+ Part VI · The test

From a specimen to a trusted result.

The CDC's test had a limit of detection of about 3 RNA copies per microliter of a sample spiked with a known amount of the target's RNA, with one automated extraction system, and of 1 copy per microliter with a manual kit. The step before the reaction set part of what the test could detect. An instrument and an assay that work in the development laboratory are not yet a test. This part follows the steps that make them one: reading RNA with a method that copies DNA, getting nucleic acid out of a specimen intact and free of inhibitors, the controls and limits that make a result believable, and the CDC's test as it was designed, made and shipped in the first weeks of 2020.

20 — From RNA

Reading RNA with a DNA method.

+ The questionMost of the viruses that matter carry RNA. How does a method that copies DNA read them?

A step before the first cycle

Taq polymerase copies DNA. It needs a DNA template, and it does not copy RNA usefully. Yet many of the targets that matter most in diagnostics are RNA: the genomes of SARS-CoV-2, influenza, respiratory syncytial virus, HIV and hepatitis C virus are all made of RNA. The activity of a gene, too, is measured through its messenger RNA, the RNA copy that a cell makes of a gene in order to use it. Reading these targets with PCR needs one extra step at the start: copying the RNA into DNA.

The enzyme that does this was discovered in 1970 in retroviruses, viruses whose RNA genomes must be copied into DNA to infect a cell, and two teams found it independently: Howard Temin with Satoshi Mizutani, and David Baltimore. The enzyme is called reverse transcriptase, because it runs the usual flow of genetic information backward, from RNA to DNA. Given an RNA template and a primer, it builds a strand of complementary DNA, or cDNA, which PCR can then amplify. The enzymes used in laboratories come from two other retroviruses, avian myeloblastosis virus and Moloney murine leukemia virus, or are engineered versions of those enzymes. They work at moderate temperatures, around 40 to 50 °C for the original enzymes, and a higher temperature helps them read RNA that folds back on itself.

49 — From RNA to cDNA
1 RNA TARGET 2 REVERSE TRANSCRIPTION 3 FIRST PCR CYCLE 5′ 3′ a fold the PCR's reverse primer also primes the cDNA N gene: the virus makes extra RNA copies of this region 5′ 3′ cDNA grows reverse transcriptase cDNA at 50 °C it reads RNA that would fold back on itself 3′ 5′ forward primer cDNA double-stranded DNA: now PCR can copy it … 25 °C 50 °C 95 °C 95 °C / 55 °C 2 min contamination control 15 min reverse transcription 2 min activate polymerase 45 cycles PCR one-step: both enzymes in one tube times not to scale
CopiesPrimers and probesRNAHeatFocal detail
Plate 49 — Taq copies RNA poorly, so reverse transcriptase first builds a complementary DNA strand along the viral RNA, primed in a one-step test by the PCR's own reverse primer. The forward primer then copies the cDNA in the first cycle. The CDC test did both in one tube: 2 minutes at 25 °C, 15 minutes at 50 °C for reverse transcription and 2 minutes at 95 °C before its 45 cycles.

One tube or two

The reverse transcription and the PCR can be run in separate tubes or in one. In a two-step method, the RNA is first copied into cDNA in its own reaction, with its own buffer, and portions of that cDNA are then used in separate PCRs. Each step can be optimized, and the cDNA can be stored and used for several tests. In a one-step method, both enzymes and both sets of conditions share one tube: the reaction is held at the reverse-transcription temperature for some minutes, then heated, and the PCR cycles begin. Fewer steps mean less handling and less risk of contamination, which is why one-step methods suit diagnostic testing.

The reverse transcriptase needs a primer too. In a one-step diagnostic test, the PCR's own reverse primer usually serves: it binds the viral RNA and is extended into cDNA, which the forward primer then copies in the first PCR cycle. The CDC test was one-step. With the TaqPath master mix, its profile began with 2 minutes at 25 °C, a step for the contamination control described at the end of this chapter. Then came 15 minutes at 50 °C for reverse transcription and 2 minutes at 95 °C to activate the polymerase, before the 45 cycles began.

Solved example: the share of the CDC run spent on reverse transcription

With the TaqPath master mix, the CDC test's profile held 2 minutes at 25 °C, 15 minutes at 50 °C and 2 minutes at 95 °C, so 2 + 15 + 2 = 19 minutes passed before the first cycle. Each of the 45 cycles then held 3 seconds at 95 °C and 30 seconds at 55 °C, or 33 seconds. Over the run that makes 45 × 33 = 1,485 seconds, about 25 minutes. Counting only the time at set temperatures, and leaving out the time the block spends changing temperature, the holds came to about 19 + 25 = 44 minutes, while the whole run, ramps included, took about 81 minutes in standard mode before any read time (Chapter 16). Reverse transcription took 15 of those 44 minutes, so reading RNA cost about a third of the time this run spent at its set temperatures, and a little under a fifth of the whole run.

The test's two viral targets lay in the gene for the nucleocapsid protein, the protein that packages the viral RNA, and this gene is called the N gene. In their paper describing the test, the CDC's scientists gave one reason for the choice. During infection, the virus makes many extra RNA copies of its N gene region, and these copies may make the target more abundant than other regions and the test more sensitive.

Why gene activity needs a control without the enzyme

When RNA is the target, any DNA in the sample with the same sequence will also be amplified, with or without reverse transcription. A reaction run without the reverse transcriptase, called a no-RT control, shows whether the signal came from RNA or from contaminating DNA. The 2025 MIQE revision asks for such controls in studies that measure gene activity with reverse transcription and real-time PCR, where the gene's DNA is present in every cell.

RNA is fragile, and enzymes that destroy it are everywhere

RNA is easier to destroy than DNA. The enzymes that destroy it are called ribonucleases, or RNases, and they are everywhere: in tears, saliva, mucus and sweat and on skin and hair, and bacteria and fungi make them too. One supplier of laboratory reagents describes RNases in a technical note as small, compact proteins held together by several internal cross-links. The same supplier states that they are robust enough to refold after being heated and to keep substantial activity even after autoclaving, which sterilizes laboratory equipment with steam under pressure.

A specimen can lose its target RNA before the test begins: in the swab itself, in a tube touched with a bare hand, or through a pipette tip that was not certified free of RNase. These are the reasons for gloves, certified plastics, inhibitors of RNase in the reaction mix and cold storage of specimens.

Getting the nucleic acid out

Between the swab and the reaction comes extraction, the step that recovers the nucleic acid from the specimen. Its jobs are to break open the virus and the cells, to separate the nucleic acid from everything else in the specimen, including RNases and substances that would slow the polymerase, and often to concentrate it. The most common method breaks open the particles with a strong salt solution of guanidinium, which also unfolds, or denatures, the proteins. The nucleic acid is then bound to silica, on a small column or on magnetic beads. The silica is washed with buffers containing ethanol, and the purified nucleic acid is released in a small volume of water or buffer.

50 — Extraction on silica
1 LYSE 2 BIND 3 WASH 4 ELUTE guanidinium lysis buffer virus opened, RNA released proteins, RNases denatured silica RNA binds column membrane or magnetic beads ethanol wash to waste proteins, salts, inhibitors removed water eluate purified RNA in a small volume traces of buffer carried into the reaction delay it the limit of detection depended on extraction: about 3 copies/µL of spiked sample with one automated system, 1 copy/µL with a manual kit (CDC instructions) carry-over: about 10 % wash buffer, or 0.3–1 % lysis buffer, delayed Cq by at least half a cycle (2020 study)
RNAFocal detail
Plate 50 — Guanidinium breaks open the virus and denatures proteins, including RNases; the nucleic acid binds silica, is washed with ethanol buffers and is released in a small volume. Extraction sets part of the test's sensitivity, and traces of its buffers carried into the reaction, as little as 0.3 to 1 % of lysis buffer, delay amplification.

Extraction is part of the test's performance, not a preliminary to it. The CDC test's instructions listed the extraction kits and automated systems that the test was validated with. The test's limit of detection differed between them: about 3 RNA copies per microliter of the spiked sample with one automated system, and 1 copy per microliter with a manual kit. A spiked sample is one into which a known amount of the target's RNA has been mixed. The instructions also allowed a simple heat treatment in place of extraction, with a warning that the heat treatment had been tested on few clinical specimens and that some loss of sensitivity could not be ruled out.

Extraction can also add a problem of its own. A 2020 study measured how traces of the extraction kits' buffers, carried into the reaction, delay amplification. Ethanol wash buffer making up about 10 % of the reaction delayed the Cq by at least half a cycle, and so did as little as 0.3 to 1 % of the guanidinium buffer that breaks the particles open.

Inhibitors

Many substances in clinical and environmental samples slow or stop the reaction. Such substances are inhibitors, which lower the efficiency of amplification (Chapter 13). The authors of one review classify them by where they act: on the nucleic acid, on the polymerase, on primer binding, or on the fluorescence itself.

InhibitorTypical sourceHow it interferes
Hemoglobin and hemered blood cellslowers polymerase activity; can quench fluorescence
Lactoferrinwhite blood cellsreleases iron; interacts with nucleic acids
Immunoglobulin Gplasmabinds single-stranded DNA, hindering primer binding
Heparin, EDTAblood collection tubesheparin inhibits the reaction; EDTA removes the magnesium the polymerase needs
Humic and fulvic acidssoil and sedimentact on the polymerase; quench DNA-binding dyes
Ethanol, guanidiniumcarry-over from extraction kitsdelay amplification

An inhibitor lowers the efficiency, so the curve rises later, and a strong enough inhibitor keeps it from rising at all. In a test without an internal control, a reaction stopped by an inhibitor looks exactly like a negative result. Additives can help. In a study of inhibitors from blood, the most effective additive at relieving them was bovine serum albumin, a protein purified from cattle blood.

An inhibited reaction looks like a negative one

A specimen that inhibits PCR gives no signal for the target, just like a specimen without the target. Only a control amplified from the same specimen distinguishes the two. When the control shares the target's well, it also shows that this very reaction could amplify. Telling the two apart is the job of the internal control described in Chapter 21. It is also why a specimen with no signal for the target and none for the internal control is called invalid rather than negative.

Keeping old product out of new reactions

Every positive PCR makes billions of copies of its target. A laboratory that runs thousands of reactions accumulates those copies on benches, pipettes and gloves, and a trace of old product in a new reaction is a perfect template that gives a false positive. Laboratories separate the rooms where reactions are prepared from the rooms where amplified product is handled, and real-time reactions help as well, because their tubes stay closed and are never opened after amplification.

There is also a chemical defense, described in 1990 by scientists at Life Technologies, a supplier of products for life-science research. Every PCR is run with dUTP, a building block carrying uracil, in place of dTTP, the usual building block carrying thymine, so that every product contains uracil instead of thymine. Before each new PCR, an enzyme called uracil-DNA glycosylase, or UNG, is allowed to act. It removes uracil from any DNA, and removing the uracil damages carried-over product so that it can no longer be copied. The specimen's own DNA contains thymine, so the enzyme leaves it untouched. The enzyme is then inactivated by heat, before the new uracil-containing copies are made. In the CDC test's TaqPath profile, the first 2 minutes, at 25 °C, are that step.

51 — Destroying old product
1 BEFORE THE RUN 2 UNG, 2 min AT 25 °C 3 HEAT U U U U U carried-over product made with dUTP: U in place of T T T T T T the specimen's own DNA T, as in all natural DNA UNG U old product cut where uracil sits can no longer be copied T T T T T untouched UNG removes only uracil UNG inactivated by heat before new copies old product in pieces not copied PCR goes on new product carries U too: the next run is protected from it in the CDC test's TaqPath profile, this is the 2 min at 25 °C before reverse transcription
Template DNACopiesHeatFocal detail
Plate 51 — Running every PCR with dUTP in place of dTTP marks all product with uracil. Before each new run, uracil-DNA glycosylase removes uracil from any carried-over product, so it can no longer be copied, while the specimen's own DNA, which carries thymine, is untouched. Heat then inactivates the enzyme before new product is made. In the CDC test's TaqPath profile the UNG step is the first 2 minutes, at 25 °C.
Treat extraction and contamination control as part of the design

For a test developer, the extraction method, the specimen types, the RNase controls and the contamination defenses are part of the test system as much as the primers. Their performance must be established with the rest of the test. Any change to them is a change to the test: a new extraction kit, a new transport medium for the swabs, or a heat step instead of extraction.

+ What this chapter established
  • Reverse transcriptase, discovered in 1970, first copies an RNA target into complementary DNA, either in a reaction of its own or, in the one-step methods that suit diagnostic testing, in the same tube as the PCR, often with the PCR's reverse primer as the primer.
  • RNases are robust and everywhere, so RNA needs careful handling, and extraction releases and purifies the nucleic acid and sets part of the test's sensitivity.
  • Inhibitors from the specimen or the extraction slow or stop the reaction, and without an internal control an inhibited reaction cannot be told apart from a negative.
  • Carried-over product is kept out by separate work areas and by dUTP with UNG, which damages old product before each run so that it cannot be copied.