28 — United States

Clearance after 2010.

+ The questionWhat does FDA ask of a new pump, and how did the 2010 initiative change that?

A warning letter about changed pumps

On 4 April 2025, FDA sent ICU Medical a warning letter about two pumps that FDA had not reviewed in their present form. Inspectors had visited the former Smiths Medical site in Minneapolis between 23 July and 9 August 2024. They found that the Medfusion 4000 syringe pump, cleared in 2011, and the CADD-Solis VIP ambulatory pump, cleared in 2013, had been changed without new premarket notifications. One of the changes was a Medfusion software update of 28 April 2023. The letter said that, without those notifications, the pumps were adulterated and misbranded under US law, terms that here cover devices sold without the clearance their changes required. A pump in the United States is cleared as it was described to FDA, and these pumps no longer matched their descriptions.

FDA's rules for a new pump, which draw on the standards of Chapter 27, have four parts: the class FDA gives a pump and the comparison a 510(k) makes; the guidance in force since 2 December 2014; the design controls of a quality regulation in force since 2 February 2026; and the cybersecurity duties added in 2023. The guidance began as a draft issued on the day FDA announced its 2010 Infusion Pump Improvement Initiative, and the chapter ends with what FDA says that initiative changed.

Class II, cleared by comparison

The federal regulation for infusion pumps, 21 CFR 880.5725, dates from 21 October 1980 and keeps its 1980 wording. It describes a device used in a health care facility to pump fluids into a patient in a controlled manner, and says the device may detect faults, such as air in the line or a blockage, and sound an alarm. It places the pump in Class II, the middle of FDA's three risk classes. Seventeen product codes sit under it, from FRN for the general pump to PHC for infusion safety management software. All of them need a 510(k), and none may be reviewed by an accredited third party.

A 510(k) is a premarket notification that compares a new or changed device with one already legally sold, called its predicate. It takes its name from the section of the Federal Food, Drug, and Cosmetic Act that requires it. FDA clears a 510(k) when it finds the two devices substantially equivalent; it does not approve it. The predicate need not work the same way. The 2019 summary for the Ivenix pump named Baxter's SIGMA Spectrum and ICU Medical's Plum 360 as predicates, although the Ivenix pump measures the volume in its pumping chamber with the gas law, a principle neither of them uses (Chapter 4). What a pump's 510(k) must contain is set out in a guidance of its own.

72 — A pump's path through FDA
21 CFR 880.5725 CLASS II 1980 regulation 17 product codes 510(k) COMPARED WITH A PREDICATE PRE-CLEARANCE INSPECTION new technology, no inspection in 2 years, or filed to fix failures CLEARED SUBSTANTIALLY EQUIVALENT IN USE reports and recalls (Ch 31) 524B plan continues SAFETY ASSURANCE CASE · 2014 GUIDANCE HAZARD ANALYSIS · 4 SYSTEM HAZARDS PERFORMANCE TESTING · TIR101 SOFTWARE · STATIC ANALYSIS HUMAN FACTORS VALIDATION CYBERSECURITY · 524B · FROM 29 MAR 2023 A CHANGE THAT NEEDS IT RETURNS AS A NEW 510(k) · WARNING LETTER, 4 APR 2025 Schematic. 21 CFR 880.5725; FDA guidance of 2 Dec 2014; FD&C Act sec. 524B.
Software and dataHuman input and use errorReadoutFocal detail
Plate 72 — A US pump is cleared as it is described in its 510(k), with a safety assurance case at the center; a later change that needs a new 510(k) sends it back along the same path.

The 2014 guidance asks for a safety case

FDA issued that guidance in draft on 23 April 2010, the day it announced the initiative, and in final form on 2 December 2014 as "Infusion Pumps Total Product Life Cycle". At its center is a safety assurance case, a structured argument that the pump is safe for its intended use. The case is made of claims, the arguments that link them and the evidence behind each. It starts from a hazard analysis, which the maker must argue is complete. The analysis is organized under four system hazards: infusion delivery error, incorrect therapy, contamination and traumatic injury. Their causes are grouped as operational, environmental, electrical, hardware, software, mechanical, biological or chemical, and use.

The rest of the guidance says what evidence to bring. Accuracy must hold over the whole infusion, at the pump's operating limits and across the conditions of Chapter 8. It must be shown in every configuration, or in representative ones the maker can justify. The software is treated as a major level of concern, and all of it is to be checked by static analysis. The guidance expects a report of a human factors validation study (Chapter 18), and a wireless pump's submission must describe its radio technology, data and security. The guidance gives no numerical test conditions, does not cite IEC 60601-2-24 and does not mention trumpet curves.

Much of that evidence is test reports, and the standards behind them can come from FDA's list of recognized consensus standards, which a maker can cite in a submission. IEC 60601-2-24 is not on the list: FDA's database held no record of it in October 2026. AAMI TIR101:2021 has been recognized since 30 May 2022. The broader standards a pump shares with other devices are recognized in their current editions, from IEC 60601-1 as amended in 2020 to ISO 14971:2019 and ISO 80369-1:2025. Which of them a 510(k) summary cites varies by maker (Chapter 27).

The guidance also lets FDA look at the factory before clearance. FDA may conduct a pre-clearance inspection when a pump has new technological characteristics or the 510(k) was filed to fix failures. It may also inspect when the maker has gone two years without an inspection, or when the last inspection did not cover similar products or found that action was needed. FDA may withhold clearance until the inspection is done. Neither the guidance nor FDA's account of the initiative says how many such inspections were made.

Parts of the 2014 guidance have been overtaken

The guidance's "major level of concern" belongs to a scheme that FDA replaced in June 2023 with software documentation levels. A notice now carried in the guidance says infusion pumps should generally meet the enhanced level. Its human factors section predates FDA's two human factors documents of 2026 (Chapter 18), and its wireless section predates section 524B. A submission in 2026 takes its structure from the guidance and its detail from the newer documents.

Design controls and the trace from hazard to test

How a maker designs a pump is regulated too, under rules called design controls. Until 1 February 2026 they were 21 CFR 820.30: design planning, inputs, outputs, review, verification, validation, transfer to production and control of changes, recorded in a design history file. On 2 February 2026 the Quality Management System Regulation, published in 2024, replaced FDA's earlier Quality System Regulation. The new regulation incorporates ISO 13485:2016, the international quality standard for medical devices, whose requirements FDA judged substantially similar to the old ones. Design now follows the same elements in clause 7.3 of ISO 13485, and a design and development file for each device type or family replaces the design history file.

Verification and validation keep the meanings given in Chapter 23, and traceability joins them. Clause 7.3.2 asks the design plan to state how outputs will be traced to inputs. FDA's 2014 pump guidance notes that "A traceability analysis is useful in demonstrating that all identified hazardous situations have at least one corresponding control".

The trace also decides what happens after clearance, because it shows which risk controls a change touches. Clause 7.3.9 asks that a change be reviewed for its effect on the device's "constituent parts". FDA's two final guidances of October 2017 set out when a changed device needs a new 510(k). The software guidance asks, among other questions, whether a change creates or modifies a risk control for a hazard that could cause significant harm. Its own pump example is the occlusion alarm of Chapter 12: splitting one alarm into upstream and downstream alarms is likely to need a new 510(k), because it modifies an existing control. A change made solely to strengthen cybersecurity, by contrast, is in many cases not likely to need one. Either way the decision is documented and the change verified. In FDA's view, the 2023 Medfusion update named in the warning letter needed a new 510(k).

Software and cybersecurity since 2023

Section 524B was added to the Federal Food, Drug, and Cosmetic Act by the Consolidated Appropriations Act, signed on 29 December 2022. It applies to a cyber device: one that includes software validated, installed or authorized by the sponsor, can connect to the internet, and has characteristics that could be vulnerable to cybersecurity threats. A sponsor that submits one for premarket review must include a plan to monitor, identify and address vulnerabilities after sale, including coordinated vulnerability disclosure, in which outside researchers report a vulnerability to the maker so that it can be fixed before the details are made public. It must keep processes that provide patches on a regular cycle it can justify, and out of cycle for critical vulnerabilities. It must also provide a software bill of materials, the list of commercial, open-source and off-the-shelf components in the device. FDA applied the section to submissions from 29 March 2023, and its transition period ended on 1 October 2023.

FDA's current premarket guidance on cybersecurity is the final version of 27 June 2025, which added a section on 524B. DEKA's 510(k) summary for the RemunityPRO+, cleared on 24 July 2026, shows what this means for a pump. It lists the changes from the predicate as two-way cloud communication, secure pairing, software updates over the air and cybersecurity controls for use on phones or tablets the user supplies, with no change to clinical performance. Each is a route by which the pump's instructions can change after it leaves the factory (Chapter 22), so the plan that section 524B requires lasts as long as the pump is in use.

What FDA says changed after 2010

FDA's account of the initiative lists three lines of action: requirements for makers, help with device improvements and awareness among users. By its own account, FDA inspected all 20 infusion pump manufacturers registered with it as of November 2011. It kept firms with serious findings under surveillance and put the rest on a two-year cycle. It reports an overall improvement in findings but gives no numbers. A pilot that tried safety assurance cases on 30 510(k)s found longer response times for sponsors, more clearances, fewer withdrawals and no significant change in review time, again without counts. The number of FRN clearances, by contrast, can be counted from FDA's data.

73 — FRN 510(k) clearances by year, 2015 to 2026
0 3 6 9 CLEARED 510(k)s UNDER FRN GUIDANCE · 2 DEC 2014 524B APPLIES · 29 MAR 2023 3 2015 3 2016 3 2017 3 2018 2 2019 1 2020 2 2021 6 2022 7 2023 5 2024 8 2025 TO 11 OCT 2 2026 26 OF 45 · 58 % Cleared decisions under product code FRN only; openFDA 510(k) data, updated 28 Sep 2026. Excludes PCA (MEA), elastomeric (MEB), enteral (LZH) and safety software (PHC).
ReadoutFocal detail
Plate 73 — Clearances of general infusion pumps rose after 2021: 26 of the 45 since 2015 fell in 2022 to 2025. The count includes changed pumps as well as new designs.
Solved example: what the FRN clearance count shows

FDA's 510(k) data list 45 cleared decisions under FRN from 1 January 2015 to late September 2026, or 46 counted from the guidance date of 2 December 2014. From 2015 to 2021 the yearly counts are 3, 3, 3, 3, 2, 1 and 2: 17 in seven years, about 2.4 a year. From 2022 to 2025 they are 6, 7, 5 and 8: 26 in four years, or 6.5 a year. 2026 has 2 so far, and 17 + 26 + 2 = 45. The four years from 2022 hold 26 of the 45, or 58 %. The counts use FDA's dataset as updated on 28 September 2026. They cover FRN only, leaving out PCA, elastomeric and enteral pumps and safety software. A clearance can cover a changed pump as well as a new one, so the rise from 2022 is a rise in filings, not a count of new designs.

Plan a US pump file around the safety case

Choose the predicate early and list every difference from it, because each difference needs a claim, an argument and evidence in the safety assurance case. Build the hazard analysis under the guidance's four system hazards and its cause categories, test flow to AAMI TIR101, and write the 524B vulnerability plan and software bill of materials with the design. After clearance, keep the trace from each hazard to its verified control up to date. Put every change through FDA's 2017 questions on whether it needs a new 510(k), and file the decision; the 2025 warning letter turned on that question.

+ What this chapter established
  • Infusion pumps are Class II under 21 CFR 880.5725 and reach the US market through a 510(k) that FDA clears against a predicate, with evidence built into the safety assurance case of FDA's 2014 guidance.
  • Since 2 February 2026 a pump's design controls follow clause 7.3 of ISO 13485 under the Quality Management System Regulation, tracing each hazard to a verified control, and FDA's 2017 guidances decide when a change needs a new 510(k).
  • FDA recognizes AAMI TIR101:2021 but not IEC 60601-2-24, and a connected pump submitted for premarket review on or after 29 March 2023 must also meet section 524B, with a vulnerability plan and a software bill of materials.
  • FDA cleared 45 pump 510(k)s under FRN from 2015 to September 2026, 26 of them in 2022 to 2025, and reports improvement after 2010 without publishing outcome numbers.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.