+ Part IV · Sensing
Sensing trouble.
A syringe pump running at 1 mL an hour can push into a blocked line for about an hour before it raises an alarm. B. Braun's manual for its Perfusor Space pump gives that figure for the maker's own 50 mL syringe with the alarm at its middle setting of 487 mmHg: up to about 63 minutes. Nothing in the pump has failed. It is waiting for pressure to build in a syringe and a line that stretch under load. This part asks what a pump can sense, how soon, and what it does next: a blockage in Chapter 12, air in Chapter 13 and fluid running with no pump in control in Chapter 14. Chapter 15 turns to the alarms that report all three, on units where alarm signals can number several hundred per patient a day.
A blocked line at 1 mL an hour.
+ The questionHow does a pump know a line is blocked, and why can it take an hour to notice?
Pressure is the only sign of a blockage
In 2020 the intensive care units of Mount Sinai Morningside raised the occlusion limit on their pumps from 300 to 500 mmHg. The pumps had been moved outside the patients' rooms, and each line now ran through two 210 cm microbore extensions, whose resistance Chapter 2 worked out. On a simple proportional model, the higher limit makes the pump take about two-thirds longer to detect a blocked line. The benefit of the higher limit, and its cost, follow from what a pump can observe. A pump counts the motion of its own mechanism, the steps of a motor or the turns of a screw, and that count does not change when the line beyond the pump closes: the motor keeps stepping as if fluid were still leaving.
What does change is pressure. Liquid barely compresses, so a pump that keeps moving against a closed line can make room for each new microliter only by stretching whatever lies between its mechanism and the block: the syringe barrel and plunger seal, the tubing, any filter. As these parts stretch, the pressure climbs. A sensor reads it, either as the force on the syringe plunger or as the pressure in the tube just beyond the mechanism. When the reading crosses a set limit, the pump stops and raises an occlusion alarm.
A block downstream, between the pump and the patient, raises the pressure the pump pushes against. A block upstream, between the container and the pump, such as a closed clamp or an empty bag, starves the mechanism instead, and a pump with a sensor on its inlet side sees the pressure there fall. BD lists the two as separate alarms for its Alaris neXus GP large-volume pump. Both rest on the same principle: the pump infers a blockage from pressure, never from a measured absence of flow.
Two failures therefore go unseen. In extravasation, the catheter has slipped out of the vein and fluid leaks into the surrounding tissue. MHRA, the UK regulator, noted in guidance it has since withdrawn that extravasation does not trigger an occlusion alarm, because the tissue offers too little resistance for the pressure to reach the limit. A disconnected line raises no pressure either, and the Mount Sinai team's 2020 report states that pumps detect occlusion but not disconnection. In both cases the drug leaves the pump at the programmed rate and goes somewhere other than the vein.
No alarm means only that the pressure has not yet reached the limit. At low rates and high limits that can take more than an hour, and all the while the display goes on counting volume the patient has not received. A slipped catheter or a parted line may never raise the pressure at all.
The give in the line sets the delay
Compliance, introduced in Chapter 5, is the volume a syringe, a set or a filter takes up for each unit of pressure. The 2015 metrology study of Chapter 5 measured it for syringes at about 0.65 bar, or about 490 mmHg, close to typical occlusion limits. A 10 mL syringe took up 0.21 mL per bar and a 50 mL syringe 1.54 mL per bar; adding an in-line filter raised the figures to 0.52 and 2.10 mL per bar. A pump pushing into a block has to fill this give at its programmed rate before the pressure reaches the limit. The time to alarm is therefore roughly the compliance multiplied by the limit and divided by the rate, which makes it long whenever the syringe is large, the limit high or the rate low.
Makers publish the result as a table. B. Braun's manual for the Perfusor Space gives the longest time to an occlusion alarm for each listed syringe at three of its nine pressure levels: P1 at 75 mmHg, P5 at 487 mmHg and P9 at 900 mmHg. With the maker's own 50 mL syringe at 1 mL/h, the times are about 13, 63 and 91 minutes. At 5 mL/h they shrink to 2.5, 9.5 and 16 minutes, and at 0.01 mL/h every syringe in the table takes more than 3 hours.
Smaller syringes stretch less, so they reach the limit sooner. At 1 mL/h and P5, the same manual gives about 11 minutes for a BD 3 mL syringe, 12.5 minutes for 5 mL, 25 minutes for 20 mL and 41 minutes for 30 mL. The 50 mL syringe takes almost six times as long as the 3 mL one, so the brand and size of syringe belong in any statement of how fast a pump detects a block. At microliter rates the wait stretches further: DEKA states 8 hours below 100 µL/h for its RemunityPRO+ subcutaneous pump, cleared in July 2026.
- A 50 mL syringe takes up 1.54 mL per bar (the 2015 study). One bar is 750 mmHg, so its compliance is 1.54 ÷ 750 = 0.00205 mL, about 2 µL, per mmHg.
- At the P5 limit of 487 mmHg the syringe stores 487 × 0.00205 = 1.00 mL before the alarm. At 1 mL/h the pump needs 1.00 ÷ 1 = 1.0 h, or 60 minutes, to store it; the maker's table gives at most 62.7 minutes for its own 50 mL syringe. At 5 mL/h the estimate is 1.00 ÷ 5 = 0.2 h, or 12 minutes, against 9.5 minutes in the table.
- Mount Sinai's change, applied to the same syringe for illustration: 300 mmHg stores 300 × 0.00205 = 0.62 mL and 500 mmHg stores 1.03 mL. At 5 mL/h the wait grows from 7.4 to 12.3 minutes, and at 1 mL/h from 37 to 62 minutes.
- Raising the limit by two-thirds lengthens both the wait and the volume held behind the block by about two-thirds.
A limit set between false alarms and late ones
The alarm limit is a compromise between two failures. A limit close to the pressure the infusion needs trips on every surge, while a high limit finds a real block later and stores more volume behind it. The adult line analyzed in Chapter 2 loses 141 mmHg across its parts, so its limit must sit above that to avoid false alarms. BD states that its Alaris Pump Module allows a maximum limit of 525 mmHg at 30 mL/h and above, decreasing in a straight line toward 50 mmHg as the rate approaches 1 mL/h. The ceiling is therefore lowest at low rates, where a high limit would take longest to reach.
Mount Sinai's long lines made false alarms likely. Microbore tubing needs more pressure for the same flow, and in 2020 ECRI's laboratory saw frequent occlusion alarms at 300 mL/h on long microbore extensions, while the Mount Sinai report warned of false alarms above that rate. Raising the limit from 300 to 500 mmHg gave the pumps headroom against false alarms but delayed the detection of real blocks. ECRI's alert of April 2020 on long extension sets found occlusion alarms delayed beyond the makers' specifications below 5 mL/h. It judged that these limitations "may be unacceptable for some high alert drugs", such as vasopressors, and advised keeping those pumps in the room.
Parts added to the line can also look like a block. A valve stays closed until the pressure behind it passes its opening pressure, and the sensor reads that rise as it would read a block. The authors of the bench study in Chapter 11 inferred that a syringe pump set to its lowest limit, 75 mmHg, would alarm on anti-siphon valves alone, with nothing blocked; they did not test that setting. In a Class II recall, Fresenius Kabi corrected Ivenix pumps that raised occlusion alarms at low flow when a check valve was added.
A carrier, the steady flow of plain fluid that takes a drug toward the vein, adds its own flow into any block beyond the junction where the lines meet. A 2014 bench study with a 1,000 mmHg limit found that joining lines moves detection in both directions. A block that a 10 mL/h line alone took 416 seconds to detect raised an alarm in 133 seconds once a 50 mL/h line joined it, because the faster pump reached the limit first. For the faster pump, the same junction lengthened detection from 81 to 133 seconds.
The bolus waiting behind the block
Whatever the pump pushed into the compliance before the alarm is still there, under pressure, when the alarm sounds. If the block then clears, because a clamp is opened, a kink straightened or a stopcock turned, the stored fluid leaves at once as a post-occlusion bolus. MHRA's guidance said the bolus grows with the alarm pressure set, and listed three ways to reduce it: smaller syringes, plungers that need little force, and back-off, in which the mechanism reverses slightly after the alarm to relieve the pressure. B. Braun's manual gives a maximum bolus of 0.2 mL for its 50 mL syringe at all three pressure levels on small-bore and microbore tubing, and up to 0.6 mL in all on standard-bore tubing.
A 2013 bench study measured what happens when an occluded line is released the wrong way, by opening a closed stopcock with the pressure still in the line. With a 50 mL syringe and the alarm set to High, one Terumo syringe pump flushed 1.99 mL and another 0.67 mL; the volume was smaller with a low alarm setting or a 10 mL syringe. At 1 mL/h, 1.99 mL is about two hours of infusion arriving at once.
Each of these numbers belongs to one combination of pump, syringe and set. BD's Class I recall of Alaris sets in July 2025, described in Chapter 8, named the time to occlusion alarm and the post-occlusion bolus among the ranges the sets might miss, and a second set recall followed in August 2026. The time to alarm is therefore a property of the whole line, not of the pump alone. A hospital that changes the syringe, the set or the extension is running a configuration the maker's table does not describe.
For a low-rate infusion of a short-acting drug, look up the maker's time to occlusion alarm for the brand and size of syringe in use, at the planned rate and pressure level. Use the smallest syringe that holds the dose, keep the limit as low as the line's normal pressure allows, and count every filter, valve and extension. A pump maker, in India or elsewhere, should publish the time to alarm and the bolus for every syringe and set on its compatible list at the lowest rates it offers.
- A pump cannot see flow stop; it infers a blockage when the pressure it reads crosses a set limit, so it cannot detect extravasation or a disconnection.
- Time to alarm is roughly compliance times limit divided by rate: at 1 mL/h and 487 mmHg, one maker's 50 mL syringe may take over an hour and a 3 mL syringe about 11 minutes.
- The limit trades false alarms against late ones; Mount Sinai's rise from 300 to 500 mmHg on long microbore lines gave headroom against false alarms at the price of slower detection at low rates.
- The volume stored before the alarm can reach the patient as a bolus when the block clears, from 0.2 mL in one maker's table to 1.99 mL in a bench test of a wrongly released line.
Let's Talk