§ Glossary
Glossary.
Terms are defined as they are used in this guide.
- 510(k)
- A US premarket notification showing that a device is substantially equivalent to a legally marketed device; a cleared 510(k) allows marketing.
- Accuracy (temperature)
- How close the average temperature of the wells is to the set point; distinct from uniformity.
- Amplicon
- The stretch of DNA between and including the two primers, which PCR copies exponentially; also the product itself.
- Analyte
- The substance or marker a test detects or measures, such as a virus's RNA in a respiratory specimen.
- Analytical performance
- A test's ability to detect or measure its analyte correctly: limit of detection, specificity, precision and related characteristics.
- Annealing
- The re-pairing of complementary strands on cooling; in PCR, the step in which primers bind their targets.
- Baseline
- The background fluorescence of the early cycles, subtracted from each curve before a threshold is applied.
- BCR-ABL1
- The fusion gene of chronic myeloid leukemia, formed by an exchange between chromosomes 9 and 22, whose RNA is measured to monitor treatment.
- cDNA
- Complementary DNA: DNA copied from an RNA template by reverse transcriptase.
- CDSCO
- The Central Drugs Standard Control Organisation, which publishes the class of each diagnostic device type in India and through its central licensing authority licenses the manufacture of class C and D devices.
- CLIA
- The US Clinical Laboratory Improvement Amendments, which regulate clinical laboratories and sort tests into waived, moderate-complexity and high-complexity categories.
- Clinical performance
- How well a test's results correspond to the clinical condition in the intended population: sensitivity, specificity and predictive values.
- Clinical utility
- Whether using a test improves outcomes for patients; it matters to doctors and payers but is generally not required for market authorization.
- Common specifications
- EU requirements set by implementing regulation for the performance of certain class D diagnostic devices.
- Cq (Ct)
- The fractional cycle at which a well's fluorescence curve crosses the threshold; MIQE's quantification cycle, called threshold cycle (Ct) or crossing point (Cp) by some manufacturers.
- CRISPR detection
- Detection using a guided bacterial enzyme that, on binding its target sequence, cuts nearby reporter molecules to produce a signal.
- De Novo
- The US route for a new type of low-to-moderate-risk device with no predicate; a granted request creates a new classification.
- Denaturation
- Separating the two strands of DNA, in PCR by heating to 94 to 98 °C.
- Digital PCR
- PCR in which the sample is divided into thousands of partitions that are each read as positive or negative, giving an absolute count of molecules.
- dNTP
- Deoxynucleoside triphosphate: one of the four building blocks the polymerase adds to a growing DNA strand.
- Dual Track
- In the FDA's 510(k) records, a submission that combines a 510(k) with an application for a CLIA waiver.
- dUTP and UNG
- A contamination defense: products are made with uracil, and uracil-DNA glycosylase destroys any carried-over product before the next PCR.
- Efficiency
- The fraction of target molecules copied in one cycle; at 100 % the target doubles each cycle.
- US authorization of an unapproved product during a declared emergency, lasting until the declaration ends or the authorization is revoked.
- Endpoint PCR
- PCR in which the product is examined only after the last cycle, typically on a gel.
- Error budget
- The allocation of a requirement, such as a temperature tolerance, among the separate sources of error that contribute to it.
- EU reference laboratory
- A laboratory designated under the IVD Regulation to verify the performance of class D devices and their compliance with common specifications, and to test their batches.
- Exclusivity
- A test's failure to react with organisms other than its target; tested as cross-reactivity.
- Extension
- The step in which the polymerase lengthens a bound primer along its template.
- Extraction
- Releasing and purifying nucleic acid from a specimen, removing inhibitors and often concentrating it.
- Fluorophore
- A molecule that absorbs light of one color and emits light of a longer wavelength.
- Heated lid
- The plate above the tubes held at about 105 °C to prevent condensation and to press the tubes into the block.
- Hot start
- Any method that keeps the polymerase inactive until the first heating step, to prevent wrong products forming during set-up.
- Hydrolysis probe
- A probe carrying a fluorophore and a quencher that the polymerase cuts as it copies the target, releasing fluorescence; often called a TaqMan probe.
- Inclusivity
- A test's ability to detect all variants of its target organism.
- Inhibitor
- A substance carried in with a specimen or from extraction that slows or stops amplification.
- Instrument family
- Instruments from one manufacturer with the same general architecture, design, tolerances and capabilities, for the purpose of extending a cleared assay.
- Intended use
- The written statement of what a device measures, in what specimen, for whom, by whom, where and for what decision.
- Interface
- A boundary between two parts of a system, with parameters both sides must agree on.
- Internal control
- A target amplified from every specimen to show that the specimen and extraction were adequate, and amplification too in any well it shares; in the CDC test, RNase P, in a well of its own.
- Isothermal amplification
- Copying nucleic acid at a single temperature, using enzymes that separate strands as they copy.
- IVDR
- Regulation (EU) 2017/746 on in vitro diagnostic medical devices, applicable since 26 May 2022.
- Laboratory-developed test
- A test made and used within one laboratory.
- LAMP
- Loop-mediated isothermal amplification: amplification at about 65 °C with a strand-displacing polymerase and four to six primers.
- Legacy device
- In the EU, a diagnostic device placed on the market under the directive that preceded the IVD Regulation, allowed to stay on the market until a deadline set by its class if its maker applied to a notified body in time.
- Limit of blank
- The highest result expected from a sample containing no analyte.
- Limit of detection
- For a qualitative molecular test, the lowest concentration detected in at least 95 % of replicates.
- Long product
- A strand made in the early cycles from an original template, with one end defined by a primer and the other ragged.
- Lower limit of quantification
- The smallest amount a quantitative test can measure with acceptable precision.
- Master mix
- The ready-made mixture of polymerase, dNTPs, magnesium and buffer to which primers, probes and sample are added.
- MDCG 2020-16
- The Medical Device Coordination Group's guidance on applying the IVD Regulation's classification rules; it is not legally binding.
- Melt curve
- A plot of how fluorescence falls as the product is heated slowly; peaks mark the melting temperatures of the products present.
- Melting temperature (Tm)
- The temperature at which half of a DNA duplex has separated into single strands; it depends on sequence, salt, magnesium and strand concentration.
- MIQE
- Minimum Information for Publication of Quantitative Real-Time PCR Experiments: reporting guidelines published in 2009 and revised in 2025.
- Mispriming
- A primer binding to and being extended from a partly matching site.
- Multicomponent analysis
- Separating the contributions of several dyes to measured light using each dye's calibrated spectrum.
- Multiplex
- Detecting several targets in one reaction, usually with a different fluorophore for each.
- No-template control
- A reaction with every reagent but no sample, which should show no amplification; any signal reveals contamination or a product the reagents make on their own, such as a primer-dimer.
- Notified body
- An organization designated in the EU to assess the conformity of devices in classes that require third-party assessment.
- Oligonucleotide (oligo)
- A short synthetic strand of DNA; primers and probes are oligonucleotides.
- Open platform
- A real-time PCR instrument that runs any assay written for it, a kit maker's or a laboratory's own.
- Partition
- One of the thousands of small separate reactions into which digital PCR divides a sample.
- Passive reference
- An inert dye such as ROX added to every well, by which the reporter signal is divided to correct for volume and optical differences.
- Peltier module
- A thermoelectric device that pumps heat from one face to the other in the direction set by its current.
- Performance evaluation (India)
- Testing of a new class B, C or D diagnostic device on three batches made from different lots of key raw materials, by a laboratory set up or designated by the central government or accredited by the national accreditation board.
- Plateau
- The final phase of a PCR, in which product stops increasing; reached at similar levels from very different starting amounts.
- Poisson distribution
- The statistics of items scattered at random, which govern how many molecules land in a reaction or partition.
- Polymerase
- The enzyme that builds a new DNA strand by adding nucleotides to a primer, copying a template.
- Positive and negative percent agreement
- Agreement of a new test with a comparator that is not a reference standard; not to be reported as sensitivity and specificity.
- Positive control
- A reaction containing the target at a known level, which must be positive for the run to be valid.
- Predicate
- The legally marketed device to which a 510(k) shows a new device to be substantially equivalent.
- Predictive value
- The probability that a positive (or negative) result is correct; it depends on the prevalence in the people tested.
- Prequalification (WHO)
- WHO's assessment of a named diagnostic product, separate from a WHO recommendation that a type of test may be used for a purpose.
- Primer
- A short designed piece of DNA that binds the template and gives the polymerase a starting point; PCR uses two, facing each other.
- Primer-dimer
- A short product formed when two primers pair with each other and are extended.
- Probe
- A short labeled oligonucleotide that binds inside the target and reports its presence by fluorescence.
- QMSR
- The US Quality Management System Regulation, 21 CFR 820, in force since 2 February 2026, incorporating ISO 13485:2016.
- Quencher
- A molecule that absorbs a nearby fluorophore's energy and keeps it dark until they are separated.
- Ramp rate
- The rate of temperature change, in degrees per second; defined variously for the block or the sample, as a maximum or an average.
- Real-time PCR
- PCR in which fluorescence is read in every cycle, so the cycle at which the signal rises can be used to infer the starting amount.
- Research use only
- A label for products that may not be used to diagnose patients.
- Reverse transcriptase
- An enzyme that copies RNA into complementary DNA.
- RNase
- An enzyme that degrades RNA; robust and widespread.
- RNase P
- A human gene amplified as the CDC test's internal control.
- RPA
- Recombinase polymerase amplification: isothermal amplification at about 37 °C using recombination proteins to insert primers.
- Sample-temperature model
- The controller's calculation of the liquid's temperature from the block's history, the volume and the tube's time constant.
- Scientific validity
- The association of an analyte with a clinical condition; in the EU's IVD Regulation, the first of the three questions a diagnostic's evidence answers.
- Sensitivity (clinical)
- The fraction of people with the condition whom a test calls positive, measured against a reference standard.
- Special controls
- Requirements written into a US classification regulation that every class II device of that type must meet, such as the studies and labeling required for SARS-CoV-2 multi-target tests.
- Specificity (clinical)
- The fraction of people without the condition whom a test calls negative, measured against a reference standard.
- Stacking
- The attraction between neighboring base pairs along the helix, which provides most of a duplex's stability.
- Standard curve
- A plot of Cq against the logarithm of known starting amounts, used to convert an unknown's Cq into an amount and to estimate efficiency.
- Standard evaluation protocol
- A protocol published by ICMR and CDSCO for evaluating one type of diagnostic kit; a compendium of 39 was released in September 2025.
- Syndromic panel
- A test that checks one specimen for the many pathogens that can cause the same symptoms, such as a respiratory infection.
- Taq polymerase
- The heat-stable DNA polymerase of the bacterium Thermus aquaticus, used in most PCR since 1988.
- Test system
- The assay, instrument and software together, evaluated and authorized as one.
- Thermal cycler
- The instrument that heats and cools PCR reactions on a programmed schedule; often called a PCR machine.
- Threshold
- The fluorescence level, set above background in the exponential region, at which Cq is read.
- Time constant
- For a first-order lag, the time the sample takes to cover about 63 % of a step in temperature.
- Uniformity (temperature)
- How closely the wells of a block agree with one another at a set point; distinct from accuracy.
- V model
- The systems-engineering picture of decomposing needs into requirements and designs, then verifying and validating upward.
- Validation
- Showing that a device meets its users' needs and intended use.
- Verification
- Showing that a design's outputs meet its inputs: that the product was built as specified.
- Waived test
- Under CLIA, a test that meets separate criteria of simplicity and low risk of error; other tests are scored as moderate or high complexity.