05 — Syringe

A screw and a plunger.

+ The questionHow does a syringe pump deliver 0.1 mL an hour to a newborn, and what makes it stall and then surge?

A screw turns steps into distance

A 50 mL syringe emptied at 0.1 mL/h lasts 500 hours, nearly three weeks, and the pump must move its plunger too slowly for the eye to see. Chapter 4 showed that a peristaltic pump depends on the bore and springiness of a tube. A syringe pump takes the tube out of the calculation: it clamps a filled syringe in place and drives the plunger forward with a carriage on a screw. Smiths Medical's 2017 guide to its Medfusion 3500 gives a lowest rate of 0.01 mL/h with 1 mL syringes, and Baxter's 2022 specification for its Novum IQ syringe pump gives a range of 0.01 to 1,200 mL/h. Each rate comes from turning the screw by a counted amount.

The motor is a stepper motor, which turns through a fixed angle each time its driver sends a pulse. Ossila, which makes laboratory syringe pumps, gives typical figures on an undated web page: a step of 1.8 degrees, or 200 steps a turn, driving a lead screw, a threaded rod whose nut and carriage advance a fixed distance for each turn. A screw cut with four interleaved threads 2 mm apart advances 8 mm a turn, so each step moves the plunger 0.04 mm. Drivers can also divide each step into microsteps, commonly 1/32 or 1/64 and sometimes 1/256, though at 1/64 the torque for one microstep is under 2.5 % of the maximum. These are laboratory figures, and the step sizes of clinical pumps are not known.

Distance becomes volume through the bore of the barrel. The volume pushed out equals the barrel's inside cross-section times the plunger's travel, so the pump has to know which syringe it holds. It looks the bore up in a table of syringes its maker has tested, after the user confirms the brand and size, and some pumps also sense the syringe's size. The pump's knowledge of the volume delivered therefore rests on two numbers it never measures during the infusion: the distance its screw has moved, counted from motor steps, and the bore it was told.

12 — From a motor step to microliters
STEPPER MOTOR 1.8° A STEP · 200 A TURN LEAD SCREW · 4 THREADS 2 mm APART → 8 mm A TURN NUT AND CARRIAGE PLUNGER STOPPER TO PATIENT CROSS-SECTION 5.5 cm² (ASSUMED) 1 STEP = 0.04 mm = 22 µL 1 MICROSTEP (1/64) = 0.000625 mm ≈ 0.34 µL 0.1 mL/h = 0.18 mm AN HOUR about 290 microsteps an hour, not 4.5 pulses of 22 µL step figures from a laboratory pump maker (Ossila); barrel cross-section assumed by the guide
Fluid and drugPressure and forceReadoutFocal detail
Plate 12 — Motor steps become plunger travel through the lead screw and volume through the barrel's cross-section. With laboratory figures and an assumed 5.5 cm² barrel, a full step pushes out 22 µL and a 1/64 microstep about 0.34 µL, so 0.1 mL/h becomes about 290 small steps an hour instead of 4.5 pulses.
Solved example: from a motor step to microliters

The illustration assumes a 50 mL barrel with an inside cross-section of 5.5 cm², which is 550 mm² and about 26.5 mm across; makers publish exact bores only in their pumps' syringe tables. With the laboratory figures above, one full step moves the plunger 0.04 mm and pushes out 550 × 0.04 = 22 mm³, or 22 µL. One microstep at 1/64 moves 0.04 ÷ 64 = 0.000625 mm and pushes out 550 × 0.000625 = about 0.34 µL. A rate of 0.1 mL/h is 100 µL/h, so the plunger travels 100 ÷ 550 = 0.18 mm an hour. In full steps that is 100 ÷ 22 = 4.5 steps an hour, a pulse of 22 µL about every 13 minutes. In microsteps it is 100 ÷ 0.34 = about 290 an hour, one every 12 seconds or so. Microstepping turns a pump of pulses into one that flows.

The syringe is part of the calibration

Because the cross-section grows with the square of the diameter, the arithmetic of Chapter 4 applies again: a barrel 3 % wider than the table assumes delivers about 6 % more for every millimeter of travel. A syringe of one brand selected as another, or a brand the pump was never tested with, changes the dose while the plunger moves exactly as commanded. BD, which makes syringes, states in a 2021 brochure that syringes not validated for a pump, or a wrong syringe type selected on it, have produced over-delivery of more than 20 %. The same brochure cites FDA's advice to choose the smallest compatible syringe for high-risk drugs at rates below 5 mL/h. Makers' accuracy figures for syringe pumps are often tighter than those for large-volume pumps, and hold only with the syringes and rates they name (Chapter 8).

Slack to take up before the drug flows

When a syringe pump starts, the plunger moves at once and the drug does not. Before fluid leaves the far end of the line, the drive must take up the play in its gears and nut and overcome the friction of the stopper in the barrel, and the stopper, the barrel and the tubing must be squeezed and stretched until the pressure inside can push fluid through the catheter. The property that stores this volume is compliance, the volume a syringe or line takes up for each unit of pressure. A 2015 study by metrology institutes measured 0.21 mL per bar for a 10 mL syringe and 1.54 mL per bar for a 50 mL syringe, about seven times as much, and more again with a filter in the line. One bar is 750 mmHg, so at an occlusion limit of 500 mmHg the 50 mL syringe stores about 1 mL.

Bench studies show how long that slack takes to fill. In a 2007 study of one syringe pump with two brands of syringe, the time to reach the set flow ranged from 3.6 minutes for a 10 mL syringe at 1 mL/h to 74.5 minutes for a 50 mL syringe at 0.1 mL/h. Start-up was slower at the lower rate, with the larger syringe and with one brand, and the authors attributed it to gear engagement in the drive and compliance of the syringe assembly. A 2020 study of seven current pump and syringe combinations at 1 mL/h found 89 to 1,622 seconds, about 1.5 to 27 minutes, before flow was steady. Its authors traced the low-rate problems mainly to the working principle of the syringe pump.

Plunger travel runs ahead of the drug at start-up

A syringe pump counts volume from the first step of its motor, so in the first minutes at a low rate it reports drug that is still being stored in the syringe and line. With a 50 mL syringe at 0.1 mL/h, the set flow may not arrive for more than an hour. A slow response read as a drug that has failed, and answered with a higher rate or a second drug, can stack doses on a line that has not yet delivered the first.

The slack can be taken up before the infusion starts. A 2022 nursing study timed the first drop reaching a balance for two unnamed pumps with a 50 mL syringe and a 150 cm line. At 1 mL/h, with the line primed by hand, the first drop took 25.5 minutes with one pump and 34.7 with the other; when the pump's own priming bolus was given after the syringe was loaded, it took 0.2 and 3.0 minutes. At 0.3 mL/h the times fell from 74.2 and 97.7 minutes to 6.7 and 19.3. At 20 mL/h, primed by hand, the first drop came in 0.9 minutes with either pump. Priming by hand fills the line but leaves the drive's play and the syringe's compliance for the infusion to take up.

13 — Priming with the pump removes most of the wait
outline: line primed by hand · filled: pump's priming bolus after loading 0.3 mL/h PUMP A PUMP B 1 mL/h PUMP A PUMP B 20 mL/h PUMP A PUMP B 0 20 40 60 80 100 MINUTES TO FIRST DROP 74.2 6.7 97.7 19.3 25.5 0.2 34.7 3.0 0.9 0.9 pump priming not reported at this rate 25.5 min by hand, 0.2 min with the pump's bolus two unnamed pumps, 50 mL syringe, 150 cm line; nursing study, 2022 (mean times)
ReadoutFocal detail
Plate 13 — Priming by hand fills the line but leaves the drive's play and the syringe's compliance to be taken up during the infusion. In a 2022 study the pump's own priming bolus cut the wait at 1 mL/h from 25.5 to 0.2 minutes with one pump and from 34.7 to 3.0 with the other.

Stall and surge come from the same stored volume

Compliance does more than delay the start: whatever volume the syringe and line take up under pressure they give back when it falls, so anything that raises the pressure makes delivery stall and anything that lowers it makes delivery surge. In a blocked line the plunger keeps advancing and the syringe absorbs drug until the pressure reaches the alarm limit, which with a 50 mL syringe at 1 mL/h can take about an hour; Chapter 12 shows that the syringe's compliance accounts for most of that hour. When the block clears, the stored volume comes out at once, and Chapter 12 gives the volumes measured on release. Height works the same way. Lowering a syringe pump below the patient raises the pressure the syringe must hold, so delivery pauses while its compliance fills, and raising it releases a bolus. Chapter 10 follows height changes and syringe changeovers through drugs whose effect fades within minutes.

14 — Compliance stores what a blocked line withholds
10 mL SYRINGE 0.21 mL PER bar 50 mL SYRINGE 1.54 mL PER bar ALARM LIMIT 487 mmHg = 0.65 bar 0.14 mL 1.0 mL ≈ AN HOUR AT 1 mL/h held back while blocked, released when the block clears PRESSURE RISES → STORES (STALL) PRESSURE FALLS → RELEASES (SURGE) maker's figure: 1 h 02 min 40 s to reach 487 mmHg, Omnifix 50 mL at 1 mL/h schematic; compliance from a 2015 metrology study; stored volumes are the guide's arithmetic
Fluid and drugPressure and forceReadoutFocal detail
Plate 14 — Compliance is volume stored per unit of pressure. At a 487 mmHg limit a 50 mL syringe stores about 1.0 mL, an hour's delivery at 1 mL/h, against 0.14 mL for a 10 mL syringe, and gives it back when the pressure falls.

FDA warned in August 2016 of flow-continuity problems with syringe pumps below 5 mL/h (Chapter 10). The newborn of this chapter's question is the patient for whom start-up, stall and surge matter most. The newborn's rates are the lowest, and in the 2020 study of seven assemblies, which modeled a newborn's epinephrine infusion, the occlusion-alarm delays and the flow irregularities after height changes gave deviations of more than 25 % in the simulated level of the drug in the blood with every assembly.

Control the syringe list and the low-rate set-up

For biomedical engineers and pharmacists: configure each pump's syringe table to the brands and sizes the hospital stocks and remove the rest, so that nobody can confirm a syringe the pump was not tested with, and review the table whenever purchasing changes supplier. For drugs below 5 mL/h, agree the smallest syringe that holds the dose and a concentration that keeps the rate up, as one pediatric unit did (Chapter 10). Teach priming with the pump after loading, and keep the pump level with the patient during transfers.

+ What this chapter established
  • A syringe pump turns motor steps into plunger travel with a lead screw and into volume through the bore of the syringe, so microstepping sets how smooth a low rate is and the syringe table sets how large it is.
  • A syringe of another brand or size than the one selected changes the dose while the plunger moves as commanded, and makers' tight accuracy figures hold only with the syringes and rates they name.
  • At low rates the plunger runs ahead of the drug while the play in the drive and the compliance of syringe and line are taken up, for more than an hour with a 50 mL syringe at 0.1 mL/h, and priming with the pump shortens the wait.
  • The same stored volume makes delivery stall when pressure rises and surge when it falls, after a blocked line clears or the pump is moved in height.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.