How a measurement becomes a result.
On 2 February 2026 an amended version of the FDA's quality rule for device makers took effect under a new name, the Quality Management System Regulation, which incorporates the international standard ISO 13485:2016. The rule governs how device makers design, make and check what they sell, and it applies to the makers of diagnostic test kits. Part VI looks at an omics test in the same way, as a designed system. Chapter 16 takes the instrument and its subsystems, Chapter 17 the chain from the blood tube to the signed report and the people who own each step, Chapter 18 the reading of a result, and Chapter 19 the evidence that a test measures what it claims before anyone relies on it. Chapter 20 takes what keeps that evidence true once the test is in use: control of suppliers and of changes, complaints and corrective action, reports of wrong results, field actions and post-market follow-up.
One architecture, many instruments.
+ The questionWhat happens inside the instrument between loading a sample and writing a data file?
Three instruments, one architecture
A short-read sequencer is a cabinet of pumps, a heated glass flow cell, lasers, a camera and a computer. A mass spectrometer and a flow cytometer look nothing like it from the outside, yet seen as systems the three share one architecture. Each has a way of handling the sample and delivering it to the measurement, a physical conversion that turns molecules into something detectable, a sensor, detection electronics, a controller that keeps every step in time, and software that turns detector counts into an answer. Instrument engineers will recognize the signal chain behind every measuring instrument, sense → condition → digitize → compute → connect → decide, and an omics instrument is that chain applied to a biological signal.
The subsystems of a short-read sequencer show the architecture at work. Fluidics pump reagents across the flow cell in a fixed sequence, while the flow cell holds the chemistry at a controlled temperature. In each cycle, lasers excite the dyes incorporated in that cycle and a camera images the surface, so that image processing can locate every cluster and measure its intensity in each color, and base calling then converts those intensities into letters, each with a quality score (Chapter 4).
A mass spectrometer has the same shape: pumps and a chromatography column deliver peptides over time, electrospray converts them into ions, ion optics and mass analyzers sort the ions, and a detector counts them, after which acquisition software writes spectra and search software turns them into identities. In both instruments no single component explains the result. The coordination of the components does.
Timing limits read length and quantification
In a sequencer, every cycle must add one base to every strand, image the whole surface and remove the blocking group before the next cycle begins. If a small fraction of the strands in a cluster miss an addition, they fall behind the rest, a problem known in sequencing chemistry as phasing, a different sense of the word from the phasing of variants in Chapter 5. Strands that add two bases in one cycle run ahead instead, which is called prephasing. Either way the cluster's signal becomes a blend of the current position and its neighbors, and because phasing and prephasing accumulate cycle by cycle, base quality declines along a read and read length is limited. The controller's job is therefore to hold flow, temperature and timing tightly enough that each cycle is as close to identical as the chemistry allows.
In a mass spectrometer, timing decides whether a peptide can be quantified at all. Each peptide emerges from the chromatography column over a few seconds, and to measure its amount the instrument must sample it several times across that peak. Every other ion the instrument is also trying to measure competes for the same seconds, so each measurement added to the instrument's cycle lengthens the cycle and leaves fewer samples of each peak.
A peptide elutes from the column over 6 seconds, and one acquisition cycle of the instrument takes 0.6 seconds: one MS1 scan of the kind described in Chapter 8, plus its set of fragment scans. Dividing 6 seconds by 0.6 seconds gives 10 points across the peak, enough to trace its shape and measure its area, and one instrument maker states in a 2024 technical note on data-independent acquisition that each peptide should be sampled at least 10 times across its peak. If more fragment windows are added to reach more peptides, the cycle stretches to 1.5 seconds, and 6 seconds divided by 1.5 seconds gives only 4 points, too few for reliable quantification. Depth and precision therefore trade against each other through time.
The data-conversion chain
The most useful way to look at any omics instrument is to follow one fact as each module re-represents it. In sequencing, a DNA fragment becomes a cluster of copies, the cluster becomes a flash of light in each cycle, the flash becomes pixel intensities, and the intensities become a base call with a quality score. The calls then become a read, the read an alignment, and alignments a count or a variant call. In proteomics, a protein becomes peptides, the peptides become ions, the ions become a spectrum, and the spectrum becomes a peptide match with a score. Matches then become an inferred protein, and signal becomes a quantity.
At no point does the instrument measure the quantity of interest directly. It measures a proxy and converts it six or seven times over, which means that the result is only as trustworthy as the weakest conversion. Each conversion has a characteristic failure mode, and the quality metrics that accompany an omics result exist to watch those conversions one by one.
| Conversion | What can go wrong | The metric that watches it |
|---|---|---|
| Library preparation | uneven or biased sampling of molecules | duplicate rate, insert size, library yield |
| Signal to base call | phasing, faint or crowded clusters | fraction of bases at Q30 or above |
| Read to alignment | repeats, missing reference sequence | mapping rate, mapping quality |
| Alignment to call | too few reads, systematic errors | coverage distribution, call quality |
| Ionization and detection | suppression by co-eluting molecules | signal of spiked internal standards |
| Spectrum to peptide | wrong match above threshold | false discovery rate from decoys |
| Peptide to protein | shared peptides, ambiguous identity | protein groups, unique peptides per protein |
Instrument output is often called raw data, but even the earliest file most users see is the product of models and settings. Base calls come from a model applied to images or currents, and spectra are centroided, reduced from a continuous trace to one point per peak, and filtered before they are written. Updating basecalling or acquisition software can change results from the same physical run, so the software version is part of the measurement and belongs in the record.
Controls travel the whole chain
A metric describes one conversion, while a control tests all of them at once. A control is a known quantity passed through every step alongside the samples, so that anything lost or distorted along the way shows up as a difference between what went in and what came out. Each kind of instrument has its own kind of control: sequencing runs carry a control genome whose sequence is known, RNA-seq experiments can include synthetic RNAs at known concentrations spanning the expected range, targeted mass spectrometry adds isotope-labeled copies of each measured peptide, and cytometers run beads of certified brightness.
The logic is the same in every case: the only way to know how well a chain of conversions worked is to push something known through it. The instrument, though, is only one link in a longer chain, which runs from the blood draw to the signed report. Chapter 17 follows it, with the people who own each step and the faults beyond the instrument that controls must also catch.
In instrument and assay development, map the conversion chain directly onto design controls (Chapter 19). Treat each conversion as a requirement with its own verification test, each quality metric as an acceptance criterion, and each control as a system-level verification carried into routine use. Place signal-processing and calling software under the software-lifecycle process, because it is part of the device. In a laboratory, turn the same list into the run record: instrument, software versions, metrics and controls for every batch.
- Sequencers, mass spectrometers and cytometers share one architecture: handle, convert, sense, detect, control, compute.
- Timing limits the system: phasing and prephasing limit read length, and the number of points across each peak limits quantification.
- Every result passes through a chain of conversions, each with its own failure mode and a metric that watches it.
- Controls are known quantities pushed through the whole chain, and software versions are part of the measurement.