24 — Fleet

Thousands of pumps, one library.

+ The questionWhat does a hospital have to keep doing, pump by pump and library by library, once the pumps arrive?

The assumptions a hospital keeps

A 50 mL graduated cylinder is one of the two instruments that the maker's 2009 service manual for the SIGMA Spectrum pump names for its yearly flow check. The technician runs the pump at 200 mL/h into the cylinder, or onto a balance, and compares what arrived with what the pump says it moved. Chapter 4 showed why that comparison is needed: the pump counts the motion of its mechanism and converts it into milliliters through assumptions about the tubing it squeezes. Chapter 23 left the hospital owning the system once the pumps arrive, network and library included. Neither the pump's count nor its network can tell whether the mechanism still matches the numbers it was built with.

Every number on a pump's screen rests on assumptions that hold for the pump as tested and drift in the pump as used. The mechanism wears, a battery loses capacity, a keypad seal lifts, and the pharmacy replaces the drug library. None of these changes shows on the screen, because the pump checks its dose against the same parts that changed. Keeping each assumption true falls to the biomedical engineers, also called clinical engineers, who maintain the equipment, and to the pharmacists who build the drug library. Their unit of work is the fleet, every pump of a kind the hospital owns, and fleets are large: a conference paper from one engineering service in British Columbia counted over 18,000 pumps, an estimated 4 % or more of them misplaced.

Checks on a schedule, and checks on a risk

How often a pump returns to the workshop is set first by its maker, and makers differ. The Spectrum manual recommends preventive maintenance, the scheduled check made whether or not anything seems wrong, at least once a year and after any drop. A 2014 service manual for BD's Alaris GP volumetric pump asks for it at least every three years. In the United States, a 2013 memorandum from the Centers for Medicare & Medicaid Services lets a hospital replace maker schedules with an alternate equipment management program, provided it is documented, based on risk and followed strictly. A 2013 study that modeled maintenance histories of infusion pumps and ventilators found that makers' intervals did not match observed failure rates.

The checks test the parts that the dose depends on. In the Spectrum procedure the flow test runs at 200 mL/h and passes within ±5 %, by weighing a 50 mL delivery or by collecting 40 mL in the cylinder. The occlusion test confirms that the downstream alarm trips within 13 ± 6 psi, about 360 to 980 mmHg, and the air test that a bubble at least 1 inch long, about 25 mm, raises an alarm. A battery test completes the checks, and a pump outside any band is taken out of service.

63 — One maker's preventive maintenance checks
LARGE-VOLUME PUMP FLOW · 200 mL/h PASS 47.5 TO 52.5 mL (± 5 %) ON A 50 mL DELIVERY or 40 mL collected in a 50 mL cylinder the check on motion → volume, made on the bench DOWNSTREAM OCCLUSION ALARM TRIPS AT 13 ± 6 psi about 360 to 980 mmHg (1 psi ≈ 51.7 mmHg) AIR IN LINE ALARM ON A BUBBLE ≥ 1 in (ABOUT 25 mm) BATTERY AT 125 mL/h NEW ≥ 8 h (ABOUT 1,000 mL) · USED ≥ 6 h (ABOUT 750 mL) NOT TESTED: the ward's set, bag height, low rates, the line beyond the pump SIGMA Spectrum service manual, Rev W, 2009 · maker's figures
Fluid and drugPressure and forceAirReadoutFocal detail
Plate 63 — A yearly check tests the mechanism and sensors against the maker's bands, on the bench and at one rate. It does not test the set, the height or the line a pump will meet on the ward.

A pass shows that the mechanism converts motion into volume within its band on the bench, at one rate, with a set from the workshop drawer. It says nothing about the next set's tubing, the bag height on the ward, a rate of 1 mL/h or the line beyond the pump (Chapters 8 and 9): it checks the pump, not any infusion. The workload is large. For Brigham's 3,800 large-volume pumps, a yearly schedule would mean about 15 checks every working day over 250 working days, and a three-yearly one about 5, though the two makers' intervals were written for other models.

Batteries and keypads wear out in use

The pump cannot fully judge its battery. The Spectrum test runs a charged pump at 125 mL/h: a new battery must last at least 8 hours, about 1,000 mL, and a used one passes at 6 hours, about 750 mL. B. Braun's 2013 service bulletin says batteries in its Space pumps should be tested or replaced after four years, because an aged one may not supply the current for a syringe change even when its gauge shows full. The gauge is an estimate. In 2023, with an update in 2024, ICU Medical corrected its instructions for one supplier's batteries in Plum pumps, whose capacity could fall faster than expected, and FDA classified this correction as a Class I recall. The company puts the low-battery alarm at an estimated 30 minutes of running time.

Cleaning is the other daily stress. Pumps are wiped down between patients, and the agents an infection-control team prefers are not always those the maker tested; the Spectrum manual allows 10 % bleach and isopropyl alcohol but forbids phenolic cleaners, sprays aimed at the pump and immersion. In 2020 BD recalled front-case keypads for its Alaris PC unit and Alaris Pump Module because keys could stick or stop responding after fluid ingress, liquid getting inside the case, delaying or interrupting infusions; FDA classified the recall Class I. A follow-on recall in March 2021 covered about 160,000 units with keypads made in 2019 that could lift and let fluid in, and FDA recorded the root cause as the supplier's process design.

A library update reaches the last pump late

A drug library changes more often than any part. Pharmacists add drugs, correct limits and adjust concentrations, and each change makes a new version that must reach every pump. A 2017 quick guide from the AAMI Foundation (Chapter 21) advises defined update intervals, monthly or quarterly for example, separate routes for urgent changes, and wireless distribution where possible. It measures the result as library currency, the share of pumps carrying the current library, and suggests that hospitals updating by cable mark each updated pump with a colored sticker.

Wireless distribution is faster than a cable, and still slow. A 2019 study of BD Alaris pumps in two hospital systems followed each pump from noncurrent through pending to current: half a fleet became current five to nine days after a release, sooner for busy pumps, later for those left switched off. A 2018 study by the same group timed the delay from a library's replacement to the last alert raised by the old version, and across 49 hospitals in 12 health systems found medians of 22 to 192 days and a longest of 661 days. A 2024 report from an 11-hospital system using ICU Medical's Plum 360 gave about 75 % of the fleet updated within 72 hours; two authors had received ICU Medical research grants.

64 — A library release reaching a fleet
NONCURRENT PENDING CURRENT first switched on after the release time powered on (busier pumps sooner) switched off or stored: stays noncurrent 0 % 25 % 50 % 75 % 100 % 0 5 10 15 20 25 30 SHARE OF FLEET CURRENT DAYS AFTER RELEASE 5 TO 9 DAYS tail across 12 systems: medians 22 to 192 days, longest 661 days (2018) half the fleet current schematic curve · 50 % at 5 to 9 days from Hsu et al. 2019 (BD Alaris, two hospital systems) · tail from DeLaurentis et al. 2018
Software and dataReadoutFocal detail
Plate 64 — A pump takes the new library only when it is switched on and connected, so half a fleet was current after five to nine days, and the last pumps of some fleets took months.
Solved example: how long a library update takes to reach a fleet

This example uses Brigham's 3,800 large-volume pumps and the 2019 finding that half a fleet is current five to nine days after a release. On those days 3,800 × 0.5 = 1,900 pumps still carry the previous version, and the AAMI measure of currency is 1,900 ÷ 3,800 × 100 = 50 %. At the 11-hospital system's pace, 3,800 × 0.25 = 950 pumps are still on the old version after three days. The 2018 medians of 22 to 192 days are about 3 to 27 weeks, and the longest, 661 days, is 661 ÷ 365 = 1.8 years. A limit corrected in the pharmacy stays uncorrected on hundreds of pumps for days and on a few for months. The result combines inputs from different hospitals.

Wireless updates leave a tail of pumps on old libraries

Releasing a library over the network changes the file on the hospital's server, not every pump. Pumps that are switched off, out of range or in storage keep the version they last received, and in the 2018 study the slowest still raised alerts from a replaced library more than a year and a half later. Library currency is a figure to measure pump by pump.

Replacing a fleet, and changing one

Sometimes a hospital changes every pump at once. Brigham and Women's had used Alaris large-volume pumps since 2003, and its one-day swap to Baxter's Spectrum IQ took twelve teams, about 120 hospital staff and 40 Baxter representatives, with wireless programming and documentation through the health record as the stated aim (Chapter 22). More often a recall decides. After the final order of 13 July 2010 that closes the opening story, Baxter offered SIGMA Spectrum pumps or a payment in exchange for its Colleague pumps; it charged $588 million against its 2010 results and reported the recall substantially complete in July 2012.

After BD's Class I recall of February 2020 (Chapter 26), BD shipped new Alaris pumps in the United States only for medical necessity until FDA cleared a new 510(k) in July 2023. The recall and the clearance were three and a half years apart. Through that time hospitals with Alaris fleets kept, repaired and updated pumps they could not renew in the ordinary way, and the clearance brought a second task: remediation, meaning correction in place, or replacement of every device, which BD said would follow.

65 — Two recalls that set a fleet's timetable
COLLEAGUE · BAXTER Jul 2005 shipments stopped 30 Apr 2010 initial order 13 Jul 2010 final order ABOUT 200,000 PUMPS IN USE (BAXTER'S COUNT) Jul 2012 recall substantially complete ALARIS · BD 4 Feb 2020 Class I recall 870,070 UNITS IN COMMERCE US shipments only for medical necessity 21 Jul 2023 510(k) K211218 cleared remediate or replace (BD) ABOUT 3.5 YEARS ONE HOSPITAL Brigham and Women's: about 3,800 pumps replaced in one day 2005 2010 2015 2020 2024 Baxter Form 10-K 2010 and 2012 · FDA recall Z-1359-2020 · FDA K211218 · BD 10-K 2020 and release of 21 Jul 2023 · Brigham Bulletin 2019
ReadoutFocal detail
Plate 65 — A recall, not the hospital, often sets when a fleet is replaced: Colleague pumps stayed in use for years after shipments stopped in 2005, and Alaris fleets waited about three and a half years for a cleared fix.

In 2020 the intensive care units of Mount Sinai Morningside made a third kind of fleet decision: they changed how every pump was used. Pumps for patients with COVID-19 moved outside the rooms on long extension lines, with higher occlusion limits and a carrier on every line, so one decision moved the lag of Chapter 3 and the alarm limits of Chapter 12 for every infusion together. Its paper reports no patient numbers, no pump brand and no outcomes. The decision was the hospital's, and so was the checking.

Keep fleet measures beside the infusion measures

For biomedical engineering and pharmacy: after every library release, list the pumps that have not been switched on since, because they are the ones stored or out of range, and report currency pump by pump. Record battery age and measured run time at each check, not only pass or fail. Check the maker's list of cleaning agents against what infection control uses, and test any new agent before it reaches the wards. When a recall or an emergency changes how every pump is used, list the accuracy and alarm assumptions it touches and measure them on the bench, as ECRI did in 2020 (Chapter 12).

+ What this chapter established
  • A pump checks its dose against its own parts, so a hospital's biomedical engineers and pharmacists keep the assumptions behind those parts true across the fleet.
  • Preventive maintenance tests flow, occlusion, air detection and batteries against maker bands at intervals that differ several-fold between makers, and a pass shows the pump is sound, not that every infusion is.
  • Batteries and keypads wear in use and in cleaning, and the Plum battery and Alaris keypad recalls show that either can put infusions at risk.
  • A library update reached half of one fleet in five to nine days and the last pumps of some fleets months later, while recalls and emergencies such as Mount Sinai's in 2020 can change every pump at once.