One reading principle, three changes
A Sanger instrument reads up to about a thousand bases from each template, one template per capillary. A short-read instrument reads 100 to 300 bases from each of millions to billions of fragments at once, and a nanopore instrument can read a single DNA molecule more than 100,000 bases long. The three numbers describe three different capabilities, which is why laboratories still run all three kinds of sequencer: each kind answers a different question about the same DNA.
Every sequencer in routine use relies on the same fact: a base is identified by what it pairs with, or by the distinctive signal it produces as it passes a detector. Most machines copy the template strand one base at a time and detect each addition, while nanopore machines detect each group of bases as it moves through a pore. What changed across three generations of instruments is not that principle. It is how many molecules are read at once, whether they are copied before reading, and how long a stretch of one molecule can be read in one go. Each change answered a sharper question, and the questions are the clearest way to understand the instruments.
Generation one: what is this sequence?
Frederick Sanger and two colleagues at the Medical Research Council's Laboratory of Molecular Biology published the chain-termination method in 1977, demonstrating it on the DNA of a virus that infects bacteria. The method copies a single purified template in the presence of a small amount of modified bases that stop the copying wherever they are incorporated, so the result is a set of copies ending at every possible position. In 1977 the copies carried a radioactive label, and they were separated by length on a gel and read from X-ray film.
Later instruments replaced the radioactive label with a dye on each terminating base, one color for each of the four bases, and separated the fragments by length in a thin capillary, so the fragments pass a detector in order and the colors read out the sequence directly. Sanger sequencing reads up to about a thousand bases with very high accuracy, but only one template per capillary, so it answers one question well: what is the sequence of this particular piece of DNA? The Human Genome Project was built largely on it, at a cost of billions of dollars over more than a decade. Sanger sequencing is still the standard way to confirm a single variant found by other methods.
Generation two: what are all the sequences, and how many of each?
The step change came from parallelism. Short-read instruments attach millions to billions of library fragments to a glass surface, copy each fragment in place into a small cluster of identical molecules, and then read every cluster at the same time. Each chemical cycle adds one dye-labeled base to every growing strand and blocks any further addition. A camera then images the whole surface, the block is removed and the next cycle begins, so that after 100 to 300 cycles every cluster has produced a read.
The individual read is short and the chemistry has an error rate per base, but the output is enormous. The makers of current high-throughput instruments specify several terabases, or several trillion bases, per run, with most bases at Q30 or better. Parallelism did two things. It collapsed the cost, and it turned sequencing into a counting instrument, because when an instrument reads hundreds of millions of fragments, the number of reads that come from each gene, each allele or each organism is itself a measurement. RNA sequencing, single-cell methods and liquid biopsy all depend on that counting property more than on reading any one sequence.
Generation three: what is the whole molecule?
Single-molecule instruments skip the copying step. One approach watches a single polymerase at the bottom of a tiny well as it builds a partner strand, recording each incorporated base as a flash of light. Reading the same circular molecule many times over gives consensus reads. The maker of the instruments that work this way states that these reads are typically 15,000 to 20,000 bases long, with an accuracy of Q30 and above. The other approach pulls a single strand through a protein pore in a membrane and reads the changing electrical current as the bases pass. Its reads can exceed 100,000 bases, and in its own runs the maker of the nanopore instruments has reported reads of millions of bases. As of October 2026, that second maker reports a raw single-read accuracy of 99.75 %, or Q26, with the latest of its basecalling models, the software that turns the current into letters.
Reading a native molecule has a second benefit that is easy to miss. Because nothing is copied, chemical marks on the original DNA are still there to be detected, and both approaches report methylated cytosines directly from the signal, so a long read can carry sequence and epigenetic state together (Chapter 6). A long read can also show which variants sit together on the same parental copy of a chromosome, which geneticists call phasing. The question this generation answers is the one short reads struggle with: what does this whole stretch of molecule look like, in one piece, across repeats and structural variants, marks included?
The evolution at a glance
| Generation | Working principle | Question it answers | Typical read | Strength | Blind spot |
|---|---|---|---|---|---|
| Sanger | chain termination, one template per capillary | What is this sequence? | up to about 1,000 bases | accuracy on a single target | throughput, cost per base |
| Short read | cluster amplification, one base per cycle, all clusters at once | What are all the sequences, and how many of each? | 100 to 300 bases | cost, counting, scale | repeats, structural variants, phasing |
| Long read | single native molecule, real-time optics or nanopore current | What is the whole molecule? | 15,000 to more than 100,000 bases | structure, phasing, native methylation | cost per base, historically raw accuracy |
Each generation is a tool matched to a different question, and none is a strict successor to the one before. Sanger still confirms single variants. Short reads remain the cheapest way to count molecules and to find single-base variants across many samples. Long reads resolve structural variants, repeats and phasing, and they read methylation without conversion. New chemistries keep arriving, including a sequencing-by-expansion method that reached the market in 2026 and hybrid approaches that add long-range information to short reads, and the right choice still follows from the question.
The cost of a genome, 2001 to 2022
The practical consequence of parallelism shows in the cost of a genome. The US National Human Genome Research Institute tracked the cost of producing a human genome sequence at the sequencing centers it funded, from about $95 million in 2001 to about $525 in 2022. Its figure is a production cost. It includes labor, reagents, amortized instruments and overheads, but it leaves out quality control, project management and any data analysis beyond the first processing of the raw data. The fall accelerated abruptly in 2008, when the centers switched from capillary instruments to massively parallel ones.
In announcements from 2022 to mid-2026, manufacturers gave figures of $80 to $345 per genome for their newest systems. The maker of the highest figure states that it is a list price for sequencing reagents only, quoted at 20× coverage rather than the usual 30×. The makers of the two lower figures do not say what their figures include or at what coverage they apply. None of the figures is comparable with the institute's production cost, but the direction of the fall is not in doubt. Chapter 23 gives the makers' figures as they stood in October 2026.
The institute's production cost fell from $95.3 million in September 2001 to $525 in May 2022, a span of 20 years and 8 months, or about 20.7 years. Dividing $95.3 million by $525 gives a fall of about 181,000-fold. At Moore's-law pace, a halving every two years, 20.7 years would allow about 10.3 halvings, which together make a fall of about 1,290-fold. The actual fall was therefore about 181,000 divided by 1,290, or about 140 times larger than Moore's-law pace would have produced. A fall of 181,000-fold takes about 17.5 halvings, and 17.5 halvings in 20.7 years means the cost halved roughly every 14 months instead of every 24.
Choose the generation from the question, not from the newest brochure. To confirm a known variant, use Sanger. To count molecules across many samples, such as expression, allele fraction or pathogen load, use short reads. To resolve structure, phasing or repetitive regions, or to read methylation directly, use long reads. When a supplier quotes a cost per genome, ask what it includes, whether reagents only or also instrument time, labor, analysis and storage, and at what coverage the figure was quoted.
- All three generations read the order of bases, and they differ in parallelism, in whether molecules are copied before reading, and in read length.
- Parallel short reads collapsed the cost of sequencing and made the sequencer a counting instrument.
- Long single-molecule reads resolve structure and phasing and read methylation without conversion.
- In the series kept by the US National Human Genome Research Institute, the production cost of a human genome fell about 181,000-fold from 2001 to 2022.