+ Opening

The empty wells.

On 4 February 2020 the US Food and Drug Administration authorized, for emergency use, a test for the new coronavirus that the Centers for Disease Control and Prevention had designed in the weeks since the virus's genome sequence was published. Until then, every test in the country had been run in the CDC's own laboratories. The next day the CDC began shipping kits: about 200 for public health laboratories in the United States and a similar number for laboratories abroad, each kit enough for 700 to 800 specimens. Each kit held the primers and probes for four reactions. Two, called N1 and N2, looked for stretches of the new virus's genome. A third, N3, looked for a broader group of related coronaviruses, and the fourth for a human gene, as a check on the specimen and its extraction.

Before a public health laboratory could report a single patient's result, it had to verify the kit on its own instrument, and the CDC expected that to take a few days. Every run included no-template controls: wells holding everything except a template, which should give no signal in any of the 45 cycles. Within days of the first shipments, laboratories reported light in wells that held no template. NPR later reported that the New York City public health laboratory began verifying the kit at once and, within hours, had inconclusive results from two of its reagents. On 12 February the CDC told reporters that some states had found inconclusive results while verifying the kit, and that one reagent was not performing as it should. On 28 February it told reporters that the third reaction, the cause of the inconclusive results, could be left out.

When CDC scientists later studied material from that first lot, false signals appeared in 97 % of the N3 no-template wells made with the kit's reagents, between cycles 34 and 39. The test called a target detected if its light crossed the detection line before cycle 40, so a well with no virus in it read as detected.

The faults that the CDC's scientists identified lay in the design of N3 and in how the reagents were handled in bulk before they were dispensed into kits, not in the instruments. Each instrument had heated and cooled its wells on schedule, read their light and reported a cycle number. Each reaction had done what every PCR does: it copied whatever its primers started, doubling it cycle after cycle until the light crossed a line. In the N3 wells, the reverse primer and the probe had paired with each other, and the reaction copied that pair. The episode shows the whole subject in a few wells: how two short primers find one sequence among billions, how heating and cooling copy it, how the cycle at which a well's light crosses a line counts it, and what evidence a test needs before anyone trusts its count.

Created with Claude (Anthropic).
Educational reference material, not engineering, regulatory or clinical advice. Product and company names referenced are the property of their respective owners.