02 — Resistance

The line pushes back.

+ The questionWhat pressure does it take to push fluid through a catheter, and what does the vein push back with?

Flow follows pressure, divided by resistance

BD rates its 20-gauge catheter, 29.5 mm long, at 61 mL/min under gravity, its 18-gauge of the same length at 95 mL/min and its 16-gauge at 193 mL/min. A catheter here is the short plastic tube that stays in the vein and carries the infusion into it, and its gauge is a size number that runs backward: the higher the gauge, the narrower the bore. BD gives the bore of its 20 G catheter as 0.76 to 0.84 mm, of its 18 G as 0.94 to 1.02 mm and of its 16 G as 1.32 to 1.40 mm. A bore less than twice as wide carries about three times the flow, and the reason begins with how pressure, resistance and flow relate.

For a steady flow through a tube, the flow is proportional to the pressure difference between its ends and inversely proportional to a property of the tube called its resistance. Doubling the pressure doubles the flow, and doubling the resistance halves it. The relation is the same as the one between voltage, current and resistance in an electrical circuit, and as in a circuit, the resistances of parts joined end to end add. In the gravity drip of Chapter 1 the pressure is fixed by the head, so resistance decides the flow, and the roller clamp works by adding resistance.

A pump reverses the roles. It fixes the flow and lets the pressure rise to whatever the resistance of the line demands. MHRA, the UK regulator, put it in one sentence in its 2013 guidance on infusion systems, since withdrawn: "The pump detects any resistance to flow and increases the pumping pressure." For a pump, then, the useful question is how much pressure it must make to hold a rate, and at what pressure that effort becomes a sign that something is wrong. Answering it needs the resistance of each part of the line.

Why the radius counts four times

For smooth, steady flow in a narrow tube, the resistance is given by Poiseuille's law. It rises in proportion to the liquid's viscosity and to the length of the tube, and it falls with the fourth power of the radius. Halving the radius raises the resistance sixteenfold, so at the same pressure the narrow tube carries one-sixteenth of the flow. Length enters simply: a tube twice as long has twice as much wall dragging on the liquid, so it needs twice the pressure for the same flow. The fourth power comes from two effects of width that multiply.

Liquid touching the wall does not move, and each layer further in slides past its neighbor, so the speed rises from zero at the wall to a maximum on the axis. A wider tube has more area, in proportion to the square of its radius. It also has more liquid far from the wall, where the wall's drag is felt least, so under the same pressure the average speed across the tube also rises with the square of the radius. Flow is area times average speed, so it grows as the square times the square: the fourth power. The narrowest part of a line therefore dominates its resistance far more than its share of the length would suggest.

03 — Half the radius, one-sixteenth of the flow
RADIUS R RADIUS R/2 SAME PRESSURE DIFFERENCE AREA 1 MEAN SPEED 1 FLOW 1 AREA 1/4 MEAN SPEED 1/4 FLOW 1/16 RESISTANCE ×16 Schematic. Poiseuille's law: smooth, steady flow; liquid at the wall does not move.
Fluid and drugPressure and forceReadoutFocal detail
Plate 03 — Width counts twice: once in the area and once in the average speed, because liquid far from the wall moves fastest. Halving the radius cuts the flow at the same pressure to one-sixteenth.

Viscosity is a liquid's internal friction, and it multiplies the resistance directly: a liquid twice as viscous needs twice the pressure for the same flow. It also changes with temperature. NIST's reference values for the viscosity of water are 1.00 mPa·s at 20 °C and 0.69 mPa·s at 37 °C, so at body temperature water is about 31 % less viscous than at room temperature, and the same pressure pushes about 45 % more of it through the same tube. The worked numbers in this chapter use water at 20 °C. Drug solutions have viscosities of their own, and Chapter 7 describes a pump that relies on viscosity to set its rate.

Resistances add along the line

If the catheter were the only resistance in BD's test, its ratings would follow the fourth power. The 16 G bore is 1.39 times the 18 G bore, taking the middle of each range, which predicts 3.7 times the flow; the ratings give 2.0 times. BD's 20 G catheter also comes in a 47.8 mm length, and Poiseuille's law predicts that the 29.5 mm version carries 1.62 times as much; the ratings give 61 against 54 mL/min, or 1.13 times. Something else in the test resists. A gravity test runs through a container, tubing and connectors as well as the catheter, and their resistances add to the catheter's in series, so the ratings spread far less than the catheters alone would predict.

04 — Catheter ratings against the fourth-power law
0 100 200 300 400 GRAVITY FLOW (mL/min) maker's rating dashed outline: fourth-power prediction from 18 G 61 43 20 G 0.76–0.84 mm 95 95 18 G 0.94–1.02 mm 193 354 16 G 1.32–1.40 mm THE TEST PATH ADDS RESISTANCE Maker's figure: BD Insyte Autoguard BC, 29.5 mm (1.16 in). Prediction: the guide's arithmetic, middle of each bore range.
ReadoutFocal detail
Plate 04 — If the catheter were the only resistance, the 16 G would pass 354 mL a minute; its rating is 193. The tubing and connectors of the test add resistance in series, so the ratings spread far less than the bores predict.
A catheter's flow rating describes its test as much as the catheter

BD's product pages do not state the conditions behind the ratings. A rating made with one head and one set does not carry over to another set, a long extension, a pump or a more viscous liquid, because the fourth-power law applies to the whole line and the catheter is only part of it. The ratings rank catheters within one maker's table; they do not give the flow of any particular line.

The same arithmetic shows how little a catheter resists at the rates pumps use. For the 20 G catheter alone, 29.5 mm long with a bore of 0.80 mm, the middle of BD's range, Poiseuille's law gives a pressure drop of 0.61 mmHg at 100 mL/h and 1.8 mmHg at 300 mL/h. Its gravity rating of 61 mL/min is 3,660 mL/h. At pump rates a peripheral catheter is a small part of the line's resistance. The parts that matter are long and narrow, the kind of tubing added when a pump must sit away from the bed.

What the vein pushes back with

The fluid enters a vein whose blood is under pressure, and the bag or the pump must exceed that pressure before anything flows. The pressure in a peripheral vein is neither fixed nor well predicted from elsewhere. In a 2023 study of 60 patients during high-risk abdominal surgery, pressure measured through an 18 G catheter at the elbow was compared with the central venous pressure, the pressure in the large veins near the heart. The two agreed only loosely: the 95 % limits of agreement, the range expected to hold 95 % of the differences between paired readings, ran from about 6 mmHg one way to 9 mmHg the other. For the gravity drip of Chapter 1, driven by 58.8 mmHg from an 80 cm head, an uncertainty of that size in the back pressure is a tenth or more of the push. For a pump, such uncertainties are small beside the pressure it can make.

Pumps push much harder, up to a limit. The occlusion limit, the subject of Chapter 12, is the pressure at which a pump stops and alarms on the assumption that the line is blocked. BD states that the limit on its Alaris Pump Module can be set from 50 to 525 mmHg, with a ceiling of 525 mmHg at 30 mL/h and above and lower ceilings below it. Fresenius Kabi states 50 to 750 mmHg for its Agilia VP MC. The limit must sit above what the line itself needs: MHRA's guidance worked an adult example in which the pressure drops along the line add to 141 mmHg, and a neonatal one in which they add to 22 mmHg.

Long extensions meet the pressure ceiling

An extension set is a length of tubing added between the administration set and the catheter. BD classes extension tubing with a bore under 2 mm as microbore. Its 213 cm microbore extension has a priming volume, the fluid needed to fill it, of 0.7 mL, which implies a bore of about 0.65 mm: narrower than the 20 G catheter and more than seventy times as long. By the fourth-power law such a tube resists far more than the catheter at its end. ECRI, a device-evaluation organization, issued a COVID-19 alert on long extension sets on 1 April 2020. In it, ECRI reported that its laboratory, testing 20 ft of microbore tubing, saw "frequent occlusion alarms, which made the pump unusable" at 300 mL/h.

05 — Pressure to push 300 mL/h through the line
0 100 200 300 400 500 600 700 PRESSURE NEEDED AT 300 mL/h (mmHg) 525 mmHg: HIGHEST OCCLUSION LIMIT BD ALARIS PUMP MODULE (MAKER'S FIGURE) MOUNT SINAI LIMITS 300 → 500 mmHg (PUMP NOT NAMED) 20 G CATHETER, 29.5 mm 1.8 mmHg ONE 213 cm MICROBORE EXTENSION 310 mmHg TWO IN SERIES 620 mmHg ALARM WITH NOTHING BLOCKED Poiseuille's law, water at 20 °C; bore 0.65 mm from BD's 0.7 mL priming volume. The guide's arithmetic; Mount Sinai did not name its tubing.
Pressure and forceReadoutFocal detail
Plate 05 — At pump rates the catheter hardly resists; long microbore tubing does. Two 213 cm extensions need more pressure at 300 mL/h than one maker's pump will make before it alarms.
Solved example: the pressure to push 300 mL/h through 213 cm of microbore tubing

The bore comes from the maker's volume: 0.7 mL is 700 mm³, spread along 2,130 mm, a cross-section of 0.3286 mm². A circle of that area has a radius of 0.3234 mm, a bore of about 0.65 mm. The flow, 300 mL/h, is 300,000 mm³ in 3,600 s, or 83.33 mm³ a second. Poiseuille's law gives the pressure drop as 8 × viscosity × length × flow, divided by 3.142 × radius to the fourth power. With water at 20 °C, 1.0016 mPa·s, that is 8 × 1.0016 × 2,130 × 83.33 = 1,422,000, divided by 3.142 × 0.01094 = 0.03437, which gives 41,380,000 mPa, or 41,380 Pa. Divided by 133.3 Pa per mmHg, that is 310 mmHg; at 100 mL/h it would be a third as much, 103 mmHg. Two extensions in series double the length and the pressure, to about 620 mmHg at 300 mL/h, above the 525 mmHg ceiling BD states for its pump module. That pump would alarm for occlusion with nothing blocked. The calculation uses BD's published tubing geometry; Mount Sinai did not name its tubing. The priming volume includes the connectors, so the true bore is, if anything, narrower and the pressure higher.

The intensive care units of Mount Sinai Morningside, whose long lines Chapter 3 describes, met the problem in 2020: they raised their occlusion limits from 300 to 500 mmHg and warned of false occlusion alarms at high rates. If the tubing matched BD's microbore geometry, the arithmetic shows why: a line that needs hundreds of millimeters of mercury just to flow leaves little room below the ceiling. Raising the limit restores that room at high rates, at a cost that falls on low rates, where the pump must build a higher pressure before it alarms. Chapter 3 follows that line through the door, and Chapter 12 measures the cost.

Add up the line before changing a pressure limit

For a biomedical engineer asked to support long lines, compute the pressure the tubing needs at the highest rate that will run through it, from each maker's length and priming volume, and compare it with the pump's ceiling at that rate. Where the sum comes close to the ceiling, expect nuisance occlusion alarms at high rates. Where the limit is raised to prevent them, check what the higher limit does to the time a pump takes to detect a real blockage at low rates (Chapter 12). Record the tubing, rates and limits chosen, so that a later investigation can reconstruct them.

+ What this chapter established
  • A gravity drip has a fixed pressure and lets resistance set the flow, while a pump fixes the flow and raises its pressure until the line's resistance is overcome.
  • Resistance rises with length and viscosity and falls with the fourth power of the radius, so halving a tube's radius raises its resistance sixteenfold.
  • Resistances in series add, which is why BD's catheter ratings spread less than the fourth-power law predicts and why a peripheral catheter barely resists at pump rates.
  • Calculated on BD's tubing geometry, one 213 cm microbore extension needs about 310 mmHg at 300 mL/h, so two in series exceed a 525 mmHg pump ceiling, consistent with ECRI's alarms and Mount Sinai's raised limits.
Created with Claude (Anthropic).