Seeing what can't be weighed.
+ The questionHow does an instrument watch copies form inside a sealed tube, and tell several targets apart?
Light in, different light out
A real-time instrument cannot weigh the product or look at it under a microscope. It can only shine light into the well and measure the light that comes back. The method that makes such a measurement possible is fluorescence. Some molecules, called fluorophores, absorb light of one color and give off light of a longer wavelength, a slightly redder color, a few billionths of a second later.
The instrument shines light of the absorbed color, the excitation, through a filter that passes only that band, and measures the returning light through a second filter that passes only the emitted band. Such a pair of filters, one for the light going in and one for the light coming out, makes one color channel. Because the two bands differ, the detector sees the fluorophore's glow and very little of the much brighter excitation light. What the instrument measures, then, is not DNA but light from a dye, and the task of real-time detection chemistry is to make that light rise in proportion to the copies of the target. Two families of chemistry do this.
Dyes that light up on any double-stranded DNA
The simplest approach adds a dye whose fluorescence increases strongly when it binds double-stranded DNA. Higuchi's group at Roche used ethidium bromide (Chapter 11). Most laboratories now use dyes such as SYBR Green I. As product accumulates, there is more double-stranded DNA for the dye to bind, and the well glows more brightly. SYBR Green I is often called an intercalating dye, one that slides between base pairs, but a study of how it binds found that it intercalates only at low ratios of dye to DNA and gives most of its signal by binding the surface of the helix.
The weakness of such a dye is that it reports any double-stranded DNA, including primer-dimers and other wrong products. The standard check is a melt curve. After the last cycle, the instrument heats the product slowly while watching the fluorescence. As each kind of product reaches its melting temperature (Chapter 2), its strands separate and the dye lets go, so the fluorescence drops. Plotted against temperature, the rate of that drop shows a peak at the melting temperature of each product: one sharp peak suggests one product, and a second, lower peak often reveals primer-dimers. One supplier states in a technical note that a single product with regions of different G–C content can give two peaks, so a melt curve is an indicator, not proof, and a gel or sequencing confirms the product.
Probes that light up only on the target
Diagnostic tests mostly use a probe, a third short piece of synthetic DNA that binds inside the target between the two primers. Like the primers, it is an oligonucleotide: a short strand of DNA made to order. The probe carries a fluorophore at one end and a quencher at the other. A quencher is a molecule that absorbs the energy of a fluorophore close to it. While the probe is intact, its two ends are close enough for the quencher to absorb that energy, and the probe is dark.
The most widely used design relies on the activity of Taq described in Chapter 7. As the polymerase extends a primer along the template, it meets the probe bound in its path and cuts the probe apart with its 5′ to 3′ exonuclease activity. The fluorophore, now separated from the quencher, glows. In each cycle the polymerase cuts probes on every newly copied target, so the fluorescence rises with the product, but only with the product the probe recognizes.
Scientists at Cetus showed the method with a cleavable probe in 1991. By 1993 scientists at Perkin-Elmer's Applied Biosystems division had designed probes that carried a fluorescent reporter and a quencher, and in 1996 researchers at Genentech and at the same division, which made the instrument they used, described real-time quantitative PCR with these probes (Chapter 11). Such a probe is often called a TaqMan probe or a hydrolysis probe. Other designs work differently, and molecular beacons, introduced in 1996, are one example: hairpins that open and glow when they bind a matching target.
A probe must bind as well as both primers, so a probe test has a third check on identity, and that check makes it far less likely than a DNA-binding dye to report primer-dimers or wrong products. Even so, a poorly designed probe can still produce signal without target (Chapter 22).
In its form authorized from March 2020, the CDC test used one hydrolysis probe for each of its two viral targets and one for a human control gene, and each carried the fluorophore FAM at one end and a quencher called Black Hole Quencher 1 at the other. All three probes used the same fluorophore, so each specimen was run in three separate wells, one per target. The instrument read the fluorescence at the 55 °C step of each cycle, when the probes are being cut. In March 2020 the FDA also accepted probes with an additional internal quencher. The test's instructions told users to set the quencher to "none" in the instrument's software, but that setting is for a quencher that emits no light of its own, not for a probe without a quencher.
The hardware of a reading
An optical system needs a light source, filters and a detector, and real instruments combine them in several ways. The instrument named in the CDC test uses a 75-watt halogen lamp, a set of five excitation filters and five emission filters, and a CCD camera that images the whole plate at once. Another instrument uses a 100-watt xenon lamp and a cooled camera.
Newer designs use light-emitting diodes, or LEDs. One uses a white LED and a CMOS camera with six excitation and six emission filters. Another uses six colored LEDs and six filtered photodiodes, mounted on a small carriage that moves over the plate and reads each well in turn, so that every well is read through the same light path. According to its maker, this carriage takes about 12 seconds to read all channels of a 96-well plate. A rotor instrument reads each tube as it spins past a photomultiplier tube, a very sensitive detector.
Each architecture has a cost. A camera reads all wells at once but views each well from a slightly different position and angle, so it needs calibrations that locate every well on the image and correct for each well's position. A scanning carriage reads every well identically but takes time, which must fit inside the cycle step during which the reading is valid. Lamps give broad, bright light, but they age. LEDs are stable and long-lived, and each colored LED gives one narrow band.
Every number from a real-time instrument begins as light from a dye, measured through a filter and corrected by calibrations. A dirty filter, a fading lamp or a bubble in the well changes the light without changing the copies, and so does a wrong pure-dye calibration, which the next section describes. That is why the calibrations are scheduled, why instruments are verified with a reference plate, and why Chapter 15 traces every conversion between the copies and the reported result.
Several targets in one well
An instrument with five or six channels can, in principle, read five or six fluorophores in the same well, one in each channel. One instrument shows the pattern. Its first channel excites at 450 to 490 nanometers and reads 510 to 530, for FAM. Its second excites at 515 to 535 and reads 560 to 580, for a dye called HEX, and the others continue in the same way up to the near infrared. The CDC's later multiplex test detected four targets in one well, influenza A, influenza B, SARS-CoV-2 and the human control, with four different fluorophores.
In the form authorized from March 2020, the CDC test detected three targets in each specimen, and all three probes carried FAM (Chapter 22). Three signals of the same color in one well could not be told apart, so each specimen took three wells, and a 96-well plate held at most 96 ÷ 3 = 32 specimens before any wells went to controls. The CDC's later multiplex test detected four targets with four different fluorophores, so each specimen took one well. On a 96-well plate that is up to 96 specimens before controls, three times as many, and each well gives four readings instead of one. The price is that four dyes share one well, so the instrument needs a channel for each of them and must separate their light, as the next paragraphs describe.
Multiplexing has a cost: fluorophores emit over broad bands, so a bright dye leaks some light into a neighboring channel. Instruments correct for this leak with calibration. The maintenance guide for one family of instruments lists the calibrations its owner must run. A region-of-interest calibration maps where each well falls on the camera image, every six months or after a lamp change. A background calibration measures the instrument's own fluorescence, monthly. A pure-dye calibration records each dye's spectrum on that instrument, every six months. During a run the software uses the pure-dye spectra to work out how much of the measured light came from each dye, a step called multicomponent analysis.
Many master mixes also contain an inert reference dye, ROX, which serves as a passive reference. Some instruments divide the reporter signal by it, and the division corrects for differences in volume, bubbles and condensation. The CDC test's instructions had users switch that correction off by setting the passive reference to "none", although its original master mix contained ROX.
Choosing fluorophores for a test is an optical design decision. Each dye must sit in a channel the instrument has, be calibrated on that instrument, and be far enough in wavelength from its neighbors that the unmixing is reliable at the brightest signal the test will produce. The instrument's list of calibrated dyes is therefore an input to assay design, not an afterthought.
- Real-time instruments measure fluorescence: light of one color in, longer-wavelength light out, separated by filters.
- DNA-binding dyes glow with any double-stranded DNA and need a melt curve as a check, while hydrolysis probes glow only when the polymerase cuts them on the intended target.
- Lamps or LEDs, filters and cameras, photodiodes or photomultipliers combine in different architectures, each with its own corrections.
- Several dyes in one well need distinct channels, pure-dye calibration and unmixing, and the instrument always reads light, never the copies directly.