accepted range for each drug, as Chapter 21 shows, but it can reject only what falls outside its limits.
The pump in a fatal case from Alberta in 2006 had no dose limits at all. ISMP Canada's root-cause analysis, published in 2007, describes fluorouracil, a cancer drug, ordered at 5,250 mg as a continuous infusion over four days. The intended rate was 1.2 mL/h. The pump was programmed at 28.8 mL/h, "24 times greater," and the drug ran in about four hours instead of 96. The factor of 24 is the number of hours in a day: 28.8 mL is what the pump should have delivered in a whole day. The pump did not show how long the infusion would take, so nothing on its screen contradicted the entry. The patient died on 22 August 2006, and Chapter 25 returns to the pump's data history, which confirmed the rate afterward.
Bedside studies count different things
Studies that watch infusions at the bedside show how often rates go wrong, and how rarely the cause is a slip at the keypad alone. A 2018 study in 16 English hospital trusts observed 2,008 infusions and found errors in 231 of them, 11.5 %, and discrepancies, which it counted separately, in 1,065, or 53.0 %. Only 23 infusions, 1.1 %, carried errors judged potentially harmful. Deviations in rate appeared in 152 infusions, 7.6 %, split almost evenly between errors and discrepancies.
A 2005 study at Northwestern Memorial Hospital found at least one error in 285 of 426 pump infusions, 66.9 %, but of its 37 rate deviations only 3 were programming mistakes, and its authors judged that the smart pumps of the day would have prevented only one error without further features. A 2017 study in ten US hospitals observed 1,164 intravenous administrations and found at least one error in 699 of them, or 60 %. Two of its authors were on the staff of BD, a pump maker, and the CareFusion Foundation and AAMI funded it. Its count included deviations from hospital policy, such as labeling. Its most frequent errors were unauthorized medication, infusions run without the smart pump's drug library, and wrong rates.
The 60 %, 11.5 % and 66.9 % above are not three estimates of one quantity. One study counted deviations from hospital policy as errors, one set errors apart from discrepancies, and one found that only 3 of its 37 rate deviations were programming mistakes. None is a rate of keypad slips, and an error rate can be compared only with another built on the same definitions.
The pattern across the three studies holds even though their totals differ. Wrong rates are common, but a slip at the keypad is one source among several: a bag of another strength, a rate carried over from an earlier order, a library skipped, an order never written. The pump sees only the final number. Chapter 17 asks which ways of entering it make a wrong one more likely, and Chapter 18 how a maker shows that its own screens do not invite such errors.
For pharmacists building a drug library, enter each concentration exactly as the label states it, noting whether it is base or salt, and in the same unit as the order. Keep the number of strengths per drug small, because each added strength is another factor by which a carried-over rate can be wrong. Give each drug entry one dose unit, per minute or per kilogram per minute, so that the two conventions cannot meet on one screen. Check that the pump shows the expected duration beside the rate, and add that check to the library's acceptance tests.
- A pump receives a rate in milliliters an hour, and the dose rate in the order reaches it only through a weight, a concentration and the factor of sixty, each supplied by a person or a library.
- The concentration on the bag decides the number keyed: the same norepinephrine dose runs at 30 mL/h from a 16 mcg/mL bag and at 7.5 mL/h from a 64 mcg/mL bag.
- Because each written digit is worth ten times the next, a moved decimal point, a lost digit or a doubled zero changes a rate by exactly ten, and 3 or 300 mL/h lies inside one maker's stated range.
- Bedside studies find wrong rates often but keypad slips rarely among them, and their headline error rates count different things.
Let's Talk