+ Part VIII · The field in 2026
The pump, the set and the software.
On 19 August 2026 BD wrote to its customers that its Alaris Pump Module 8100 "may perform worse than previously disclosed" with certain infusion sets at low flow rates or small bolus volumes. It told them to discard the listed sets, starting in critical care, neonatal, infant and pediatric areas. FDA posted the recall as Class I on 15 September 2026. Its record covers about 94 million sets under about 50 catalog numbers and gives the reason as "discrepancies in specific attributes in the performance data for infusion sets". No pump was removed. The recall concerned the published behavior of the plastic that turns the pump's motion into milliliters. Part VIII describes who makes and sells infusion pumps as of October 2026, sorted by pump type and by layer: the pump, the set, the drug-library software and the connection to the hospital network, with India included.
Who makes infusion pumps.
+ The questionWho builds and sells each kind of pump, and what has changed since 2020?
A clearance shows who filed, not who sells
Chapter 25 answered this guide's question: a pump knows how far its mechanism has moved and what it was told. Each part of that answer is a product with a maker. The mechanism is the pump. The set's bore and plastic turn the mechanism's motion into milliliters, which is why makers state accuracy for named sets and syringes (Chapter 8). The drug library that checks what the pump was told is software, released and updated on its own schedule (Chapter 21). The connection that carries orders, libraries and updates from the hospital network is a separate layer with software of its own (Chapter 22). This chapter sorts the makers by pump type and by these layers, using FDA's records and the companies' own statements, and ranks none of them.
From January 2025 to late September 2026, FDA cleared ten 510(k)s under FRN, the product code it gives to the general infusion pump, and they went to six companies: BD, Baxter, ICU Medical, Iradimed, Koru Medical and DEKA. Fresenius Kabi, which sells the Ivenix pump in the United States, received none of them, yet it appears in 15 of FDA's recall records since 2025, seven of them Class I. A count of clearances shows who brought a new or changed pump to FDA, not who sells pumps or how many.
Large-volume pumps after 2020
BD makes the Alaris system through its subsidiary CareFusion 303. The system shows how long a large-volume pump can take to return after a recall. On 4 February 2020 BD began a Class I recall of Alaris point-of-care units and modules for five problems, among them a low-battery alarm that could fail and errors in programming custom concentrations; FDA's record counts about 870,000 units in commerce. BD committed to seeking a comprehensive new 510(k), and FDA cleared it on 21 July 2023 for the Alaris system with Guardrails Suite MX version 12, about three and a half years after the recall began. A further Alaris 510(k) followed on 25 April 2025. Chapter 24 follows what the wait and the promised remediation meant for hospital fleets.
Baxter sells two families. Its Spectrum IQ and SIGMA Spectrum systems received 510(k) clearances on 28 July 2025. Its newer Novum IQ large-volume pump was the subject of a Class I correction begun on 14 July 2025, described in Chapter 31. Baxter first disclosed a pause in distribution with its second-quarter 2025 results, at the end of July 2025. In its report for the quarter ended 30 June 2026 it said it had elected to temporarily stop distributing and installing the Novum LVP in the United States and Canada except where medically necessary. Whether the pause has since ended is not known.
ICU Medical added two pump lines by acquisition. In January 2022 it completed its purchase of Smiths Medical, whose products included the Medfusion syringe pumps and the CADD ambulatory pumps, and added them to the Plum cassette pumps it already sold. On 2 April 2025 FDA cleared the new single-channel Plum Solo and an updated version of the dual-channel Plum Duo, first cleared in August 2023, both with the company's LifeShield software. Two days later FDA sent ICU Medical a warning letter saying that changes to the Medfusion 4000 and the CADD-Solis VIP had needed new 510(k)s, a question Chapter 28 takes up. The company says it filed 510(k)s for a next-generation Medfusion and a refreshed CADD pump in July 2025; no clearance of either is known as of October 2026.
ICU Medical's announcement claims "±3% accuracy in real-world conditions" for the Plum Solo. The company's 510(k) summary in FDA's record gives ±3 % for 0.1 to 999 mL/h under the standard conditions of AAMI TIR101 (Chapter 9), ±5 % under non-standard conditions, and ±5 % for the predicate, the Plum Duo. In an illustrative order of 1,000 mL at 125 mL/h, the pump counts the volume down over 8 hours from the motion of its own mechanism. Under standard conditions the volume actually delivered when the count reaches zero lies between 1,000 × 0.97 = 970 mL and 1,000 × 1.03 = 1,030 mL, a spread of 60 mL. Under non-standard conditions it lies between 950 and 1,050 mL, a spread of 100 mL. Which band applies depends on the conditions of use, and only the summary names them.
Fresenius Kabi completed its purchase of Ivenix on 4 May 2022, weeks after FDA cleared Fresenius Kabi's own Agilia Connect pumps on 1 March 2022. The Ivenix pump is one of the two exceptions to the rule that pumps infer volume rather than measure it: its 2019 clearance describes a pneumatic pump that measures the volume in its pumping chamber with the gas law, as Chapter 4 explained. B. Braun's Space system pairs the Infusomat Space large-volume pump with the Perfusor Space syringe pump and the DoseGuard drug library. For the MRI suite, B. Braun's SpaceStation MRI, cleared in 2020, holds up to four Space pumps, and Iradimed's MRidium 3870 was cleared in May 2025.
Syringe, ambulatory, elastomeric and feeding pumps
Syringe pumps come from the same large firms, as modules or standalone units: the Alaris syringe module, ICU Medical's Medfusion, B. Braun's Perfusor Space and Baxter's Novum IQ syringe pump. Baxter's syringe pump had Class I corrections, begun on 4 August 2025, for a blank run screen and false motor-movement errors. Ambulatory and PCA pumps include ICU Medical's CADD-Solis, Moog's CURLIN 6000 and 8000, and Eitan Medical's Sapphire. Eitan says it launched its Avoset system for regional anesthesia in the United States on 16 April 2026. DEKA's RemunityPRO+, cleared on 24 July 2026, is built for one drug, subcutaneous treprostinil. Its 510(k) summary describes acoustic volume sensing as feedback for monitoring delivery accuracy, which makes it the second exception.
Elastomeric pumps, which have no motor (Chapter 7), are cleared under a product code of their own, MEB, with close to 90 records in FDA's 510(k) database by October 2026. Among the holders are Halyard for the ON-Q pump, which Avanos now sells, Baxter for its Infusor, Nipro for SureFuser and, most recently, Epic Medical of Singapore for SMARTeZ, cleared on 24 April 2026. A 510(k) holder is not always the present seller, as the ON-Q shows. Enteral feeding pumps have another code, LZH, with 34 records, among them Cardinal Health's Kangaroo OMNI of 2022 and Luminoah's FLOW system, cleared in April 2026. Interoperable insulin pumps have yet another code, QFG, held by eight companies from Tandem and Insulet to Beta Bionics; Chapter 6 covers these pumps.
The set, the library and the network
The recalls of 2025 and 2026 reached beyond the pump. On the set side, BD began a Class I recall on 8 July 2025 because the Alaris Pump Module with compatible sets might perform outside its established ranges, which Chapter 8 lists; FDA's record counts about 33 million sets. A further Class I recall of Alaris sets began in September 2025, and the August 2026 recall that opens this part covers about 94 million. The two large counts belong to separate recalls, and FDA's records do not say whether any set is counted in both. ICU Medical's IV sets with drip chambers had a Class II recall in May 2026 for black specks in the plastic, covering about 100 million units.
Library and connection software have recalls of their own. In February 2025 BD began a Class I recall of its Alaris Systems Manager, the software that relays auto-programming requests to the pumps (Chapter 22), because it could send an outdated request to the point-of-care unit. ICU Medical recalled its LifeShield library software in November 2025 (Class II) because its change-summary review failed after a library migration. In April 2026 it recalled Plum Solo and Duo pumps that might not show a soft-limit alert before delivery started. On the network side, Alaris labeling did not say which version of the TLS encryption protocol had to be enabled for secure Wi-Fi to the pumps (Class II, February 2026, about one million units).
The quantity in an FDA recall record counts whatever that record counts: pumps in commerce, modules with repair kits, sets or hospital systems. The Alaris units of 2020, about 870,000, and the sets of 2026, about 94 million, differ by a factor of about 108, but a set serves one patient and is thrown away, while a pump stays in service for years. The ratio says nothing about relative risk or market size. Fresenius Kabi's Class I software recall of May 2026 counts "32 systems", which may mean hospital servers rather than pumps; FDA does not define the term.
India: makers, a launch and an estimate
India's pump makers are documented mostly by their own websites, so what follows is what the companies say about themselves. AKAS Infusions of Chennai, founded in 1996, has factories in Tiruchirapalli and Chennai and lists four syringe pumps, among them the Syru 4004, and two infusion pumps. It says that more than 50,000 of its products are installed and calls two models "CE approved". Skanray Technologies of Mysuru sells the floSKAN 3000 syringe pump, listed at 0.1 to 2,000 mL/h with a library of 1,000 drugs; its page does not say where the pump is made. BPL Medical Technologies of Bengaluru lists an infusion pump category, and Narang Medical of New Delhi calls itself a manufacturer and exporter of volumetric pumps, but neither page names a factory. Poly Medicure's home page lists infusion sets, not pumps.
Multinationals sell alongside them. In August 2025 a trade report described Terumo India's launch of its Terufusion range, a syringe pump, a volumetric pump and monitoring software, and the distribution list of BD's September 2025 set recall shows that Alaris sets reached India. The size of the market is known only from estimates. The market-research firm IMARC puts India's syringe-pump market at 157.5 million US dollars in 2025 and 230.6 million in 2034, growing at 4.19 % a year from 2026. Those three figures do not quite agree, since 157.5 million growing at 4.19 % for nine years gives 227.9 million, and none of them counts pumps. Which licenses these companies hold from CDSCO, India's national device regulator, is not known; Chapter 30 explains how licensing works.
Before buying, list for each candidate pump its authorization and date, the catalog numbers of the sets its maker lists, the versions of its library and server software, and the network settings it requires. Search FDA's recall database for every layer before purchase and again at each update; in 2025 and 2026 each layer had recalls of its own. Ask in which document, and under which conditions, a maker states its accuracy before comparing it with another's. An Indian maker preparing a technical file can define its product the same way: the pump, the sets validated with it, the software versions and the connection.
- Ten 510(k)s under FRN, the code for the general infusion pump, from January 2025 to late September 2026 went to six companies, while Fresenius Kabi received none and appeared in 15 recall records, so clearances show who filed rather than who sells.
- Since 2020, BD's Alaris system has been recalled and, about three and a half years later, cleared under a new 510(k) in 2023; ICU Medical has bought Smiths Medical and Fresenius Kabi has bought Ivenix; and Baxter's Novum IQ has had corrections and a pause in distribution.
- The recalls of 2025 and 2026 reached the set, the library and the network as well as the pump, and FDA's quantity fields count different units.
- India's pump makers are documented mainly by their own websites, multinationals such as Terumo sell alongside them, and the only market size is an estimate whose figures do not quite agree.
Let's Talk