+ Part IX · Quality and regulation, as of October 2026
What each regulator reviews, and when.
On 23 September 2026, from nine in the morning until six in the evening, an FDA advisory panel weighed a blood test that reads methylation patterns in cell-free DNA to screen adults aged 50 and older for many kinds of cancer. The test was GRAIL's Galleri, and the application before the panel was for premarket approval, the route the FDA reserves for devices in its highest risk class. A panel advises the FDA without binding it. The FDA's decision will close its review of the test's safety and effectiveness, the review that premarket approval requires, and as of October 2026 the agency had not decided. The FDA's summary of the meeting records three votes, each for the patients who meet the test's criteria. The panel voted 10 to 0 that the data show a reasonable assurance that the test is safe, 6 to 4 that they show a reasonable assurance that it is effective, and 7 to 2, with one abstention, that its benefits outweigh its risks. GRAIL reported the same counts that day. Chapters 19 and 22 set out two kinds of evidence for a test: a quality system that turns a claim into traced requirements and validated measurements, and a ladder of evidence that climbs from analytical validity through clinical validity to clinical utility. The three chapters of this part set each jurisdiction's rules against those two chapters, with their dates. Chapter 25 covers the United States, where a test reaches patients either as a device the FDA reviews or as a laboratory's own test. Chapter 26 covers the European Union's In Vitro Diagnostic Regulation, and Chapter 27 India's Medical Devices Rules. All three give regulatory status as of October 2026.
Device or laboratory test.
+ The questionDoes an omics test reach US patients as a device the FDA has reviewed or as a laboratory's own test, and what has each route checked before the first result is reported?
Two routes, after a court vacated the 2024 rule
On 31 March 2025 a federal court in the Eastern District of Texas vacated the FDA's rule on laboratory-developed tests, in a case brought by the American Clinical Laboratory Association. The FDA had published the rule on 6 May 2024. It had added a phrase to the regulation that defines in vitro diagnostic products, "including when the manufacturer of these products is a laboratory", so that a laboratory designing and running its own test became the manufacturer of such a product. On 19 September 2025 the FDA published a final rule that removed the phrase, "reverting to the text of the regulation as it existed prior to the effective date of the Rule". It called the step "ministerial in nature", because the court had already vacated the rule. As of October 2026 the FDA's page on laboratory-developed tests shows no later action, and a search for this guide found no action of 2026 by the FDA or by Congress.
The regulation left in place defines in vitro diagnostic products as "those reagents, instruments, and systems intended for use in the diagnosis of disease or other conditions". The definition turns on what a product is intended for, not on how it works, so a reagent kit, an instrument such as a sequencer, or a system that joins them with its analysis software can each be one. A product made and sold for diagnosis is a device, and its class decides what the FDA reviews before it is sold. A test that a laboratory designs and runs itself is what the FDA calls a laboratory-developed test, and it falls under CLIA, the law for US clinical laboratories (Chapter 17). As of October 2026, an omics test therefore reaches US patients by one of two routes, as a device or as a laboratory's own test.
The line between the two routes was argued long before the 2024 rule. A trade report of 2005 records that in 2004 the FDA had determined the serum proteomic test of Chapter 22 to be a medical device subject to premarket review, although its developer had hoped to offer it as a laboratory test under CLIA.
Products that are not yet sold for diagnosis carry one of two statements set out in the FDA's labeling regulation. A product in the laboratory research phase is labeled "For Research Use Only. Not for use in diagnostic procedures." A product being tested before full commercial marketing is labeled "For Investigational Use Only. The performance characteristics of this product have not been established." The second statement thus tells every user, on the label itself, that the product's performance has not yet been shown, and Chapter 27 returns to it.
Three classes, three routes, three verbs
The FDA's regulation on classification, 21 CFR 860.3, defines three classes by the controls each needs. "Class I means the class of devices that are subject only to the general controls" of the law. "Class II means the class of devices that is or eventually will be subject to special controls", the requirements the FDA writes for one type of device (Chapter 21). Class III is the class "for which premarket approval is or will be required". The FDA describes general controls as the basic provisions of the Medical Device Amendments of 28 May 1976, and they apply to devices in all three classes unless a device is exempted. They include the provisions on adulteration and misbranding, registration and listing, premarket notification, records and reports, and good manufacturing practice. The FDA's classification regulations for in vitro diagnostics sit in Parts 862, 864 and 866 of Title 21, and each names a type of device and the class it belongs to.
The class sets the route, and each route ends in a decision with its own verb. In a 510(k), or premarket notification, the maker compares its device with a predicate, a legally marketed device that is not subject to premarket approval, and when the FDA finds the two substantially equivalent, its decision is a clearance. A novel device without a predicate can be classified through a De Novo request, and when the FDA agrees, it grants the request. The FDA describes the route as "a pathway to classify novel medical devices" for which general controls, alone or with special controls, give reasonable assurance of safety and effectiveness. A class III device needs premarket approval, or PMA, the FDA's review of its safety and effectiveness, and when the review succeeds the FDA approves the maker's application. In each case the agency decides, and the company receives the decision.
All three routes carry omics tests. In 2007 the FDA granted a De Novo request for a 70-gene expression signature, and it has approved by PMA the blood tests for colorectal cancer and the companion diagnostics of Chapter 21. On 12 May 2026 it cleared by 510(k) a test classified as a "Next Generation Sequencing Based Tumor Profiling Test", under 21 CFR 866.6080 (Chapter 23). Each test's route follows from its claim, through the class that the FDA assigns to a device of its type. For some companion diagnostics that type is now defined by technique, which the next section describes.
For sequencing tests, two final guidances that the FDA issued in April 2018 set out the evidence it expects, and they correspond to the first two rungs of Chapter 22's ladder. The first covers the design, development and analytical validation of sequencing tests that aid the diagnosis of suspected germline disease, and Chapter 19 sets out its measures. The second covers the use of public databases of human genetic variants to support clinical validity, for genetic and genomic tests in general, and the FDA runs a voluntary process for recognizing such databases. Neither guidance had been withdrawn as of October 2026.
Companion diagnostics, reclassified one technique at a time
Companion diagnostics for cancer drugs have been class III devices in the United States, approved by PMA, and the FDA's guidance on in vitro companion diagnostics dates from 2014. The claims on a companion diagnostic's label can be extended one drug at a time, each through a supplement that the FDA approves. In September 2026 the FDA approved two supplements to the PMA of one blood-based panel, Guardant360 CDx, and each added a companion claim for a gene called ESR1. The first, on 4 September, paired the gene with the drug camizestrant, and the second, on 18 September, with imlunestrant given with abemaciclib.
The FDA has begun to move these tests down a class, one technique at a time, by orders that reclassify a type of device. On 17 August 2026 it published a final order that places test systems based on in situ hybridization, the probe-pairing method of Chapter 21, in class II with special controls when they are used with an approved cancer drug. The order is codified as 21 CFR 864.1890 and covers four product codes, the three-letter codes by which the FDA groups devices of one type. A policy blog reports that it took effect on 16 September 2026, and that the FDA had proposed it on 11 June 2025.
The parallel order for nucleic-acid tests, sequencing panels among them, is still a proposal. The FDA published it on 25 November 2025. It would move nucleic-acid-based test systems used with an approved cancer drug from class III to class II with special controls, "subject to premarket notification", which is the 510(k) route. No final order had been found as of October 2026, and a policy blog of September 2026 reported none. One of the four product codes the proposal covers, PQP, is the code under which the FDA approved two companion diagnostics by PMA in May 2026, a test for molecular residual disease on 15 May and a blood panel covering 741 genes on 19 May (Chapter 21). Those approvals came about six months after the proposal, and until an order is final, the route it proposes does not apply.
The FDA also grants some devices a Breakthrough Device designation before it decides on them. The FDA describes its Breakthrough Devices Program as a voluntary program for devices that provide more effective diagnosis or treatment of a life-threatening or irreversibly debilitating disease and that meet at least one further criterion, such as having no approved or cleared alternative. A maker whose device is designated can interact with FDA experts during development and can expect prioritized review of its later submissions, but the device must still meet the FDA's standards of safety and effectiveness. By 30 June 2026 the FDA had granted 1,320 designations and had issued 210 marketing authorizations for designated devices.
Two omics tests in this guide show that a designation and a decision on the test can be years apart. The company that then made the EV-based prostate test of Chapter 14 announced the test's designation in 2019, and a search for this guide found no EV-based test cleared, granted or approved by October 2026. A news report of September 2026 states that Galleri was designated in 2018. Its PMA was still awaiting a decision eight years later. A designation is therefore not a decision on the test: by designating a device, the FDA has not cleared, granted or approved it.
A laboratory's own test under CLIA
CMS, the federal agency that runs the CLIA program, defines three categories of test complexity. Its regulation lists waived tests, tests of moderate complexity, including a subcategory of microscopy performed by the provider, and tests of high complexity. The FDA's center for devices assigns the category, scoring a test system on seven criteria from 1 to 3 each, and a total of 12 or less makes the test one of moderate complexity. The FDA states that "Test systems that have not been CLIA categorized default to high complexity until they have been categorized." CMS states that laboratory-developed tests are considered tests of high complexity, and CLIA's regulation treats any test that is not on the lists of categorized tests as one of high complexity until it has been reviewed.
The category sets what the laboratory needs before it may run the test. A laboratory that runs tests of high complexity must hold a certificate for them and must have a laboratory director, a technical supervisor, a clinical consultant, general supervisors and testing personnel who meet the qualifications the regulation sets for that category. For tests of moderate complexity, by contrast, the regulation accepts testing personnel with a high school diploma and documented training. A laboratory that builds its own omics test therefore needs the certificate and the staff that tests of high complexity require.
What the laboratory must show is analytical. Before it reports any patient result from a test it built or modified, the laboratory establishes the performance specifications listed in Chapter 19, from accuracy and precision to the reference intervals. CMS reviews them at its routine survey every two years, after testing has started. For an unmodified test that the FDA has cleared or approved, the laboratory instead verifies that it obtains figures comparable to the maker's (Chapter 17). CMS states that its CLIA program does not address the clinical validity of any test (Chapter 19). For a laboratory's own test, the federal check reaches the first rung of Chapter 22's ladder, and it does so after use has begun.
A test described as "CLIA-approved" has had no such approval, because the certificate under CLIA belongs to the laboratory, not to any of its tests. The certificate therefore says nothing about the clinical evidence for a test the laboratory built. No CLIA review asks whether the result corresponds to a condition in the intended patients, or whether acting on it helps them. Those are the clinical rungs of Chapter 22, and for a laboratory's own test the evidence on them must come from studies that no CLIA survey asks for.
The two routes carry the two quality systems of Chapter 19. The maker of a device, such as a kit or an instrument sold for diagnosis, works under the Quality Management System Regulation, which took effect on 2 February 2026 and incorporates ISO 13485:2016. On that day the FDA also stopped inspecting device makers with its older Quality System Inspection Technique and began to inspect them under an updated compliance program. A laboratory's own test rests instead on the laboratory's validation, checked by CMS, and on any accreditation the laboratory adds, such as the checklist requirements of CAP for sequencing tests (Chapter 19).
Galleri and Shield in the FDA's records
The FDA describes Galleri, the test before the panel at the opening of this part, as a "qualitative, next-generation sequencing (NGS)-based in vitro diagnostic test". Its bulletin of 21 September 2026 adds that the test is for prescription use and detects cancer-specific methylation patterns in cell-free DNA from whole blood, and it gives the intended use as "screening for the early detection of multiple types of cancer in adults aged 50 years or older". The meeting of the FDA's Molecular and Clinical Genetics Panel was public. Comments received in the meeting's public docket by 8 September went to the panel, and the docket closed on 16 September.
GRAIL states that it submitted the PMA on 29 January 2026. In August 2026 the trade publication 360Dx reported the company's account of the evidence in the application. By that account, the application drew on the first 25,000 participants of GRAIL's PATHFINDER 2 study, on the first year of NHS-Galleri, the randomized trial in England whose results Chapter 22 reads, and on an analysis that bridges those data to an updated version of the test. That evidence reaches past the analytical rung. NHS-Galleri enrolled about 142,000 adults and was built to show whether screening reduced the cancers diagnosed at late stages, which is a question of clinical utility. In September 2026 its investigators reported in The New England Journal of Medicine that the trial missed that primary endpoint (Chapter 22). GRAIL funded the trial, and three of the paper's 13 authors were affiliated with the company. After the vote GRAIL wrote that the FDA "is expected to make a final decision on the Galleri PMA in the coming months", and the 360Dx report of August had given the company's expectation of approval in the "first part of 2027". As of October 2026 no decision had been posted.
Shield, Guardant Health's blood test for colorectal cancer, shows how an approved test changes. The FDA received its PMA on 10 March 2023 and approved it on 26 July 2024 for adults aged 45 and older at average risk, with the statement that "Shield is not a replacement for diagnostic colonoscopy or for surveillance colonoscopy in high-risk individuals". On 14 November 2025 the FDA received a supplement, and on 22 July 2026 it approved "a change to the Shield EP1 classification model which streamlines Guardant Shield to a methylation-only workflow, removing fragmentomic and somatic components". The model that the 2024 approval had locked changed only through a new FDA decision. The decision was made on a 180-day supplement, the kind of supplement the FDA requires for a significant change in design, specification or software. The FDA reviews such a supplement in depth and approves it before the change is made. The performance figures in Chapter 22 belong to the version approved in 2024.
A smaller change took a lighter form. On 24 July 2026 the FDA approved the third blood test for colorectal cancer screening, which Chapter 21 describes. By 26 August 2026 the test's record also held a 30-day notice for a change in a supplier's reagent process. A 30-day notice is the route for changes to manufacturing methods, and the maker may put the change into effect 30 days after the FDA receives the notice, unless the FDA finds the notice inadequate. In both records the change went to the FDA before it was made, but only the new classification model needed the FDA's approval. Chapter 20 opens with the supplier's change to the third colorectal test and describes the review that a maker itself gives such a change.
The FDA's records give the day each application arrived and the day it was decided, and the gap is counted here in calendar days. Shield's original PMA arrived on 10 March 2023 and was approved on 26 July 2024, 504 days later, or about 16.6 months. The 2026 colorectal test's PMA arrived on 1 August 2025 and was approved on 24 July 2026, 357 days later, or about 11.7 months. Shield's methylation-only supplement took 250 days, from 14 November 2025 to 22 July 2026. Galleri's PMA, counted from GRAIL's own date of 29 January 2026, had waited 268 days by 24 October 2026. Taking the first part of 2027 as January to June, the approval GRAIL expects would come 337 to 517 days after submission. That span contains both 357 and 504 days, so the company's expectation is in line with the times the two recent colorectal PMAs took.
Write the intended use first, and decide from it whether the test will be sold as a product or run as the laboratory's own test. For a product, find the classification regulation in Parts 862, 864 or 866 that describes it, read its class, and check on the date of submission whether any reclassification order for its kind of test is final or still proposed. Plan the evidence for every rung of Chapter 22's ladder, and mark which rungs the chosen route's review will examine. For a laboratory's own test, establish the performance specifications before the first report, and assemble the clinical evidence that no CLIA survey will ask for. Write the FDA's decisions with its own verbs: cleared, granted, approved.
- A court vacated the FDA's rule on laboratory-developed tests on 31 March 2025 and the FDA restored its earlier definition on 19 September 2025, so an omics test still reaches US patients as a device the FDA reviews or as a laboratory's own test under CLIA.
- A device's class sets its route: the FDA clears a 510(k), grants a De Novo request and approves a PMA, and in 2026 it moved companion diagnostics based on in situ hybridization to class II, while its order for nucleic-acid tests remained a proposal in October 2026.
- Under CLIA a laboratory establishes the analytical performance of its own test and CMS checks it at a survey after testing has started, with no CLIA review of clinical validity.
- GRAIL's PMA for Galleri awaited the FDA's decision in October 2026, after the panel of 23 September 2026, and Shield's change to a methylation-only model needed a PMA supplement, approved on 22 July 2026.
Let's Talk