Evidence before anyone relies on it.
+ The questionBefore the first result reaches a patient, what must the team behind a test be able to show, and how does it judge whether the risk that remains is acceptable?
A test is proven against a claim stated exactly
In August 2014 a consortium coordinated by the FDA reported one kind of measurement that the RNA sequencing it had tested did not deliver: an accurate absolute amount of each transcript. The Sequencing Quality Control project, known as SEQC or MAQC-III, measured reference samples at several laboratory sites on three platforms, Illumina's HiSeq, Life Technologies' SOLiD and Roche's 454, and produced more than 100 billion reads. Its authors included staff of Illumina, Life Technologies and other companies that make gene-expression products. Neither RNA sequencing nor the microarrays it was compared with gave accurate absolute measurements of expression. Relative expression, the ratio between two samples, was accurate and reproducible across sites and platforms "if specific filters are used", and the consortium wrote that "Measurement performance depends on the platform and data analysis pipeline".
The finding divides one claim into two. A claim about how much a gene's expression changed between two samples passed the consortium's comparison, on stated platforms, with a stated pipeline and stated filters. A claim about how many copies of each transcript a sample holds did not. Proving a test therefore starts before any experiment, with a claim stated exactly enough that an experiment could fail it: the specimen, the analyte or substance measured, the kind of result, the platform and the pipeline. Everything the team later shows is shown against that claim, and against nothing wider.
For a device maker in the United States, the route from such a claim to a finished design is set by the Quality Management System Regulation, or QMSR, the current form of 21 CFR Part 820. FDA issued the final rule on 31 January 2024, the Federal Register published it on 2 February 2024, and it took effect on 2 February 2026 in place of the Quality System Regulation. The QMSR incorporates by reference the requirements of ISO 13485:2016, the international standard that sets quality-system requirements for device makers, which ISO reviewed and confirmed in 2025. FDA states that a certificate of conformance to ISO 13485 "will not exempt a manufacturer from an FDA inspection".
Clause 7.3 of ISO 13485, on design and development, sets out what US practice calls design controls, the name of the corresponding section of FDA's earlier Quality System Regulation. Under the QMSR it applies to devices in class II and class III, the two higher of FDA's three risk classes (Chapter 25), and to the class I devices that are automated with computer software or named in a list in the rule. These requirements bind the maker of a device, such as a kit sold to laboratories. A US laboratory that builds its own test answers for it under CLIA, whose legal frame Chapter 25 sets out together with what became of FDA's 2024 rule on such tests. A laboratory can apply the same discipline of written requirements and kept evidence to a test it builds.
Every requirement points to its evidence
FDA's training material on the QMSR sets out what clause 7.3 asks, step by step. A plan documents the stages of the work, its reviews, its responsibilities, its resources and its traceability. Design inputs, the requirements, must be complete, unambiguous, able to be verified or validated and free of conflict, and they must "Account for all pertinent risk management outputs". Design outputs must meet the inputs and contain or reference their acceptance criteria. For an omics test the outputs include its protocols, its pipeline, its specifications and its labeling, the documents and software that embody the design. Reviews check the work at planned points.
Verification shows that each output meets its input, and validation shows that the finished design meets its users' needs and its intended use. Validation runs on representative product to a documented plan and is completed before release. Transfer then moves the design into manufacturing, and every later change passes through change control before it is implemented: the change is reviewed for its effects, and the studies it touches are run again as the review requires (Chapter 20). All of it is recorded in the design and development file, kept for every device type or family, which must "include or reference records that demonstrate conformity". Traceability makes the file usable: each requirement points to the outputs that meet it and to the records that prove it, so a reader can start from any claim and reach its evidence.
Chapter 17 introduced FDA's final guidance of April 2018 on sequencing tests for inherited disease, with the test elements it asks a developer to specify. The same guidance shows what design inputs look like for an omics test, and it covers germline tests only, not tests of tumors or of cell-free DNA. Beside the elements Chapter 17 listed, the developer specifies the methods for DNA quantity and quality and for library preparation and enrichment. The developer then sets thresholds for accuracy, precision and the limit of detection, the smallest amount of a variant the test reliably finds, and run-quality thresholds for average and minimum depth, uniformity of coverage and base quality. The guidance gives no number for any of those run-quality thresholds, neither a depth such as 30× nor a share of bases at Q30, so the developer sets each one and justifies it.
The specimen requirement belongs among the inputs as well. Chapter 17 showed that the tube, the time to centrifugation and the storage conditions change what the instrument receives, so a claim made for blood drawn into one tube and processed within one window holds only for specimens handled that way. The input then names those limits, the output is the specimen instruction, and verification has to show that the result holds at the edges of the window the instruction allows.
A wrong result becomes dangerous when someone acts on it
Some design inputs come from the claim, and others from an analysis of what could go wrong. The standard for that analysis is ISO 14971:2019, which ISO confirmed in a review of 7 March 2025 and lists as current as of October 2026. Its catalogue entry describes a process for the risk management of medical devices across their life cycle, with in vitro diagnostic devices and software included. The QMSR does not incorporate ISO 14971 by reference, and FDA's training on design and development does not name the standard. The training states instead that design inputs must account for all pertinent outputs of risk management, the process that ISO 14971 sets out.
ISO's guidance on applying the standard, ISO/TR 24971:2020, carries an annex for in vitro diagnostic devices, and ISO registered a project on 6 July 2026 to replace that guidance with a technical specification, ISO/TS 24971-1, which as of October 2026 is at an early stage. ISO's page for the project notes that ISO 14971 does not require its process to sit inside a quality management system, and that applying it can meet some of the risk requirements of ISO 13485.
ISO 14971's terms, as Health Canada and FDA reproduce them, separate the source of harm from the harm itself. A hazard is a potential source of harm, and a hazardous situation is a circumstance in which people, property or the environment are exposed to one or more hazards. Harm is injury or damage to the health of people, or damage to property or the environment. Risk combines the probability of occurrence of harm with the severity of that harm. FDA's introductory training on risk reproduces the chain the standard draws between these terms: a hazard, a sequence of events, a hazardous situation and harm. The probability of occurrence of harm is then the product of two probabilities, that the hazardous situation occurs and that it leads to harm.
For an omics test, which works on a specimen already taken, the path to harm runs through the result that someone acts on. This guide sorts wrong results into three kinds: a false negative, which misses a condition that is present; a false positive, which reports one that is absent; and a delayed result, correct but too late for the decision it was meant to inform. Regulators' documents use broader words. IMDRF, the international forum of device regulators, speaks in its 2025 document on software risk of "information-related hazards" and of harm that includes "reduction of effectiveness". The EU's guidance on post-market surveillance speaks of "expected erroneous results", a term Chapter 20 takes up. Whatever the words, this guide reads ISO 14971's definitions to mean that a wrong result becomes a hazardous situation only when it reaches a decision about a patient.
Index hopping, measured in Chapter 17, can be traced through every step of this chain. The hazard is the presence of reads from another patient's library in a pooled run. The sequence of events is that some of those reads carry a variant, the pipeline calls it above its filters, and interpretation reports it as relevant. The hazardous situation is a false-positive result in front of a clinician who must decide on treatment, and the harm is a treatment or a procedure the patient did not need. Each step of the sequence is a place where a control can stop it.
Chains such as this one are found by systematic analysis, and two general methods each have an international standard. A failure modes and effects analysis, or FMEA, starts from the parts and steps of a design or a process, asks how each might fail and follows each failure to its effects. IEC's catalogue entry for the 2018 edition of IEC 60812 states that the purpose of the method "is to establish how items or processes might fail", and that "Failure modes can be prioritized to support decisions about treatment." Treatment there means what is done about a risk, not the care of a patient. A fault tree analysis works in the opposite direction to an FMEA, from one undesired event back through the events and failure modes that could combine to produce it, and IEC 61025, of 2006, gives guidance on it. Both standards are in their current editions as of October 2026, and both come from IEC's committee on dependability rather than from medical-device work.
FDA's training on risk management under the QMSR lists a process FMEA, which "Assesses manufacturing process risks", among a maker's risk documents. On the hopping chain, an FMEA of the pooling step would list index hopping as one way that step can fail and follow it to a false call. A fault tree drawn down from a false-positive report would set hopping beside a sample swap, the other single fault of Chapter 17, as events that can each lead to it.
A risk is accepted only against criteria set in advance
Finding and estimating the risks does not yet decide whether a test may be released with them. FDA's training on risk management under the QMSR names the risk documents in which that decision is made. The risk management plan "Outlines risk management scope, responsibilities, and criteria", so the criteria for accepting a risk are fixed before any risk is estimated, and each estimated risk is then judged against them. A risk that fails the criteria needs a control.
ISO 14971 ranks the controls. FDA reissued its guidance on human factors on 3 August 2026, and the guidance lists them "in order of preference and effectiveness": inherent safety by design, then protective measures in the device itself or in the manufacturing process, then information for safety. On the hopping chain, the design measure is unique dual indexes, which remove most hopped reads before any call is made. A protective measure is a check that stops a result, such as a run that must pass its quality thresholds before its calls are released. An identity fingerprint that must match is the same kind of check against the sample swaps of Chapter 17. Information for safety is the report's statement of the test's limitations, which FDA's 2018 guidance asks for. Each control then becomes a requirement in its own right and enters the trace with its own verification.
In the Broad Institute's HiSeq X data of Chapter 17, an average of 0.89 % of the reads in PCR-free genomes came from other samples when the reads were sorted by one index, and 0.13 % when they were sorted by both unique indexes. The second index therefore removed about 85 % of the foreign reads and left about one in seven of them. Double swaps, which form another valid pair of indexes and so pass the check, account for only part of what remains: Costello and colleagues at the Broad estimate them at about 0.01 % of reads, a thirteenth of the 0.13 %, and their paper does not attribute the rest to hopping. The design measure removes most of the measured contamination but not all of it, so the protective checks and the report's statement of limitations are still needed on the hopping chain.
What a control leaves is the residual risk. FDA's human factors guidance reproduces the definition of ISO 14971: "Risk remaining after risk control measures have been implemented." The guidance states that "some residual risk will remain" in any device, and that "The remaining residual risk is then evaluated for acceptability", which means a second judgment against the plan's criteria. The guidance is written about the risks of using a device, and the same reasoning applies to a wrong result. On the hopping chain, the foreign reads left after dual indexing can still produce a false call, so that residual risk is judged with the run's quality thresholds and the report's limitations in place.
The last judgment covers all the residual risks together. FDA's training on risk management under the QMSR names two documents for it: a benefit-risk analysis, made when residual risks remain after the controls, that "Justifies why benefits outweigh residual risks", and a risk management report that "Justifies acceptability of individual and overall risk". The overall residual risk, all the residual risks taken together, is therefore weighed against the benefit of the test. FDA's human factors guidance asks a submission in the same way to "describe how the overall benefits of using the device outweigh the residual risks."
Software that measures has a life cycle of its own
In an omics test, software does much of the measuring. Base-calling software turns images or currents into letters (Chapter 16), and the pipeline aligns, counts and calls (Chapter 17). The international standard for the life cycle of device software is IEC 62304, in force as of October 2026 in its edition 1.1, the 2006 text with its 2015 amendment. It sets requirements for developing and maintaining software that is itself a medical device or an integral part of one. A normative reference is another standard whose requirements a standard takes in, and the only one IEC 62304 names is ISO 14971, so the software process of IEC 62304 depends on risk management carried out to ISO 14971.
IEC 62304 sorts software into three safety classes, A, B and C, and it sets requirements in several clauses for software of unknown provenance, or SOUP. The Johner Institute, a company that sells regulatory training and consulting, reproduces the standard's definition: SOUP is software already developed and generally available but not developed for incorporation into the device, or software developed earlier for which adequate records of its development are not available. IEC's committee has drafted a design specification for a second edition. That edition would apply "to the life cycle of health software", not only to device software, and would replace the three classes with two levels of process rigor, so that "Current Class A is equivalent to level I, Classes B and C are intended to be level II". As of October 2026 the second edition is a proposal, and edition 1.1 is the standard in force.
FDA issued final guidance on the content of premarket submissions for device software functions on 14 June 2023, and the guidance is current as of October 2026. It asks for one of two documentation levels. The Enhanced level applies where a failure or flaw of the software "could present a hazardous situation with a probable risk of death or serious injury", and the Basic level applies otherwise. The level decides what a maker submits, not how the software is built, and for its development practices a maker may submit a declaration of conformity to the FDA-recognized version of IEC 62304.
FDA calls a component outside the maker's full control off-the-shelf software, "a generally available software component used by a device manufacturer for which the manufacturer cannot claim complete software life cycle control". On that wording, an open-source aligner or variant caller would fit the definition. For a germline sequencing test, FDA's 2018 guidance treats the pipeline as part of the test. It asks the developer to document "software versions and traceability, reference sequence assembly", to include "the finalized bioinformatics pipeline" in the end-to-end validation of the test, and to document every modification, "including software updates and other modifications to the bioinformatics pipeline".
A pipeline validated this way is locked: every tool, version, parameter and reference is fixed before validation, so the pipeline that runs on patients' samples is the one that was tested. For a minor change, and only for one, the guidance allows "comparing results from the new pipeline to the existing test pipeline".
A pipeline built from open-source aligners and callers changes whenever one of its tools, parameters or reference files changes, even when none of the laboratory's own code does. FDA's 2018 guidance asks for every such modification to be documented and for the finalized pipeline to be validated as part of the test, so an automatic update of a dependency is a design change. It needs change control and, at the least, the comparison with the existing pipeline that the guidance allows for a minor change.
IMDRF's 2025 document adds a principle for judging the risk of any medical device software. Where possible, it suggests setting "the probability of software failure to 1", so that the analysis assumes the software will fail and judges the consequence. It also suggests categorizing that risk "prior to implementation of risk controls", before any control is credited.
Analytical validation, measure by measure
The word validation also has a laboratory sense, different from the design validation above. Analytical validation shows that the test measures its analyte correctly, and CLIA's regulation lists the measures. Before reporting patient results from a test it built or modified, a US laboratory must establish, as applicable, accuracy, precision, analytical sensitivity, analytical specificity including interfering substances, the reportable range, the reference intervals and any other characteristic the test requires. For an unmodified cleared or approved test, the laboratory verifies the maker's accuracy, precision and reportable range and checks the reference intervals against its own patients. CMS, which runs the CLIA program, reviews that validation at its routine survey every two years, after testing has started, and its 2013 answers on laboratory-developed tests state that the program "does not address the clinical validity of any test".
CLSI, a standards organization for laboratory medicine, publishes a consensus document for each measure in the table below, which lists the documents as CLSI's catalogue describes them in October 2026.
| Measure | What the study shows | CLSI document |
|---|---|---|
| Precision | How closely repeated measurements agree within a site and between sites, over days | EP05, 4th edition, 11 December 2025; quantitative procedures |
| Reportable range | The interval over which results are linear, between the lower and upper limits of quantitation | EP06, 2nd edition, 24 November 2020 |
| Analytical specificity | Whether other substances in the specimen change the result, screened first and then confirmed in patient samples | EP07, 3rd edition, 30 April 2018, reaffirmed October 2022; written for clinical chemistry |
| Accuracy of a quantitative result | The bias between two measurement procedures on the same patient samples | EP09, 3rd edition, 20 June 2018 |
| Analytical sensitivity | The limits of blank, detection and quantitation | EP17, 2nd edition, 18 June 2012, reaffirmed June 2017 |
| Qualitative results | The imprecision of a positive or negative call near its cutoff, and its clinical sensitivity and specificity | EP12, 3rd edition, 7 March 2023 |
| Sequencing tests | Design, development, validation, reporting and continual quality management of sequencing-based genetic tests | MM09, 3rd edition, 20 April 2023 |
CLSI's pages list EP06, EP07, EP09, EP12 and EP17 as recognized by FDA, and they do not state whether FDA recognizes EP05's new edition or MM09. EP05, EP06 and EP09 are written for quantitative procedures, while EP12 covers a binary call such as a variant present or absent. EP07 was written for clinical chemistry, so applying it to the interferents of an omics test, such as hemoglobin released by hemolysis, is the developer's own extension. MM09 introduces bioinformatics and defers detailed guidance on it to a future CLSI document.
For a germline sequencing test, FDA's 2018 guidance names its own measures. Accuracy is reported as positive percent agreement, the share of known variants the test calls; negative percent agreement, the share of known reference positions the test leaves uncalled; and technical positive predictive value, the share of the test's calls that are true. The developer sets a threshold for each, on both the point estimate, the value the study measured, and the lower bound of its 95 % confidence interval, the lowest value of the range that the data support at that level of confidence. Precision is tested across "multiple samples, runs, reagent lots, days, and operators", and the limit of detection for each variant type is the level at which "at least 95% of positive calls" are obtained. A reference material such as RM 8398 can supply the known variants, within the regions its benchmark covers (Chapter 17).
The three figures answer different questions. Negative percent agreement is counted over every reference position the test reads, and those far outnumber the positions that carry a variant, so a few false calls barely move it. Technical positive predictive value is counted over the calls the test makes, so every false call lowers it, and it shows directly what share of the variants in a report are true.
In September 2023 the FDA granted a De Novo request for Invitae's Common Hereditary Cancers Panel, a 47-gene germline sequencing test for inherited cancer risk (Chapter 25 describes the route). FDA's decision summary reports the maker's precision study: 43 samples, each run 14 times across operators, sequencers, reagent lots and days, with every replicate's call compared with the majority call at its position. For single-nucleotide variants the replicates made 28,619 of 28,633 expected variant calls, a positive percent agreement of 99.95 %, and 35,645,506 of 35,645,541 expected reference calls, a negative percent agreement of 99.9999 %. At positions where the majority call was the reference, the other 35 replicate results were therefore not reference calls, and they move negative agreement by about 1 in a million. If all 35 were calls of a variant, they would join the 28,619 agreeing variant calls to make 28,654 variant calls, of which the 35 would be about 1 in 820. The summary gives no technical positive predictive value, and agreement with a majority call measures precision, not accuracy against a benchmark.
The measures CLIA lists and FDA's guidance names are all analytical: each asks whether the test measures its analyte correctly. Whether the result tracks a condition in the intended patients, and whether acting on it helps them, are the clinical rungs of the evidence ladder in Chapter 22, and no analytical study answers them.
The laboratory's own quality system surrounds these studies. The College of American Pathologists, CAP, an accreditation body, added 18 requirements for sequencing tests to the 2014 edition of its molecular pathology checklist. Aziz and colleagues, led from CAP, described them in a report published online in 2014 and in print in 2015. The requirements cover the laboratory bench work and the bioinformatics, and the committee behind them weighed documentation, validation, confirmatory testing, the monitoring of upgrades, version traceability, variant interpretation and reporting, and the storage and transfer of data. CAP's current checklist is not public, so the count describes the 2014 edition.
ISO 15189:2022 is the fourth edition of the international standard for quality and competence in medical laboratories, and it applies to point-of-care testing as well. In a 2026 comparative review, Gupta and Agarwal, writing from a Delhi institute, find that the 2022 edition makes risk management a requirement throughout every process, more strongly than the 2012 edition did. In India, NABL, the National Accreditation Board for Testing and Calibration Laboratories, posted its notice of the move to the 2022 edition on 29 December 2022. It required its specific criteria for medical laboratories, NABL 112A, to be in place by 31 December 2025, checked at each laboratory's next assessment. How each jurisdiction uses these systems is the subject of Part IX.
The evidence assembled before release has to stay true afterwards, as reagent lots, instruments, software and the tested population change, and Chapter 20 follows how: through the control of suppliers and of changes, complaints and corrective action, reports of wrong results, field actions and post-market follow-up.
State the claim in one sentence that names the specimen and its handling limits, the analyte, the kind of result, relative or absolute, and the platform and pipeline version. Give every design input an identifier, an acceptance criterion and the study that will test it, and make every output, risk control, verification record and report line cite the identifiers it serves. Write the criteria for accepting a risk into the risk management plan before the analysis begins, and record for every risk its residual risk after the controls, then the judgment of the overall residual risk against the benefit. Lock the pipeline before validation by fixing every tool version, parameter and reference, and validate it end to end with the laboratory steps.
- A test can be proven only against a claim stated exactly: in the FDA-coordinated SEQC project of 2014, RNA sequencing measured relative expression accurately, with specific filters, but not absolute expression.
- Under the QMSR, in force from 2 February 2026, the design and development clause of ISO 13485 requires the maker of a class II or class III device, and of some class I devices, to turn the claim into inputs, outputs, verification and validation, traced to their records in a design and development file.
- Read with ISO 14971's definitions, a wrong result, which this guide sorts into false negatives, false positives and delayed results, becomes a hazardous situation once it reaches a decision, and the risk left after the controls is judged against criteria set in advance and, as a whole, against the benefit.
- Analytical validation measures precision, range, interference, accuracy and detection with the locked pipeline inside the test, and each agreement figure is counted over its own denominator.