Fast, even and exact.
+ The questionWhy is changing a block's temperature quickly and evenly, without overshoot, the hard part of the engineering?
The sample is behind the block
The controller measures the block. The chemistry happens in the liquid. Between them is a plastic wall a fraction of a millimeter thick and the liquid's own bulk, and heat takes time to cross both. When the block jumps to a new temperature, the liquid follows it smoothly, covering a fixed fraction of the remaining difference in each interval of time. Engineers describe this as a first-order lag and characterize it by a time constant: the time the liquid takes to cover about 63 % of a step. In Perkin-Elmer's 1992 patent (Chapter 9), the inventors measured a time constant of about 9 seconds for 100 microliters in a standard tube.
In this example the block jumps from 55 to 95 °C, a step of 40 degrees, and the instrument holds it there for 3 seconds, starting the clock when the block arrives. With a time constant of 9 seconds, the liquid covers 63 % of a step in 9 seconds, leaving 37 %, and 95 % in 27 seconds. It covers the same share of the remaining gap in every 3 seconds, and 0.72 × 0.72 × 0.72 is about 0.37, so each 3 seconds leaves about 72 % of the gap and covers about 28 %. In the 3-second hold it therefore covers 28 % of 40 degrees, about 11 degrees, and reaches only about 55 + 11 = 66 °C before the block is cooled again. At that temperature no target strand would separate. The 9-second figure is for 100 microliters, and a 20-microliter reaction in a thin-walled plate, like the CDC test's, responds faster. The principle is the same: a 3-second denaturation is meaningful only if the liquid itself reaches 95 °C.
Two engineering responses solve the problem of the lag, and modern instruments use both. The first is to calculate the sample's temperature instead of measuring it. The controller knows the block's temperature history, the volume of liquid the user has entered and the time constant of the tube, and from these it computes, moment by moment, where the liquid's temperature must be. That calculation is the sample-temperature model. The hold timer starts when the calculated sample temperature reaches the set point, not when the block does.
The second response is to overshoot. To make the liquid arrive sooner, the controller drives the block deliberately past the set point for a moment and then brings it back, so that the liquid approaches its target quickly without passing it. The early patent describes exactly this: a model of the sample with a single time constant, a block driven past its set point, and a control loop running on the instrument's processor. The loop is of the proportional-integral-derivative kind used throughout industry.
Current instruments leave the choice to the user. One manufacturer's cyclers offer a "calculated" mode, which uses the sample-temperature model, and a "block" mode, which simply holds the block. Sample-temperature control of this kind is what makes a 3-second hold meaningful. Such control also makes the reaction volume entered in the software part of the temperature control: if a user enters 50 microliters for a 20-microliter reaction, the model is wrong.
How fast is fast
Speed is quoted as a ramp rate, in degrees per second (Chapter 9). The number on a datasheet is easy to misread, because manufacturers define it differently: some quote the maximum rate the block reaches for a moment, some the average rate of the block over a temperature change, and some the average rate of the calculated sample.
For the instrument named in the CDC test, the manufacturer's brochure gives a peak block heating rate of 5.5 °C per second. The same brochure gives a sample ramp rate of 3.5 °C per second in the instrument's fast mode and 1.6 °C per second in its standard mode. Another manufacturer gives a maximum of 5 °C per second and an average of 3.3, without saying whether either refers to the block or the sample. A third gives a maximum block rate of 6.5 and an average sample rate of 3.66.
These numbers are not comparable unless their definitions match, and even then a maximum rate may be held only briefly. One instrument maker makes the practical point in its own educational article: the honest test of speed is to run the same protocol on two instruments and time it. For the user, what matters is the time the liquid spends getting from one temperature to the next, added up over every rise and fall in the run: two in each cycle of a two-step protocol such as the CDC test's, and three in a three-step one.
The ramp rate on a datasheet usually describes the block, often at its fastest moment. The liquid follows more slowly, and the difference grows with reaction volume and tube wall thickness. A fair comparison states which rate is quoted, over which temperature range and with which plate and volume, and it times runs under one protocol instead of comparing headline numbers.
Every well the same
A 96-well block is a slab of metal several centimeters across, and its edges lose heat to the surrounding air and to the parts that hold it. Left alone, the corner and edge wells would run cooler than the center when the block is hot and warmer when it is cold. Designers counter the edge loss with geometry and with extra heat. The design in the early patent had a groove cut around the perimeter of the block, a guard band that limited heat flow from the central wells to the edge. It also had separately controlled heaters along the edges to replace the heat lost there.
The result is specified as temperature uniformity, which says how far apart the wells are from one another. It is a different property from accuracy, which says how close the average well is to the set point: a block can be perfectly uniform and two degrees off, or accurate on average with a cold corner. For uniformity, too, the definitions on datasheets differ. One manufacturer states ±0.4 °C well to well, measured within 10 seconds of the block arriving at 90 °C. For its later model it states ±0.3 °C, with the hottest and coldest wells no more than 0.6 °C apart, measured 10 seconds after the block reaches its target. Between the two models the same maker changed the stated moment of measurement and added a limit on the hottest-to-coldest spread, so even its own two figures do not share one definition. The brochure for the instrument named in the CDC test states a uniformity of ±1 °C and an accuracy of ±0.5 °C without stating the conditions.
Some instruments make the block uneven on purpose. A gradient block deliberately holds a range of temperatures across the block, so that a developer can test several annealing temperatures in one run, and in one design the range can be up to 24 °C wide. Another design divides its block into six independently controlled zones that can differ by up to 5 °C.
When temperature goes wrong
Temperature errors are not hypothetical. In a study published in 2005 in a clinical microbiology journal, researchers tested six new 96-well Peltier cyclers. They measured thirteen positions in each block with a measuring system certified to ±0.3 °C, and then ran a diagnostic PCR on each cycler. The errors were largest in the first 15 seconds of the denaturing step. When the hold timer started, the liquid in the accurate instruments was 0.5 to 1.5 °C from the set point, while in two less accurate instruments it was still 13 to 20 °C short. With those two the diagnostic test gave no positive results until the denaturation temperature and time were increased. The authors named the critical parameters: holds timed too early, undershoot and overshoot, and differences in heat transfer across the block.
A cool well at the annealing step lets primers bind to near-matches, which can produce wrong products (Chapter 5). A cool well at the denaturing step leaves strands paired, lowering the efficiency, and every cycle of lower efficiency delays the curve (Chapter 6). A hot well uses up the enzyme's life faster (Chapter 7). The effects compound over forty cycles, so a well that is consistently a degree or two off can give a result that differs from its neighbors', and a well that is grossly off can give no result.
Measuring what the wells see
Since the controller reads only the block, the wells must be measured separately. Temperature verification kits place calibrated probes in selected wells and record what they read through a test program. One manufacturer measured eleven cycler models from six manufacturers with such a kit, whose probes were calibrated traceably to national standards. It found steady-state errors between −0.33 and +0.39 °C, and measured performance that did not always match the published specifications. The manufacturer that ran the study competes with the makers of most of the instruments tested, so the study is best read as an illustration of method, not as a ranking.
Calibration services offer the same measurements: steady accuracy, well-to-well spread, overshoot and undershoot, and stability during holds. Accredited services work under ISO/IEC 17025, the international standard for the competence of testing and calibration laboratories. One such service states an overall accuracy better than 0.1 °C and recommends calibration once a year or every 250 runs. CYCLERtest sells cycler calibration systems, and in a metrology paper presented in 2017 two of its staff and a coauthor argued that methods which judge a cycler by running a PCR on it cannot be traced to physical units and carry large uncertainties. They also argued that acceptance limits should be derived from the needs of the user's own PCR method, not taken from the manufacturer's specification. The point generalizes: how much error a cycler may have is a property of the test it runs.
When specifying or buying a cycler, ask for accuracy and uniformity with their conditions: at which temperature, how long after arrival and with which plate. Ask for ramp rates with their definitions, and for the assumptions of the sample-temperature model. At installation and at intervals afterward, measure the wells with traceable probes and compare the results with limits derived from the tests the instrument will run.
- The liquid lags the block with a time constant of seconds, so instruments calculate the sample's temperature, overshoot the block deliberately and start hold timers when the sample arrives.
- Ramp rates are defined in several incompatible ways, and the speed that matters is how fast the liquid moves between temperatures, best compared by timing the same protocol.
- Edge wells lose heat, so blocks use guard bands and edge heaters, and uniformity and accuracy are distinct specifications, each with conditions attached.
- Real instruments can start a hold before the liquid arrives: in one study, the liquid in two of six new cyclers was still 13 to 20 °C below the denaturing temperature when the timer started, so wells must be measured with traceable probes against limits set by the test.
Let's Talk