02 — Why some molecules are easy to count

Why some molecules are easy to count.

+ The questionIf a whole genome can now be read for a few hundred dollars, why can a drop of blood not be measured for every protein?

Signal per molecule is the whole problem

A sequencer can read a person's whole genome from a blood sample, yet no method measures every protein in that same blood as completely and as cheaply as sequencing reads the genome, and most of the difference lies in the molecules themselves. Every detector, whether a camera, a photomultiplier or an ion counter, needs a minimum amount of signal before it can report anything above its own noise. One molecule on its own produces very little: a few photons if it carries a fluorescent dye, or one ion if it survives the trip through a mass spectrometer. The central engineering question of every omics measurement is therefore the same. How can an instrument get enough signal from each molecule of interest, and how can it tell that the signal came from that molecule and not from something else?

There are only a few ways to answer that question. A method can make more copies of the target until the signal is large. It can attach a label, a tag such as a fluorescent dye that is easy to detect, to a probe or binder that recognizes the target and nothing else, and make that label bright. Or it can use a detector sensitive enough to see individual molecules and count them one at a time. Nucleic acids happen to allow the first two answers cheaply and almost universally. That single fact explains most of the difference between the layers of the omics stack in cost, in breadth and in depth. Breadth is how many different molecules a method reaches, and depth is how scarce a molecule it can still detect.

Nucleic acids can be copied, and proteins cannot

DNA is a template for itself. A DNA polymerase moves along one strand and builds its partner, adding the base that pairs with each one it reads. Two primers, short pieces of DNA that pair with the sequence on either side of the target, give the polymerase its starting points. When a reaction is supplied with the two primers, the four building blocks and heat cycling to separate the strands, each cycle can at most double the number of copies. The method is called the polymerase chain reaction, or PCR. Thirty cycles of perfect doubling would turn one starting molecule into about a billion, because 2 multiplied by itself 30 times is about 1.07 billion. Real reactions do not copy every molecule in each cycle, and they eventually level off, so a billion is the ceiling. Even so, a signal that was invisible becomes large enough to measure with ordinary optics.

RNA can join in with one extra step. An enzyme called reverse transcriptase copies RNA into DNA, called complementary DNA or cDNA, and from there the DNA machinery takes over, which is why transcriptomics runs on the same instruments as genomics. Proteins cannot be copied. Information in the cell flows from nucleic acid to protein and never back, so there is no enzyme that reads a protein and builds another one like it. Metabolites cannot be copied either, because they are made by chains of enzymes, not from templates. The proteins and metabolites in the tube are all that any measurement of them will ever have to work with.

03 — Two tricks only nucleic acids allow
COPYING PAIRING 1 2 4 8 ≈ 1 billion copies 30 perfect doubling cycles: 1 molecule → about 1 billion (2³⁰ ≈ 1.07 × 10⁹; real reactions copy less) 1 molecule stays 1 molecule — no enzyme copies a protein A T T A G C C G C G G C T A A T 5′ 3′ 3′ 5′ TARGET PROBE the probe is designed from the sequence alone each protein needs its own binder, raised, selected or designed, and validated separately schematic
DNAProteinFocal detail
Plate 03 — Nucleic acids can be copied by enzymes and recognized by a partner written from their own sequence. Proteins and metabolites allow neither, so every measurement of them must work with the molecules that are already there.

Nucleic acids can be recognized by design

The second trick, pairing, is just as important and less often noticed. Because A pairs with T and G pairs with C, a probe for any nucleic-acid sequence can be written from the sequence alone. If a target reads ATGCCGTA, its partner is known before anything is made: it can be synthesized chemically within days, and its binding strength can be predicted from its composition. The same chemistry works for every target. Laboratories run probes or assays in sets designed to be used together, called panels, and a panel of ten thousand DNA probes amounts to ten thousand entries in a design file.

Proteins are recognized by shape, not by a code. A protein-binding reagent such as an antibody has to be raised, selected or designed for each target, one at a time, because no rule like base pairing turns a protein's sequence into the sequence of its binder. Each reagent must then be tested to show that it binds the right protein and not a similar one. A panel of five thousand proteins needs five thousand validated reagents, and ten thousand if the method uses two binders per protein. Every one of those reagents can fail in its own way, and its specificity has to be proved, not predicted.

The asymmetry in one line

Nucleic acids can be multiplied, and they can be recognized by a partner designed from their sequence alone. Proteins and metabolites allow neither, so every measurement of them must work with the molecules already present and must recognize them by mass or by a binder made for each one.

The range of abundance in plasma

A second difference compounds the first. Nearly every nuclear gene is present at two copies per cell, one from each parent, so nearly every genomic target starts at the same abundance, the number of copies of a molecule in the sample. The genome is flat, and blood plasma is the opposite. Albumin circulates at around 40 mg/mL, while signaling proteins such as interleukin-6 sit at a few picograms per milliliter in a healthy person. That is a span of about ten orders of magnitude, and the rare proteins are often the informative ones. The span from the most to the least abundant molecule in a sample is its dynamic range, and an instrument has a dynamic range of its own: the span between the largest and the smallest signal it can record at once.

Solved example: how far albumin outnumbers interleukin-6

Albumin circulates at about 40 mg/mL. For interleukin-6 the example takes about 5 pg/mL, the pooled mean that Said and colleagues found in a 2020 meta-analysis in the Journal of Medical Virology, across 3,166 values from 57 studies of healthy adult donors. A milligram is a billion picograms, so 40 mg is 40 billion pg, and by weight albumin exceeds interleukin-6 by 40 billion to 5, or 8 billion to 1. An albumin molecule has a molecular weight of about 66,500 and an interleukin-6 molecule about 21,000, so each albumin molecule is about 3.2 times heavier, and a gram of albumin holds about 3.2 times fewer molecules. Dividing 8 billion by 3.2 gives the ratio by number: for every molecule of interleukin-6 in healthy plasma there are about 2.5 billion molecules of albumin.

A span of about ten orders of magnitude in the sample decides what an instrument that measures everything at once can see. A mass spectrometer receives ions in proportion to what is in the sample, and it has a limited capacity per second and a limited dynamic range in one scan. If albumin and a few dozen other abundant proteins take almost all of that capacity, the rare proteins are never counted at all. That is why plasma proteomics spends so much effort removing or damping the abundant proteins before measurement, and why it still reaches far fewer proteins per sample than a sequencer reaches genes.

04 — Ten orders of magnitude in one tube
100 fg 1 pg 10 pg 100 pg 1 ng 10 ng 100 ng 1 µg 10 µg 100 µg 1 mg 10 mg 100 mg per mL of plasma, log scale albumin 40 mg/mL immunoglobulin G 10 mg/mL fibrinogen 3 mg/mL transferrin 2.5 mg/mL C-reactive protein (healthy) 1 µg/mL insulin (fasting) 0.5 ng/mL cardiac troponin I (healthy) a few pg/mL interleukin-6 (healthy) 5 pg/mL the rare, informative signal all levels approximate ≈ 10 orders of magnitude molar ratio albumin : IL-6 ≈ 2.5 billion : 1
ProteinFocal detail
Plate 04 — Plasma proteins span about ten orders of magnitude in concentration. Any method that reads all molecules at once spends almost all of its capacity on the most abundant few, which is why the informative, rare proteins are the hardest to see.

Copying, pairing and abundance predict the order of difficulty

Copying, pairing and dynamic range together predict a ranking of the layers that matches practice. The genome is the easiest layer, because it can be copied and paired and every target starts at the same abundance. The transcriptome is next: it can be copied after reverse transcription and paired, but its molecules span a wider range of abundance and change quickly. The proteome is harder, because nothing in it copies or pairs, its abundances span ten orders of magnitude, and each protein comes in modified forms. The metabolome is the hardest to identify, because nothing copies or pairs, its molecules are chemically diverse, and there is no template from which to predict what should be there.

Direct single-molecule detection could change the ranking

The ranking is a claim about signal per molecule, not a law of nature. Any method that detects single molecules directly, with a specificity that does not depend on copying or pairing, removes the advantage nucleic acids enjoy. Single-molecule fluorescence counting, digital assays and early single-molecule protein sequencing all move in that direction. If one of them reached the whole proteome in plasma, down to its scarcest proteins, at sequencing-like cost, the ordering in this chapter would change, and as of October 2026 none has.

The ranking also explains why so many methods for proteins and other molecules work by converting them into DNA. If two antibodies that recognize a protein each carry a short DNA tag, the tags can be joined and then copied and counted like any other DNA. A method of that kind borrows the two tricks for a molecule that does not have them, and Chapter 9 describes one in detail.

In practice

When an assay claims to measure thousands of analytes, ask two questions before anything else. Where does the signal per molecule come from: amplification, a bright label, or single-molecule detection? And how does the method know which molecule produced the signal: sequence, mass, or a binder whose specificity was tested for that analyte? The answers predict the breadth and depth, the failure modes and the validation burden more reliably than the headline number of analytes.

+ What this chapter established
  • The breadth and depth of a measurement are set by the signal an instrument can get from each molecule and by how surely it can attribute that signal.
  • Nucleic acids can be copied and recognized by partners designed from their sequence alone, and proteins and metabolites allow neither.
  • Plasma proteins span about ten orders of magnitude, so an instrument's capacity is consumed by the abundant few.
  • Together these predict the ranking: genome easiest, then transcriptome, proteome and metabolome.