26 — The European Union

The IVDR's classes and evidence.

+ The questionWhat does the European Union's regulation ask of an omics test before it reaches a patient, who checks the answer, and when may a hospital laboratory make its own?

Four classes, and the rule that places three kinds of omics test in class C

The European Union's In Vitro Diagnostic Regulation sorts every test into one of four classes of risk, from A to D, and three kinds of test that recur in this guide fall in the third. Under Rule 3 of the regulation's Annex VIII, which holds its classification rules, human genetic tests, companion diagnostics and tests for the screening, diagnosis or staging of cancer are class C, unless Rule 1 or Rule 2 places them in class D. The regulation is Regulation (EU) 2017/746, known as the IVDR, and Chapter 20 describes its rules for surveillance after sale. A sequencing panel for inherited variants, a test that selects a cancer drug and a blood test that screens for cancer therefore all fall in class C under Rule 3, whichever technique each one uses. Classes A and B hold tests of lower risk and class D those of the highest, and the class decides who must assess a test before it is placed on the market.

Chapter 21 introduced the term companion diagnostic, and the regulation defines it in its own words as "a device which is essential for the safe and effective use of a corresponding medicinal product". The United States and the EU reach the class of such a test from different starting points. The FDA classes a type of device by regulation and by order, and in 2026 it moved companion diagnostics based on in situ hybridization to class II by an order that names the technique (Chapter 25). The EU's rule names a purpose instead, so a companion diagnostic is class C whether it pairs a probe, sequences a panel or measures a protein.

Who assesses a test: the notified body

For every class but the lowest, and for sterile tests in the lowest, the IVDR requires an assessment by a body other than the test's maker. That body is a notified body, a conformity assessment body designated under the regulation. The IVDR defines conformity assessment as "the process demonstrating whether the requirements of this Regulation relating to a device have been fulfilled". A maker whose test needs no notified body declares conformity itself: it draws up the test's technical documentation and issues an EU declaration of conformity, which states that the regulation's requirements "have been fulfilled". Once the assessment its class requires is complete, the maker affixes the CE marking, by which it "indicates that a device is in conformity with the applicable requirements", before the test is placed on the market. Where a notified body took part, its identification number follows the marking.

Regulation (EU) 2024/1860 amended the IVDR's transitional rules in 2024. It lists the tests whose makers could declare conformity themselves under the old Directive, 98/79/EC, and which now need a notified body: tests in classes D, C and B, and class A tests supplied sterile. Class A tests that are not sterile are the only ones missing from that list, and under Article 48 of the IVDR their makers declare conformity themselves. For a sterile class A test, Article 48 limits the notified body's part to establishing, securing and maintaining the sterile conditions. Under the IVDR's own rules a class C omics test therefore passes through a notified body, and the chapter's final section sets out the transition for tests already on the market under the Directive.

A test's clinical evidence is assembled in its performance evaluation, the subject of the next section. The plan for surveillance after sale that Chapter 20 describes is part of the test's technical documentation. The certificates that notified bodies issue are now kept in EUDAMED, the EU's database for devices. Its module for notified bodies and certificates became mandatory on 28 May 2026, at the same time as three others, for registering manufacturers and other actors, for registering devices by their unique identifiers, and for market surveillance. For records that predate the modules, the HPRA, Ireland's regulator, gives two later dates: 28 November 2026 for registering devices already on the market, and 28 May 2027 for entering certificates issued earlier.

54 — IVDR classes, and who assesses each
CLASS A CLASS B CLASS C CLASS D NOT STERILE maker's own declaration STERILE notified body, for the sterile conditions only notified body notified body notified body RULE 3, ANNEX VIII human genetic tests companion diagnostics cancer screening, diagnosis or staging the omics tests of Rule 3: class C RISK RISES IN-HOUSE, ARTICLE 5(5) a health institution's own test: exempt except Annex I's safety and performance requirements; eight conditions now, among them: not transferred to another legal entity, a quality system, a laboratory compliant with EN ISO 15189 or national provisions; from 31 Dec 2030 also a justification that no equivalent device on the market meets the patients' needs at the appropriate level of performance notified bodies' certificates kept in EUDAMED, module mandatory from 28 May 2026 IVDR Annex VIII and Article 5(5); Regulation (EU) 2024/1860; status as of October 2026; schematic
Focal detail
Plate 54 — The IVDR places human genetic tests, companion diagnostics and cancer tests in class C, and every class but non-sterile class A goes to a notified body. A health institution's own test can be exempt, but only on the conditions of Article 5(5): eight apply now, among them a laboratory compliant with ISO 15189 or with national provisions, and the ninth from 31 December 2030.

Performance evaluation, and the rung it leaves out

The IVDR defines performance evaluation as "an assessment and analysis of data to establish or verify the scientific validity" of a test's analyte, together with the test's analytical performance and, where applicable, its clinical performance. Each of the three elements has its own definition. The scientific validity of an analyte is "the association of an analyte with a clinical condition or a physiological state". Analytical performance is "the ability of a device to correctly detect or measure a particular analyte". Clinical performance is "the ability of a device to yield results that are correlated with a particular clinical condition". The Medical Device Coordination Group, whose guidance documents interpret the regulation, builds the clinical evidence for a test on these three elements in its guidance of January 2022. It calls performance evaluation "a continuous process", so the evaluation is updated through the test's life.

The three elements map onto the ladder of Chapter 22 only in part. Analytical performance matches the first rung, analytical validity, and the CLSI studies of Chapter 19 supply the evidence for it under either name. The second rung, clinical validity, is split in two. Scientific validity asks whether the analyte is associated with the condition at all, and clinical performance asks whether this test's own results correlate with the condition. The third rung, clinical utility, has no element of its own among the three. None of them asks whether acting on a result improves outcomes for patients, although the regulation does ask the maker to confirm, through post-market performance follow-up, that the balance of benefit and risk stays acceptable over the device's lifetime (Chapter 20).

The split matters most for an omics signature. A methylation signature for cancer screening needs evidence that its pattern is associated with cancer, and separate evidence that the test's own results, from its own specimen, assay and pipeline, correlate with cancer in the people it is meant to screen. The signatures of Chapter 22 that failed on independent samples showed an association only in the samples used to find them, so on new samples they had neither kind of evidence. The IVDR's performance evaluation asks for both, and neither is evidence of the third rung.

A test made inside a health institution

The IVDR exempts some tests that a health institution makes and uses itself. Under Article 5(5), such in-house devices are exempt from the regulation's requirements, except the relevant general safety and performance requirements of its Annex I, when the article's conditions are met. The article sets nine conditions, and the first is that the device is not transferred to another legal entity. The second is that it is made and used under an appropriate quality management system, and the third is that the institution's laboratory is compliant with EN ISO 15189, the laboratory standard of Chapter 19, or with national provisions. The fourth condition requires the institution to justify that the needs of the patient group it serves "cannot be met, or cannot be met at the appropriate level of performance by an equivalent device available on the market". Regulation (EU) 2022/112 deferred that condition, and Regulation (EU) 2024/1860 moved its date to 31 December 2030, so as of October 2026 an in-house test must meet the other eight.

The last five conditions concern disclosure and records. The institution gives information to its competent authority on request, makes a public declaration about the devices, keeps documentation for any class D device, makes each device according to that documentation, and reviews the experience gained from the devices' clinical use. The exemption therefore asks for the two systems of Chapter 19, a quality system and a laboratory that complies with ISO 15189 or with national provisions, and leaves out the notified body and the certificate.

Set beside the US route for a laboratory's own test, the exemption is narrower and asks more. In the United States a laboratory establishes the analytical performance of its own test under CLIA, and CMS checks that work at a survey after testing has begun (Chapter 25). An EU in-house test must still meet Annex I's requirements and comes from a laboratory compliant with ISO 15189 or with national provisions. From 31 December 2030 it will also need a justification that no equivalent test on the market meets the patients' needs at the appropriate level of performance. The exemption also applies only inside the institution, because the test cannot pass to another legal entity. A law firm's summary of the Commission's proposal of December 2025 describes three changes to these conditions: they would become more flexible, the condition about an equivalent device would be removed, and central laboratories that make tests only for clinical trials would be added.

Transition dates, and a proposal still under examination

Tests that were on the market under the old Directive, known as legacy devices, move to the IVDR on a timetable that Regulation (EU) 2024/1860 sets. A test that holds a certificate under the Directive may be placed on the market until 31 December 2027. A test whose maker declared it conforming under the Directive, and which now needs a notified body, may be placed on the market until 31 December 2027 in class D, 31 December 2028 in class C, and 31 December 2029 in class B or as a sterile class A test. Each date holds only "provided the conditions set out in paragraph 3c of this Article are met". The conditions begin with a quality management system, which the maker had to have running by 26 May 2025. The maker must also apply formally to a notified body, by 26 May 2025 for class D, 26 May 2026 for class C and 26 May 2027 for class B and sterile class A, and must hold the notified body's written agreement by 26 September of the same year.

The last of the class C deadlines passed on 26 September 2026. A legacy class C test whose maker had no signed agreement with a notified body by that day does not meet the conditions for the extension to 2028. The class D deadlines passed in May and September 2025, and as of October 2026 the class B deadlines lie ahead, in May and September 2027.

Solved example: the windows the transition gives each class

For class C, the formal application was due by 26 May 2026 and the written agreement by 26 September 2026, four months later, or 123 days. From the agreement deadline to the end of the transition, 31 December 2028, is 2 years, 3 months and 5 days, or 827 days, about 27 months. A test whose maker met both deadlines may be placed on the market through those 27 months, and by the end of them it needs its IVDR certificate. Class D ran the same two windows a year earlier, with 123 and 826 days, one day fewer because its second window ends before 29 February 2028, and class B runs them a year later, with 123 and 827 days. The regulation therefore staggers the three classes by one year and gives each the same four months to sign and about 27 months to finish.

55 — The IVDR transition for tests already on the market
2025 2026 2027 2028 2029 CLASS D about 27 months 31 Dec 2027 application agreement CLASS C about 27 months 31 Dec 2028 application agreement CLASS B, STERILE A about 27 months 31 Dec 2029 application agreement DIRECTIVE CERTIFICATE 98/79/EC 31 Dec 2027 26 Sep 2026: class C agreement deadline, passed quality system in place by 26 May 2025 October 2026 Regulation (EU) 2024/1860; each date holds only if the conditions are met; the simplification proposal of 16 December 2025 had not been adopted as of October 2026
ReadoutFocal detail
Plate 55 — Regulation (EU) 2024/1860 staggers the classes by one year and gives each the same windows: four months from application to written agreement, and about 27 months from agreement to the end of the transition. By October 2026 the class D and class C deadlines had passed.

On 16 December 2025 the European Commission proposed one regulation to amend the IVDR, the regulation on medical devices and two other regulations, "as regards simplifying and reducing the burden of the rules on medical devices and in vitro diagnostic medical devices". The same law firm's summary describes what the proposal would change for tests. Beyond the in-house conditions, performance studies that involve only routine blood draws would no longer need prior authorization, sterile class A tests would no longer need a notified body, and the five-year limit on a certificate's validity would go. Most of the changes would apply six months after the regulation enters into force. The same summary notes that its provisions may change during the legislative process.

The Council and the European Parliament were still examining the proposal separately as of October 2026, and the two institutions had not begun to negotiate a joint text. The Council's working party began to examine it article by article on 23 March 2026. In a progress report dated 5 June 2026, the Council's Presidency reported that "additional work with a view to refining and complementing the proposal will be necessary to reach a General Approach", the position the member states must agree before talks with the Parliament. The working party was due to discuss the Presidency's third compromise text on 26 and 27 October 2026. In the Parliament the public health committee, SANT, leads, and the same report gives early 2027 as the Parliament's aim for a vote on its position. Two law firms report that the Commission expects adoption by the second quarter of 2027. The European Economic and Social Committee gave its opinion on 29 April 2026, by 191 votes to none with 4 abstentions, and the opinion does not bind either institution.

Planning the EU file for an omics test

Classify the test against Annex VIII from its intended purpose, and record which rule places it in which class. Build the performance evaluation in its three parts, scientific validity, analytical performance and clinical performance, and keep the evidence on clinical utility in a section of its own, because none of the three parts requires it. For a test already on the market under the Directive, check the dates for its class in Regulation (EU) 2024/1860 and keep the notified body's written agreement on file. For an in-house test, document each of the eight conditions of Article 5(5) that apply now, prepare the justification against equivalent devices that applies from 31 December 2030, and check the proposal's status again before relying on any condition it would change.

+ What this chapter established
  • Under Rule 3 of the IVDR's Annex VIII, human genetic tests, companion diagnostics and tests for the screening, diagnosis or staging of cancer are class C, and a notified body assesses every class above A and the sterile conditions of sterile class A tests.
  • Performance evaluation rests on scientific validity, analytical performance and clinical performance, which split Chapter 22's clinical validity in two and give clinical utility no element of its own.
  • A health institution's in-house test is exempt from most of the IVDR only under the conditions of Article 5(5): eight apply now, among them a laboratory compliant with ISO 15189 or with national provisions, and the remaining one, a justification that no equivalent device on the market meets the patients' needs, applies from 31 December 2030.
  • Regulation (EU) 2024/1860 lets legacy tests be placed on the market to the end of 2027, 2028 or 2029 by class, provided their makers meet its conditions; the class C agreement deadline passed on 26 September 2026, and the simplification proposal of December 2025 had not been adopted by October 2026.