Licenses, laboratories and research rules.
+ The questionUnder India's rules, what turns a reagent into a licensed diagnostic, and which bodies answer for the laboratory that runs a test and for the research behind it?
The class and the place of manufacture decide who licenses
A class B test kit made in India is licensed by the authority of the state in which it is made, while the same kit made abroad is licensed by India's central regulator, and a class C kit is licensed centrally wherever it is made. India regulates in vitro diagnostics, or IVDs, under the Medical Devices Rules, 2017, and the Rules divide the licensing that way. CDSCO, the Central Drugs Standard Control Organisation, classifies these tests from class A to D by risk, and it publishes the classification in lists of IVDs dated 23 July 2021 and, as updated, 25 October 2023. On 6 February 2026 it added a provision for the risk classification of IVDs on its online system for medical devices.
The license to make a test follows its class. A maker of a class A or B test applies to the State Licensing Authority, the licensing body of the state where the test is made, on Form MD-3 and receives its license on Form MD-5. A maker of a class C or D test applies to the Central Licensing Authority, which is CDSCO, on Form MD-7 and receives its license on Form MD-9. Imports of every class go to the Central Licensing Authority, with the application on Form MD-14 and the license issued on Form MD-15. The Rules also provide test licenses, on Forms MD-12 and MD-13 for manufacture and MD-16 and MD-17 for import, for small quantities made or imported for clinical investigation, test, evaluation, demonstration or training.
A new IVD, one that CDSCO has not yet approved for manufacture or import, takes its own route through the Rules. Its sponsor applies on Form MD-24 for permission to run a clinical performance evaluation, a study of the test's performance on specimens from human participants, and CDSCO grants the permission on Form MD-25. The maker or importer also applies on Form MD-28 for permission to make or import the new IVD, which CDSCO grants on Form MD-29 before it issues the license. Other devices take the parallel route of a clinical investigation, on Forms MD-22 and MD-23, and a permission for a device without a predicate, on Forms MD-26 and MD-27.
In the terms of this part, the licensing authority does for an Indian test what the FDA does for a US device and a notified body for an EU test: its decision comes before sale, and in India the class and the place of manufacture decide which authority makes it. A sequencing panel made in India for a class C purpose goes to CDSCO, as does every sequencing panel imported, whatever its class. CDSCO publishes the classes in its lists. Its updated list of 25 October 2023 places a gene analyzer, an instrument that analyzes the sequence of nucleic acids, in class A, and software that interprets human genome data from molecular genetic testing in class C, but it names no sequencing panel. A maker therefore looks for its kind of test in the lists and notes the date of the list before it plans the license, and a maker whose test is not listed applies to CDSCO for its class through the online system.
Licensing dates, and the tests notified earlier
Some IVDs had been notified earlier, which means that a government notification under the Drugs and Cosmetics Act had already brought them into licensing. Among them were blood-grouping reagents and the tests for HIV, hepatitis B surface antigen and hepatitis C, and those were licensed before the dates that follow. For the rest, licensing came in two steps. For devices not already notified, licensing became mandatory for classes A and B on 1 October 2022, following the health ministry's notification of 11 February 2020. It became mandatory for classes C and D on 1 October 2023, under a CDSCO notification of 12 April 2023. CDSCO's answers to frequently asked questions about IVDs give registration for classes C and D until 30 September 2023 and licensing from the next day. Since 1 October 2023, therefore, a test in any class needs a license before it is made or imported for diagnosis.
Research use only, decided by intended use
Products for research sit outside the Rules on one condition. CDSCO's answers to frequently asked questions about IVDs state that products for research use only are not regulated under the Rules provided they are "not intended to be used in human or animals for diagnosis of any disease or disorder". The condition is the product's intended use. The same reagent, instrument or software is outside the Rules while it is intended for research, and inside them, in the class its diagnostic purpose sets, once it is intended for diagnosis. On 13 March 2026 CDSCO added a second addendum to the current edition of these answers. The addendum does not mention research use, and it answers procedural questions instead, among them that the applications on Forms MD-24 and MD-28 may be filed together and that each change to the version of approved software needs an application for a post-approval change (Chapter 20).
The United States draws the line in the same place and marks it on the label. The FDA's labeling regulation requires a product in the research phase to state on its label that it is "Not for use in diagnostic procedures" (Chapter 25). For omics the line matters because the same sequencers, kits and analysis programs serve research and diagnosis. A team that moves a research method toward patients therefore moves it across this line, and in India a reagent, instrument or program made or imported for diagnosis then needs a license in its class.
Software, performance evaluation and the drafts of 2026
CDSCO issued its guidance document on medical device software under the Rules by circular on 21 July 2026. The document itself carries its number but no date. CDSCO's list of public notices dates it 21 July 2026, as do a law firm and two consultancies that have reported on the guidance under the same document number. The guidance covers software that meets the definition of a medical device, and CDSCO states that its term for such software includes software for in vitro diagnostics unless otherwise specified. Software that drives or influences a hardware device takes the class of that device. Standalone software is classed on two scales: the state of the healthcare situation, critical, serious or non-serious, and the significance of the software's information to the healthcare decision, from treatment or diagnosis down to informing clinical management. The combinations run from class D down to class A.
For changes to software, the guidance asks that change management be documented as part of the risk management plan. It states that an Algorithm Change Protocol "may be devised, wherever applicable, based on the nature and risks", so the protocol is an option, not a requirement. The protocol it describes has five plans, for data management, for performance evaluation and monitoring, for retraining the algorithm where that applies, for software updates and for rollback. CDSCO states that the guidance "should not be misconstrued as a new regulatory control" on such software. One of the consultancies dates the draft that preceded it to 2025. The guidance reaches the software of Chapter 19, the aligners, variant callers and pipelines that do much of an omics test's measuring, when that software meets the definition of a device.
CDSCO's page on IVDs also lists a series of documents on performance evaluation. On 17 September 2025 it published standard protocols for evaluating the performance of IVDs, prepared with ICMR, the Indian Council of Medical Research. The next day it launched a compendium of those protocols and a portal for validating IVD kits, a launch that CDSCO describes as a "Global first initiative". On 29 June 2026 it updated its list of laboratories for performance evaluation. Two guidances of 2026 are drafts for comment as of October 2026: one for the import of IVDs, dated 30 January 2026 and posted on 2 February 2026, and a "Draft Guidance document for Manufacturers of In-vitro diagnostic Medical device for stakeholders comment", posted on 9 October 2026.
An assessment outside the three jurisdictions, WHO's prequalification of in vitro diagnostics, covers listed analytes only. Its eligibility tables, last updated on 27 September 2026, name tests for HIV, hepatitis B and C, malaria, syphilis, tuberculosis, HPV by nucleic-acid testing, G6PD, SARS-CoV-2, point-of-care glucose, HbA1c and hemoglobin, and gonorrhea, chlamydia and trichomoniasis. The only sequencing entry is targeted sequencing for drug resistance in tuberculosis. No genomic, multi-omics or protein signature test is eligible, and no cancer screening other than for HPV. The tables list what may be assessed, not tests that WHO has prequalified.
Laboratories, accreditation and research ethics
Chapter 19 described NABL's move of India's medical laboratories to ISO 15189:2022. NABL announced the move on 29 December 2022 and set 31 December 2025 as the deadline for its criteria NABL 112A. Between those two dates, NABL issued an addendum to its earlier criteria, NABL 112, for the new edition on 13 February 2024, and on 18 December 2024 it published NABL 112A with its guidance document, NABL 112B. ILAC, the international cooperation of laboratory accreditation bodies, had set a transition of three years from the standard's publication on 8 December 2022, which also ended in December 2025. A laboratory that cites the 2012 edition in October 2026 therefore cites an edition that ISO has superseded and NABL no longer accredits against.
ISO 15189 appears in two of this part's jurisdictions in different roles. In India it is the standard against which NABL accredits medical laboratories, and in the EU a health institution's laboratory must comply with it, or with national provisions, before the institution may use its own tests under the in-house exemption (Chapter 26). Some Indian genomics laboratories also state that they hold accreditation from CAP, the US body of Chapter 19 (Chapter 23).
Research on people follows a separate set of rules. ICMR issued its National Ethical Guidelines for Biomedical and Health Research Involving Human Participants in October 2017. Their scope is research for health conducted in India that involves human participants. Their Section 10, on human genetics testing and research, has subsections on genetic counseling, informed consent, the results of genetic testing, and genetic diagnosis, testing and screening. An omics study that recruits participants in India therefore works under them, from consent to the handling of results.
A reader who finds genetic counseling and informed consent among the subsections of ICMR's Section 10 may take the guidelines as rules for every genetic test offered in India. Their stated scope is research involving human participants. ICMR's handbook of 2018 notes that genetic research and genetic services overlap. The handbook is ICMR's own summary of the guidelines, and it gives Section 10 as recommendations: pre- and post-test non-directive counseling by qualified persons is important, written consent should be obtained for genetic screening, next-generation sequencing and prenatal or carrier testing, and laboratories offering genetic testing should take part in quality assurance programs for genetic testing. These are recommendations in guidelines written for research ethics, and the license for a test kit comes from the Medical Devices Rules.
Prenatal diagnosis is governed by a statute of its own, the Pre-Conception and Pre-Natal Diagnostic Techniques (Prohibition of Sex Selection) Act, 1994. The Act defines a prenatal diagnostic test to include any test or analysis of the blood of a pregnant woman conducted to detect genetic or metabolic disorders or chromosomal abnormalities. The words of that definition cover a cell-free DNA test of maternal blood, such as the prenatal screen of Chapter 14, and a search for this guide found no court ruling or ministry statement on such tests. A genetic laboratory, genetic clinic or genetic counseling center may take part in prenatal testing only once it is registered under the Act. The Act allows the tests only to detect the abnormalities it lists, and only when the person conducting them records in writing that one of its conditions is met, such as the pregnant woman being above 35 years of age. It prohibits telling the woman, her relatives or anyone else the sex of the fetus.
Seven percent, and the denominator a regulated figure carries
The first reports of the opening's 7 percent counted it against the wrong whole: NASA's page of January 2018 put it in terms of the astronaut's genes, and the wire story of March in terms of his DNA. NASA's statement of 15 March 2018 named the right whole, the gene expression that had changed during spaceflight (Chapter 7). A share whose whole is wrong or missing can be read against any whole the reader supplies, and in 2018 it was read against three.
Read against the DNA, the haploid human reference genome of about 3.1 billion bases, 7 % would be about 217 million bases, since 0.07 times 3.1 billion is 217 million. Read against the genes, of which about 20,000 encode proteins, 7 % would be about 1,400 genes. Read against NASA's whole, the gene expression that changed during spaceflight, 7 % is a share of a list whose length NASA's statement did not give, so the figure names no number of genes at all. One number therefore made three different claims, a change to about 217 million letters of code, a change to about 1,400 genes, or a share of an unstated list, and only the denominator decides which claim was made.
The records of this part show what a regulated test's label and the documents that govern it state in place of a bare share. A US product still under test carries on its label the sentence "The performance characteristics of this product have not been established" (Chapter 25). The FDA's summary of the 2024 colorectal approval gives its sensitivity and its false-positive rate each with the group in which it was counted, and with the number of people studied (Chapter 22). The FDA's guidance for sequencing tests asks the developer to set each accuracy threshold on the lower bound of a 95 % confidence interval as well as on the point estimate, and that bound moves with the number of samples behind it (Chapter 19). The IVDR defines analytical and clinical performance as abilities "of a device", so its figures belong to one test, with its own specimen and pipeline (Chapter 26). The investigational label states plainly that no figure has yet been established, and the FDA's summary, its sequencing guidance and the IVDR's definitions fix the whole against which a figure is counted.
The 7 percent also holds the question this guide began with. What the flight changed was a state of the cells, the expression of genes. The genome does not record that state, and the inherited sequence did not change (Chapters 1 and 15). That state could be counted because RNA can be copied and paired, the property that makes the nucleic-acid layers the easiest to measure and leaves proteins and metabolites the hardest (Chapter 2). Turning such a count into a result someone can act on takes the rest of the book. It takes a chain from tube to report whose every link is controlled, the subject of Chapter 17, and a claim stated exactly, proven against traced evidence and kept proven after release, the subjects of Chapters 19 and 20. It takes a climb from analytical through clinical validity to clinical utility, which Chapter 22 describes. It also takes the review that each jurisdiction of this part sets for the route a test takes, by the FDA or CMS, a notified body or a licensing authority. Each of those steps fixes something the 7 percent left open when it was first read: what was measured, in whom, by which version of the test, and against what whole.
Find the test's class in CDSCO's lists of IVDs, and record the list and its date, or request a class through CDSCO's online system if the test is not listed. Decide where the test will be made, because a class A or B test made in India goes to its state's licensing authority, while a class C or D test and every import go to the central one. Keep a product intended for research labeled and supplied for research, and bring it into licensing before it is intended for diagnosis. Check that the laboratory that will run the test is accredited by NABL to ISO 15189:2022, and run any study on people under ICMR's 2017 guidelines. Read the draft guidance for IVD manufacturers of 9 October 2026 and comment while it is open, and state every performance figure with its specimen, its population and its denominator.
- Under the Medical Devices Rules, 2017, CDSCO classes IVDs from A to D; state authorities license the manufacture of classes A and B, and the central authority licenses classes C and D and every import.
- For IVDs not notified earlier, licensing began for classes A and B on 1 October 2022 and for classes C and D on 1 October 2023, and a product for research use only stays outside the Rules while it is not intended for diagnosis.
- CDSCO issued its software guidance on 21 July 2026 and posted a draft guidance for IVD manufacturers on 9 October 2026, NABL accredits medical laboratories to ISO 15189:2022, and ICMR's 2017 guidelines are written for research on people and recommend counseling and written consent for genetic testing.
- Read against three denominators, the 7 percent of the opening made three different claims, and a regulated test's figures come with the whole they are counted against: specimen, population, version and number of samples.