19 — Verification

Verifying the system.

+ The questionHow do you show that every part, and the whole, does what it was specified to do?

Two questions on the right side of the V

A thermal block that runs cold in one corner can make a specimen with little virus cross the threshold too late, or not at all, and the result is then a false negative. Faults of that kind are what verification looks for.

The right side of the V in Chapter 16 asks two different questions. Verification asks whether the product was built as specified: does each part, and the assembled whole, meet its written requirements? Validation asks whether the specified product is the right one: does it meet the user's needs in the intended use? For a PCR instrument and its test, verification runs from the smallest part up to the integrated system, and validation shows that the test gives correct results on real specimens. The two meet in the analytical studies of the assay on the finished system, which belong to both.

46 — What is verified, and how
REAL SPECIMENS correct results in the intended use ASSAY ON THE FINISHED SYSTEM limit of detection, precision, specificity, carry-over analytical studies belong to both INTEGRATED SYSTEM reference plate: 5,000 vs 10,000 copies at 99.7 % SOFTWARE FDA 2023 guidance; IEC 62304 FIRMWARE event timing: the read falls where it should ELECTRONICS noise; drive currents MOTION positioning and repeatability OPTICAL well location, background, uniformity, dye response, channel separation THERMAL traceable probes in the wells: accuracy, uniformity, overshoot, hold stability, ramp time whole parts VERIFICATION built as specified VALIDATION the right product
Focal detail
Plate 46 — Verification climbs from each subsystem, measured directly against its requirements, to the integrated instrument, tested as a whole with a reference plate. Validation shows correct results on real specimens. The assay's analytical studies on the finished system, limit of detection, precision, specificity and carry-over, belong to both.

Subsystems first

Each subsystem is verified against its own requirements, by methods that measure the property directly.

The thermal subsystem is measured in the wells with traceable probes (Chapter 10). The measurements cover accuracy and uniformity at each set point, overshoot and undershoot, stability during holds, and ramp times under the conditions the specification states, with the plates and volumes the assays will use. The optical subsystem is verified for its ability to locate each well, its background, its uniformity across the plate, its response to known amounts of each calibrated dye, and the separation of dyes in neighboring channels. Motion is verified for positioning and repeatability, the electronics for noise and for the drive currents the modules need, and the firmware for the timing of events within each cycle, so that the read falls where it should every time.

Software is verified as carefully as hardware, because it computes the result. In June 2023 the FDA finalized its guidance on the content of premarket submissions for device software functions. The guidance applies to device software, including firmware and software that controls a device, and it asks for documentation at one of two levels: a basic level, or an enhanced level where a software failure could present a probable risk of death or serious injury. It also allows a manufacturer to declare conformity with IEC 62304, the recognized standard for the software life cycle, as part of that documentation.

The current edition of IEC 62304 dates from 2006, with an amendment in 2015. A second edition is in development and, as of October 2026, was not expected before 2028. In vitro diagnostic devices are the tests and instruments that examine specimens taken from the body, and in the European Union they fall under the IVD Regulation (Chapter 26). Its general safety and performance requirements include requirements for software, and a manufacturer can use IEC 62304 to show that its software meets them.

Then the system

Assembled subsystems can pass every test and still fail together, which is why the integrated instrument is verified as a whole. The reference-plate test of Chapter 15 shows the kind of whole-system test needed. It is an installation and periodic check, and its one plate of 5,000 and 10,000 copies of RNase P exercises the thermal, optical, calibration and software parts together.

Beyond the reference plate, the assay's analytical studies are run on the finished system. They cover the limit of detection and the precision across runs, days, operators, reagent lots and instruments. They also cover the ability to tell targets from near relatives, and carry-over: whether a strongly positive sample can contaminate the negative sample run after it. The FDA's guidance on moving assays between instruments asks for carry-over studies in every migration of a nucleic acid test for exactly that reason (Chapter 17).

Verification covers the whole system and every change to it

Subsystems verified separately can still fail together: a plate that fits the block but fogs under the lid, an optical read that drifts into the ramp when the firmware timing changes, a software update that moves the baseline. The integrated system, with its consumables and its assays, must be verified as a whole, and re-verified when any part of it changes.

Risk: how a wrong number becomes harm

A diagnostic instrument rarely hurts anyone directly. It can harm through its result: a false negative that sends an infected person home untreated, a false positive that leads to an unnecessary treatment or isolation, a wrong quantity, a result that arrives too late to matter. Risk management for a diagnostic device therefore traces each possible fault forward to a wrong result and on to its consequence. ISO 14971:2019 is the international standard for medical device risk management, and it applies to diagnostic devices; ISO confirmed it in March 2025. Its companion guidance, ISO/TR 24971:2020, includes an annex on applying the standard to diagnostic devices, and in July 2026 ISO marked that guidance for revision.

47 — From a hazard to a harm
HAZARD an incorrect result SEQUENCE OF EVENTS a fault: one corner runs cold; a low-level specimen reads negative HAZARDOUS SITUATION a false negative is reported and acted on HARM a missed diagnosis cold corner CONTROLS DESIGN edge heaters, guard zones VERIFICATION measure every corner, at installation and at intervals DETECTION positive control fails if affected; internal control not detected before cut-off → invalid INFORMATION labeling states the limits of a negative the internal control can turn a false negative into an invalid result when its own well is hit ISO 14971:2019; ISO/TR 24971:2020 annex on diagnostics
HeatFocal detail
Plate 47 — A diagnostic instrument harms through its result. A fault, a cold corner at the denaturing step, delays amplification, a low-level specimen in that corner reads negative, the false result is acted on, and a diagnosis is missed. Controls act at every step: design and verification prevent the fault, the internal control can turn a false negative into an invalid result only when its own well is affected, and labeling states what a negative cannot exclude.

A worked chain, in ISO 14971's terms, shows the method. For a diagnostic device the hazard is usually taken to be an incorrect result. A sequence of events starts with a fault: the block runs cold in one corner at the denaturing step, some target strands in those wells do not separate, and amplification is delayed or fails, so a specimen with little virus crosses the threshold too late or not at all, giving a false-negative result. A hazardous situation: that result is reported and acted on. A harm: a missed diagnosis.

Controls can act at each step. The design adds edge heaters and guard zones. Verification measures the corners at installation and at intervals. Detection catches gross failures. The positive control fails if it sits in an affected well. Every specimen is also tested for its internal control, a human gene (Chapter 16). If the internal control's own well is among the cold ones, a specimen whose control is not detected before the cut-off is reported invalid rather than negative, but a modest delay that leaves the internal control inside the cut-off passes unnoticed.

The last control is information for the user: the labeling states the limits of a negative result. For multi-target tests that include SARS-CoV-2, the FDA sets special controls: requirements that every device of that type must meet (Chapter 25). Under them, the labeling must state that a negative result does not rule out infection and that sequence variants can cause incorrect results. The special controls also require continuous monitoring for new variants, and results suggesting that a variant affects the stated performance must be sent to the agency immediately, with the evaluations submitted within 48 hours if the FDA asks for them.

Solved example: what the internal control costs on a plate

In the CDC test as authorized from March 2020, each specimen takes three wells, one each for N1, N2 and the internal control, RNase P. The run's three controls take three wells each, which is 3 × 3 = 9 wells. A 96-well plate then has 96 − 9 = 87 wells left for specimens, and at three wells each it holds 87 ÷ 3 = 29 specimens, as in Chapter 16. The specimens' internal controls fill 29 of the plate's 96 wells, about 30 %. Almost a third of a full plate therefore goes to the internal control.

Why the internal control earns its well

In the CDC test, a specimen with no viral target is reported negative only if the human control gene, RNase P, is detected (Chapter 13). That rule converts several silent failures into an invalid result, which prompts a repeat, instead of a false negative, which does not. The failures include a swab with too few cells, a failed extraction and an inhibitor that stopped the reaction. A well that never reached temperature is caught only when the control shares that well, as in a multiplex test, and in the CDC test RNase P ran in a well of its own (Chapter 21). One extra target per specimen turns a hazard that would otherwise pass unnoticed into one the system detects.

Safety, electromagnetic compatibility and the user

A laboratory instrument must also be safe to use and compatible with the other electrical equipment around it. The IEC 61010 family of standards covers the safety of electrical equipment for measurement, control and laboratory use, against shock, fire, heat and mechanical hazards. Part 1 gives the general requirements. Its edition of 2010 was amended in 2016 and consolidated in 2017, and a second amendment had completed its final vote in September 2026 but had not been published as of October 2026. Particular parts add requirements: Part 2-010 for equipment that heats materials, Part 2-081 for automatic laboratory analyzers, and Part 2-101, from 2018, for in vitro diagnostic equipment. Which parts of IEC 61010 apply to a given instrument is a judgment made from its functions. For example, one manufacturer states that its air-heated cycler conforms to Parts 1 and 2-101.

Electromagnetic compatibility is the instrument's ability to work without disturbing other equipment or being disturbed by it. IEC 61326-1 covers it, and a particular part for diagnostic equipment, IEC 61326-2-6, was published in a new edition in 2025. IEC 62366-1 covers usability engineering, which addresses the errors users can make, such as entering the wrong volume or loading a plate the wrong way round, and the labeling standard ISO 18113-3 covers the information supplied with diagnostic instruments for professional use.

48 — The standards around one instrument
QUALITY SYSTEM ISO 13485:2016, confirmed October 2025 US QMSR incorporates it from February 2026 RISK ISO 14971:2019 ISO/TR 24971:2020, annex for diagnostics SOFTWARE IEC 62304:2006 + A1:2015 FDA guidance, June 2023 USABILITY IEC 62366-1: wrong volume, plate loaded the wrong way round ELECTRICAL SAFETY IEC 61010-1 (2010, amended 2016, consolidated 2017) particular parts: -2-010 heating; -2-081 automatic analyzers; -2-101 IVD equipment (2018) which parts apply is judged from what the instrument does EMC IEC 61326-1 IEC 61326-2-6, new edition 2025 LABELING ISO 18113-3, professional-use instruments PCR INSTRUMENT AND TEST
Focal detail
Plate 48 — One instrument answers to many standards: a quality system to ISO 13485, risk management to ISO 14971, software to IEC 62304, electrical safety to the IEC 61010 family, compatibility to IEC 61326, usability to IEC 62366-1 and labeling to ISO 18113-3. Which particular parts apply is a judgement made from what the instrument does.

Production and the field

Verification does not end when the design is approved. Every unit built is tested before it ships, with its thermal and optical performance checked against limits, and each unit is calibrated at installation and checked at intervals in use. One accredited calibration service recommends calibration once a year or every 250 runs. The Peltier modules, the fan, the lamp and other parts with finite lives are inspected or replaced on a schedule. Complaints and service records are collected and analyzed for trends.

All of this happens inside a quality management system. For medical devices the international standard for such a system is ISO 13485:2016, which was confirmed without change after review in October 2025. In the United States, the Quality Management System Regulation of Chapter 16 has been in force since February 2026 and incorporates ISO 13485 (Chapter 25). On the day the regulation took effect, the FDA stopped using its Quality System Inspection Technique and began inspecting under a new compliance program.

One evidence core

The thermal mapping, optical verification, software documentation, risk file, safety and compatibility test reports, analytical studies and quality records described here are one body of evidence. When it is organized around the system's requirements from the start, the same core serves design reviews, customers' audits, the investigation of field problems and regulatory submissions in several jurisdictions (Chapters 25 and 26).

+ What this chapter established
  • Verification shows that the product was built as specified, from subsystems up to the integrated system, while validation shows that it meets users' needs, and analytical studies on the finished system belong to both.
  • Thermal, optical, motion, electronic, firmware and software subsystems are each verified directly, and software documentation follows the FDA's 2023 guidance, which allows a declaration of conformity with IEC 62304.
  • A diagnostic instrument harms through wrong or late results, so risk management traces faults to wrong results and adds controls at each link, and the internal control turns several silent failures into invalid results.
  • The safety standards of IEC 61010, the compatibility standards of IEC 61326 and standards for usability and labeling apply, and verification continues in production and the field under a quality system.