The chemistry in motion.
+ The questionWhy is the layer closest to the phenotype also the hardest to identify?
Small molecules, enormous variety
The Human Metabolome Database listed 217,920 metabolites in its 2022 release, yet most of its entries are molecules expected or predicted to occur in people rather than molecules detected in human samples. Metabolites are the small molecules of life, typically under about 1,500 daltons: sugars, amino acids, fatty acids and other lipids, nucleotides, vitamins, hormones, and the intermediates of every pathway that makes or breaks them. A blood or urine sample also carries molecules the body did not make, such as drug residues, food components, pollutants and the products of gut microbes.
Two features set this layer apart. Metabolites are the direct substrates and products of enzymes, so they reflect what the cell's machinery is actually doing rather than what it could do. They are the closest molecular layer to the phenotype, the observable outcome of a cell or an organism, and a change in diet, drug or disease shows up here within minutes. Metabolites also have no template: nothing in the genome encodes a metabolite the way it encodes a protein's sequence, so the list of what should be present cannot be fully predicted from the genome. Every identification has to rest on evidence from the chemistry of the molecule itself.
In October 2026 the Human Metabolome Database's statistics page listed 253,245 metabolites, more than the 217,920 of its 2022 release. Of these, 3,444 had been detected and quantified in human samples and 20,924 detected but not quantified, so 3,444 plus 20,924, or 24,368, had been detected at all. Dividing 24,368 by 253,245 gives about 0.096, which means that fewer than one entry in ten has been detected in a human sample. The remaining entries, about 229,000, are listed as expected or as predicted rather than detected.
Three instruments, three trade-offs
Liquid chromatography coupled to mass spectrometry, LC-MS, covers the widest range of metabolites, because different columns retain polar or fatty molecules and running the ion source in positive and negative mode captures molecules that prefer to gain or to lose a proton. Gas chromatography with mass spectrometry, GC-MS, handles molecules that are volatile or can be made volatile by chemical derivatization, a reaction that attaches groups which make them evaporate more easily. Its fragmentation is so reproducible that large reference libraries of spectra exist for matching.
Nuclear magnetic resonance, or NMR, works on a different principle entirely. It is the main exception to the five physical properties of Chapter 3, and its spectrum reports the atomic environments in a molecule. NMR is less sensitive than mass spectrometry, typically reaching micromolar rather than nanomolar concentrations. It is quantitative without standards for each compound, it does not destroy the sample and it is highly reproducible between laboratories, so it suits very large studies of the most abundant metabolites and lipoproteins.
Identifying a metabolite takes more than its mass
Chapter 3 noted that mass alone rarely identifies a molecule, and metabolomics is where that limit matters most. Glucose, galactose and fructose, the three sugars of this part's opening, all have the formula C₆H₁₂O₆ and the same exact mass, 180.0634 daltons. A mass spectrometer that weighs them to within about a thousandth of a dalton still cannot tell them apart, because they are isomers: the same atoms, arranged differently. The same problem appears across the whole metabolome, since many lipids differ only in where a double bond sits and many drug metabolites differ only in which position carries a hydroxyl group.
Identification therefore needs a second property. Chromatography separates many isomers, so a molecule's retention time, the moment it emerges from the column, helps to identify it, and so does the way it fragments. The decisive evidence comes from running an authentic standard, a pure sample of the suspected compound, under the same conditions and showing that it matches on at least two independent properties.
An untargeted LC-MS run records whatever ionizes rather than a chosen list, and it detects features: signals each defined by an m/z and a retention time. A typical dataset contains tens of thousands of them. Most are not distinct compounds. The same molecule appears as several features because it picks up different ions such as sodium or potassium, because of natural carbon-13 isotopes, and because some molecules break apart in the ion source, so a count of features overstates the number of compounds present.
Chi and colleagues at the Jackson Laboratory for Genomic Medicine set out to separate real compounds from these duplicate features. They reanalyzed 61 public LC-MS datasets from blood, using the patterns that isotopes and added ions make to group the features that come from one compound. They estimated that each dataset detects about 1,000 to 2,000 compounds with high confidence, that more than half of these compounds are unknown, and that fewer than half match an entry in the Human Metabolome Database. The analysis was posted as a preprint in 2025 and published in a journal in 2026.
In 2007 Sumner and colleagues proposed confidence levels for reporting identifications on behalf of a working group of the Metabolomics Standards Initiative, and those levels remain the working vocabulary:
| Level | Meaning | Evidence |
|---|---|---|
| 1 | Identified | matches an authentic standard, run in the same laboratory, on two or more independent properties such as retention time and fragment spectrum |
| 2 | Putatively annotated | matches a spectral library or literature, without a standard |
| 3 | Putatively characterized compound class | properties consistent with a chemical class, such as a phosphatidylcholine |
| 4 | Unknown | detected and quantified, not identified |
Untargeted results often list compound names without saying how each was assigned. A level 2 annotation can be wrong about which isomer is present, and a level 3 annotation names only a class. A metabolite reported as a biomarker should carry its identification level, and one that will support a decision should be confirmed at level 1.
Targeted metabolomics: the version already in routine use
The opposite approach, targeted metabolomics, measures a predefined set of metabolites, each with an isotope-labeled internal standard of the kind Chapter 8 described, and reports absolute concentrations. It is less exploratory and far more reliable, and it is the form in which metabolomics entered routine medicine long before the word existed. Newborn screening measures amino acids and acylcarnitines, the forms in which carnitine carries fatty acids, in a dried blood spot by tandem mass spectrometry, the two-stage measurement of Chapter 8, to detect inherited metabolic disorders. Laboratories developed the method in the early 1990s, and public screening programs adopted it from the late 1990s. A screening panel is the list of conditions a program tests every newborn for. As of October 2026 the method covers a large share of the conditions on panels such as the US recommended panel.
Handling is part of the measurement
The metabolome changes within seconds to minutes, and it keeps changing after the sample is taken. Blood cells in an unprocessed tube continue to consume glucose and release lactate, red cells that break release their contents, and repeated freezing and thawing degrades unstable molecules. Biological context adds further variation, because fasting state, time of day, recent meals, exercise and medication all move metabolite levels. A metabolomics study that does not standardize collection, processing time, temperature and storage therefore measures its own sample handling as much as the biology of its subjects.
Lipids and glycans
Two sub-fields deserve separate mention because they matter in both biology and manufacturing. Lipidomics measures the thousands of lipid species in membranes, stores and signaling pathways; it uses mass spectrometry and faces an extreme version of the isomer problem. Glycomics measures glycans, the branched sugar chains attached to proteins and lipids. No template encodes glycans: enzymes build them, which makes them heterogeneous, and their branching produces isomers that share a mass and often emerge from the column together. A standard workflow releases glycans from proteins with enzymes before chromatography and mass spectrometry. In therapeutic antibodies the glycan pattern affects how the drug behaves in the body, which is why manufacturers monitor it during production as a critical quality attribute, a property that should stay within set limits to ensure the product's quality.
Write pre-analytical handling into the protocol as acceptance criteria: time from collection to processing, temperature, anticoagulant, freeze–thaw limits, and fasting status. When commissioning untargeted work, ask for identification levels per compound and for the fraction of features annotated at each level. For any metabolite that will support a clinical or release decision, move to a targeted, internally standardized method.
- Metabolites are the layer closest to phenotype, change within seconds to minutes, and have no template to predict them from.
- Isomers share a mass, so identification needs retention time, fragmentation and ultimately an authentic standard.
- Most untargeted features are not distinct compounds, and in blood datasets more than half of the compounds detected with high confidence are unknown.
- Targeted, internally standardized metabolomics has been routine in newborn screening since the late 1990s.