§ Glossary
Glossary.
Terms are defined as they are used in this guide.
- 16S ribosomal RNA gene
- A gene that every bacterium carries, about 1,500 bases long, with parts nearly identical across bacteria and nine hypervariable regions that differ between lineages; amplified and sequenced, it shows which bacteria a sample holds.
- 180-day supplement
- The PMA supplement the FDA requires for a significant change in design, specification or software of an approved device, which it reviews in depth and approves before the change is made.
- 30-day notice
- The route for a change to the manufacturing methods of an approved device: the maker may make the change 30 days after the FDA receives the notice, unless the FDA finds the notice inadequate.
- 510(k) (premarket notification)
- A submission showing that a device is substantially equivalent to a legally marketed predicate not subject to premarket approval; the FDA's decision on it is a clearance.
- Abundance
- The number of copies of a molecule in a sample.
- Adapter
- A short synthetic DNA sequence attached to the ends of each fragment so that the sequencer can grip and read it.
- Advisory panel (FDA)
- A committee of outside experts that discusses an application in public and votes; its vote advises the FDA without binding it.
- Affinity reagent
- A molecule, usually an antibody or aptamer, that binds a target by shape. Its specificity must be shown for each target.
- Algorithm Change Protocol
- In CDSCO's software guidance of July 2026, an optional protocol for changes to medical device software, which a maker may devise where applicable as part of the software change management documented in its risk management plan. It has five plans: data management, performance evaluation and monitoring, retraining where it applies, software updates and rollback.
- Alignment
- Placing each sequencing read at the position in a reference genome that it best matches.
- Analyte
- The substance a test is meant to measure.
- Analytical performance (IVDR)
- The ability of a device to correctly detect or measure a particular analyte.
- Analytical sensitivity
- An instrument's or assay's ability to detect a weak signal; a different quantity from diagnostic sensitivity, the share of true cases that a test detects.
- Analytical validation
- The studies that show a test measures its analyte correctly: as applicable, accuracy, precision, analytical sensitivity, analytical specificity including interfering substances, the reportable range and the reference intervals.
- Analytical validity
- Whether a test measures its analyte accurately and reproducibly in the specimens it will receive.
- Aptamer
- A short single strand of DNA or RNA, often chemically modified, selected to fold into a shape that binds a particular target.
- ATAC-seq
- A method that maps open chromatin by letting an adapter-inserting enzyme act only where DNA is accessible.
- Authentic standard
- A pure sample of a suspected compound, run under the same conditions as the sample to show that the two match on at least two independent properties.
- Base calling
- The conversion of a sequencer's raw signal, light or current, into letters with quality scores.
- Batch
- A set of samples processed together, which leaves a small systematic shift on every measurement.
- Batch effect
- The systematic shift that laboratory conditions, reagent lots and personnel leave across samples processed together; across thousands of features it can be larger than the biology, and a balanced design is its control.
- Becoming aware (US reporting)
- The start of an FDA reporting clock. For a 30-day report, and for a 5-day report the FDA has requested, a maker becomes aware when any employee learns of information that reasonably suggests a report is required. For a 5-day report on an event that needs remedial action, it becomes aware when an employee in a particular role learns that the action is needed: FDA's guidance names a manager or supervisor of staff with regulatory, scientific or technical duties, and its wording leaves open whether an employee whose duties concern the collection and reporting of adverse events also counts.
- Biomarker
- A molecule whose level indicates a state of the body, such as a disease or a response to treatment.
- BIOSECURE Act
- A US law enacted on 18 December 2025 within the defense authorization act for 2026, whose prohibitions on biotechnology companies of concern take effect only after the federal acquisition rules are revised.
- Bisulfite conversion
- A chemical treatment that turns unmethylated cytosine into uracil, read as T, while methylated cytosine stays C.
- Bottom-up proteomics
- Proteomics in which proteins are first digested, usually with trypsin, into peptides that are then measured.
- Breakthrough Device designation
- A designation the FDA grants, under a voluntary program, to some devices that provide more effective diagnosis or treatment of a life-threatening or irreversibly debilitating disease and meet at least one further criterion, such as having no approved or cleared alternative. It brings interaction with FDA experts and prioritized review, and the device must still meet the FDA's standards; a designation is not a clearance, a grant of a De Novo request or an approval.
- cDNA
- Complementary DNA: a DNA copy of RNA made by reverse transcriptase, so that RNA can be amplified and sequenced.
- CDSCO
- The Central Drugs Standard Control Organisation, India's national regulator for drugs and devices; under the Medical Devices Rules it is the Central Licensing Authority and publishes the IVD classification lists.
- CE marking
- The marking by which a maker indicates that a device conforms to the applicable requirements of the IVDR. It is affixed before the device is placed on the market, once the conformity assessment its class requires is complete, and the notified body's identification number follows it where one took part.
- Cell barcode
- A DNA sequence shared by every molecule captured from one cell, used to sort pooled reads back into cells.
- Cell-free DNA (cfDNA)
- Short fragments of DNA in plasma, released when cells die and mostly about 166 base pairs long, about the length of DNA on one nucleosome.
- Under India's Medical Devices Rules, CDSCO in its role of licensing the manufacture of class C and D devices and the import of devices of every class.
- CFIUS
- The Committee on Foreign Investment in the United States, which reviews foreign investment in US businesses.
- Change control
- The process through which every change to a released design passes before it is implemented: the change is reviewed for its effects, and the studies it touches are run again as the review requires. ISO 13485 requires design and development changes to be controlled before they are implemented (clause 7.3.9).
- Change review (design change)
- The step of change control that judges what a change affects before it is implemented; under clause 7.3.9 of ISO 13485 it includes evaluating the change's effect on the device's constituent parts, and the trace of the design shows which studies the change requires to be run again.
- Chromatin accessibility
- How far a stretch of DNA is free of nucleosomes and other proteins, and so reachable by the transcription machinery.
- Circulating tumor DNA (ctDNA)
- The fraction of cell-free DNA released by a tumor.
- Clearance
- The FDA's decision on a 510(k). A cleared device is not described as approved.
- CLIA
- The US law for clinical laboratories, applied through a program that CMS runs. It certifies laboratories, sorts tests by complexity and requires a laboratory to establish the analytical performance of a test it builds or modifies; it does not address clinical validity.
- CLIA complexity
- The category of a test under CLIA: waived, moderate or high. FDA scores a test system on seven criteria from 1 to 3; 12 or less is moderate, and a test system not yet categorized defaults to high.
- Clinical performance (IVDR)
- The ability of a device to yield results that are correlated with a particular clinical condition.
- Clinical performance evaluation
- Under India's Medical Devices Rules, a study of a new IVD's performance on specimens from human participants, run with CDSCO's permission (applied for on Form MD-24, granted on Form MD-25).
- Clinical utility
- Whether acting on a test result improves outcomes compared with not testing.
- Clinical validity
- Whether a test result corresponds to the condition of interest in the intended population.
- Cluster (cell data)
- A group of cells with similar measurements, formed by an algorithm; a statistical grouping, so calling it a cell type needs independent evidence.
- Co-isolates
- Materials isolated along with extracellular vesicles, such as lipoproteins, platelets or soluble proteins; reporting how far EVs were enriched over them is how a study states its preparation's purity.
- Coefficient of variation (CV)
- The spread of repeated measurements as a percentage of their mean, measured on a pooled quality-control sample run at intervals through a batch.
- Companion diagnostic
- A test essential for the safe and effective use of a particular drug. The IVDR defines it as a device essential for the safe and effective use of a corresponding medicinal product.
- Compensation
- The mathematical correction of fluorescence spillover between detectors in flow cytometry.
- A member state's regulator for devices; under the IVDR it receives the maker's reports of serious incidents and evaluates the risks.
- Complaint record (QMSR)
- The maker's record of a complaint under 21 CFR 820.35: the device's name, the date received, the unique device identifier or other identifiers, the complainant's contact details, the nature and details of the complaint, any corrective action and reply, and the records of review and investigation, with a justification when no investigation was done.
- Conformity assessment
- In the IVDR's words, "the process demonstrating whether the requirements of this Regulation relating to a device have been fulfilled"; a notified body takes part for every class but non-sterile class A, whose makers declare conformity themselves.
- Confounded
- Of two factors that cannot be separated in the data, such as treatment and processing day when every treated sample was processed on one day and every control on another.
- Control
- A known quantity passed through every step alongside the samples, so that anything lost or distorted shows up as a difference between what went in and what came out.
- Copying
- The first of the two properties that make nucleic acids easy to measure: enzymes copy DNA, and RNA once it has been written into DNA, so a few molecules can be multiplied until an instrument detects them.
- Correction (21 CFR 806)
- The repair, modification, adjustment, relabeling, destruction or inspection of a device, including patient monitoring, without moving it from its point of use. A maker or importer reports one to the FDA when it is made to reduce a risk to health or to remedy a violation of the law that may present one.
- Corrective action
- In ISO 13485 (clause 8.5.2), action on a nonconformity so that it does not recur, proportionate to the effects of the nonconformity. The IVDR defines it more widely, as action to eliminate the cause of a potential or actual non-conformity or other undesirable situation.
- Count table
- The output of RNA-seq: genes down one side, samples across the top, and in each cell the number of sequenced fragments, not molecules.
- Coverage
- The number of sequencing reads overlapping a position, or its average across a region or genome, written as 30× and so on.
- CpG
- A cytosine followed by a guanine in the DNA sequence: the usual site of DNA methylation in human cells.
- Critical quality attribute
- A property of a product that should stay within set limits to ensure its quality, such as the glycan pattern of a therapeutic antibody.
- Cross-reactivity
- Accidental binding, such as a detection antibody meeting the wrong capture antibody in a mixture, which produces false signal and limits how many antibody pairs a conventional multiplexed immunoassay can mix.
- Cytomics
- The measurement of many properties of individual cells across very many cells.
- Data-dependent acquisition (DDA)
- A mass spectrometry mode that selects the most intense ions in each MS1 scan for fragmentation; the choice is partly random, so a peptide measured in one run may be skipped in the next.
- Data-independent acquisition (DIA)
- A mass spectrometry mode that fragments all ions within successive mass windows rather than selecting the most intense.
- De Novo request
- A route for classifying a novel device that has no predicate, where general controls, alone or with special controls, give reasonable assurance of safety and effectiveness; the FDA grants the request.
- Demultiplexing
- Sorting the reads of a pooled sequencing run back to their samples by their indexes.
- Design and development file
- The record kept for each device type or family that includes or references the evidence of conformity to the design requirements.
- Design controls
- The design and development requirements of ISO 13485 clause 7.3: planning, inputs, outputs, review, verification, validation, transfer, change control and the design and development file.
- Design input
- A requirement the design must meet, complete, unambiguous, able to be verified or validated, and taking account of risk management outputs.
- Design output
- A protocol, software, specification or label that embodies the design; it must meet the inputs and contain or reference acceptance criteria.
- Detection limit
- The smallest amount or signal an instrument or assay can distinguish reliably from background.
- Device class (US)
- One of three classes in 21 CFR 860.3: class I under general controls only, class II under special controls, and class III, for which premarket approval is required.
- Differential expression
- The comparison of each gene's level between conditions; the normalization methods used for it estimate a scaling factor from the genes that did not change.
- Differential result
- A table that compares two groups of samples feature by feature, giving for each feature a fold change, a p-value, the p-value adjusted for the number of features tested, and the feature's abundance.
- DNA methylation
- A small methyl group attached to cytosine, in most cell types almost always at a CpG site, and copied through cell division.
- Doublet
- A droplet that captured two cells, which looks in the data like a hybrid cell type that does not exist.
- Duplex mode (sequencing by expansion)
- The sequencing-by-expansion mode in which the two strands of each molecule are linked during library preparation, so that their reads can be combined for accuracy.
- Dynamic range
- The ratio between the largest and smallest quantities present in a sample, or measurable by an instrument in one run.
- Electrospray
- The usual way of ionizing proteins and peptides for mass spectrometry: liquid from a chromatography column is sprayed from a fine needle at high voltage, the charged droplets shrink as the solvent evaporates, and the dissolved molecules are left carrying extra protons.
- Epigenome
- The chemical marks on DNA and on the proteins that package it, which record which genes a cell keeps open for use, often through many divisions.
- EUDAMED
- The EU's database for medical devices and IVDs; four of its modules, among them the one for notified bodies and certificates, became mandatory on 28 May 2026.
- Event
- In cytometry, a pulse that rises above the instrument's threshold, usually one cell, recorded as one row of measurements.
- Exome
- The protein-coding regions of the genome, less than 1.5 % of it, which can be sequenced alone.
- Exon
- A segment of a gene that is kept in a messenger RNA; a cell can join different subsets of a gene's exons into different messenger RNAs.
- Expected erroneous result (IVDR)
- A wrong result of the kind that the product information clearly documents and quantifies. It is left out of serious-incident reports and counted instead in trend reporting.
- Expression (of a gene)
- A gene is expressed when it is transcribed, and its expression is usually measured as the amount of its RNA, or of the protein made from it.
- Expression signature
- A panel, a fixed set of genes, measured in a tumor to estimate the risk that the cancer will return.
- Extracellular vesicle (EV)
- A particle enclosed by a lipid bilayer, released by cells, carrying protein, RNA and lipid cargo.
- Failure modes and effects analysis (FMEA)
- A risk analysis that starts from the parts and steps of a design or process, asks how each might fail and follows each failure to its effects; IEC 60812 describes it.
- False discovery rate (FDR)
- The expected share of false findings among all results declared significant.
- FASTQ
- The common file format for sequencing reads, holding each read's sequence with a quality score for every base.
- Fault tree analysis
- A risk analysis that starts from one undesired event, such as a false-positive report, and works back through the events and failure modes that could combine to produce it; IEC 61025 describes it.
- Feature
- Anything a study measures, such as a gene, a protein or a metabolite. In untargeted metabolomics, a signal defined by a mass-to-charge ratio and a retention time, which may or may not be a distinct compound.
- Field safety corrective action (FSCA)
- Under the IVDR, a corrective action that a maker takes for technical or medical reasons to prevent or reduce the risk of a serious incident; it is reported before it is undertaken, except in urgent cases.
- Field safety notice
- Under the IVDR, the maker's communication to users or customers about a field safety corrective action; it must explain the risk clearly without understating it, state every action users must take, and be publicly accessible.
- Fold change
- The ratio of a quantity between two conditions, usually reported on a log₂ scale.
- Forward scatter
- Light scattered by a cell at small forward angles, which grows with the cell's size.
- Foundation model (single-cell)
- A model trained on many single-cell datasets at once.
- Gate
- A boundary drawn on a two-dimensional plot of one measured parameter against another to select a population of cells or particles.
- Gating
- Selecting a population of cells or particles by drawing boundaries on plots of their measured parameters.
- GC-MS
- Gas chromatography with mass spectrometry, for molecules that are volatile or can be made volatile by chemical derivatization; its reproducible fragmentation supports large libraries of reference spectra.
- General controls
- The basic provisions of the US Medical Device Amendments of 28 May 1976, among them those on adulteration and misbranding, registration and listing, premarket notification, records and reports, and good manufacturing practice; they apply to devices in all three classes unless a device is exempted, and class I devices are subject to them alone.
- Genome
- The complete DNA sequence a person inherits, the same in nearly every cell of the body.
- Glycomics
- The measurement of glycans, the branched sugar chains attached to proteins and lipids.
- Half-life
- The time in which half of the molecules present are destroyed.
- Harm
- In ISO 14971, injury or damage to the health of people, or damage to property or the environment.
- Hazard
- In ISO 14971, a potential source of harm.
- Hazardous situation
- In ISO 14971, as Health Canada and FDA reproduce it, a circumstance in which people, property or the environment are exposed to one or more hazards; this guide reads it, for a test, as a wrong result once it reaches a decision about a patient.
- Health technology assessment
- An evaluation of a medicine, device or test that informs a health system's decisions on paying for it. In the EU, Regulation (EU) 2021/2282 has applied since January 2025: it provides for joint clinical assessments, for which tests of the highest risk class may be selected, and leaves pricing and reimbursement to each member state.
- Hemolysis
- The bursting of red cells in a blood specimen, which releases their contents into the plasma.
- Heterozygous
- Of a variant: present on only one of the two parental copies of a chromosome, so that it appears in roughly half the reads at its position.
- Histone modification
- A chemical mark on the histones, the proteins that make up each nucleosome, that recruits or repels the transcription machinery.
- Homozygous
- Of a variant: present on both parental copies of a chromosome, so that it appears in nearly all the reads at its position.
- Hydrodynamic focusing
- Narrowing a sample stream inside a faster sheath flow so that cells pass a detector in single file.
- Hypervariable regions
- The nine parts of the 16S ribosomal RNA gene that differ between bacterial lineages.
- In vitro diagnostic product (US)
- In FDA's regulation, the reagents, instruments and systems intended for use in the diagnosis of disease or other conditions; the definition turns on intended use.
- In-house device (IVDR)
- A test that a health institution makes and uses itself without transferring it to another legal entity. Under Article 5(5) of the IVDR it is exempt from the regulation's requirements, except the relevant general safety and performance requirements of Annex I, when the article's conditions are met: eight of the nine apply as of October 2026, among them an appropriate quality management system and a laboratory compliant with EN ISO 15189 or with national provisions, and the ninth, a justification that no equivalent device on the market meets the patient group's needs at the appropriate level of performance, applies from 31 December 2030.
- Incident (IVDR)
- Any malfunction or deterioration in the characteristics or performance of a device on the market, including use error due to ergonomic features and inadequate information from the maker, and any harm that follows from a medical decision, or an action taken or not taken, on the basis of the device's result.
- Index hopping
- The appearance of a read carrying the index of a library it did not come from, so that it is assigned to the wrong sample.
- Information value
- A value NIST characterizes for a reference material without certifying it; it cannot be used to establish metrological traceability.
- Instruction set
- The genome seen as what constrains what a cell can become, as against its state.
- Interface
- The place where one link of a test hands its output to the next, often as a file in an agreed format.
- Internal standard
- A known amount of an isotope-labeled copy of the measured molecule added to a sample, which turns signal into an absolute amount.
- Investigational Use Only (IUO)
- The US labeling statement for a product tested before full commercial marketing: "For Investigational Use Only. The performance characteristics of this product have not been established."
- ISO 15189
- The international standard for quality and competence in medical laboratories, in force in its fourth edition of 2022; a basis for NABL's accreditation in India and a condition of the in-house exemption in the EU.
- Isotope labels that weigh the same, so labeled samples are measured together and told apart by reporter ions released on fragmentation.
- Isomers
- Molecules with the same atoms, and therefore the same mass, arranged differently.
- IVDR
- Regulation (EU) 2017/746, the European Union's In Vitro Diagnostic Regulation, which sorts tests into classes A to D by the rules of its Annex VIII.
- Laboratory-developed test (LDT)
- A test that a single laboratory designs and runs itself rather than one sold as a kit; in the United States it falls under CLIA.
- Layer
- One whole set of molecules of one kind that omics measures, such as the genome, the transcriptome or the proteome.
- LC-MS
- Liquid chromatography coupled to mass spectrometry, which covers the widest range of metabolites.
- Legacy device (IVDR transition)
- A test placed on the market under the old Directive 98/79/EC, which may stay on the market until a date set by class in Regulation (EU) 2024/1860, provided its maker met the regulation's conditions.
- Library
- The collection of adapter-tagged fragments prepared from a sample for sequencing.
- Library preparation
- The laboratory steps that turn extracted DNA, or cDNA, into a library: fragmenting it and fitting each fragment with adapters at both ends.
- Lipidomics
- The measurement of the thousands of lipid species in membranes, stores and signaling pathways, by mass spectrometry.
- Liquid biopsy
- Testing the material that tissues release into the blood, such as cell-free DNA and extracellular vesicles, as a sample of tissues otherwise reachable only by a tissue biopsy.
- Locked
- Fixed before validation, so that what runs on patients' samples is what was tested: for a pipeline, every tool, version, parameter and reference; for a signature, every step of its processing and its decision threshold.
- Log₂ fold change
- A fold change counted in doublings: 1 is a doubling, −1 a halving, and every further unit another factor of two.
- Long read
- A sequencing read from a single native molecule, typically thousands to hundreds of thousands of bases long.
- Low-biomass sample
- A sample with little microbial DNA of its own, such as blood, spinal fluid or a tissue biopsy, in which traces of DNA from reagents can dominate the result.
- Major change (India)
- A change listed as major in the Sixth Schedule of India's Medical Devices Rules; the license holder needs the licensing authority's prior approval, and the change is deemed approved if the authority does not answer within 45 days.
- Malfunction (US reporting)
- The failure of a device to meet its performance specifications or otherwise perform as intended, where the performance specifications include all claims made in the device's labeling. It is reportable when it would be likely to cause or contribute to a death or serious injury if it recurred, and FDA treats the failure of a diagnostic device in the same way.
- Marker
- A molecule, such as a protein, whose level distinguishes a cell type or state.
- Mass cytometry
- Cytometry that labels antibodies with metal isotopes and reads them by mass instead of fluorescence.
- Mass-to-charge ratio (m/z)
- The quantity a mass spectrometer actually measures: an ion's mass divided by its charge.
- Materiovigilance Programme of India (MvPI)
- India's national program that collects reports of adverse events from medical devices. By a compliance-software firm's account of June 2024, a CDSCO circular directs reports of suspected unexpected serious adverse events to it.
- Medical Device Reporting (21 CFR 803)
- The US regulation under which makers and importers report deaths, serious injuries and malfunctions, and user facilities report deaths and serious injuries, within set periods after becoming aware of them: 30 calendar days for a maker, or 5 work days when the FDA designates the event or it requires remedial action.
- Medical Devices Rules, 2017
- The rules under which India regulates medical devices, in vitro diagnostics included, with four risk classes, A to D, and licenses for manufacture and import.
- Metabolites
- The small molecules of life, typically under about 1,500 daltons: sugars, amino acids, fatty acids and other lipids, nucleotides, vitamins, hormones and the intermediates of every pathway that makes or breaks them.
- Metabolome
- The full set of small molecules in a cell or sample, such as sugars, amino acids and fats, recording the chemistry that proteins carry out.
- Metagenomics
- Sequencing the mixed DNA of a microbial community without culturing its members.
- Metatranscriptomics
- Sequencing the RNA of a microbial community, which shows which genes the community is actively using.
- Methylation level
- The fraction of reads at a site that carry a methyl mark: a population average, not a property of one cell.
- MISEV2023
- The International Society for Extracellular Vesicles' guidelines on nomenclature, isolation, characterization and reporting.
- Mock community
- A sample of microbes of known composition processed alongside real samples as a positive control.
- Molecular residual disease
- The traces of cancer that remain after treatment and can be detected only by molecular tests.
- MS1 and MS2
- The two spectra of tandem mass spectrometry: MS1 records the ions present at one moment, and MS2 the masses of the fragments of selected ions broken by collision with gas.
- Multiple testing
- The inflation of chance findings when many hypotheses are tested at once.
- Multiplexing
- Measuring many analytes, such as many proteins, in one well or tube.
- mzML
- The Proteomics Standards Initiative's open file format for raw mass spectrometer output.
- NABL
- The National Accreditation Board for Testing and Calibration Laboratories, which accredits Indian medical laboratories against ISO 15189:2022 through its criteria NABL 112A.
- Nanoparticle tracking analysis (NTA)
- A method that sizes and counts particles by following the scattered or fluorescent light of each as it drifts.
- Negative percent agreement (NPA)
- The share of a reference's known non-variant positions that a test leaves uncalled.
- Nominally significant
- Passing the conventional threshold of 0.05 before any allowance is made for the number of endpoints tested.
- Non-notified device (India)
- A device that had not been notified under the Drugs and Cosmetics Act before the Medical Devices Rules; such IVDs came under licensing on 1 October 2022 for classes A and B and on 1 October 2023 for classes C and D.
- Nonconformity
- A failure to meet a requirement, such as a result outside a test's specification.
- Notified body
- A conformity assessment body designated under the IVDR. It assesses tests in classes B, C and D before they are placed on the market, and for a sterile class A test only the sterile conditions.
- Nuclear magnetic resonance (NMR)
- A method that reads the spin of atomic nuclei: in a strong magnetic field certain nuclei absorb radio waves at frequencies set by their chemical surroundings, so a spectrum reports the atomic environments in a molecule. It is the main exception to the five properties most instruments read.
- Nucleosome
- A spool of histone proteins around which DNA is wound; a short stretch of linker DNA lies between one spool and the next.
- Off-the-shelf software
- FDA's term for a generally available software component whose complete life cycle the device maker cannot control.
- Omics
- The measurement of a complete set of molecules of one kind, a whole layer, in a single experiment.
- Outsourced process
- Part of a maker's work that an outside party carries out for it. ISO 13485 requires its controls to be proportionate to the risk involved and to the outside party's ability to meet requirements.
- Overall residual risk
- All the residual risks of a device taken together, weighed against its benefit.
- p-value
- How often chance alone would produce a difference at least as large as the one observed, if there were truly no difference.
- Pairing
- The second of the two properties: a strand of DNA or RNA binds a partner strand letter by letter, so a probe for any sequence can be designed from the sequence alone.
- Panel
- A set of probes, assays, antibodies or genes chosen to be measured together; the expression signatures of Chapter 21 measure a fixed set of genes.
- Patterned flow cell
- A sequencing flow cell whose surface Illumina, its maker, describes as nanowells at fixed, evenly spaced locations, in which clusters grow by a method it calls exclusion amplification.
- PCR-free library
- A sequencing library prepared without amplification.
- Peptide
- A short chain of amino acids; in bottom-up proteomics, a piece of a protein cut by trypsin, typically roughly 7 to 30 amino acids long.
- Performance evaluation
- Under the IVDR, the assessment and analysis of data that establishes or verifies a test's scientific validity, analytical performance and, where applicable, clinical performance.
- Periodic safety update report (PSUR)
- Under the IVDR, the post-market surveillance report that a maker prepares for a class C or D device.
- Pharmacogenomics
- Using a person's genotype to choose a drug or its dose.
- Phasing (sequencing)
- Loss of synchrony in a sequencing cluster when some strands miss an addition and fall behind the rest; with prephasing, it degrades base quality along a read.
- Phasing (variants)
- Showing which variants sit together on the same parental copy of a chromosome, which a long read can do.
- Phenotype
- The observable outcome of a cell or an organism.
- Phosphoproteomics
- The enrichment and measurement of phosphorylated peptides, to map which proteins in signaling pathways carry a phosphate.
- Positive percent agreement (PPA)
- The share of a reference's known variants that a test calls.
- Positive predictive value
- The share of positive test results that are true positives. It depends strongly on how common the condition is.
- Post-market performance follow-up (PMPF)
- Under the IVDR, the follow-up that confirms, over a device's expected lifetime, that its safety, performance and clinical evidence remain valid and its balance of benefit and risk acceptable; MDCG guidance calls it a continuous process that updates the performance evaluation.
- Post-market surveillance
- A maker's systematic collection and analysis of information from its devices on the market. Under the IVDR it is a planned system whose plan is part of the technical documentation, with a surveillance report for classes A and B and a periodic safety update report for classes C and D.
- Pre-analytical phase
- Everything that happens to a specimen before the analytical test: the tube, transport, time to centrifugation, processing and storage. ISO's standards call it pre-examination.
- Predetermined change control plan
- A plan, allowed under US law since 2022, that sets out in advance what may change in a device after release, such as a model in its software, and how each change will be verified.
- Predicate
- A legally marketed device, not subject to premarket approval, to which a 510(k) shows substantial equivalence.
- Premarket approval (PMA)
- The FDA's review of the safety and effectiveness of a class III device; the FDA approves the application, and changes to an approved device can come back to it as supplements.
- Prephasing
- Loss of synchrony in a sequencing cluster when some strands add two bases in one cycle and run ahead of the rest.
- Preventive action
- In ISO 13485 (clause 8.5.3), action on a potential problem before it occurs, proportionate to the effects of the potential problem.
- Primer
- A short piece of DNA that pairs with the sequence on one side of a target and gives a polymerase its starting point; PCR uses two, one on either side.
- Principal component analysis (PCA)
- A method that finds the directions of greatest variation among samples, used to spot batches and outliers.
- Protein group
- The set of proteins that the same identified peptides could have come from, reported in place of a single protein when peptides are shared.
- Protein inference
- Deducing which proteins were present from the peptides identified; ambiguous when peptides are shared.
- Proteoform
- A specific molecular form of a protein, including its sequence variant, splicing and modifications.
- Proteome
- The full set of proteins in a cell or sample, recording which proteins have been made and in what modified forms.
- Proximity extension assay
- An affinity method in which two DNA-tagged antibodies must bind the same protein for their tags to form a countable barcode.
- Quality Management System Regulation (QMSR)
- FDA's rule for device makers' quality systems, 21 CFR Part 820, in force since 2 February 2026; it incorporates ISO 13485:2016 by reference.
- Quality score (Q)
- A logarithmic estimate of a base call's error probability: Q30 is one error in 1,000.
- Read
- The sequence that an instrument reports for one piece of DNA.
- Recall (India)
- Under India's Medical Devices Rules, any action by the maker, its authorized agent or a supplier to remove a device from the market or retrieve it from those it was supplied to, because it is hazardous to health, fails to conform to a claim about its quality, safety or efficacy, or does not meet the Act and the Rules.
- Reclassification order
- An FDA order that moves a type of device from one class to another, first proposed and then made final; a proposed order changes nothing until it is final.
- Reference genome
- An agreed genome sequence used as the coordinate system for aligning reads.
- Reference material
- A material whose answer is known in advance. It tests the links of a test's chain from the point where it enters, and only where its values are known, because the difference between the known answer and the report measures the error of those links.
- Relative abundance
- An organism's share of the reads in a community sample; the shares sum to 100 %, so a rise in one organism lowers the shares of all the others.
- Removal (21 CFR 806)
- Moving a device from its point of use to another location.
- Research Use Only (RUO)
- In the US, the labeling statement "For Research Use Only. Not for use in diagnostic procedures." for a product in the laboratory research phase. In India, CDSCO treats research-use products as outside the Medical Devices Rules provided they are not intended for diagnosis.
- Residual risk
- The risk remaining after risk control measures have been implemented (ISO 14971, as FDA reproduces it), judged again against the criteria of the risk management plan.
- Retention time
- The time a molecule takes to emerge from a chromatography column, used as an identity coordinate.
- Reverse transcriptase
- An enzyme that copies RNA into DNA.
- Risk (ISO 14971)
- The combination of the probability of occurrence of harm and the severity of that harm.
- Risk management plan
- The document that sets out the scope, responsibilities and criteria of risk management, so that the criteria for accepting a risk are fixed before any risk is estimated.
- RNA-seq
- RNA sequencing: measuring the levels of RNA molecules by converting them to cDNA, sequencing it and counting the reads that land on each gene.
- SAM and BAM
- File formats for reads aligned to a reference: SAM is tab-delimited text, BAM the same data compressed.
- Sample index
- A short known sequence in a library's adapters that lets many libraries share one flow cell and be sorted back by sample.
- Sandwich immunoassay
- An assay, often called an ELISA, in which one antibody fixed to a surface captures a protein and a second antibody, against a different part of it, binds on top and carries an enzyme that amplifies the signal.
- Scientific validity (IVDR)
- The association of an analyte with a clinical condition or a physiological state.
- Segmentation
- The step in spatial methods that decides which molecules belong to which cell, usually by drawing cell boundaries from a nuclear or membrane stain.
- Sensitivity
- The share of true cases that a test detects, such as the share of people with a condition whom the test calls positive; also called diagnostic sensitivity, to set it apart from analytical sensitivity.
- Sequencing depth (RNA-seq)
- The total number of reads a sample receives; genome laboratories use depth for the coverage of each position.
- Serious incident (IVDR)
- An incident that directly or indirectly led, might have led or might lead to a death, a temporary or permanent serious deterioration in a person's state of health, or a serious public health threat. The maker reports it within 15 days of becoming aware, 10 days for a death or an unanticipated serious deterioration, and 2 days for a serious public health threat.
- Service laboratory
- A company laboratory that sequences or measures samples for clients and returns data or reports, rather than selling the instruments or kits.
- Shotgun metagenomics
- Sequencing all the DNA in a mixed sample, without amplifying a marker gene, so that reads are assigned to species and strains and the community's genes are cataloged.
- Side scatter
- Light scattered by a cell to the side, which grows with its internal complexity, such as its granules and the shape of its nucleus.
- Signature
- A pattern across many measurements that separates one group of people or samples from another.
- Simplex mode (sequencing by expansion)
- The sequencing-by-expansion mode that Roche describes as built for throughput, for which it states reads of up to 1,500 bases under suitable sample and library conditions.
- Single-nucleotide polymorphism (SNP)
- A position in the genome at which people commonly carry one of two different letters; a panel of them read by two independent assays exposes a sample swap.
- Software of unknown provenance (SOUP)
- In IEC 62304, software already developed and generally available but not developed for incorporation into the device, or software developed earlier for which adequate records of its development are not available.
- Sparse (single-cell data)
- Of data in which most genes read zero in any given cell, because only a fraction of each cell's RNA is captured and converted.
- Spatial methods
- Methods that keep each measurement's position in a tissue, either by barcoded capture arrays or by imaging molecules in place.
- Special controls
- Requirements the FDA writes for one type of class II device, in addition to the general controls.
- Specificity
- The share of samples without the target, or of people without the condition, that a test correctly reports as negative.
- Spectral cytometry
- Cytometry that records each cell's full emission spectrum and unmixes the dyes computationally.
- Spike-in
- A known quantity of a reference organism, molecule or DNA added to every sample, to test and calibrate the measurement chain or to turn relative amounts into absolute ones.
- Spillover
- The light from one fluorescent dye caught by a detector meant for another, because each dye emits across a broad band of wavelengths.
- State (of a cell)
- What a cell is at a given moment: which of its genome's instructions it is using, how intensely and with what result.
- Under India's Medical Devices Rules, the authority of an Indian state that licenses the manufacture of class A and B devices made there.
- Structural variant
- A genomic change of more than about 50 base pairs: a deletion, duplication, inversion or insertion.
- Supplier
- A company from which a device maker obtains a part, material, software component or service that goes into the device. Under ISO 13485, which the QMSR incorporates, the criteria for evaluating and selecting suppliers are proportionate to the risk associated with the device.
- Suspected unexpected serious adverse event (India)
- The event that a license holder under India's Medical Devices Rules reports to its licensing authority, with any action taken, including a recall, within 15 days of learning of it (Rule 26(ii)).
- Targeted metabolomics
- The measurement of a predefined set of metabolites, each with an isotope-labeled internal standard, reported as absolute concentrations.
- Technical positive predictive value (TPPV)
- The share of a test's variant calls that a reference confirms.
- Test (as a chain)
- The whole chain from the patient's specimen to a signed report: specimen, pre-analytical handling, library or assay preparation, instrument and primary analysis, pipeline, interpretation and report.
- Third-party review (510(k))
- A route in which an organization accredited by the FDA reviews a 510(k) submission before the FDA makes the decision.
- Top-down proteomics
- Mass spectrometry that weighs and fragments intact proteins, measuring proteoforms whole, at lower throughput than bottom-up work.
- Traceability
- The links from each requirement to the outputs that meet it and the records that prove it.
- Trajectory
- An ordering of single cells caught at different stages of a continuous process, such as differentiation.
- Transcript length
- The length of an RNA molecule, which affects its count, because a longer transcript is cut into more fragments and collects more reads at the same number of molecules.
- Transcriptome
- The complete set of RNA molecules in a cell or sample at a given time.
- Trend reporting (IVDR)
- The maker's watch, under Article 83, over non-serious incidents and expected erroneous results for a statistically significant increase in their frequency or severity; MDCG guidance asks the maker to define the methods, thresholds and observation period in its post-market surveillance plan.
- Tumor genome profiling
- Reading tens to hundreds of genes in a biopsy to find mutations that match a targeted drug.
- UKCA marking
- The marking under which a device is sold in Great Britain.
- Unique dual indexes
- A pair of indexes, one at each end of a library, neither shared with any other library in the pool, so that a read whose indexes disagree can be set aside.
- Unique molecular identifier (UMI)
- A random tag attached to each captured molecule so that amplification copies are counted once.
- Untargeted (metabolomics)
- Of a run that records whatever ionizes rather than a chosen list of compounds, detecting features.
- User facility (US reporting)
- In FDA's reporting rules, a hospital, ambulatory surgical facility, nursing home or outpatient diagnostic or treatment facility; it reports device-related deaths to the FDA and the maker, and serious injuries to the maker, within 10 work days.
- Validation (design)
- Evidence, on representative product and before release, that the finished design meets its users' needs and intended use.
- Variant calling
- Identifying positions where the reads consistently differ from the reference, with a confidence for each call.
- VCF (variant call format)
- The file format for variant calls, with one data line per genomic position and one column per sample.
- Verification (design)
- Evidence that each design output meets its design input.
- Volcano plot
- A plot of every feature by its log₂ fold change across the page and by its significance, in tenfold steps of the p-value, up the page.
- WHO prequalification (IVDs)
- WHO's assessment of in vitro diagnostics for listed analytes. Its eligibility tables, last updated on 27 September 2026, list what may be assessed, and their only sequencing entry is targeted sequencing for drug resistance in tuberculosis.
- Work day (US reporting)
- In FDA's rules for medical device reporting, Monday through Friday except federal holidays; with no holiday, 5 work days from any weekday end on the seventh calendar day.
- y ions
- The fragments of a peptide that contain its end; they form a ladder whose rungs differ by the mass of one amino acid, so the gaps read the sequence from the end.