+ Part VII · The system

The pump as a system.

A pump on the shelf of a hospital's biomedical engineering workshop is checked on its own, for flow, occlusion alarm and battery. On a ward it is one part of something larger: it drives a set from its maker's list, carries a library that the hospital's pharmacists wrote, may take orders from a server on the wireless network and is programmed by a nurse, and each of these has its own owner. Brigham and Women's Hospital in Boston replaced about 3,800 large-volume pumps on its main campus in one day, 19 May 2019. Part VII treats the pump as part of a system. Chapter 23 asks who owns each interface between the pump and the rest, Chapter 24 follows the fleet through maintenance, batteries, cleaning and library updates, and Chapter 25 reads the records pumps keep and answers the central question of this guide.

23 — Systems

Who owns the interfaces.

+ The questionIf the pump, the set, the drug library and the health record each work as designed, who makes sure they work together?

A pump evaluated as a system

In its guidance of 2 December 2014, FDA told infusion pump makers that a pump "should be evaluated as a system", and it drew the system wide. Besides the pump, it includes the set for the whole fluid path, to be identified "whether manufactured by you or another firm", and accessories such as a power cord or a wireless controller. It includes the network, meaning any device or system connected to the pump, and the patient, the user and the environment of use. Chapter 22 found failures in the path between pump and health record, and FDA's definition puts that path inside the maker's evidence. That gives half of the answer: before sale, the pump maker answers for the whole system.

After sale the system has a second owner. The guidance asks for testing "with the complete infusion pump system", but a maker's file holds only the configurations it tested. The hospital decides the rest: library limits, server software version, wireless settings, health record and link, the sets it buys and the extensions added at the bedside, each changing on its own schedule. When Mount Sinai Morningside ran its intensive care lines through long extensions in 2020 (Chapters 3 and 12), it changed the system without touching a pump. Making the parts work together is shared work, and the discipline that organizes it is systems engineering.

INCOSE, the international council of systems engineers, defines a system as "a set of elements that work together to produce results that the individual elements couldn't achieve alone." No part of an infusion gives a dose alone: the pump moves, the set turns motion into milliliters, the library and the record supply numbers, and a nurse starts it. Systems engineering, in INCOSE's description, integrates many disciplines across a system's life, from its realization to its retirement, and ISO/IEC/IEEE 15288, revised in 2023, is the international standard for those life-cycle processes.

From hazards to requirements to parts

The work begins with hazards. The hazard analysis behind a pump's 510(k) (Chapter 28) names the ways the system could harm a patient, and each hazardous situation needs at least one control. Each control is written as a requirement, a statement of what the system must do that a test, an inspection or an analysis can check. For example, a requirement might read: the system shall alarm and stop delivery when the line downstream of the pump is blocked, within the time its labeling states for each listed set. How it does so is decided in the architecture, the division of the system into parts and the connections between them.

Each requirement is then allocated, assigned to the parts that will meet it, and the occlusion requirement goes to five parts. A pressure sensor reads the line: hardware. Firmware compares the reading with a limit and raises the alarm: software. The give of the set or syringe decides how long the pressure takes to build (Chapter 12): the set. The hospital decides which limits its units may use: the hospital. A nurse must hear the alarm and act on it (Chapter 15): the user. The labeled time holds only while all five do their part, and two of them lie outside the maker's hands.

Two kinds of evidence close the loop. Verification asks whether each part, and the whole, meets its requirements, such as whether the alarm sounds within the stated time with each listed set at the lowest rate offered. Validation asks whether the requirements were the right ones, by trying the finished system with real users in its intended use, as the human factors validation test of Chapter 18 does. A trace links each hazard to its controls, each control to its requirements, and each requirement to the parts that carry it and the tests that verified it, so that a change to any part shows which hazards it can reach.

60 — From a hazard to the tests that verify it
HAZARD drug not delivered: line blocked downstream CONTROL detect the blockage, alarm and stop REQUIREMENT alarm and stop within the labeled time, for each listed set example: the guide's own ALLOCATION HARDWARE pressure sensor SOFTWARE limit check and alarm SET its give sets the delay HOSPITAL limits its units may use USER hears the alarm, acts above: the maker below: outside the maker's hands VERIFICATION bench test: alarm time with each listed set, at the lowest rate VALIDATION real users, intended use (human factors validation, Chapter 18) A CHANGE FOLLOWS THE TRACE BACK TO ITS HAZARDS schematic
Fluid and drugPressure and forceSoftware and dataHuman input and use errorReadoutFocal detail
Plate 60 — Each hazard needs a control, each control becomes a requirement, and each requirement is allocated to parts and checked; the trace that links them shows which hazards a change can reach, and two of the five parts here lie outside the maker's hands.

Four interfaces, each with two owners

An interface is a boundary across which two parts exchange something: force and fluid, data, or what a screen shows a person. Four decide a dose. Between pump and set the exchange is mechanical, and the set's bore and give turn motion into milliliters (Chapter 8). The pump maker lists the sets it tested. ICU Medical's 2025 510(k) summary for the Plum Solo states compatibility with "currently marketed Plum Administration Sets" without listing them. The hospital buys the sets and adds filters, valves and extensions. BD's Class I recalls of Alaris sets in 2025 and 2026 sat on this interface (Chapters 8 and 26), as did an Ivenix correction for low-flow occlusion alarms when a check valve was added (Chapter 12).

Between pump and drug library the exchange is data that pharmacists write with the maker's software. Baxter's 2025 510(k) summary for the SIGMA Spectrum describes a stand-alone application, on a computer the hospital provides, that builds one library file for every compatible pump. The contents are the hospital's (Chapter 21), and so, largely, is the time a new version takes to reach each pump (Chapter 24). Between pump, network and health record the exchange is orders and status. The maker supplies the pump's side and often the server; the hospital runs the rest. Two recalls described in Chapter 22 concerned this interface: BD's Alaris Systems Manager, which could send an outdated programming request, and Alaris labeling that did not say which TLS version to enable.

Between pump and user the exchange is what the screen shows and what the keys or touchscreen take in. The maker's evidence is the human factors file of Chapter 18; the hospital owns the training, staffing and policies that shape how the screen is used. Two Ivenix corrections concerned this interface: the rate with two leading zeros that froze the screen, and the loosened touchscreen cable that registered touches nobody made (Chapter 17). None of the four interfaces has a single owner, and many of the recalls of 2025 and 2026 sat on an interface rather than in the pumping mechanism alone.

61 — The pump system and the owners of its interfaces
THE SYSTEM, AS FDA'S 2014 GUIDANCE DRAWS IT PUMP SET MAKER lists the sets it tested HOSPITAL buys sets; adds filters, valves, extensions PATIENT Alaris sets, Class I, 2025 and 2026; Ivenix check-valve alarms (Ch 8, 12, 31) EVERY INTERFACE HAS AN OWNER ON EACH SIDE MAKER library software HOSPITAL limits; release; reach to each pump DRUG LIBRARY library tool on a hospital computer (Ch 21, 24) MAKER pump side, often the server HOSPITAL network, record, link NETWORK · HEALTH RECORD Alaris Systems Manager, Class I, 2025; TLS labeling, Class II, 2026 (Ch 22) MAKER screen, keys, human factors file HOSPITAL training, staffing, policy USER Ivenix two leading zeros; phantom touches (Ch 17, 31) ENVIRONMENT OF USE FDA 2014 guidance, sec. 3 (scope) · recalls from FDA records, 2025 to 2026 · schematic
Fluid and drugSoftware and dataHuman input and use errorFocal detail
Plate 61 — FDA asks that a pump be evaluated as a system with its set, accessories, network, user and environment; each of the four interfaces that decide a dose has the maker on one side and the hospital on the other, and recalls of 2025 and 2026 sat on each.

An interface written as a specification

FDA's guidance of 6 September 2017 on interoperable devices treats each electronic interface as a specification for its maker to publish. Makers should "clearly set forth in device labeling the functional and performance requirements of their electronic interface": its purpose, the devices and software versions it is meant to connect with, what it needs from them, its timing, how it tolerates faults and fails safe, and its known limits. The guidance asks for risk analysis of data that arrive late, corrupted, misformatted or out of step, and for tests under abnormal conditions. Its examples are infusion pumps.

Such a specification also allocates work, and part of it falls to the nurse. In its 510(k) summary for its 2023 Alaris clearance, BD states that if communication is interrupted the pump "will continue to function as programmed". IHE International, where health professionals and industry agree how standards are used together, publishes profiles for particular exchanges of data. Its profile for sending infusion orders to pumps states that "The clinician must always verify all settings and start the infusion directly on the pump." Both leave the last check to the nurse at the pump.

A recognized standard names the two owners outright. ANSI/AAMI/UL 2800-1 is a family of interoperability standards, and FDA recognized its 2022 edition on 19 December 2022. It has one part for the life cycle of developing an interoperable product (2800-1-2) and another for integrating such products into a working system (2800-1-3). The first is the maker's work. The second belongs to whoever assembles the system, often the hospital or the integrator it hires. None of four recent pump 510(k) summaries cites the family.

Testing the versions a fleet runs at once

A maker can test its pump with its own sets, server and library tool, in the versions it ships, but not the limits a hospital writes, the extensions a ward adds or the mix of versions a fleet runs. Integration testing, the testing of parts together across their interfaces, therefore happens twice: by the maker against the configurations it labels, and by the hospital against the configuration it runs. The second must cover every combination of versions in use, and since a library update spreads unevenly (Chapter 24), the count grows by multiplication.

Solved example: how many configurations one fleet runs during a software correction
  • Library versions. One 11-hospital system released a library about every 60 days, according to a 2024 report whose authors include recipients of ICU Medical research grants. In the 2018 study of Chapter 24, a replaced version kept raising alerts for a median of 22 to 192 days, depending on the health system, and for at most 661 days. In the system with the slowest median, 192 ÷ 60 = 3.2, so about the three latest replaced versions run beside the current one, which makes 4 library versions in use. On the longest delay, 661 ÷ 60 = 11.0, so 11 replaced versions run beside the current one: 11 + 1 = 12.
  • Pump and server software in the Ivenix correction of November 2025: pumps on 5.10.1 or 5.10.2, servers before or after 5.2.2, which went first. Of 2 × 2 = 4 pairs, new pump software on the old server is ruled out by that order, so 3 remain.
  • Across the correction: 4 × 3 = 12 configurations, or 12 × 3 = 36 on the long tail. Once the server is updated, both pump versions run on it: 4 × 2 = 8 to 12 × 2 = 24 at the same moment.

A fleet built from these figures runs 8 to 24 combinations of library, pump and server software at once, and 12 to 36 across one correction, before set catalog numbers multiply them. The three sources do not report these combinations: they come from joining their figures and reading the installation order as ruling out one pair.

62 — Configurations one fleet runs during a correction
a library about every 60 days; old versions alert up to a median 192 days (4), at most 661 (12) AFTER THE SERVER UPDATE, AT ONCE 5.10.1 SERVER BEFORE 5.2.2 5.10.1 SERVER 5.2.2 5.10.2 SERVER 5.2.2 5.10.2 SERVER BEFORE 5.2.2 current replaced 1 replaced 2 replaced 3 replaced 4 replaced 11 ruled out: server first 4 × 3 = 12 12 × 3 = 36 UP TO 36 CONFIGURATIONS IN ONE CORRECTION 4 × 2 = 8 TO 12 × 2 = 24 AT ONCE cadence: 2024 report (authors received ICU Medical grants) · delays: DeLaurentis et al. 2018 · pairs: Ivenix correction, Nov 2025 · combination and ruled-out pair are the guide's own
Software and dataReadoutFocal detail
Plate 62 — Four to twelve library versions, multiplied by the three pump and server pairs a correction passes through, give 12 to 36 configurations, of which 8 to 24 run at once after the server update.

Every change restarts the count. A change impact analysis asks, before release, which requirements, interfaces and parts a change can reach, and the trace makes the question answerable. Chapter 20 follows the Ivenix pump through five software corrections, and Chapter 28 sets out when FDA treats a change as needing a new 510(k). A hospital faces the same question with every library release, server patch and new set catalog number.

Give each interface an owner, a specification and a test

For a pump or set maker, an Indian maker in particular, record each requirement with the parts it is allocated to, set, hospital and user included, and the test that verifies it. Label each electronic interface as FDA's 2017 guidance asks, with the sets, software versions and systems it was tested with, what it needs from them and what the pump does when the link fails. For the hospital team that owns the integration, name an owner for each interface, keep a register of the library, pump, server and set combinations in use, and retest those combinations after each change.

+ What this chapter established
  • FDA's 2014 guidance asks that a pump be evaluated as a system with its set, accessories, network, user and environment, so the maker answers for the whole before sale and the hospital owns its configuration after.
  • Systems engineering derives requirements from hazards, allocates each to hardware, software, set, hospital or user, verifies the parts against them and validates the whole with its users.
  • Each interface, from pump to set, library, network and user, has an owner on each side; FDA's 2017 guidance asks makers to label each electronic interface, and the UL 2800-1 family separates the maker's work from the integrator's.
  • Joining one system's library cadence, a published update delay and the Ivenix correction gives 8 to 24 configurations running at once and 12 to 36 across the correction.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.