16 — Requirements

From intended use to specification.

+ The questionHow does a need such as "detect a few copies in about an hour" become numbers that the instrument's parts must meet?

The intended use comes first

One sentence in the CDC test's instructions stated what the test was for: the qualitative detection of nucleic acid from SARS-CoV-2 in upper and lower respiratory specimens, in laboratories certified for high-complexity testing. That sentence is the test's intended use. Every instrument and test begins with such a statement, and in a regulated product it is written down before anything is designed. It names what is measured, in what kind of specimen, by whom, in what setting, to inform what decision, and what kind of result is given.

The CDC test's intended use decides much of the system. Qualitative means a cut-off and rules, not standards and a quantity. SARS-CoV-2 is an RNA virus, so nucleic acid from SARS-CoV-2 means handling that keeps the RNA intact and a reverse transcription step (Chapter 20). Respiratory specimens means a sample type with its own inhibitors and variability, and therefore an internal control: a target amplified from every specimen to show that the specimen and the extraction were adequate, and that amplification worked in any well it shares. In the CDC test it is the human gene RNase P, in a well of its own (Chapter 21). High-complexity laboratories means trained staff, a separate extraction step and an instrument that can be fed plates by hand. A test intended for a clinic and run by a nurse in twenty minutes would share the chemistry and almost nothing else.

The V

Systems engineers draw the path from a need to a working product as a V. The picture is known as the V model and is credited to Kevin Forsberg and Harold Mooz in 1991. Down the left side, the need is decomposed: user needs into system requirements, system requirements into requirements for each subsystem, and those into detailed designs. At the bottom the parts are built. Up the right side they are assembled and checked: each part against its own requirements, the assembled subsystems against theirs, the whole system against the system requirements, and finally the finished product against the original user needs.

38 — The V
USER NEEDS AND INTENDED USE VALIDATION detect SARS-CoV-2 RNA in respiratory specimens, in high-complexity labs clinical performance study SYSTEM REQUIREMENTS SYSTEM VERIFICATION detect about 5 copies per reaction; run in about an hour verification plate SUBSYSTEM REQUIREMENTS SUBSYSTEM VERIFICATION ramp rate, uniformity, read time, channels thermal and optical measurements DETAILED DESIGN PART VERIFICATION block, modules, filters, firmware module and board tests BUILD THE PARTS validation: the right product (meets user needs) verification: built as specified (outputs meet inputs) validation answers the intended use Forsberg and Mooz, 1991; FDA design controls, 1997; QMSR with ISO 13485:2016 from 2 Feb 2026
Focal detail
Plate 38 — Down the left side a need is decomposed into system and subsystem requirements and designs; up the right side the parts are assembled and checked against them. Verification shows the product was built as specified; validation, at the top, shows it is the right product for the user's needs and intended use.

The two checks on the right side have names that regulators use precisely. The FDA's 1997 guidance on design controls defines verification as confirming that the design's outputs meet its inputs, that the product was built as specified. It defines validation as confirming that the device meets the user's needs and intended uses, that the right product was built. The same guidance asks that design inputs be developed to an engineering level of detail, and it notes that real development loops back between phases. Together these documented steps, from design inputs to verification and validation, are called design controls, and in the United States they are now required through the Quality Management System Regulation, the FDA's rule for the quality systems of device makers. The regulation took effect on 2 February 2026 and incorporates the international quality standard ISO 13485:2016, including its clause on design and development.

Requirements flow down

The left side of the V is where a need such as "detect a few copies in about an hour" becomes numbers. Each system requirement splits into requirements on several subsystems, and most subsystems answer to several system requirements. The example below is illustrative and is built from figures in earlier chapters.

39 — Requirements flow down
SYSTEM REQUIREMENTS WHAT EACH ASKS OF THE PARTS SENSITIVITY about 5 copies per reaction TIME TO RESULT about an hour THROUGHPUT specimens per run SPECIFICITY AND ROBUSTNESS EXTRACTION recover RNA, remove inhibitors OPTICS see 10¹⁰–10¹¹ copies above background; read time; channels ASSAY AND SAMPLE INPUT 5 µL extract in 20 µL; 45 cycles FIRMWARE AND SOFTWARE sample-temperature model THERMAL sample ramp rate; accuracy and uniformity at anneal and denature PLATE LAYOUT 3 wells per specimen + 3 controls → 29 specimens per 96 wells the optics answer to three requirements illustrative, built from figures in earlier chapters
Focal detail
Plate 39 — Each system requirement splits into requirements on several subsystems, and most subsystems answer to several system requirements. Sensitivity reaches the assay, extraction, optics and cycle count; time reaches the ramps, the sample-temperature model and the read; throughput becomes a plate layout: three wells per specimen leave room for 29 specimens on a 96-well plate.

Sensitivity. A requirement to detect about five copies per reaction, the figure of Chapter 1, sits just above the sampling floor of about three copies, below which no limit can go (Chapter 6). The assay must therefore put enough of the specimen into each reaction: in the CDC test, 5 microliters of extract in a 20-microliter reaction. The extraction must recover the RNA efficiently and remove inhibitors. The optics must distinguish the light of 10¹⁰ to 10¹¹ copies of product from the background reliably enough that the threshold can sit low. The run must also have enough cycles for a handful of copies to cross that threshold at realistic efficiency. In the CDC test the cut-off was at cycle 40 of 45.

Time to result. The run time is the number of cycles times the cycle time, plus the initial steps. The cycle time is the hold times plus the time the liquid takes to move between temperatures twice, plus any time the read adds. So the time requirement becomes a requirement on the sample ramp rate, on the read time and on the sample-temperature model of Chapter 10, with which the controller brings the liquid itself to each set point so that a short hold is a real one.

Throughput. A requirement for a number of specimens per run becomes a plate format and a layout. If each specimen takes three wells, as in the CDC test from March 2020, and three controls take three wells each, a 96-well plate holds 29 specimens. A multiplex test that detects all targets in one well with different dyes needs one well for each specimen and each control, so it would hold 93, about three times as many, at the cost of more optical channels, more calibration and the separation of overlapping dye signals (Chapter 12).

Specificity and robustness. The assay's tolerance to temperature error becomes a requirement on accuracy and uniformity at the annealing and denaturing steps, and the instrument's guaranteed accuracy and uniformity become a requirement on the assay's robustness.

Solved example: the CDC test's run time in standard and fast mode

The CDC test's cycle holds 3 s at 95 °C and 30 s at 55 °C, so the liquid swings 40 degrees each way. The instrument's maker states in its brochure that the sample ramps at 1.6 °C per second in standard mode and 3.5 °C per second in fast mode. In standard mode each ramp takes 40 ÷ 1.6 = 25 s, so a cycle takes about 3 + 30 + 25 + 25 = 83 s, and 45 cycles take about 62 minutes. In fast mode each ramp takes 40 ÷ 3.5, or about 11 s, so a cycle takes about 56 s and 45 cycles about 42 minutes. The TaqPath mix adds 19 minutes of initial holds, so the run takes about 81 minutes in standard mode or about 61 in fast mode, not counting the read time. In standard mode the two ramps take 50 of the 83 seconds, more of the cycle than the holds, while in fast mode the holds take the larger share. The test's instructions required the standard mode, so that is how the test had to run.

Error budgets

Where a requirement depends on several contributions, engineers allocate it among them in an error budget. Temperature at the annealing step is a clear case. The actual temperature of a given well differs from the set point by the instrument's accuracy, plus that well's departure from the average, plus any residual error from the sample-temperature model, plus the uncertainty of the sensor's calibration.

One manufacturer specifies ±0.2 °C accuracy and ±0.4 °C uniformity within 10 seconds of arrival, both stated at 90 °C. Those two figures alone put a well on that instrument up to about 0.6 °C from its set point at 90 °C before the model and calibration errors are counted, and the specification does not say what holds at the annealing step. The maker of the CDC's instrument gives ±0.5 °C accuracy and ±1 °C uniformity in its brochure, figures that allow about 1.5 °C. The brochure does not say at what temperature or how soon after arrival they were measured, so the two makers' figures cannot be compared directly. Engineers often combine independent errors by adding their squares and taking the square root, which gives a smaller, more likely figure, but a diagnostic test must work in the worst well.

40 — Budgets for time and temperature
CYCLE TIME, CDC PROFILE STANDARD MODE 1.6 °C/s ramp down 25 s 55 °C hold 30 s ramp up 25 s 95 °C hold 3 s = 83 s; ×45 ≈ 62 min; with 19 min of initial holds ≈ 81 min FAST MODE 3.5 °C/s 11 s 30 s 11 s ≈ 56 s; ×45 ≈ 42 min; ≈ 61 min in all 0 10 20 30 40 50 60 70 80 90 s ignoring read time; the instructions required standard mode ramps take 50 of 83 s TEMPERATURE ERROR, WORST WELL CDC TEST'S INSTRUMENT brochure accuracy ±0.5 uniformity ±1 ≈ 1.5 °C ANOTHER MAKER 90 °C, 10 s after arrival ±0.2 ±0.4 ≈ 0.6 °C the assay's tolerance must exceed this, with margin 0 0.5 1.0 1.5 °C
HeatFocal detail
Plate 40 — In the CDC test's cycle, two 40-degree ramps take 50 of the 83 seconds in standard mode, so in standard mode the ramps take more of the cycle than the holds: about 81 minutes in all, against about 61 in fast mode. Temperature error adds up the same way: brochure accuracy and uniformity allow a well to sit about 1.5 °C from its set point, and the assay's tolerance must exceed that worst case.

The temperature budget closes on the assay side. A test developer establishes how far the annealing temperature can move before the test loses sensitivity or specificity, typically by running the test deliberately at temperatures above and below the set point. That tolerance must exceed the instrument's worst-case error with margin. The optical budget closes the same way. Background, noise and calibration error set the lowest signal the instrument can distinguish, and that signal sets where the threshold can sit. The position of the threshold in turn sets how precisely Cq is determined, and that precision is what the verification plate of Chapter 15 checks.

Trade-offs

Requirements pull against each other, and the systems engineer's job is to make the trade explicit rather than let it happen by accident. Speed pulls against uniformity, because a lighter block changes temperature faster and spreads heat less evenly. More optical channels allow more targets per well, but they cost more, demand more calibration and leak more light between channels. More cycles reach lower copy numbers but lengthen the run. A sealed cartridge prevents contamination and simplifies the user's work, but it fixes the assay and the volume. Every one of these trade-offs is a decision about which requirement matters more for the intended use, and every one shows up later as a specification someone must verify.

Trace every number to a need

Each figure in an instrument specification should trace up to a system requirement and from there to the intended use, and down to a verification method that will show it is met. That holds for a ramp rate, a uniformity, a read time and a channel count alike. A number that traces to nothing is either unnecessary or a sign of a requirement nobody wrote down.

+ What this chapter established
  • The intended use states what is measured, in what specimen, by whom, where and for what result, and it decides much of the system before any design begins.
  • The V model decomposes needs into system and subsystem requirements and then checks upward, with verification showing that the product meets its specifications and validation showing that it meets the user's needs.
  • Sensitivity, time, throughput and robustness each flow down to several subsystems, and in the CDC test's standard mode the ramps take more of each cycle than the holds do.
  • Error budgets allocate a requirement among its contributors and close against the assay's tolerance, and trade-offs between speed, uniformity, channels and cycles are made explicitly.