How one becomes a billion.
+ The questionDoubling thirty times makes a billion. Why do real runs need closer to forty cycles?
Products become templates
In each cycle, every target strand in the tube can bind a primer and be copied, and every new strand contains the binding site for the other primer, so in the next cycle it is a template too. If every target is copied in every cycle, the count doubles: 1, 2, 4, 8, 16. Ten cycles give 2¹⁰ = 1,024, twenty give about a million, thirty give 2³⁰, about 1.07 × 10⁹, and forty give about 1.1 × 10¹². This arithmetic is behind the claim that PCR makes a billion copies in thirty cycles, and it is the forward reading of the fact stated at the start of this guide: a cycle can at most double the number of targets.
Doubling is the ceiling, and real reactions sit below it. The fraction of targets actually copied in a cycle is the efficiency (Chapter 4). At 100 % efficiency every target is copied and the count doubles. At 90 % efficiency, nine targets in ten are copied and the count is multiplied by 1.9. Each cycle multiplies the count by one plus the efficiency, so after a run of cycles the reaction holds the starting number multiplied by that factor once for every cycle.
Efficiency compounds
A shortfall of a few percent per cycle sounds harmless. Compounded over thirty cycles, it is not.
At 100 % efficiency each cycle multiplies the target by 2, so 30 cycles multiply it by 2³⁰, about 1.07 × 10⁹. At 90 % efficiency each cycle multiplies it by 1.9, so 30 cycles multiply it by 1.9³⁰, about 2.3 × 10⁸. Dividing the first figure by the second gives about 4.7, so the reaction at 90 % ends with about 4.7 times fewer copies. In 1985 the first published PCR, by scientists at Cetus (Chapter 3), copied about 85 % of the target per cycle. Its 20 cycles multiplied the target by 1.85²⁰, about 220,000, where perfect doubling would have given 2²⁰, about a million.
Doubling has a shape that is worth learning now, because Part IV depends on it. Plotted on an ordinary scale, an exponential curve hugs zero for most of its length and then shoots up, so the early cycles look like nothing is happening. Plotted on a scale where each step up the axis is a factor of ten, the same curve is a straight line. Perfect doubling climbs one tenfold step every 3.32 cycles, because 3.32 doublings make a factor of ten. At 90 % efficiency the line climbs one step every 3.59 cycles. The steepness of the line is set by the efficiency, and a sample that starts with ten times more target runs along a parallel line, 3.32 cycles ahead at perfect doubling and 3.59 cycles ahead at 90 %.
Efficiency falls short of 100 % for several reasons that recur through this guide. Some strands are not fully separated in a short denaturing step, some primer sites are not occupied at the annealing temperature, some extensions do not finish, and substances carried in from the sample slow the enzyme (Chapter 20). Because every cycle applies the per-cycle factor again, the effect of any difference in efficiency between two reactions grows with every cycle. That matters for counting, because two samples with the same starting amount but different efficiencies reach the detection line at different cycles (Chapter 13).
Where the forty comes from
The number of cycles a test needs is set by three things: how few molecules it must detect, how many molecules the detector needs to see, and the efficiency in between.
A real-time instrument's detection line sits at around 10¹⁰ to 10¹¹ copies of product per reaction (Chapter 1). From a single copy, perfect doubling would reach 10¹⁰ in about 33 cycles. Real reactions do not double perfectly, and they rarely start from exactly one copy. The field's reference for reporting real-time PCR is a set of guidelines called MIQE, short for Minimum Information for Publication of Quantitative Real-Time PCR Experiments. Their 2025 revision gives worked figures: at 90 % efficiency, 10 starting copies cross a typical detection line at about cycle 35 and 3 copies at about cycle 37, while at 70 % efficiency the same inputs cross at about cycles 42 and 45.
A run of 40 to 45 cycles therefore covers the handful of molecules a sensitive test must detect, even with an efficiency below perfect, and little is gained by going further. The CDC test uses 45 cycles and counts a target as detected only if it crosses the line before cycle 40.
Adding cycles beyond about forty does not lower the floor that random sampling sets on detection, and it gives rare side reactions, such as primer-dimers and traces of contamination, more chances to produce a signal. That is one reason tests set a cut-off below the last cycle, as the CDC test does at cycle 40 of 45, and why a late signal in a no-template control invalidates a run.
The floor no instrument can lower
There is a limit below which no number of cycles helps, and it is set not by chemistry but by sampling. When a few microliters of extract are pipetted into a reaction, the number of target molecules that land in the tube is a matter of chance. If the extract contains an average of one target molecule per 5 microliters, some reactions receive none, some one, and some two or three.
The chance that a reaction receives no molecules follows a simple rule, the Poisson distribution, which describes items scattered at random. With an average of one molecule per reaction, 37 % of reactions get none and must be negative however good the chemistry. Each additional molecule in the average multiplies that chance by 0.37 again: about 14 % for an average of two, about 5 % for three. A test's limit of detection is usually defined as the lowest amount of target that it detects in at least 95 % of repeated reactions.
For an average of 1 copy per reaction, the chance that a reaction receives no target is 37 %. For 2 copies it is 0.37 × 0.37, about 14 %. For 3 copies it is 0.37³, about 5 %, and for 5 copies it is 0.37⁵, about 0.7 %. The exact factor is 0.368, which is one divided by the number e, about 2.718. A test that must be positive at least 95 % of the time at its limit of detection therefore cannot have a limit below about three copies per reaction, and the MIQE guidelines state this floor explicitly. The CDC's paper on its test reported detecting about five copies per reaction (Chapter 21).
The same chance governs reactions that do receive molecules. A reaction that receives two molecules instead of four starts one doubling behind, so it crosses the detection line about one cycle later. At very low concentrations, replicate reactions therefore scatter by a cycle or more, for reasons that have nothing to do with the instrument or the chemistry. For this reason, the MIQE guidelines put the lower limit of reliable quantification of a good assay at about ten copies, above the limit of detection.
So a test that is "as sensitive as possible" is one that reliably detects a few copies per reaction, and no extra cycles, brighter dyes or better optics can change that limit. The only way to detect a rarer target is to put more of the sample into the reaction, by extracting from a larger volume or concentrating the extract.
For the instrument, the number of cycles multiplies every second of the cycle. A test of 45 cycles spends 45 times the duration of one cycle, so a saving of two seconds per cycle saves a minute and a half per run. Time to result is one of the requirements that most shape an instrument (Chapter 16). It is set mainly by how quickly the liquid in every well can be moved between two temperatures, cycle after cycle.
- Every copy becomes a template, so at 100 % efficiency the count doubles each cycle, and 30 cycles give 2³⁰, about 1.07 × 10⁹.
- Efficiency compounds: at 90 % per cycle, 30 cycles give about 4.7 times fewer copies than perfect doubling.
- Reaching the detection line from a few copies at realistic efficiency takes about 35 to 45 cycles, which is why tests run 40 to 45.
- Sampling sets a floor: with an average of three copies per reaction, 5 % of reactions receive none, so no PCR can reliably detect fewer.