+ Part III · Accuracy

How accurate, measured how.

Two syringe pumps can carry the same accuracy claim, such as the ±2 % that BD states for its Alaris syringe pump at 1 mL/h and above, and both can meet it. Yet when a 2020 laboratory study ran seven syringe pump assemblies at 1 mL/h, the time each took to reach a steady flow ranged from 89 seconds to 27 minutes. A claim calculated after the pump has settled says nothing about how it got there, and for a drug whose effect fades within a minute or two of the flow stopping, those first minutes count most. Part III asks what an accuracy claim measures and leaves out, how a laboratory weighs flow to draw a trumpet curve, why low rates, syringe changes and a pump moved up or down matter for short-acting drugs, and what happens to each drug when several pumps share one vein.

08 — Accuracy

Two numbers for one pump.

+ The questionWhat does a pump's accuracy claim mean, and what does it leave out?

The mechanism and the system

A supplier's listing for one large-volume pump, the B. Braun Infusomat Space, gives two accuracies for the same machine, both taken from the maker's specification. The first is a "mechanical accuracy" better than 0.5 %. The second is ±5 % "according to IEC/EN 60601-2-24", the international standard for infusion pumps. The two need not conflict, because they describe different things, and the gap between them is the subject of this chapter. Part II showed that a pump counts the motion of its own mechanism and converts that motion into milliliters through assumptions about the tubing or the syringe. The first number describes the motion. The second describes the volume that leaves the set, which is what the patient receives.

Mechanical accuracy is how closely the fingers, the cam or the lead screw move as commanded: a motor told to turn a certain number of steps turns that number, and the parts it drives travel the distance the design says. Modern drives do this very well, which is why the figure is small. System accuracy is how closely the volume delivered matches the volume programmed, with a real set or syringe in place and under stated conditions. Chapter 4 showed that a 2.5 % change in the bore of a pumping segment changes the volume per stroke by about 5 %. A mechanism that moves perfectly can therefore be several percent out once the tubing is a little wider, softer or warmer than the design expected.

Syringe pumps show the same split. Two staff of the Therapeutic Devices Branch of Australia's Therapeutic Goods Administration noted that the figure quoted for a syringe pump is often the accuracy of the plunger's travel. In the same review, undated on the agency's web page, they added that a single-use syringe can bring up to 4 % of further deviation, because the syringe, not the pump, sets how much volume each millimeter of travel pushes out. BD's instructions for its Alaris syringe pump state ±2 % "typical by volume" at 1 mL/h and above, measured with the syringes BD recommends, and warn that accuracy "may be compromised" below 1 mL/h. The claim describes a pump and a named syringe together, never the pump alone.

20 — Two accuracies for one pump
MOTOR STEPS FINGER TRAVEL BORE SET: TUBING BORE VOLUME DELIVERED SYSTEM ACCURACY: ±5 % (IEC 60601-2-24 TEST) MECHANICAL ACCURACY: BETTER THAN 0.5 % a 2.5 % change in bore gives about 5 % in volume (Chapter 4) Maker's figures for one large-volume pump (B. Braun Infusomat Space), via a supplier listing. Schematic.
Fluid and drugPressure and forceReadoutFocal detail
Plate 20 — Mechanical accuracy covers only the motion of the mechanism; system accuracy covers the volume that leaves the set, so the tubing alone can move the second figure tenfold beyond the first.

Even a pump that measures volume makes a claim of the same kind. The Ivenix large-volume pump of Chapter 4 works out the volume in its pumping chamber from air pressure, yet the summary its maker filed for its 2019 clearance states an accuracy of ±5 % from 0.5 to 1,000 mL/h, the same form of claim as a peristaltic pump's. Measuring the chamber removes one assumption, about the size of each stroke, but it leaves the rest of the line, from the chamber's outlet to the vein, exactly as uncertain as before. Whatever the mechanism, the percentage on the label is a statement about the volume that reached the end of a set in a test.

A claim carries its conditions

An accuracy claim is a sentence with conditions attached, and the conditions are as much a part of it as the percentage. ICU Medical's specification for its Plum 360 cassette pump, as reproduced by a distributor, states ±5 % from 1 to 999 mL/h and ±10 % from 0.1 to 0.9 mL/h. The conditions follow: room and fluid at 22 °C, within 5 °C either way; no back pressure; sterile water as the fluid; and a container 30.5 to 61 cm above the pump. Back pressure is the pressure the pump must push against at the far end of the line, from the catheter and the vein (Chapter 2). Head height is the height of the container above the pump, which sets the pressure at its inlet; Chapter 1 measured the head from the container to the vein instead.

Each condition stands for an assumption that a ward will break. A narrow catheter and a vein supply back pressure that the test excluded, drugs are more viscous than water, and Chapter 7 showed how strongly viscosity changes with temperature. A bag hung low or high changes how fully each stroke fills. BD's syringe-pump instructions add their own allowances to the headline figure: a further ±0.5 % across 5 to 40 °C, and ±2 % more at rates above one syringe volume per hour. The plain reading of "±2 %" is therefore longer: ±2 % with this syringe, at these rates and this temperature, with stated allowances outside them.

21 — What one accuracy claim assumes
ACCURACY CLAIM ±5 % at 1 to 999 mL/h ±10 % at 0.1 to 0.9 mL/h TEMPERATURE room and fluid 22 °C ± 5 °C FLUID sterile water BACK PRESSURE 0 mmHg HEAD HEIGHT 30.5 to 61 cm AVERAGING PERIOD not stated FDA's 2014 guidance asks the label to state it Maker's figures: ICU Medical Plum 360 specification, reproduced by a distributor.
Fluid and drugPressure and forceReadoutFocal detail
Plate 21 — An accuracy claim is a percentage plus its conditions; this sheet states temperature, back pressure, fluid and head height but not the period over which the error is averaged.

FDA's guidance on infusion pumps, issued on 2 December 2014 after the initiative of 2010, turns this into a requirement. It asks makers to show accuracy at the limits of the pump's operating range and across expected changes in ambient and fluid temperature, head height, back pressure, atmospheric pressure and viscosity, with the effects "discussed quantitatively in the labeling". It also asks the label to state the period over which accuracy is specified, the effect of rate changes and boluses, and the time to steady state, that is, how long the pump takes to reach the accuracy it claims. The Plum 360 sheet quoted above gives no averaging period at all.

Every percentage has an averaging period

A percentage error needs a period, because it compares two volumes: what the pump delivered in some interval and what it should have delivered in that interval. The averaging period is that interval. Over a long period, short excesses and shortfalls cancel; over a short one they show. A pump that pauses for a minute and then catches up is exact over an hour and badly out over the minute that contained the pause. Both statements are true of the same delivery, and a single percentage chooses one of them without saying which, so the period is part of the claim.

A percentage holds only over its stated period

A claim of ±5 % computed over an hour says nothing directly about any ten minutes inside that hour. A label that gives no period leaves the reader to guess, and the natural guess, that the error is spread evenly through the hour, is the one the mechanisms of Part II give least reason to expect.

The period matters most where the drug acts fastest, the subject of Chapter 10, and at the start of an infusion, before the pump has settled. FDA's request for the time to steady state exists for the second reason. As Chapter 5 showed, a syringe pump takes up slack in its drive and compresses the syringe's stopper before fluid moves at the set rate, and a claim measured after the pump has settled leaves that delay out. The test method of Chapter 9, as secondary accounts describe it, plots the start separately for the same reason and analyzes accuracy only once the pump has run for an hour.

Solved example: what ±5 % allows at 25 mL/h over one hour and over ten minutes
  • Over one hour at 25 mL/h the pump should deliver 25 mL. A tolerance of ±5 % is ±1.25 mL, so any delivery from 23.75 to 26.25 mL meets the claim.
  • Over ten minutes the pump should deliver 25 × 10 ÷ 60 = 4.17 mL. If the claim applied to every ten-minute window, the allowance would be ±5 % of 4.17 mL, or ±0.21 mL, giving 3.96 to 4.38 mL.
  • A claim stated over the hour sets no such limit. In the extreme, the hour's whole allowance of 1.25 mL could fall in one ten-minute window: 4.17 − 1.25 = 2.92 mL, or 4.17 + 1.25 = 5.42 mL.
  • Those volumes are 30 % below and 30 % above the ten-minute target, while the hour as a whole stays within ±5 %.
  • In words, an hourly claim of ±5 % at 25 mL/h is compatible with a ten-minute error of up to ±30 %, which is why the averaging period has to be stated.
22 — Within 5 % over the hour, up to 30 % in ten minutes
−30 −20 −10 0 +10 +20 +30 ERROR (%) ONE HOUR (25 mL) TEN MINUTES (4.17 mL) 23.75 to 26.25 mL 2.92 to 5.42 mL 3.96 to 4.38 mL if the claim held for every window the hour's whole 1.25 mL allowance in one window Guide's arithmetic for a claim of ±5 % at 25 mL/h. Schematic.
ReadoutFocal detail
Plate 22 — A claim of ±5 % over an hour at 25 mL/h limits the hour to 23.75 to 26.25 mL but allows a single ten-minute window to run from 2.92 to 5.42 mL, 30 % either side of its target.

When the claim fails in the field

Two recent recalls show where accuracy claims break. Baxter's Novum IQ large-volume pump, met in Chapter 4, could underinfuse when its rate was more than doubled by a new rate or a bolus. Its Class I correction, begun in July 2025, is the case that opens Chapter 31. A claim measured at steady rates said nothing about the moment the rate changed, which is one of the items FDA asks labels to state.

The second recall shows that a claim belongs to the pump and the set together. In July 2025 BD began a Class I correction because its Alaris Pump Module, with certain compatible sets, might perform outside its published ranges for flow-rate and bolus accuracy, the time to an occlusion alarm and the bolus after an occlusion. FDA's notice linked the deviations to set features such as filters and in-line components. Another Class I recall of BD sets, for low-rate performance worse than previously disclosed, followed in August 2026 (Chapter 26).

Read a specification as a list of conditions

Before a pump enters a formulary or a tender, list what its accuracy claim assumes: the averaging period, the time to steady state, the test fluid, the temperature, back pressure and head height, and the sets or syringes it names. Check those sets against the ones the hospital actually stocks, including filters and extensions added at the bedside. A maker, in India or anywhere else, should write each item of FDA's 2014 list into the label, because a claim without its period and conditions cannot be checked against anything.

+ What this chapter established
  • Mechanical accuracy describes how exactly a pump's mechanism moves, while system accuracy describes the volume delivered through a real set or syringe, and in one maker's figures the two differ tenfold.
  • An accuracy claim holds only under its conditions, among them temperature, back pressure, fluid, head height and the named set, and FDA asks makers to state the effect of each in the labeling.
  • A percentage needs an averaging period: ±5 % over an hour at 25 mL/h allows an error of up to ±30 % in a single ten-minute window.
  • Recent Class I recalls arose at rate changes, which a steady-rate claim does not describe, and from sets that performed outside the published claims.
Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.