What a tube can borrow from a cell.
+ The questionA cell uses more than a dozen proteins to copy its DNA. How few can a tube manage with?
How a cell copies its DNA
A dividing cell copies its DNA with a team of proteins, and a PCR tube manages with just one of them. In the cell, the team works at a moving structure called the replication fork, and it copies the entire genome every time the cell divides. A helicase travels along the double helix and unwinds it, separating the two strands. Proteins that bind single strands keep them from pairing again. A primase lays down a short RNA primer on each exposed strand. A DNA polymerase then extends each primer by adding nucleotides, the units of a strand, one by one, and a sliding clamp keeps the polymerase attached. Other enzymes relieve the twisting strain ahead of the fork and replace the RNA primers with DNA, and a ligase seals the gaps.
At the center of this machinery is the polymerase, and PCR has to follow its rules. A DNA polymerase cannot start a new strand from nothing: it can only add to the free end of an existing strand that is paired to a template, the end chemists call the 3′ end. It reads the template one base at a time and adds the nucleotide that pairs with it: A opposite T, G opposite C. And it works in one direction only, extending each new strand from its 5′ end toward its 3′ end. Lengthening a primer along its template in this way is called extension. The polymerase's building blocks are deoxynucleoside triphosphates, or dNTPs, one for each base, and each one added releases a small molecule called pyrophosphate. The reaction also needs magnesium ions to proceed.
- It needs a primer: a short stretch already paired to the template, ending in a free 3′ end.
- It copies the template by pairing: each base added is the complement of the base opposite.
- It extends in one direction only, toward the 3′ end of the new strand.
Everything about how PCR is designed follows from these three rules, from the orientation of the primers to the fact that only the stretch between them is copied.
What the tube keeps and what it replaces
PCR keeps the polymerase and replaces almost everything else with something simpler. Heat replaces the helicase. Chapter 2 showed that raising the temperature above the melting temperature separates the strands without breaking them, so no unwinding enzyme is needed. No proteins are needed to hold the strands apart either, because at 95 °C they stay apart on their own.
Synthetic DNA replaces the primase. Instead of an enzyme that lays down RNA primers wherever replication starts, PCR uses two short pieces of DNA, about 18 to 30 bases long, designed to match the two ends of the stretch to be copied. These primers do two jobs at once. They give the polymerase the starting point it needs, and because they bind only where their sequence matches, they decide which stretch is copied.
The rest of the cell's team is unnecessary. A PCR target is short, typically tens to a few thousand bases, so there are no long strands to untangle and no fragments to join. The polymerase works on its own, and in most diagnostic PCR it also works without the proofreading that cells use to correct errors, a choice that Chapter 7 explains.
What remains is a short list of ingredients, which a manufacturer's protocol gives for a typical 50-microliter reaction with Taq polymerase, a polymerase that survives the heat of the cycle (Chapter 7). The protocol recommends each of the four dNTPs at about 200 micromolar, magnesium chloride at 1.5 to 2 millimolar and each primer at 0.1 to 0.5 micromolar, and it also lists about 1.25 units of polymerase, a buffer to hold the pH and salt, and the template. For the template, the protocol gives anything from a nanogram to a microgram of human DNA, or from 1 picogram to 10 nanograms of a small viral genome or of a plasmid, a small ring of DNA that bacteria carry.
A ready-made mixture of the polymerase, the dNTPs, the magnesium and the buffer is called a master mix, and the primers, any probes and the sample are added to it. Magnesium needs care, because its balance is easy to upset. The polymerase is inactive without enough free magnesium, and too much makes it less accurate and encourages wrong products. The dNTPs bind magnesium, and so do chelators, chemicals that hold metal ions tightly. A sample can carry chelators in, such as EDTA from a blood tube or citrate. What matters is the free magnesium left over, which is one reason a reaction that works with clean DNA can fail with a crude sample.
The primers win by numbers
When a reaction cools after denaturation, each single template strand can either pair with its original partner or bind a primer. The primers win because there are so many of them.
Each primer is at 0.2 µM, which is 0.2 picomoles per microliter, so a 50 µL reaction holds 0.2 × 50 = 10 picomoles of it. A picomole is about 6.0 × 10¹¹ molecules, so the reaction holds about 6 × 10¹² molecules of each primer. The template is 100 ng of human DNA, which at 3.3 picograms per genome is about 30,000 genome copies (Chapter 1), and a target that occurs once per genome has one site in each copy. Each primer therefore outnumbers each target site by about 6 × 10¹² ÷ 30,000, or 2 × 10⁸: two hundred million to one. Each dNTP at 200 µM is 200 picomoles per microliter, or 10,000 picomoles in 50 µL, which is about 6 × 10¹⁵ molecules, enough for far more product than any reaction ever makes (Chapter 8).
At those odds the template strands almost never find each other before a primer finds them. The excess also means that a normal reaction does not use up its primers, a point that matters when Chapter 8 asks why a reaction eventually stops.
The idea that made it a chain reaction
None of the ingredients was new in the early 1980s. Heating DNA to separate its strands had been studied for decades, and chemists had used DNA polymerases to extend short synthetic primers on templates before. Kary Mullis was a chemist making synthetic DNA at the Cetus Corporation in California, and most accounts date his idea to 1983. What he conceived was an arrangement: two primers, one on each strand, pointing toward each other across the target, with the steps of separating, priming and extending repeated over and over. In that arrangement the new strand made from one primer contains the binding site for the other primer, so every product becomes a template in the next round. The process works on its own output, which is why it is called a chain reaction, and why it multiplies rather than adds.
Randall Saiki and colleagues at Cetus, Mullis among them, published the first application in Science in 1985: they amplified a 110-base-pair stretch of the human β-globin gene in order to tell the normal gene from the sickle-cell variant. Their reaction used a polymerase from the bacterium E. coli and ran 20 cycles, with extension at 30 °C, and because the enzyme was destroyed at the denaturing temperature, fresh enzyme had to be added by hand at every cycle. Each cycle copied about 85 % of the target, and twenty of them amplified it about 220,000-fold. Mullis received half of the 1993 Nobel Prize in Chemistry for the invention, and the other half went to Michael Smith for oligonucleotide-based site-directed mutagenesis.
PCR is sometimes credited with a single innovation: putting heat in place of the cell's unwinding enzyme. Heat denaturation was decades old, and the enzyme in the first experiments came from bacteria, not from the cells being tested. The invention was a design that combines two primers facing each other with repeated cycles. That design makes copying exponential and confines it to one chosen stretch.
For an instrument or test developer, the reaction mix is not a commodity poured in at the end. Magnesium, dNTP and primer concentrations, the enzyme and its activation step, and the buffer all set how fast and how specifically the reaction runs, and so they also set the temperatures and hold times that the instrument must deliver. Chapter 17 shows that regulators treat the mix, the instrument and the software as one system for the same reason.
- A cell copies DNA with a team of proteins, and at its center is a polymerase that needs a primer, copies by base pairing and extends in one direction only.
- PCR keeps the polymerase, replaces the unwinding enzyme with heat and the primase with two synthetic DNA primers, and needs only dNTPs, magnesium, a buffer and the template besides.
- Primers outnumber each template site by about two hundred million to one, so they almost always bind before the template strands can pair again.
- Mullis's contribution was the arrangement of two primers facing each other and repeated cycles, which turns copying into a chain reaction.