31 — Recalls

After the pump is sold.

+ The questionHow are pump problems found and corrected once pumps are in use, and what do the recalls since 2010 show?

A recall that left the pumps in place

On 14 July 2025 Baxter wrote to hospitals that its Novum IQ large-volume pump could underinfuse when the rate was raised to more than twice its previous value, by a rate change or a bolus. It added that a wrongly loaded set could cause over- or underinfusion. In the worst case no fluid might be delivered. By 27 June 2025 Baxter had received reports of 79 serious injuries and 2 deaths. FDA relayed these as the company's figures, not as established causes. FDA classed the action as a Class I recall, and its record covers about 37,000 units and gives the cause as "device design".

The pumps stayed on the wards, although Baxter paused distribution in the United States and Canada except where medically necessary (Chapter 26). Until a fix arrived, users who could not change the pump or set without unacceptable delay were told to move the set about 13 mm downstream in the pump before a bolus or any rate increase above 100 %. Most pump recalls take this shape once a pump has passed the regulators of Chapters 28 to 30. Problems are found as reports accumulate: hospitals report to makers, and both report to the regulator. The maker then corrects them, usually in place, with a letter, a workaround, a changed instruction or new software.

In FDA's regulations a recall is a firm's removal or correction of a product. A correction includes repair, modification, adjustment, relabeling, destruction or inspection without moving the product. FDA assigns the class. Recall classes are a separate scale from the device classes of Chapter 28 and run the other way: Class I is the lowest-risk device class but the most serious recall class. Class I means "a reasonable probability" that using the product "will cause serious adverse health consequences or death"; Class II covers temporary or medically reversible harm, and Class III covers products unlikely to cause harm. FDA's recent notices call Class I "the most serious type of recall".

From a report to the record

In the United States the reports that start most recalls run on fixed clocks. A hospital must report a death linked to a device to FDA and to the maker within 10 work days, and a serious injury to the maker within 10 work days. A manufacturer must file its reports within 30 calendar days, or within 5 work days when an event calls for remedial action. A firm that starts a correction or removal must report it to FDA within 10 working days. These medical device reports, mandatory and voluntary, go into FDA's MAUDE database, which is updated monthly and searchable online for the last ten years.

FDA states that MAUDE data are not meant for judging rates of adverse events. A report has no denominator: nobody counts the infusions that ran without incident, or the pumps in use. One event can produce several reports, and each records a suspicion rather than a finding. The reports counted in the 2010 white paper, about 56,000, were of this kind. Recalls, which FDA classes after review, are firmer ground. A 2019 analysis by academic authors covered 70 Class I recalls of infusion pumps from 2001 to 2017, 12 of them for insulin pumps. Among the 66 recalls that stated a harm, it found underdosing in 36 and overdosing in 18, or 55 % and 27 %. It attributed 45 of 65 recalls to hardware faults and 17 to software, or 69 % and 26 %. Underdosing fits the mechanisms of Part IV: a pump that stops, locks up or alarms for nothing delivers less.

Sixteen years of Class I records

FDA's recall database can be searched by product code (Chapter 6). Under FRN, the code for infusion pumps, it lists 163 records of Class I recalls posted from 1 January 2010 to 11 October 2026. By year they run 10, 19, 9, 11, 7 and 8 from 2010 to 2015; 0, 1, 0 and 3 from 2016 to 2019; then 27, 11, 4, 15, 15 and 21 from 2020 to 2025, and 2 so far in 2026.

79 — Class I recall records for infusion pumps, 2010 to 2026
0 10 20 30 CLASS I RECORDS, CODE FRN 2010 2011 2012 2013 2014 2015 2016 2017 2018 2019 2020 2021 2022 2023 2024 2025 2026 to 11 Oct 27 10.7 a year 1.0 a year 15.5 a year 163 records · 1 Jan 2010 to 11 Oct 2026 NEAR-EMPTY YEARS: MAY REFLECT CODING FDA Medical Device Recalls database, product code FRN, Class 1, by posting date; query 11 Oct 2026. Records are product lines, not recall events.
ReadoutFocal detail
Plate 79 — FDA's database holds 163 Class I records under FRN since 2010, about 10.7 a year to 2015 and 15.5 a year from 2020 to 2025, but the records count product lines, and the near-empty years may reflect coding.
Solved example: what the count of Class I records can say
  1. 2010 to 2015: 10 + 19 + 9 + 11 + 7 + 8 = 64 records in six years, about 10.7 a year.
  2. 2016 to 2019: 0 + 1 + 0 + 3 = 4 records in four years, 1 a year.
  3. 2020 to 2025: 27 + 11 + 4 + 15 + 15 + 21 = 93 records in six years, 15.5 a year.
  4. Check: 64 + 4 + 93 = 161, and the 2 records of 2026 make 163.
  5. Ratio of the last period to the first: 15.5 ÷ 10.7 ≈ 1.45.

Class I records under FRN were therefore about 45 % more frequent from 2020 to 2025 than from 2010 to 2015. Each record, however, is one product line, so a recall of several models makes several records, and PCA pumps, insulin pumps and many sets carry other codes.

A recall count measures the record, not the risk

The 163 records cannot be set against the white paper's 14 Class I recalls of 2005 to 2009, because one counts product records and the other recall events. Neither has a denominator of pumps or infusions, and the near-empty years 2016 to 2018 may reflect how recalls were coded rather than a safer fleet. A rising count can mean more problems, better reporting or a recall split into more records.

What the recalls of 2025 and 2026 were about

Few of the recent records concern the pumping mechanism alone. Several concern the set. BD's Class I recalls of July 2025 and August 2026 covered about 33 million and 94 million Alaris sets whose performance might fall outside, or below, what BD had published (Chapter 26). FDA's record gives the reason for the second as discrepancies in the sets' performance data. No pump was removed. That is the point of Chapters 8 and 9: accuracy belongs to a pump and its set together.

Others concern software, screens, alarms and power. Fresenius Kabi's Ivenix pump had Class I corrections for software that locked up and for a touchscreen cable that could loosen after a drop (Chapter 17). Another, described at the end of this chapter, concerned a misread battery capacity and a frozen screen. The software issued as its fix, version 5.10.2, was itself corrected in May 2026 because it could show a false battery reading; that recall opens Part IX. ICU Medical's CADD-Solis VIP ambulatory pump could give false upstream occlusion alarms, in a recall that covered about 186,000 units in the United States and about 128,000 elsewhere. These problems belong to the alarms, interfaces, firmware and fleet batteries of Chapters 15, 17, 20 and 24.

Field safety notices in Europe and reports in India

In the EU the duties sit in the MDR (Chapter 29). A manufacturer reports a serious incident within 15 days of becoming aware of it, a death or unanticipated serious deterioration of health within 10 days, and a serious public health threat immediately, and no later than 2 days, counting calendar days from awareness. A field safety corrective action is the maker's action to prevent or reduce the risk of a serious incident, and the field safety notice is the letter that tells users about it. The maker sends a draft of the notice first to the national authority evaluating the case, which has 48 hours to comment except in urgent cases. Until EUDAMED's vigilance module runs, national authorities publish the notices.

In India, reports go to MvPI, the materiovigilance program, on the clocks Chapter 30 sets out. The distribution list of a BD Alaris set recall that FDA classed as Class I in 2025 includes India. An Indian hospital may therefore receive a notice that began as a US correction, while its own reports go to MvPI and CDSCO.

80 — Reporting clocks after awareness: US, EU and India
US EU INDIA 0 5 10 15 20 25 30 DAYS FROM AWARENESS maker: remedial action, 5 work days user facility: death to FDA and maker, serious injury to maker correction or removal report to FDA maker: 30 calendar days serious public health threat death or unanticipated serious deterioration other serious incidents proposed for some incidents, COM(2025) 1023 serious adverse event to MvPI; serious unexpected to licensing authority non-serious event solid: calendar days · hollow: work days · dashed: proposed DAY 0: AWARENESS 21 CFR 803 and 806.10 · MDCG 2023-3 rev.2 (January 2025) · IPC MvPI guidance v1.3 (2026)
Software and dataFocal detail
Plate 80 — Every reporting clock starts on the day someone becomes aware of an event, and in the US, the EU and India the deadlines for serious events fall within 30 days.
Treat each recall notice as a change to the fleet

Match the notice's models, serial numbers, lots and software versions against the pump inventory and the set stock. Put interim instructions where users see them: the Novum IQ set move before a large rate increase, the Ivenix advice to keep pumps plugged in, and any entry the maker says to avoid. Check the drug library entries the notice touches, and check whether the fix has a notice of its own, as Ivenix version 5.10.2 did. Report events to the maker and the regulator on their clocks, and record each correction in the pump's service history (Chapters 24 and 25).

April 2010, and two zeros in 2025

The guide's opening story is set on 23 April 2010, when FDA announced its initiative with a white paper that described a key registering one press of zero as two, so that "0" became "00". A week later FDA ordered Baxter to recall and destroy its Colleague pumps, about 200,000 of them in American hospitals by Baxter's count, and FDA issued its final order on 13 July. The pumps were recalled for documented defects in their batteries, their user interface and failure codes, and a temperature-sensitive part of their timing circuit. The white paper's examples, like the recalls of 2025 and 2026, were mostly about what surrounds a pumping mechanism.

As Chapter 25 concluded, most pumps know how much they have delivered only by inference. Each counts the motion of its own mechanism, the steps of a motor or the strokes of a cassette, and converts that motion into milliliters through assumptions about the set, the line, the fluid and the numbers it was given. Its sensors test some of those assumptions, and only after a delay (Chapters 4, 8 and 12). The key bounce of 2010, which registered one press of zero as two, gave the pump a number nobody meant. A pump that worked perfectly would deliver exactly that number, which is how a faultless mechanism gives a wrong dose. The BD sets broke an assumption about the set, and the Ivenix battery reading broke one about power. Each time, the dose depended on something other than the mechanism.

Even the exceptions to that rule show it. The Ivenix pump measures the volume in its pumping chamber with air pressure and the gas law, so its milliliters are measured rather than inferred, and DEKA's RemunityPRO+ monitors delivery accuracy with acoustic volume sensing (Chapter 4). Neither measures the flow that reaches the patient, and no pump measures what it is told. On 14 November 2025 Fresenius Kabi began a Class I correction of the Ivenix pump's software for two anomalies (Chapters 15 and 17). A misread battery capacity could shut the pump down, and a rate entered with two leading zeros above a drug library limit could freeze the interface. Until the fix, users were to keep the pumps plugged in and to avoid programming two leading zeros. Fifteen years after one press of zero became two, a pump again took a zero differently from what was meant.

81 — Zeros the software did not expect, fifteen years apart
APRIL 2010 · FDA WHITE PAPER software defect · no pump or maker named NOVEMBER 2025 · IVENIX LVP SOFTWARE 5.10.1 Z-0885-2026 · Class I · 30 units · 2 serious injuries reported 1 2 3 4 5 6 7 8 9 . 0 C ONE PRESS 00 KEY BOUNCE: REGISTERED AS TWO 0 0 _ _ TWO LEADING ZEROS RATE ABOVE A DRUG LIBRARY LIMIT → INTERFACE FREEZES interim advice: avoid programming infusates with two leading zeros when possible FIFTEEN YEARS APART schematic · FDA white paper, April 2010; FDA recall Z-0885-2026
Software and dataHuman input and use errorReadoutFocal detail
Plate 81 — In 2010 a key bounce turned one zero into two; in November 2025 a pump that measures its own chamber volume could freeze on a rate entered with two leading zeros, and users were told to avoid such entries.
+ What this chapter established
  • Most pump recalls are corrections in place, by letter, workaround or new software, and FDA assigns the class, with Class I for a reasonable probability of serious harm or death.
  • Reports to MAUDE, EU authorities and MvPI run on clocks of days, but without a denominator they show where to look, not how often harm occurs.
  • FDA's database holds 163 Class I records under product code FRN from 2010 to October 2026, a count of product records shaped by coding, not a measure of risk.
  • The recalls of 2025 and 2026, like the examples of April 2010, mostly concerned sets, software, screens, alarms and power, which a pump's count of milliliters takes on trust.