+ Part IX · Quality and regulation, as of October 2026
Deciding when a count can be trusted.
On 29 June 2026 the Secretary of Health and Human Services terminated the declaration that had justified emergency authorization of COVID-19 diagnostic tests since 4 February 2020, with effect from 26 December 2026. In a laboratory advisory of 14 October 2026, the CDC told laboratories what follows: the FDA does not intend to object to continued distribution of a test whose marketing submission it has accepted by 26 December 2026, and laboratories may go on using tests already distributed until they expire. The earlier chapters showed why each requirement on a test exists, from the limit of detection to the controls in every run and the instrument named in every authorization. This part maps those requirements onto the rules of three jurisdictions, giving the reason for each rule before its clause, with every status dated October 2026: the United States in Chapter 25, and the European Union and India in Chapter 26.
The device and the laboratory.
+ The questionWhat must a PCR instrument and test show before US laboratories may use them on patients, and who decides?
Three questions behind every route
Within days of receiving the CDC's first kits in February 2020, some public health laboratories could not verify them, and the emergency authorization under which the kits had shipped already said what had to happen next. Each such laboratory had to notify the CDC, and the CDC had to notify the FDA (Chapter 22). That condition is a small instance of how US regulation acts on a test. It attaches obligations to a named product for a stated use, and it rests on evidence that answers three questions.
The rules differ between jurisdictions, but the evidence they ask for answers the same three questions, in order, and the European Union's regulation for diagnostic devices names them precisely. Scientific validity asks whether the thing measured, the analyte, is associated with the clinical condition: for a SARS-CoV-2 test, whether the virus's RNA in a respiratory specimen is associated with infection. Analytical performance asks whether the device detects or measures the analyte correctly, on the complete test system, instrument included. Clinical performance asks whether the results correspond to the condition in the people the test is meant for. Chapter 21 set out what each of the last two contains.
A fourth question, whether using the test improves outcomes for patients, is its clinical utility. It matters to doctors and to those who pay for tests, but market authorization generally does not require it. In the United States, two bodies of law divide the work. The FDA regulates the instrument and the test as medical devices, and a separate law, the Clinical Laboratory Improvement Amendments, or CLIA, regulates the laboratories that run them. Every status in this chapter is stated as of October 2026.
Classes and routes for the instrument and the test
The FDA sorts devices into three risk classes, from class I, the lowest, to class III, the highest, and by product type. Each type has a classification regulation that names its class and, for class II, its special controls: requirements that every device of that type must meet. A device then reaches the market by one of a few routes. A 510(k), or premarket notification, shows that the device is substantially equivalent to one already legally marketed, its predicate, and the FDA clears it. A device of a new type with no predicate, and of low to moderate risk, can instead go through a De Novo request, which the FDA grants. The grant creates a new classification, and the granted device can then serve as the predicate for later 510(k)s.
Real-time PCR instruments are classified under 21 CFR 862.2570, instrumentation for clinical multiplex test systems, as class II devices with special controls. A final order of 30 December 2019 exempted them from premarket notification, subject to the general limits on exemptions. The instrument named in the CDC's test had been cleared through a 510(k) in September 2008, before that exemption.
The exemption removes the premarket notification and nothing else. The instrument remains a class II device, its product type is not exempt from the manufacturing requirements, and the quality management system regulation described in this chapter's last section requires design controls for class II devices. The exemption also covers the instrument alone: a test sold for use on it still needs its own authorization, for its intended use with specified instruments (Chapter 17).
Tests follow the same logic. For tests that detect SARS-CoV-2 together with other respiratory pathogens, the FDA created a classification, 21 CFR 866.3981, through the De Novo route. It granted a De Novo request for BioFire Diagnostics' Respiratory Panel 2.1 on 17 March 2021, and the trade press reported it as the first SARS-CoV-2 test authorized through a premarket route rather than an emergency authorization. The classification was codified on 16 August 2024. Its special controls turn the second and third questions into required studies of the limit of detection, cross-reactivity and clinical performance. They also require labeling stating that a negative result does not rule out infection and that sequence variants can cause incorrect results, and continuous monitoring of viral variants (Chapter 19). Results suggesting an effect on performance go to the FDA immediately, and evaluations within 48 hours if the agency asks. Later tests of the same type reach the market through 510(k)s that cite earlier ones, as Cepheid's Xpert Xpress CoV-2/Flu/RSV plus did in August 2023.
Emergency use, and how it ends
An Emergency Use Authorization, or EUA (Chapter 22), lets the FDA allow an unapproved product during a declared emergency. Under section 564 of the Federal Food, Drug, and Cosmetic Act, the agency may authorize a test if the disease is serious or life-threatening, the test may be effective, its known and potential benefits outweigh its known and potential risks, and no adequate, approved and available alternative exists. An authorization lasts until the declaration behind it ends or the FDA revokes it, whichever comes first. Ending a declaration requires advance notice in the Federal Register, with reasonable time to dispose of the products. Authorizations granted outside an emergency eventually remove the fourth criterion, the lack of an adequate, approved and available alternative: on the day it granted the De Novo request for the BioFire panel, the FDA revoked that panel's emergency authorization, because an adequate approved alternative now existed.
The declaration for COVID-19 tests was made on 4 February 2020 and amended on 15 March 2023. On 29 June 2026 the Secretary of Health and Human Services, Robert F. Kennedy, Jr., terminated it, with effect from 26 December 2026. HHS judged 180 days a sufficient period for disposition of the products, and its notice observed that more than fifty COVID-19 tests held marketing authorization granted outside an emergency. The FDA states that from 26 December the COVID-19 device authorizations on its list will no longer be in effect.
The end comes with a transition. The FDA's guidance of March 2023 asks a manufacturer that intends to keep distributing an emergency-authorized test to submit a marketing application in time for the agency to accept it before the termination date. The FDA does not intend to object to continued distribution once it has accepted the submission and while it has taken no final action. Under the same guidance, single-use tests distributed before the termination date may be used until the expiry date listed on that day, which cannot be extended. In its laboratory advisory of 14 October 2026, the CDC repeated these points for laboratories. It added that the Centers for Medicare & Medicaid Services lets laboratories keep using such tests, as their authorizations allowed, until the product's expiration date, and it asked laboratories to review the emergency-authorized tests they run and plan their transition.
Laboratories: CLIA, and tests made in one laboratory
CLIA regulates the laboratory, not the product. It sorts tests into three categories by the complexity of performing them: waived, moderate complexity and high complexity. A waived test meets separate criteria of simplicity and low risk of error. The others are scored on seven criteria: knowledge; training and experience; reagent preparation; operational steps; calibration, quality-control and proficiency-testing materials; troubleshooting and maintenance; and interpretation and judgment. The FDA categorizes commercially marketed tests, during premarket review or on request, and a manufacturer can apply separately for a waiver. In the FDA's 510(k) records, a submission that combines a 510(k) with an application for a waiver is marked Dual Track. The CDC's panel was authorized only for laboratories certified for high-complexity testing.
Each of the seven criteria is scored 1, 2 or 3, so totals run from 7, the lowest grade everywhere, to 21. A total of 12 or less is moderate complexity, and anything above is high. A test of middle grade on every criterion scores 7 × 2 = 14, which is high complexity. From the minimum of 7, a test has 12 − 7 = 5 points to spend: a criterion scored 2 spends 1 point, and one scored 3 spends 2. Each criterion above the lowest grade spends at least 1 point, so at most five criteria can rise above it, and five at 2 spend all 5 and leave the other two at 1. Three criteria at 3 would spend 6, one too many, so at most two can score 3. In words: a moderate-complexity test must score the lowest grade on at least two of the seven criteria and the highest on no more than two.
CLIA also covers tests that never pass through the FDA. A laboratory-developed test is a test made and used within a single laboratory. In May 2024 the FDA issued a rule stating that such tests are devices even when the manufacturer is a laboratory, and phasing out its long-standing practice of not enforcing device requirements on them. On 31 March 2025 a federal district court in Texas vacated the rule, holding that such tests are professional services rather than the products the law's device provisions cover. On 19 September 2025 the FDA removed the rule's words from its regulations. As of October 2026, laboratory-developed tests remain subject to CLIA. CLIA's regulations require a laboratory that introduces a test it developed, or modifies one the FDA has cleared or approved, to establish the test's performance specifications before reporting patient results, including, as applicable, accuracy, precision, analytical sensitivity and specificity (Chapter 21), reportable range and reference intervals.
The quality system since 2 February 2026
The design controls of Chapter 16 were already required in the United States under the FDA's earlier quality system regulation. They are now required through the Quality Management System Regulation, or QMSR, which the FDA issued as a final rule on 2 February 2024 and which took effect on 2 February 2026. It incorporates ISO 13485:2016 by reference, with clause 3 of ISO 9000:2015 for its terms, so the design controls, risk management and production controls of a diagnostic manufacturer follow the international standard. The FDA adds its own requirements on unique device identification, complaints and medical device reports, corrections and removals, and traceability. Design controls apply to class II and class III devices and to listed class I devices. A real-time PCR instrument under the exempt classification is a class II device, so its maker carries out design controls without ever filing a 510(k).
Write the intended use first, then find the classification regulation that fits it and read its special controls as a list of studies the evidence must contain. Decide early which laboratories the test is for, because the CLIA category sets how much of the procedure the design must take from the user, and a waiver is a separate application. For an instrument under an exempt classification, plan design controls, the quality system and the labeling as for any class II device. For a test that relies on an emergency authorization, have a marketing submission accepted before the emergency ends.
- Regulators ask in turn whether the analyte is associated with the condition, whether the device measures it correctly on the complete test system, and whether its results match the condition in the intended population.
- In the US, real-time PCR instruments are class II and have been exempt from premarket notification since December 2019, while multi-target SARS-CoV-2 tests are class II under a classification created by a 2021 De Novo, whose special controls require analytical, clinical and variant-monitoring work.
- COVID-19 emergency authorizations for devices end on 26 December 2026, the FDA does not intend to object to a test staying on the market if it has accepted the test's marketing submission by then and taken no final action, and single-use tests already distributed may be used until they expire.
- CLIA sorts tests into waived, moderate and high complexity, laboratory-developed tests remain subject to it as of October 2026 after the 2024 rule was vacated, and since 2 February 2026 the QMSR has incorporated ISO 13485:2016 into US device regulation.