+ Part VII · What comes next
The cost of the cycle, and what removes it.
Abbott states that its point-of-care COVID-19 test, ID NOW, reports results in 13 minutes or less, and the test does so without thermal cycling, by amplifying its target at a single temperature. PCR's costs are time, a laboratory and an instrument that can move heat quickly and evenly. This part looks at three kinds of method that cut those costs: faster cycling, amplification at a single temperature, and enzymes that detect sequences directly. It asks what each gives up, and it ends by setting out what remains unsolved.
Faster, smaller, simpler.
+ The questionIf PCR's costs are time, a laboratory and an instrument, what do faster and simpler methods give up to remove them?
Faster cycling
In 2015 Carl Wittwer's group amplified short human DNA targets through 35 cycles in 15 to 60 seconds. The cycle takes time because heat takes time to move (Chapters 9 and 10). It shortens when the volume and the distance that heat must travel shrink, and when higher concentrations let each step finish in less time. In the group's method, which it called extreme PCR, the primer and polymerase concentrations were 10 to 20 times above normal, the samples were 1 to 5 microliters and the cycles lasted 0.4 to 2 seconds. The targets were 45 to 102 base pairs long, and they amplified at efficiencies of about 92 to 96 %.
In a 2015 demonstration of photonic PCR, blue light heated a thin gold film under the wells, and the device completed 30 cycles in 5 minutes. A 2019 commentary described the pairing of extreme PCR with rapid melting analysis as a step toward molecular diagnosis while the patient waits.
What faster cycling gives up is visible in the figures: short targets, small volumes and much more reagent per reaction. A smaller volume holds fewer target molecules at a given specimen concentration, so the reaction works closer to the sampling floor of Chapter 6, and high reagent concentrations raise the cost and the risk of side reactions. Fast cycling is also only as good as the temperature control that delivers it. A three-second hold achieves nothing if the liquid never reaches the set temperature.
The 2015 photonic device ran 30 cycles between 55 and 95 °C in 5 minutes, and its authors reported heating at about 12.8 °C per second and cooling at about 6.6 °C per second. Five minutes is 300 seconds, so each of the 30 cycles had 10 seconds. In each cycle the temperature rose and fell by 40 °C. Heating took 40 divided by 12.8, about 3.1 seconds, and cooling took 40 divided by 6.6, about 6.1 seconds, so changing temperature took about 9.2 seconds. That leaves under 1 second of each 10-second cycle for the holds, assuming the rates held across each swing and the 5 minutes went to the cycles alone. At these rates, changing temperature filled more than nine-tenths of the run, and the slower cooling alone filled about three-fifths of it.
Amplifying at one temperature
The other route removes the cycle altogether. Isothermal amplification methods copy nucleic acid at a single temperature, using enzymes that separate strands as they copy instead of relying on heat to separate them. One of them is loop-mediated isothermal amplification, or LAMP, published in 2000 by Tsugunori Notomi and colleagues, most of them staff of the Japanese company Eiken Chemical. LAMP runs at about 65 °C with a polymerase that displaces the strand ahead of it as it copies, and with four primers that recognize six distinct regions of the target. The primers are designed so that the products fold into loops that serve as new starting points, and in the authors' experiments the reaction accumulated about a billion copies in under an hour.
Recombinase polymerase amplification, or RPA, was published in 2006 by a team mostly from ASM Scientific, a company that declared a financial interest in the method, and it pushes primers into double-stranded DNA at 37 °C with proteins borrowed from the machinery that cells use to repair and recombine DNA. In that 2006 work, RPA detected fewer than ten copies of genomic DNA in under 30 minutes. Other isothermal schemes use enzymes that nick, or cut, one strand of the DNA to start each round of copying. ID NOW, the test at the opening of this part and in Chapter 18, amplifies its target isothermally, and Abbott states that it gives results in 13 minutes or less.
Removing the cycle removes the instrument's hardest job: the instrument needs only to hold one temperature, which a simple heater can do, so isothermal tests suit small, cheap, battery-powered readers. The price is paid in the chemistry. Many primers in one tube give many chances for primers to interact, and isothermal reactions are prone to amplifying something even without a target. Without cycles, there is no Cq built on doubling per cycle, so quantification is harder. And the design of four to six primers per target constrains where on a genome a test can be placed.
A method that reaches a result in minutes rather than an hour usually does so with a different chemistry, a smaller volume, fewer cycles or a different target. Its limit of detection, specificity and failure modes must be established on their own, not inferred from the slower test it replaces.
Enzymes that recognize a sequence
A third approach, CRISPR detection, adds sequence recognition by the bacterial immune enzymes known as CRISPR systems. Some of these enzymes are guided by a short RNA that matches a target, and once they bind their target they become indiscriminate cutters, cutting any nearby single-stranded nucleic acid. When the enzymes are mixed with a reporter molecule that fluoresces once it is cut, the recognition of one sequence becomes a signal.
Published in 2017, the method called SHERLOCK used the enzyme Cas13 after a pre-amplification by RPA, and its developers reported sensitivity down to attomolar concentrations, on the order of one molecule per microliter, with the ability to distinguish single-base differences. DETECTR, published in 2018, used Cas12a and distinguished two closely related types of human papillomavirus. In May 2020 the FDA issued an emergency authorization for a CRISPR-based SARS-CoV-2 test, the Sherlock CRISPR SARS-CoV-2 Kit, and its maker, Sherlock Biosciences, called it "the very first FDA-authorized use of CRISPR technology".
CRISPR detection recognizes a sequence but does not amplify it. On its own it is not sensitive enough for a few copies, so it is usually paired with an isothermal amplification step, and it inherits that step's strengths and weaknesses. What it adds is a second check of sequence identity, like a probe's, in a format that a simple fluorescence reader or a paper strip can read.
Where PCR stays
None of these methods has displaced PCR, and the reasons are the subject of this guide. PCR's chemistry is understood in detail, and its primers are simple to design. Its exponential phase gives a Cq that can be turned into a quantity, its digital form counts molecules, and its instruments, assays, controls and regulatory routes are mature. Most laboratory molecular diagnostics run on real-time PCR, and sample-to-answer cartridges have brought it close to the patient. Digital PCR measures rare variants in blood, and PCR prepares the DNA for most sequencing. The newer methods find their places where time, cost or portability matter more than quantification and the breadth of established tests.
What remains unsolved
Several problems are common to every method in this chapter, and none is solved. Sample preparation remains the slowest, least standardized and least integrated step outside the cartridges of Chapter 18: amplification takes minutes, but getting clean nucleic acid out of a specimen takes skill or an expensive disposable. Comparability remains elusive, because a Ct from one test cannot be compared with a Ct from another (Chapter 13). The field's move toward vendor-neutral raw data and efficiency-corrected quantities is recent. Evolving targets are a lasting problem: every test that detects a virus must be checked against each new variant for as long as the test is used. And cost per result still separates what is technically possible from what a laboratory, a clinic or a health system can provide at scale.
This guide asked how heating and cooling a sealed tube can find one stretch of DNA among billions, copy it a billion times and count how much of it the sample held. The chemistry answers the first two parts. The last part depends on the engineering of the instrument as a system and on the evidence built around the test, and so does the trust a laboratory places in the answer. Most of the remaining work lies in those two.
Before choosing a faster or simpler method, name which of PCR's costs matters for the intended use: time to result, the need for a laboratory, the price of the instrument, the price per test or the skill to run it. Each alternative removes some of these and adds others, and the right choice follows from the intended use, as every other decision in this guide does.
- Extreme PCR has completed amplification in a minute or less and photonic PCR in five minutes, at the cost of short targets, small volumes and high reagent concentrations.
- Isothermal methods such as LAMP at about 65 °C and RPA at 37 °C remove thermal cycling and suit simple readers, but they are harder to design, more prone to non-specific amplification and harder to quantify.
- CRISPR enzymes add sequence recognition, usually after an amplification step, and in May 2020 a CRISPR-based SARS-CoV-2 test became, by its maker's account, the first FDA-authorized use of the technology.
- PCR remains central, while sample preparation, comparability of results, evolving targets and cost per result remain unsolved.