09 — Trumpet curve

Flow, minute by minute.

+ The questionHow is flow measured on a balance, and what does a trumpet curve show that an average hides?

Weighing the flow

The instrument that settles an accuracy claim is a laboratory balance. The end of the pump's set is led into a beaker on the pan, the balance sits under a draft shield, and a computer records its reading at fixed intervals while the pump runs. Because a milliliter of water weighs about a gram, each reading is a volume, and the change between two readings is the volume delivered in between. This is the gravimetric method: flow measured by weight against a clock. It answers the question Chapter 8 left open, because once the readings are logged, the error can be computed over any period an analyst chooses, an hour or two minutes.

The method looks simple and is exacting at low rates. The sampling interval, the time between readings, is 30 seconds in IEC 60601-2-24, according to a 2024 article by staff of Sartorius, a balance maker, and B. Braun, a pump maker. At 25 mL/h each reading adds about 208 mg, but at 1 mL/h only 8.3 mg. Water evaporating from the beaker would pass for missing flow, so a 2013 metrology study that weighed gravity drip sets to 1 mg every 30 seconds covered the water with a film of oil, which kept the loss below 1 mg over five hours.

The 2024 article adds that the depth of the outlet needle in the liquid changed the measured mass by up to 0.264 g, and that two rigs measuring one 1 mL/h flow gave 0.99 ± 0.02 and 0.96 ± 0.32 mL/h. A test rig can add more error than the pump it tests claims. The balance must also resolve masses this small, and the same article tabulates the readability that AAMI's 2021 report on fluid delivery testing recommends: a balance reading to 0.1 mg for flows of 10 to 1,000 mL/h, and to 0.01 mg for 0.1 to 10 mL/h. At 1 mL/h, one step of a 0.1 mg balance is about 1.2 % of a 30-second reading, while one step of a 0.01 mg balance is about 0.12 %.

23 — Weighing the flow
PUMP UNDER TEST draft shield oil film against evaporation BALANCE xx.xxxx g LOGGER a reading every 30 s (IEC 60601-2-24, per secondary sources) or 10 s (AAMI TIR101) 25 mL/h: about 208 mg per 30 s reading 1 mL/h: 8.3 mg needle depth changed the mass by up to 0.264 g Figures: Foerster and Peissig 2024 (Sartorius and B. Braun staff); Leetang et al. 2013. Schematic.
Fluid and drugReadoutFocal detail
Plate 23 — The gravimetric method turns weight into volume at fixed intervals, and at 1 mL/h each 30-second reading is only 8.3 mg, so evaporation and the needle's depth matter as much as the balance.

The test the standard describes

The method most often cited is the one in IEC 60601-2-24, the international standard for infusion pumps. Its second edition, of 2012, is current as of October 2026, and a third edition is in draft (Chapter 27). The standard's text is sold rather than published, so the test details in this chapter come from secondary accounts, marked as such: a certifier's slides on the 2012 revision, a published evaluation that applied the method, and articles by makers of balances and test equipment.

On those accounts, a volumetric pump is tested at an intermediate rate of 25 mL/h and a syringe pump at 5 mL/h. Further tests run at the lowest rate the user can select, but not below 1 mL/h, and in the 2012 edition at the highest selectable rate too. A white paper from a maker of test equipment describes a test period of 120 minutes at the intermediate rate. The readings of the first hour are plotted as a start-up curve, flow against time from the moment the pump starts, which shows how long the pump takes to reach its set rate. The readings of the second hour, when the pump has settled, are the ones analyzed for accuracy.

From readings to a trumpet

The analysis asks a question that a single average cannot: how far off can the flow be over a window of a given length? The analysis takes an observation window, for example 2 minutes long. Starting at the first reading of the second hour, the analysis computes the average rate over the next 2 minutes and its percentage difference from the set rate, then slides the window on by one reading, 30 seconds, and repeats to the end of the hour. With 120 readings in the hour, a 2-minute window fits 117 times, and only the largest and the smallest of those 117 errors are kept. The same is done for windows of 5, 11, 19 and 31 minutes, the lengths the secondary sources give, and a 31-minute window fits 59 times.

24 — A window slides through the second hour
60 75 90 105 120 TIME FROM START (min) 2 min window, moved 30 s at a time: 117 positions each position gives one average error; keep the largest and the smallest 31 min: 59 positions other windows: 5 min: 111 11 min: 99 19 min: 83 Window lengths and 30 s interval from secondary accounts of IEC 60601-2-24. Schematic.
ReadoutFocal detail
Plate 24 — Each window length is slid through the second hour one reading at a time, so a 2-minute window yields 117 average errors and a 31-minute window 59, of which only the largest and smallest are kept.

Plotted against window length, the largest error above zero and the smallest below it form two lines that flare apart at short windows and close in at long ones, like the bell and the pipe of a trumpet: hence the trumpet curve. The overall error of the whole hour sits at its narrow end. The review by two TGA staff in Chapter 8 printed one such curve for an unnamed pump. Over the hour the error was −2 %, but over 2-minute windows it reached +7 % and −10 %. A spread of 17 points at 2 minutes sat behind a figure of 2 % for the hour.

Solved example: the largest and smallest errors over 2, 5 and 11 minute windows
  • For illustration, a pump set to 0.5 mL/h delivers nothing for 90 seconds, the longest pause Baxter declares for its pump at that rate in a 2025 510(k) summary, and then delivers the volume owed at once. Each pulse is 0.5 × 90 ÷ 3,600 = 0.0125 mL, or 12.5 µL, and with readings every 30 seconds, every third reading shows one pulse.
  • A 2-minute window spans four readings and should hold 0.5 × 2 ÷ 60 = 16.7 µL. Depending on where it starts, it holds one pulse (12.5 µL, an error of −25 %) or two (25 µL, +50 %).
  • A 5-minute window spans ten readings and should hold 41.7 µL. It holds three pulses (37.5 µL, −10 %) or four (50 µL, +20 %).
  • An 11-minute window spans 22 readings and should hold 91.7 µL. It holds seven pulses (87.5 µL, −4.5 %) or eight (100 µL, +9.1 %).
  • Over the full hour, 40 pulses deliver exactly 0.5 mL, an error of zero.
  • In words, a pump that is perfect over the hour can be 25 % short or 50 % over in a 2-minute window, and the trumpet records both.
25 — A trumpet from a pump that is exact over the hour
−30 −20 −10 0 +10 +20 +30 +40 +50 2 5 11 19 31 ERROR (%) OBSERVATION WINDOW (min) published example (review by TGA staff): +7 and −10 % at 2 min, −2 % over the hour +50 % −25 % whole hour: 0 % +50 % in the worst 2-minute window Computed from the guide's illustrative pattern: 12.5 µL every 90 s at 0.5 mL/h (pause length from Baxter 510(k) K251640, maker's figure).
ReadoutFocal detail
Plate 25 — A pump that delivers 12.5 µL every 90 seconds is exact over the hour, yet its 2-minute windows run from 25 % short to 50 % over; the trumpet shows what the hourly figure hides.

The curve answers a clinical question that the average cannot. For a drug whose effect fades within two minutes, the 2-minute end of the curve matters far more than the 31-minute end; for a feed or a bag of fluid, the reverse. A pump whose curve stays narrow at short windows delivers smoothly, while one whose curve flares delivers in pulses that the long average conceals. Two pumps with the same hourly error can have very different trumpets, and only the trumpet tells them apart.

A trumpet curve begins after the pump has settled

Because the curve is computed from the second hour, it leaves out the start, when a syringe pump may take many minutes to reach its set rate (Chapter 5). The start-up curve covers that period, and a pump chosen for short-acting drugs should be judged on both plots, not on the trumpet alone.

Two methods, two markets

In the United States FDA recognizes AAMI TIR101 rather than IEC 60601-2-24 (Chapter 28). TIR101 is a technical information report on fluid delivery performance testing for infusion pumps. It was issued in 2021, entered in FDA's database on 30 May 2022 and reaffirmed in 2026. It covers syringe, container and volumetric pumps across delivery modes, including enteral, patient-controlled and epidural use. According to the 2024 article by balance and pump makers, it reads the balance every 10 seconds rather than every 30. It sets no criteria for clinical acceptability and does not address measurement uncertainty.

At least one maker now cites it in a claim. ICU Medical's 510(k) summary for the Plum Solo, cleared on 2 April 2025, states ±3 % from 0.1 to 999 mL/h "under standard conditions" per TIR101 and ±5 % under non-standard conditions. The pair makes Chapter 8's point again: one pump, one test report, two accuracies, each true under its own conditions. Faster readings show more detail at low rates, but each reading then holds less, 2.8 mg at 1 mL/h, and the balance's own limits loom larger. A 2015 metrology study of syringe pumps met the same limit, confirming a claimed 2 % at 2 and 10 mL/h but not at 0.5 mL/h, where the uncertainty of the measurement itself reached 2 %.

Validate the rig before trusting the curve

For a maker preparing a submission, or a hospital laboratory checking pumps after repair, measure the rig first. Choose a balance whose readability suits the lowest rate tested, cover the water against evaporation, fix the depth of the outlet, and run a known flow at 1 mL/h to see what the rig itself adds. Record both the start-up and the trumpet curves, at the 30-second interval of IEC 60601-2-24 and the 10-second interval of TIR101 when the pump is meant for both markets, and state which method each claim rests on.

+ What this chapter established
  • The gravimetric method weighs the fluid a pump delivers at fixed intervals, and at low rates the balance, evaporation and the outlet's position can add more error than the pump claims.
  • Secondary accounts of IEC 60601-2-24 describe tests at 25 mL/h for volumetric and 5 mL/h for syringe pumps, readings every 30 seconds, a start-up curve from the first hour and accuracy analyzed in the second.
  • A trumpet curve plots the largest and smallest errors over windows of 2 to 31 minutes, and shows the pauses and pulses that an hourly average hides.
  • FDA recognizes AAMI TIR101 rather than IEC 60601-2-24, and at least one maker, ICU Medical for the Plum Solo in 2025, states its accuracy claim against TIR101 conditions.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.