22 — From signature to diagnostic

From signature to test.

+ The questionWhy do many omics discoveries never become tests, and what does it take for one that does?

Discovery and validation are different experiments

With 20,000 random features measured in 40 patients, a model can sort the patients perfectly into their two groups and then do no better than chance on new ones. An omics discovery study works in exactly that situation. It measures thousands of features, whether genes, proteins or metabolites, in tens or hundreds of samples, and it looks for a pattern, or signature, that separates two groups. When a study has many more features than samples, some combination of features will separate any two groups perfectly, even when every measurement is random, which means that a model fitted to such data can learn the noise of that particular set of samples as easily as the biology.

46 — How a signature fits noise
20,000 measured features · 40 patients · pure noise case (20) control (20) TRAINING SET (the data used to choose the signature) perfect separation (AUC 1.00) INDEPENDENT VALIDATION chance (AUC about 0.5) fails on new patients shaded: called a case by the signature · AUC: area under the curve, 1.00 for perfect separation, 0.5 for chance · schematic
ReadoutFocal detail
Plate 46 — With twenty thousand measurements and forty patients, some combination of features will separate any two groups perfectly — even when every number is random. Only data the signature has never seen can show whether it learned biology or noise.

The only reliable defense is data the signature has never seen. Like a pipeline in Chapter 19, the signature is fixed, or locked, with every step of its processing and its decision threshold, and it is then tested on independent samples. Those samples are ideally collected elsewhere, processed at a different time, and drawn from the population in which the test would be used. Many omics signatures fail at this step, because their performance in discovery was real only for the samples used to find them.

History has made the point expensively. In 2002 Petricoin and colleagues reported in The Lancet that a pattern of proteins in serum, read by mass spectrometry, detected ovarian cancer with near-perfect accuracy. They found the pattern in serum from 50 women with ovarian cancer and 50 without, and on 116 masked samples it identified all 50 cancers and 63 of the 66 other women. Six of the paper's 11 authors came from the US National Cancer Institute and the FDA, and two were then at Correlogic Systems, the company that planned to offer the test and prepared it for commercial launch within two years. In 2004, however, the FDA said the test was a medical device subject to premarket review. Independent reanalyses, one of them by Baggerly, raised concerns that the separating patterns reflected differences in how and when the samples had been processed, and the test never reached the market as planned.

The Duke University trials that open Part VII ran in the late 2000s on gene-expression signatures developed at Duke. Findings of research misconduct followed: in 2015 the US Office of Research Integrity found that a former Duke researcher had engaged in misconduct in research funded by the National Institutes of Health. The case led directly to a 2012 report of the Institute of Medicine, requested by the National Cancer Institute, on the evidence standards that translational omics should meet.

The evidence ladder

A widely used framework for genetic and genomic tests asks three questions in order. Analytical validity asks whether the test measures the analyte, the substance it is meant to measure, accurately and reproducibly, in the specimen type it will receive, across operators, instruments, lots and days. Clinical validity asks whether the result corresponds to the condition in the intended population, with known sensitivity, the share of people with the condition whom the test calls positive, and known specificity, the share without it whom the test calls negative (Chapter 13). Clinical utility asks whether acting on the result improves outcomes for patients, compared with not testing. Ethical, legal and social considerations run alongside all three. The framework comes from the US Centers for Disease Control and Prevention, which named it ACCE after its four parts.

47 — From signature to test
CLINICAL UTILITY does using the result improve outcomes? CLINICAL VALIDITY does the result track the condition in the intended population? ANALYTICAL VALIDITY does the test measure the analyte accurately and reproducibly? DISCOVERY thousands of features, tens of samples INDEPENDENT VALIDATION COHORT where many omics signatures fail US LDT under CLIA (lab analytical validity) FDA 510(k) / De Novo / PMA companion diagnostic claims EU IVDR classes A–D · notified body companion diagnostics and genetic tests: class C INDIA CDSCO · MDR 2017 IVD classes A–D licensing: state A/B, central C/D + imports regulators review up to clinical validity; a CLIA-only LDT: analytical performance only schematic · routes as Chapters 25 to 27 set them out
ReadoutFocal detail
Plate 47 — A signature becomes a test by climbing a ladder of evidence: it must be measured reliably, must track the condition in the people it is meant for, and must change what happens to them. Routes differ in how much of the ladder a regulator reviews.

Each rung needs a different kind of study, and the rungs get harder going up. Analytical validation is laboratory work, and Chapter 19 shows how it is done, one measure at a time. Clinical validation needs well-characterized samples from the right people, while clinical utility often needs a prospective trial in which testing changes what is done and outcomes are followed. The expression signatures of Chapter 21 entered guidelines on the strength of studies of archived trial samples. Two randomized trials in thousands of women later showed that many women whose tumors scored at low or intermediate genomic risk could forgo chemotherapy, although some younger women with intermediate scores gained from it (Chapter 21).

The arithmetic of screening

Screening healthy people raises the bar further, because the disease is rare in the population being tested, and rarity changes what a positive result means. The number that matters to a person who tests positive is the positive predictive value, the share of positive results that are true.

Solved example: what a positive screening result is worth

Suppose 100,000 people are screened and 1 % of them have undiagnosed cancer, so 1,000 have cancer and 99,000 do not. A test with a sensitivity of 50 % is positive in 500 of the 1,000 people with cancer. A specificity of 99.5 % means it is falsely positive in 0.5 % of the 99,000 without cancer, which is 495 people. The positive predictive value is therefore 500 divided by the 995 positive results, about 50 %. Half of all positive results are false, even with a false-positive rate of only one in two hundred, and half the cancers are missed.

These numbers are close to those reported for a real test. NHS-Galleri, a randomized, double-blind trial in eight regions of England, enrolled about 142,000 adults aged 50 to 77 who had not had cancer diagnosed or treated in the previous three years and were not being assessed for possible cancer. Half of them, assigned at random, had their blood tested at up to three yearly visits with a test designed to detect many cancers from one draw. The blood of the other half was stored, and both groups received usual care. The Cancer Research UK Cancer Prevention Trials Unit at Queen Mary University of London ran the trial with King's College London and Guy's and St Thomas' NHS Foundation Trust, and GRAIL, the test's maker, funded it. The investigators reported the results in two papers in September 2026, and three of their 13 authors were affiliated with GRAIL. The paper in Nature Medicine gives a positive predictive value of 52.0 % across the three screening rounds and a specificity of 99.50 % to 99.60 % in each round.

In The New England Journal of Medicine, Sasieni and colleagues reported that the trial's primary endpoint, the incidence of stage III or IV cancer of 12 prespecified types, did not differ significantly between the groups after three screening rounds. The screened group's rate divided by the control group's, the incidence rate ratio, was 1.03, with a p-value of 0.63. For stage IV cancer of those types, the key secondary endpoint, the ratio was 0.86, so the screened group's rate was about 14 % lower, but the 95 % confidence interval of the ratio ran from 0.74 to 1.00, and 1.00 is the value that means no difference. The journal's summary of the paper gives no p-value for that result and does not call it significant. GRAIL had presented it in May 2026 as nominally significant, meaning that it passed the conventional threshold of 0.05 before any allowance was made for the number of endpoints tested (Chapter 18).

On 23 September 2026 an FDA advisory panel voted on the test. A panel of this kind is a committee of outside experts whose vote advises the agency without binding it. As of October 2026 the agency's decision is pending, and the opening of Part IX gives the panel's votes (Chapter 25).

By contrast, a blood test for colorectal cancer that the FDA approved in 2024 targets one cancer. The FDA's summary of that approval reports results from 7,861 people in a study run by the maker. The test detected colorectal cancer 83 % of the time and gave a false positive to about 1 in 10 people without cancer or an advanced precancerous growth, and it is approved for average-risk adults aged 45 and older. Those figures describe the version approved in 2024, and in July 2026 the FDA approved a version that reads methylation alone (Chapter 25).

A test is judged by what follows a positive result

Sensitivity, specificity and the area under the curve, a single score of how well a test separates two groups, all overstate performance when they are measured in a discovery cohort, because the features were chosen on that same cohort. Even validated figures do not say whether a test is useful, which depends on how common the condition is in the tested population, what is done after a positive result, what that costs and risks, and whether it changes outcomes.

Routes to market differ by jurisdiction

A test reaches patients by different routes in the United States, the European Union and India, and the regulators differ in how much of the evidence ladder they review before it does. Part IX sets out the three as of October 2026. Chapter 25 covers the United States, where the FDA clears, grants or approves a test sold as a device, while a laboratory-developed test, one that a laboratory designs and runs itself, falls under the federal rules for clinical laboratories known as CLIA. Chapter 26 covers the European Union's In Vitro Diagnostic Regulation, its risk classes and the evidence that a notified body, the organization designated to assess a device, asks to see. Chapter 27 covers India's Medical Devices Rules and the licensing of tests by CDSCO, the national regulator, and by the state licensing authorities.

The pipeline is part of the test

In an omics test, the analysis software does much of the measuring: it calls the variants, normalizes the counts, applies the model and produces the result. Regulators treat the software accordingly. The FDA's 2018 guidance on sequencing-based tests for germline disease treats the bioinformatics pipeline as part of the test, to be documented, versioned and validated end to end and revalidated after changes. A signature defined by a model must also be locked before validation.

If the model is meant to change after release, US law since 2022 allows a predetermined change control plan, which sets out in advance what may change and how each change will be verified. The FDA finalized its guidance on such plans for software functions that use artificial intelligence in December 2024. India's software guidance of July 2026 asks that changes to software be documented as part of risk management, and it states that an Algorithm Change Protocol "may be devised, wherever applicable" (Chapter 27). Such a protocol can include a plan for retraining the algorithm, and the guidance makes it an option rather than a requirement.

In practice

Write the intended use before the first validation sample: the population, the specimen, the analyte, the result and the decision it informs. Everything else follows from it: the rungs of evidence needed, the size and source of the validation cohort, the risk class in each market, and whether a laboratory-developed test, a kit or a companion diagnostic is the right route. Lock the pipeline and the threshold before independent validation, version every component, and plan the regulatory file for each target market at the start. Treat regulation as a design input from the first day rather than as a last step before launch.

+ What this chapter established
  • Discovery with many more features than samples overfits easily, and only locked, independent validation shows that a signature is real.
  • Tests climb a ladder of evidence: analytical validity, clinical validity, then clinical utility.
  • In screening, rarity makes even a very specific test produce many false positives.
  • US, EU and Indian routes differ in path and in how much of the ladder a regulator reviews, and the analysis pipeline is part of the test.