§ Glossary

Glossary.

Terms are defined as they are used in this guide.

510(k)
A premarket notification, named for the section of the Federal Food, Drug, and Cosmetic Act that requires it, comparing a new or changed device with a legally sold predicate; FDA clears it when it finds the two substantially equivalent, and does not approve it (Chapters 4, 26, 28).
AAMI TIR101
AAMI's 2021 technical information report on fluid delivery performance testing for infusion pumps, reaffirmed in 2026 and recognized by FDA; it sets no clinical acceptance criteria.
adverse drug event
Harm caused by a drug. A 2005 trial of smart pumps counted 25 preventable adverse drug events, and 155 potential ones that nobody intercepted, among 219 medication errors in 744 admissions.
air embolism
Air in the circulation in an amount that can harm. For veins, a 2016 review gives 200 to 300 mL, or 3 to 5 mL/kg, as the lethal dose for adults, from adult cases and animal models; air reaching the arteries can harm in small volumes.
air-in-line detector
A sensor, ultrasonic or optical, clamped around the tubing that tells liquid from gas. A bubble appears as a gap in its signal, which the pump turns into a bubble length and, through the bore of the set, a volume (Chapter 13).
alarm
A signal from the pump that a condition needs attention, from a blocked line to an infusion about to finish. IEC 60601-1-8 sorts alarm conditions into priorities by degree of urgency.
alarm fatigue
The name hospitals give to staff learning to treat alarms as background when most of them need nothing urgent.
alert
A prompt on the pump's screen that asks the user to confirm or change an entry, such as a drug library's warning that a dose exceeds a soft limit.
alternate equipment management
A program, allowed to US hospitals by a 2013 CMS memorandum, that replaces makers' maintenance schedules with the hospital's own, provided it is documented, based on risk and followed strictly (Chapter 24).
Ambulatory pump
A pump built to be carried so that an infusion can continue at home or at work; it tracks the reservoir by subtracting its own count from an entered reservoir volume (Chapter 6).
anti-reflux valve
A one-way valve fitted to each line above a junction to prevent backflow; one model opens at 5.3 mmHg.
anti-siphon valve
A valve that opens only above about 100 to 150 mmHg, stopping a syringe from emptying by gravity, at the cost of a higher pressure in the line.
Architecture
The division of a system into parts and the connections between them, which decides how each requirement will be met.
auto-documentation
The pump sending its rate, the volume infused and its state back to the electronic health record, which writes them into the chart.
auto-programming
Sending an order's drug, concentration, dose and rate from the electronic health record, through an integration server, to the pump channel the nurse has scanned, for the nurse to confirm. At one 286-bed hospital it cut programming from a mean of 15 keystrokes to 2.
averaging period
The interval over which a percentage error is computed. An error of ±5 % over an hour at 25 mL/h allows up to ±30 % in a single ten-minute window.
back pressure
The pressure a pump must push against at the far end of the line, from catheter and vein; accuracy claims are often stated at 0 mmHg.
backflow
Fluid driven from a shared junction up a stopped or slower line instead of on to the vein, storing drug that later returns as a bolus.
Basal rate
A continuous infusion running beneath PCA demand doses. It keeps running whether or not the patient is awake (Chapter 6).
biomedical engineer
A hospital engineer, also called a clinical engineer, who maintains medical equipment, including the pump fleet, its checks and its repairs (Chapter 24).
callback
A reminder alarm that a pump has been left unattended, not a fault; 28.8 % of critical care alarms in one maker's pump logs.
care area
A group of drug library entries for one kind of unit, such as an adult intensive care unit or a neonatal unit, so that limits suit the patients there. The nurse selects it on the pump before the drug.
carrier
A plain fluid infused continuously into the same line as a drug to push it to the vein sooner. It shortens the transit time, adds fluid and ties the drug's delivery to the carrier's rate; Mount Sinai ran carriers at 50 mL/h and later at no less than 10 mL/h.
Cassette
A rigid molded piece in the set of a cassette pump, containing a pumping chamber, valves formed by a diaphragm and, in some designs, an air trap and a secondary inlet. The pump's plunger presses the diaphragm and the pump counts plunger steps (Chapter 4).
catheter
The short plastic tube that stays in the vein and carries the infusion into it.
CDSCO
The Central Drugs Standard Control Organisation, India's national regulator for drugs and devices and the seat of the Central Licensing Authority.
CE mark
The mark a maker affixes after declaring that a device conforms to the MDR, backed for all but the lowest-risk devices by a notified body's certificate; it allows sale in every EU member state.
Central Licensing Authority
The authority at CDSCO that licenses the manufacture of Class C and D devices (application on Form MD-7, license on MD-9, factory inspection within 60 days) and the import of devices of any class (MD-14, MD-15).
Change impact analysis
The review, before a change is released, of which requirements, interfaces and parts the change can reach.
Class I recall
FDA's most serious recall class: a reasonable probability that use of the product will cause serious adverse health consequences or death. Class II covers temporary or medically reversible harm; Class III, products unlikely to cause harm. The class is FDA's.
Class II (FDA)
The middle of FDA's three device risk classes, in which 21 CFR 880.5725 places infusion pumps; every pump product code under it needs a 510(k) (Chapter 28).
Clinical evaluation consultation procedure
The MDR procedure (Article 54) in which a notified body consults an expert panel on its clinical evaluation assessment report before certifying a class III implant or a class IIb active device that administers a medicinal product, such as a new infusion pump. The 2025 proposal would remove pumps from it.
cold flow
The slow deformation of a plastic under a steady load. Under a roller clamp, PVC tubing keeps creeping after the clamp is set, so the flow falls; a 2012 patent puts the change at 10 to 20 percent, half of it within about 10 minutes.
collateral standard
A standard numbered IEC 60601-1-x that adds requirements on one topic, such as alarms (60601-1-8) or electromagnetic compatibility (60601-1-2), for all medical electrical devices (Chapter 27).
committee draft
The first text of a new standard circulated to national bodies for comment, followed by a draft for vote and a final draft before publication; the third edition of IEC 60601-2-24 closed its committee-draft comments on 21 August 2026 (Chapter 27).
Compliance
The volume a syringe, line or filter takes up for each unit of pressure; one 2015 study measured 0.21 mL per bar for a 10 mL syringe and 1.54 mL per bar for a 50 mL syringe. It delays start-up and stores volume that is released as a surge when pressure falls (Chapter 5).
Configuration
One combination of library version, pump software, server software and set in use together. During one software correction a fleet can run 8 to 24 at once, on the guide's arithmetic.
continuous quality improvement (CQI) report
A report built from a drug library's records: alerts, overrides, corrected entries, the drugs that alert most, the share of infusions programmed through the library and the share of pumps on the current version (Chapter 25).
coordinated vulnerability disclosure
The practice by which a security flaw is reported to the maker, confirmed and fixed before it is made public. Flaws in Baxter's SIGMA Spectrum pumps were reported on 20 April 2022 and published on 8 September 2022, 141 days later.
Correction
A recall action that leaves the product where it is: repair, modification, adjustment, relabeling, destruction or inspection without moving it.
Critical task
In FDA's guidance, a user task that, if performed incorrectly or not performed at all, would or could cause serious harm (Chapter 18).
cyber device
Under section 524B of the US Federal Food, Drug, and Cosmetic Act, a device that includes software validated by its maker, can connect to the internet and has characteristics that could be vulnerable to cybersecurity threats; a premarket submission for one made on or after 29 March 2023 must include a plan for vulnerabilities after sale, processes to patch them and a software bill of materials (Chapters 22 and 28).
dead volume
The fluid in a line between the point where a drug enters it and the tip of the catheter. The drug reaches the vein only after this fluid has been pushed out.
debouncing
Accepting a change in a key's state only after the contact has held that state for a set period, typically 20 to 50 ms, sampled every 1 to 5 ms. The period must outlast any bounce and stay shorter than the gap between deliberate presses.
Demand dose
The fixed dose a PCA pump delivers for each honored press of the handset, for example 1 mg of morphine (Chapter 6).
Design and development file
Under clause 7.3.10 of ISO 13485:2016, and so under FDA's Quality Management System Regulation, the file for each device type or family that contains or references the records of its design.
Design controls
The regulated process by which a maker designs a device: planning, inputs, outputs, review, verification, validation, transfer to production and control of changes. In the US under 21 CFR 820.30 until 1 February 2026, and under clause 7.3 of ISO 13485 since 2 February 2026.
dose error reduction system (DERS)
The software in a smart pump that checks each entry against the drug library's concentrations and limits for the selected drug and care area.
Dose rate
The amount of drug a patient is to receive per unit of time, often scaled to body weight, such as 0.1 mcg/kg/min of norepinephrine; the pump never sees it, only the rate in mL/h derived from it (Chapter 16).
double pumping
A syringe changeover in which the new syringe is started on a second pump before the first is stopped; in a 2020 trial it was followed by more swings in blood pressure than a quick change.
downstream
In an infusion line, the stretch between the pump and the patient. A downstream block raises the pressure the pump pushes against.
drip chamber
The short clear cylinder at the top of an administration set in which liquid falls as separate drops from a small tip, so that the drops can be counted.
drop factor
The number of drops an administration set is built to make per milliliter: 20 for a standard set (0.05 mL a drop) and 60 for sets with a finer tip (about 0.017 mL). A nominal figure: a 2013 laboratory study measured drops of water 1.1 to 13.2 percent larger.
Elastomeric pump
A pump with no motor or sensor in which a stretched rubber balloon pushes the drug through a narrow flow restrictor; its rate is the balloon's pressure divided by the restrictor's resistance (Chapter 7).
electronic health record (EHR)
The hospital's electronic chart of a patient, including the orders for that patient; also called the electronic medical record.
ENFit
Trade name under which many enteral connectors made to ISO 80369-3 are sold (Chapter 19).
Enteral feeding
Delivery of liquid formula into the stomach or small intestine through a tube (Chapter 19).
Enteral feeding pump
A pump that moves feeding formula at a set rate; in the US it is regulated as an infusion pump under product code LZH, Class II (Chapter 19).
EUDAMED
The EU database of devices, makers, certificates and, in time, incidents. Its first four modules have been mandatory since 28 May 2026; its vigilance module is still in development as of October 2026.
event log
A pump's time-stamped record of what happened to it: values entered, starts, stops, alarms and alerts, and the volume it counted. Also called a history log; its maker decides its contents (Chapter 25).
extension set
A length of tubing added between the administration set and the catheter.
extravasation
Fluid leaking into the tissue around a catheter that has slipped out of the vein. It does not trigger an occlusion alarm, because the tissue offers too little resistance for the pressure to reach the limit (MHRA).
fail-stop
The state a pump enters when its software can no longer trust its own condition: it halts and alarms rather than carry on. A fault in Ivenix software 5.10.1 could freeze the screen in this state when a rate with two leading zeros exceeded a library limit (corrected from November 2025).
Field safety corrective action
In the EU, a maker's action to prevent or reduce the risk of a serious incident with a device already in use.
Field safety notice
The letter that tells users about a field safety corrective action; its draft goes first to the evaluating national authority, which has 48 hours except in urgent cases.
firmware
The software stored in a pump and written for its own processor, changed only by a deliberate update. It reads the keys, converts a dose into a rate, checks the entry against the drug library, schedules the motor steps, watches the sensors, raises alarms, estimates the battery and keeps the log; in most pumps the volume shown on the screen is its count of motor steps.
fleet
Every pump of a kind that a hospital owns, managed as one population; one British Columbia service reported over 18,000 pumps, an estimated 4 percent or more of them misplaced (Chapter 24).
Flow restrictor
The narrowest part of an elastomeric pump's fluid path, chosen to give the labeled rate. Its resistance follows the liquid's viscosity, so the rate changes with temperature (Chapter 7).
fluid ingress
Liquid getting inside a pump's case, for example through a lifted keypad during cleaning; it led to Class I recalls of Alaris keypads in 2020 and 2021 (Chapter 24).
Forcing function
A design that makes a wrong action impossible rather than unlikely, such as a connector that cannot join the wrong line (Chapter 19).
Formative evaluation
A smaller study with users during development, run to find where people stumble while the design can still change (Chapter 18).
gauge
The size number of a catheter, which runs backward: the higher the gauge, the narrower the bore. BD gives its 20 G catheter a bore of 0.76 to 0.84 mm and its 16 G 1.32 to 1.40 mm.
general standard
IEC 60601-1, the standard for the basic safety and essential performance of every medical electrical device; in October 2026 the 2005 edition as amended in 2012 and 2020 (Chapter 27).
Generic Infusion Pump project
Work by FDA researchers and engineers at the University of Pennsylvania, begun before 2010, that set out the hazards and safety requirements of a generic pump as an open reference. Its patient-controlled analgesia model listed 97 safety requirements, about 20 of which could be verified formally against the model.
good catch
An entry that crossed a drug library limit and was then corrected; the count of errors a library report can show it caught.
gravimetric method
Measuring flow by weighing the fluid a pump delivers on a balance at fixed intervals; a milliliter of water weighs about a gram.
gravity infusion
An infusion driven only by the head of liquid above the vein, with its rate set by a roller clamp and checked by counting drops.
half-life
The time the body takes to clear half of the drug present; about 2.4 minutes for norepinephrine, so a gap of a few minutes lowers the blood level markedly.
hard limit
A drug library limit marking a dose that should never be given in that care area. The pump refuses an entry beyond it, and the nurse must reprogram.
Harmonized standard
A European standard whose reference the Commission has published in the Official Journal; a device that meets it is presumed to meet the requirements it covers. As of October 2026, EN 60601-1 and EN ISO 14971 are cited under the MDR and EN 60601-2-24 is not.
head
The height of the liquid surface in the bag above the vein, which sets the pressure driving a gravity infusion: 1 mmHg for every 1.36 cm of water at 20 °C, so 80 cm gives 58.8 mmHg.
head height
The height of the fluid container above the pump, which sets the pressure at the pump's inlet; one maker's claim assumes 30.5 to 61 cm.
High-alert medication
In ISMP's list, a drug that bears a heightened risk of causing significant patient harm when used in error; intravenous adrenergic agonists such as norepinephrine are on it (Chapter 16).
Human factors validation test
Also called a summative test: representative users perform the critical tasks on the final design, with every use error recorded and traced to its cause; FDA asks for at least 15 participants per distinct user group (Chapter 18).
Incremental entry
Entering a number by stepping a displayed value up or down with arrow keys or chevrons; in a 2015 laboratory study, 1.41 % slips per number and 25.45 s per number (Chapter 17).
Integration testing
Testing parts together across their interfaces: by the maker against the configurations it labels, and by the hospital against the configuration it runs.
Interface
A boundary across which two parts of a system exchange something: force and fluid, data, or what a screen shows a person. Four decide a dose: pump and set, pump and drug library, pump and network or health record, and pump and user.
interoperability
The exchange of data between a pump and the hospital's information systems, so that orders reach the pump as data and the pump's status returns to the chart.
IS 13450 (Part 2/Sec 24):2019
The Indian standard for infusion pumps and controllers, identical to IEC 60601-2-24:2012 under dual numbering; India's device rules name a BIS standard first where one exists.
Keep vein open (KVO)
A slow flow that many pumps drop to after the volume to be infused is delivered, to stop the catheter clotting until someone attends (Chapter 16).
Large-volume pump
Also called a volumetric pump: a pump that draws fluid from a bag through an administration set and delivers it at a programmed rate by displacing a known volume with each movement of its mechanism. It counts that movement and does not measure the fluid (Chapter 4).
Lead screw
A threaded rod whose nut and carriage advance a fixed distance for each turn; a screw with four threads 2 mm apart advances 8 mm a turn (Chapter 5).
Legacy device
A device still placed on the EU market under a certificate issued under the directives that preceded the MDR; for a class IIb pump that is not implanted, possible until 31 December 2028 if the 2023 conditions were met and the design has not changed significantly.
library compliance
The share of infusions programmed through the drug library rather than in plain mL/h or without it; unrelated to the compliance of tubing. A Berlin cardiothoracic ICU measured 92.8 percent of 133,601 syringe-pump starts in 2012; the AAMI Foundation's 2017 target is at least 95 percent of primary and secondary infusions.
Lockout interval
The period after a PCA dose during which further presses are refused, for example 6 to 10 minutes for morphine in one order set (Chapter 6).
Low-dose tip
An enteral syringe design that, used with a tap to clear the recess between tip and collar, delivered small doses close to a range of plus or minus 10 % in industry testing (Chapter 19).
Luer connector
The small tapered fitting on intravenous lines and syringes, now specified in ISO 80369-7; it was also used, or matched closely enough to mate, on feeding, epidural and cuff lines (Chapter 19).
Malformed number
A keystroke sequence that is not a valid number for its field, such as two decimal points, digits beyond the field's precision or superfluous leading zeros (Chapter 17).
Materiovigilance
The collection and study of adverse events with medical devices once they are in use, as pharmacovigilance does for drugs.
MAUDE
FDA's database of medical device reports, updated monthly and searchable online for the last ten years; FDA states it is not meant for judging rates of adverse events, because reports have no denominator.
MDCG
The Medical Device Coordination Group, where the EU member states' authorities meet; its guidance MDCG 2021-24 Rev.1 (April 2026) gives the classification examples for pumps.
MDR (EU Medical Device Regulation)
Regulation (EU) 2017/745, under which a pump is classified, assessed by a notified body and CE marked in the EU. Not to be confused with US medical device reports.
mean arterial pressure
The average pressure in the arteries over a heartbeat; the measure used to count swings in blood pressure at syringe changeovers.
mechanical accuracy
How closely a pump's mechanism moves as commanded, without the set: one maker gives better than 0.5 % for a large-volume pump whose system accuracy is ±5 %.
Medical device report
In the US, a report of a death, serious injury or malfunction sent to FDA by a user facility, maker or importer, or voluntarily; reports go into MAUDE. Not the EU MDR.
Medical Devices Rules, 2017
India's device regulations, published on 31 January 2017 and administered by CDSCO, with four risk classes from A (lowest) to D (highest).
microbore
Narrow-bore tubing. BD classes extension tubing with a bore under 2 mm as microbore; its 213 cm microbore extension holds 0.7 mL, a bore of about 0.65 mm.
Microstep
A fraction of a motor step, commonly 1/32 or 1/64, produced by the motor's driver. Microstepping turns a delivery of separate pulses into a near-continuous flow at low rates (Chapter 5).
Misconnection
A line joined to a port meant for another, such as an enteral feed connected to an intravenous catheter (Chapter 19).
MvPI
India's materiovigilance program, approved on 6 July 2015 and coordinated by the Indian Pharmacopoeia Commission; 15 days for serious adverse events and 30 days for non-serious ones.
Notified body
An independent organization, designated by a national authority, that audits a maker's quality management system and assesses its technical documentation before issuing the certificate behind a CE mark. In the EU it, not a regulator, assesses a pump.
nuisance alarm
An alarm that sounds when nothing needs to be done, such as an air alarm caused by tubing not fully seated in the detector.
observation window
A period of fixed length slid through the test readings one reading at a time, over which the average error is computed; 2, 5, 11, 19 and 31 minutes in secondary accounts of IEC 60601-2-24.
occlusion alarm
The alarm a pump raises when the pressure it reads, as the force on a syringe plunger or the pressure in the tube beyond the mechanism, crosses a set limit. The pump infers a blockage from pressure; it never measures the absence of flow (Chapter 12).
occlusion limit
The pressure at which a pump stops and alarms on the assumption that the line is blocked. BD states 50 to 525 mmHg for its Alaris Pump module; it must sit above the pressure the line itself needs.
Out by ten
Wrong by a factor of ten or more between the number intended and the number the device processed, the measure used in Thimbleby and Cairns's 2010 simulation (Chapter 17).
override
Proceeding with an entry after a soft-limit alert. A 2019 multi-hospital analysis found 73.8 percent of alerts overridden for ordinary drugs and 75.8 percent for high-alert drugs.
override rate
The share of drug-library alerts that a user overrides to run the entry as programmed; a 2019 analysis found means of 75.8 percent for high-alert drugs and 73.8 percent for others (Chapter 25).
override-to-reprogram ratio
The number of alerts overridden for each one corrected; 9.92 for high-alert drugs and 7.30 for others in a 2019 multi-hospital analysis, and 49 at an override rate of 98 percent (Chapter 25).
particular standard
A standard numbered IEC 60601-2-x that amends and adds to the general standard for one type of device; for infusion pumps and controllers, IEC 60601-2-24:2012 (Chapter 27).
Patient-controlled analgesia (PCA)
Pain relief, usually an opioid, that a patient gives by pressing a handset; the pump delivers a fixed demand dose per press within a lockout interval and a cumulative limit (Chapter 6).
PCA by proxy
Activation of a PCA pump for the patient by a family member, caregiver or clinician without authorization, the subject of the Joint Commission's Sentinel Event Alert 33 of December 2004 (Chapter 6).
Peristaltic pump
A pump that pinches a length of flexible tubing and moves the closed point along it, pushing fluid forward: with rollers on a wheel (rotary) or with a row of fingers pressing the tube against the door (linear). The volume per wave depends on the tube's bore and how fully it re-opens (Chapter 4).
plug flow
The approximation that liquid moves along a narrow line as a plug, first in, first out; it gives the shortest possible transit time, and real lines are slower.
Poiseuille's law
The relation for smooth, steady flow in a narrow tube: resistance rises in proportion to viscosity and length and falls with the fourth power of the radius, so halving the radius raises it sixteenfold.
post-occlusion bolus
The volume a pump pushed into the stretched syringe and line before an occlusion alarm, released at once when the block clears: at most 0.2 mL in one maker's table for small-bore and microbore tubing, and 1.99 mL in a 2013 bench test of a wrongly released line.
pre-clearance inspection
An inspection of the maker that FDA may conduct before clearing a pump 510(k), for example when the pump has new technological characteristics or the maker has not been inspected in two years, and that may delay clearance (Chapter 28).
predicate
The legally sold device a 510(k) compares a new or changed device with; it need not work on the same principle, as the Ivenix pump's predicates show (Chapters 6, 28).
preventive maintenance
The scheduled check of a pump made whether or not anything seems wrong: flow, occlusion alarm, air detection and battery against the maker's bands. One maker recommends it at least yearly, another at least every three years (Chapter 24).
Primary and secondary infusion
The primary infusion is the main fluid running through a set; a secondary infusion is a second fluid given through the same set (Chapters 4 and 11).
priming volume
The volume of fluid needed to fill a set or extension before use, stated by its maker; it approximates the dead volume the component adds to a line.
product code
The three-letter code FDA gives to a type of device, under which its 510(k)s and recalls are filed: FRN for the general infusion pump, MEB for elastomeric pumps, LZH for enteral feeding pumps, PHC for infusion safety management software (Chapters 6, 26, 28).
Quality Management System Regulation
FDA's quality regulation for devices, 21 CFR Part 820 as amended by the rule published on 2 February 2024 (89 FR 7496) and in force from 2 February 2026. It incorporates ISO 13485:2016 by reference and replaced the quality system regulation.
quick change
A syringe changeover in which the pump is stopped, the syringe replaced and the pump restarted, in a mean of 9.48 seconds in one bench study.
Recall
In FDA's regulations, a firm's removal or correction of a product. For infusion pumps it is usually a correction in place, by letter, workaround, changed instruction or new software.
recognized consensus standard
A standard on FDA's list of standards a maker can cite in a submission as part of its evidence; AAMI TIR101:2021 is recognized, IEC 60601-2-24 is not (Chapter 28).
Requirement
A statement of what a system must do that a test, an inspection or an analysis can check. It says what, not how.
resistance
The property of a tube or line that sets how much pressure a given flow needs; flow is the pressure difference divided by the resistance, and the resistances of parts joined end to end add.
roller clamp
A plastic case on the tubing of an administration set with a wheel that pinches the tube; the only rate control of a gravity infusion, and in MHRA's words the "primary means of occlusion" of a line. Under it, PVC tubing creeps after the clamp is set (cold flow).
root cause analysis
A structured search for every cause behind an incident, not only the last action before it; ISMP Canada's 2007 analysis of a fluorouracil infusion is one (Chapter 25).
Rule 12
The MDR classification rule for active devices that administer or remove medicinal products, body liquids or other substances: class IIa, or class IIb when done in a potentially hazardous manner. Infusion pumps are class IIb; elastomeric and feeding pumps class IIa.
safety assurance case
A structured argument that links each claim about a device's safety to evidence, built on a hazard analysis. FDA's infusion pump guidance of 2 December 2014 asks for one in a 510(k) submission.
sampling interval
The time between balance readings in a flow test: 30 seconds for IEC 60601-2-24 and 10 seconds for AAMI TIR101, as secondary sources give them.
Serial entry
Entering a number by typing its digits in order on a numeric keypad, as on a calculator; in a 2015 laboratory study, 1.36 % slips per number and 4.95 s per number (Chapter 17).
set-based anti-free-flow device
A clamp built into the administration set that closes the tube automatically whenever the set is taken out of the pump, so that free-flow protection travels with the set (Chapter 14).
Single fault condition
In IEC 60601-1, a condition in which a single means for reducing a risk is defective or a single abnormal condition is present. Equipment must stay free of unacceptable risk under it, and for a pump a fault that could cause over-infusion must raise an alarm and stop delivery or cut it to the keep-vein-open rate.
single-bubble and accumulated-air limits
The two air thresholds a pump applies: one for a single bubble (2.5 cm, about 140 µL, in one maker's sets) and one for air added up over time (more than 1 mL within 15 minutes in the same pump).
siphonage
Gravity emptying a syringe held above the patient when its plunger is not held by the pump's driver. Anti-siphon valves, which open only above a set pressure, prevent it.
soft limit
A drug library limit marking a dose that is unusual but sometimes right. An entry beyond it raises an alert that the nurse may override.
software bill of materials
The list of commercial, open-source and off-the-shelf software components inside a device, required under section 524B for cyber devices submitted to FDA from 29 March 2023.
start-up curve
Flow plotted against time from the start of an infusion, showing how long a pump takes to reach its set rate; drawn from the first hour of the IEC test.
State Licensing Authority
The state authority that licenses the manufacture of Class A and B devices in India (application on Form MD-3, license on MD-5).
static analysis
Inspection of source code by tools, without running it, to flag constructions known to cause faults. FDA's 2010 white paper said the agency would apply it to pump code, and its 2014 guidance asks for static analysis of all pump software with an account of unresolved findings.
Stepper motor
A motor that turns through a fixed angle for each pulse from its driver, typically 1.8 degrees, or 200 steps a turn (Chapter 5).
Syringe pump
A pump that clamps a filled syringe and drives its plunger forward with a carriage on a lead screw turned by a stepper motor. It infers volume from plunger travel and the syringe bore in its table (Chapter 5).
System
In the plain-language definition of INCOSE, the international council of systems engineers, a set of elements that work together to produce results that none could achieve alone. For an infusion: the pump, the set, the drug library, the health record and network, and the people who use them.
system accuracy
How closely the volume delivered through a real set or syringe matches the volume programmed, under stated conditions; the figure an accuracy claim normally gives.
Systems engineering
The discipline that integrates many disciplines across a system's life, from its realization to its retirement: it derives requirements from hazards, allocates them to parts, specifies the interfaces, and verifies and validates the whole.
Tenfold error
An error in which a number is multiplied or divided by ten, because a digit was added or lost or a decimal point moved by one place; 30 mL/h becomes 300 or 3 (Chapter 16).
time to occlusion alarm
The time from a blockage to the occlusion alarm, roughly the compliance times the limit, divided by the rate. One maker gives up to 1 h 02 min 40 s for its 50 mL syringe at 1 mL/h and a limit of 487 mmHg.
time to steady state
How long a pump takes after starting, or after a change, to reach the accuracy it claims; FDA's 2014 guidance asks for it on the label.
titrate
To adjust a drug's rate step by step to hold a measured effect, such as blood pressure.
Trace (traceability)
The links from each hazard to its controls, from each control to its requirements, and from each requirement to the parts that carry it and the tests that verified it, so that a change shows which hazards it can reach.
transit time
The time a change at the pump takes to cross a line's dead volume: the dead volume divided by the total flow, in the plug-flow approximation. 20 mL at 12 mL/h is 1 hour 40 minutes.
trumpet curve
A plot of the largest and smallest average errors against observation window length, flaring at short windows and closing toward the overall error at long ones.
upstream
In an infusion line, the stretch between the container and the pump. An upstream block, such as a closed clamp or an empty bag, starves the mechanism instead of raising the pressure against it.
Usability engineering
The work of analyzing, specifying, developing and evaluating a device's usability as it bears on safety, set out in IEC 62366-1; FDA's guidance calls it human factors and usability engineering (Chapter 18).
Use error
An action, or a failure to act, by which a user brings about a result the maker did not intend or the user did not expect; a rate entered in the wrong unit is one (Chapter 18).
Validation
Evidence that the requirements were the right ones, from trying the finished system with real users in its intended use; for a pump, the human factors validation test is its core.
venous pressure
The pressure of the blood in a vein, against which an infusion must push. In a peripheral vein it follows the central venous pressure only loosely.
Verification
Evidence that each part, and the whole, meets its requirements, by test, inspection or analysis; in design-control terms, that the design output meets the design inputs.
viscosity
A liquid's internal friction, which multiplies the resistance of a tube; water's is 1.00 mPa·s at 20 °C and 0.69 mPa·s at 37 °C.
Volume to be infused (VTBI)
The volume a nurse programs after which the pump ends the infusion; divided by the rate, it gives the time the infusion should take (Chapter 16).
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.