§ Lessons
Lessons.
The chapters reduce to a short set of working rules for anyone who designs, verifies, buys, regulates or writes about PCR instruments and tests. Each one traces back to a mechanism explained earlier.
- Specify sensitivity in copies per reaction. The chemistry, the optics and the statistics of sampling act on the molecules that reach the well, not on the concentration in the specimen (Chapters 1, 6 and 16).
- Add sample, not cycles, to detect rarer targets. Below about three copies per reaction more than one reaction in twenty receives none, and no number of cycles or brightness of dye changes that (Chapters 6 and 21).
- Check every oligonucleotide against every other, at the 3′ ends above all. Any two that can pair and be extended can make a product in any tube, including those with no target (Chapters 5 and 22).
- Read quantity in the exponential phase or count it in partitions, never from one reaction's plateau. The plateau forgets the starting amount; the cycle at which a curve rises remembers it, and so does the fraction of partitions that stay dark (Chapters 8, 11, 13 and 14).
- State efficiency with every quantity. A tenfold difference is 3.32 cycles only at perfect doubling, and unequal efficiencies masquerade as unequal amounts (Chapters 6 and 13).
- Never compare Ct values across tests, instruments or thresholds. A Ct is a property of one validated test system, not a unit of viral load (Chapter 13).
- Measure temperature in the wells, and time holds from the sample's arrival. The controller reads the block; specifications need their definitions and conditions, and real instruments can start a hold while the liquid is still short of the set point (Chapters 9 and 10).
- Trace every reported number back through its conversions. Copies become light, electrons, counts, corrected values, a curve, a Cq and a call, and each step needs its own calibration and check (Chapters 12 and 15).
- Own the interfaces. Plate, seal, volume, dyes, read timing, data formats and software version are where instruments and assays fail together (Chapters 15 and 17).
- Freeze the test system and treat any change as a change to the test. Instrument, software, reagents, extraction, settings and rules were validated together and must be re-shown together (Chapters 17 and 22).
- Put an internal control in every specimen. It turns a poor swab, a failed extraction or an inhibitor from a false negative into an invalid result, and a cold well too when it shares the target's well (Chapters 19, 20 and 21).
- Read predictive value with the prevalence of the people tested. The same test can be right about two positives in three in one population and about forty-nine in fifty in another (Chapter 21).
- Build one evidence core and present it to each regulator. Analytical and clinical studies, risk, software and safety evidence serve the US, the EU and India; keep filed, cleared, granted, authorized, approved and certified apart (Chapters 19, 25 and 26).
- Read a product's status in the regulator's record, with its date. Cleared, granted, waived and authorized for emergency use are different records, a company's statement is not one of them, and an emergency authorization lasts only as long as the declaration behind it (Chapters 24 and 25).
Let's Talk