Five principles, and one exception.
+ The questionHow is a molecule detected at all if it can be neither copied nor paired?
Every instrument measures a proxy
At the moment a sequencer detects a base being added to a growing strand of DNA, what it usually records is a flash of colored light, and the base itself is inferred from the color. No omics instrument observes biology directly. Each one converts some physical property of a molecule into a signal that a detector can count, whether photons, ions or an electrical current, and the biology is inferred from that signal through a chain of conversions that Chapter 16 follows in detail. Chapter 2 showed that only nucleic acids can be copied and paired, so a protein or a metabolite has to be detected through one of its physical properties without being copied first. In practice, the field relies mostly on five physical properties, and almost every method in this guide reads one of them or a combination of them.
The five are the order of the bases in a nucleic acid, the mass of a molecule, the shape that lets a reagent bind it, the light it emits or scatters, and the speed at which it travels through a medium. Each has a characteristic strength and a characteristic blind spot, so the property a method reads indicates most of what is needed to judge the method. The main exception to the five is nuclear magnetic resonance, or NMR, which reads the spin of atomic nuclei. In a strong magnetic field certain nuclei absorb radio waves at frequencies set by their chemical surroundings, so an NMR spectrum reports the atomic environments in a molecule, and Chapter 10 describes how metabolomics uses it.
Each principle measures a property that correlates with the biology, not the biology itself. A fluorescence signal reports bound dye, not protein, and an ion count reports molecules that ionized well, not molecules that were abundant. A count of sequences reports the molecules that survived library preparation, the steps that turn a sample into fragments a sequencer can load (Chapter 4). Every quantitative claim in omics rests on the assumption that the proxy tracks the quantity of interest, and much of good experimental design is checking that assumption.
Sequence: reading order by pairing
Sequencing reads the order of bases in DNA, or in RNA converted to DNA. Most instruments do it by making a partner strand one base at a time and detecting which base was added, usually as the flash of colored light described at the start of this chapter. Others pull a single strand through a nanometer-scale pore and read the change in electrical current as each group of bases passes. Either way, sequencing depends on the pairing rule that makes nucleic acids special, which is why it is the dominant readout for the genome, the transcriptome and, after chemical conversion, the epigenome. The sequence that an instrument reports from one molecule is called a read, and Chapters 4 to 7 cover how reads are made and used.
Weigh: mass spectrometry
A mass spectrometer gives molecules an electric charge, moves them through electric or magnetic fields, and measures how they respond. The response depends on the ratio of mass to charge, written m/z, which high-resolution instruments measure to within a few thousandths of a unit. Mass is universal: any molecule that can be charged and brought into a vacuum can be weighed, and the instrument does not need to know in advance what to look for.
Thermo Fisher Scientific makes the Orbitrap Exploris 240, a high-resolution mass spectrometer, and specifies its mass accuracy as better than 3 parts per million over 24 hours with external calibration and better than 1 part per million with an internal reference mass. These are the maker's figures, and they hold under conditions the maker defines. A part per million is one millionth of the value measured, and one millionth of 1,000 is 0.001. At m/z 1,000, 3 parts per million is therefore 3 × 0.001, or 0.003 of a unit, and 1 part per million is 0.001 of a unit. At m/z 200 the same 3 parts per million is only 0.0006. Under the maker's conditions, this instrument typically places a molecule measured at m/z 1,000 within three thousandths of a unit of its true value, which is the accuracy that "a few thousandths of a unit" describes.
That universality has two costs. The instrument sees whatever is abundant first, so rare molecules compete with common ones for its capacity, as Chapter 2 showed. Mass alone also rarely identifies a molecule, because many different molecules share a mass, so identity usually comes from breaking the molecule and weighing the pieces as well. Mass spectrometry is the main readout for proteins and metabolites. Chapters 8 and 10 cover its use, and Chapter 23 sets out who makes the instruments as of October 2026.
Bind: affinity reagents
A binder, usually an antibody or a short folded nucleic acid called an aptamer, recognizes its target by shape. The binder carries a reporter: an enzyme that produces color, a fluorescent dye, a metal atom or a DNA tag. Because the reporter can be amplified after binding, affinity methods can reach very low concentrations. An enzyme reporter turns over many substrate molecules for each bound target, and thirty cycles of PCR can copy a DNA tag up to about a billion-fold, as Chapter 2 showed.
The limit is that a binder finds only what it was made for. An affinity method measures a predefined list of targets, and each measurement is only as specific as its reagent. A binder that also sticks to a related protein reports both, and nothing in the signal reveals the mistake. Chapter 9 covers affinity proteomics, and Chapters 11 and 14 rely on binders to label cells and vesicles.
Light and travel time: counting objects and separating mixtures
Light is the reporter behind most of the other methods, but it becomes a measurement principle of its own when an instrument looks at individual objects. A flow cytometer measures each cell as it passes a laser, a microscope images each molecule in a tissue section, and the most sensitive instruments detect single fluorescent molecules. When objects can be seen one at a time, the measurement changes character, because instead of estimating how much signal a mixture gives, the instrument counts events, each event being one object detected. Counting is more robust than estimating intensity, and it is the basis of the digital methods that recur throughout this guide.
Separation methods do not identify anything on their own. Liquid chromatography pushes a mixture through a packed column, and molecules emerge at different times depending on how strongly they stick to the packing. Gas chromatography does the same for volatile molecules, and electrophoresis sorts by size and charge in an electric field. The job of separation is to reduce complexity, because a detector presented with a few molecules at a time performs far better than one presented with thousands at once. The time at which a molecule emerges, its retention time, also becomes an extra coordinate for identifying it.
Real methods combine the principles
Almost every method in practical use chains two or more of these principles, and the combination matters because each principle covers the blind spot of another. Separation spares a mass spectrometer the full mixture, a binder gives a count of cells its specificity, and a DNA tag lets a protein measurement end as a count of DNA.
| Method | Principles combined | What the combination buys |
|---|---|---|
| LC-MS proteomics and metabolomics | separate + weigh | fewer molecules reach the detector at once; retention time adds identity |
| Flow cytometry | bind + see | a specific label, read one cell at a time |
| Proximity extension assay | bind + sequence (or quantitative PCR) | protein recognition, with DNA's copying and counting |
| Single-cell sequencing | sequence, after barcoding each cell | every molecule carries the address of its cell |
| Spatial imaging | bind (by pairing) + see | molecule identity with its position in the tissue |
| CITE-seq | bind + sequence | surface proteins and RNA from the same cell |
The table shows a pattern. Whenever a molecule lacks the two nucleic-acid tricks, methods try to borrow them, most often by attaching a DNA tag to a binder so that the measurement ends as a count of DNA. The proximity extension assay and CITE-seq in the table both rest on that one move.
When reading a methods section, reduce each technique to its principles before judging it. A protein assay that ends in sequencing inherits the dynamic range of counting DNA and the specificity of its antibodies. A particle count from a scatter-based instrument inherits that instrument's size limit. Stating the principles in a specification or a technical file also makes the verification plan obvious: each principle has a known failure mode, and each failure mode needs a test.
- Almost every omics method reads one of five physical properties: sequence, mass, binding, light or travel time; NMR, which reads nuclear spin, is the main exception.
- Mass is universal but favors the abundant, while binding reaches deep but only for predefined targets.
- Counting single objects is more robust than estimating the intensity of a mixture.
- Practical methods chain principles, and many convert non-nucleic-acid targets into DNA to borrow its advantages.