Where the instrument meets the chemistry.
+ The questionWhy can't an instrument be specified without its assay, or an assay validated without its instrument?
The tube is an interface
The plastic tube or plate looks like packaging. It is an engineered part of both the instrument and the test, and it sits on two interfaces at once (Chapter 15).
On the thermal side, the wall of the tube is the resistance that makes the liquid lag behind the block (Chapter 10). A thinner wall, a smaller volume or a lower-profile well shortens the lag, which is why fast instruments use their own plates. The maker of the instrument named in the CDC test states in its brochure that the instrument takes fast 96-well plates or 8-tube strips with 100-microliter wells, and reactions of 10 to 30 microliters. The tube's fit in the block and the lid's pressure on it decide how well heat crosses from metal to plastic. The volume in the tube is an input to the instrument's sample-temperature model, so a change of plate can change the temperature the liquid actually reaches without any change to the instrument's settings.
On the optical side, light enters and leaves through the cap or seal. The seal's clarity, its own faint fluorescence and any condensation on it all affect the signal, and preventing that condensation is one reason for the heated lid. The reaction volume sets the height of the liquid in the well and therefore how much of the liquid the optics see.
The chemistry is matched to the machine
The reagents are matched to the instrument as closely as the plastic is. A master mix's hot-start enzyme needs an activation step of a certain length at a certain temperature, and the instrument's profile must provide it. The CDC test's four authorized mixes used three different initial steps. Reverse transcription ran for 15 minutes at 50 °C, 10 minutes at 50 °C or 15 minutes at 45 °C, followed by 2 or 3 minutes at 95 °C.
The CDC test's instructions set the initial steps for each authorized mix before the 45 cycles. The TaqPath mix holds 2 minutes at 25 °C, 15 minutes at 50 °C and 2 minutes at 95 °C, so its initial steps take 2 + 15 + 2 = 19 minutes. Two of the other mixes share initial steps of 10 minutes at 50 °C and 3 minutes at 95 °C, 13 minutes in all. The fourth holds 15 minutes at 45 °C and 2 minutes at 95 °C, 17 minutes in all. The 45 cycles are the same for every mix and take about 62 minutes in standard mode, using the ramp rate the instrument's maker states (Chapter 16). A run therefore takes about 81 minutes with TaqPath, about 75 with either of the pair and about 79 with the fourth, before any read time. The choice of mix alone changes the run by up to 6 minutes, and the instrument's profile must provide each mix's steps.
A mix formulated for slow cycling may not finish its extension in the short holds of a fast instrument. The fluorophores on the probes must also be ones the instrument has channels for and has been calibrated with. The brochure for the CDC's instrument lists the dyes the instrument was calibrated for, beginning with FAM, the dye that all three of the test's probes carried.
The software is part of the method
Then there are the settings. They look like configuration and are in fact part of the test. The CDC's instructions specified that the instrument run in its standard mode, not its faster mode, and that the passive reference dye be set to none. In the software each target is entered as a detector, and the instructions specified that each detector be set to FAM with the quencher set to none (Chapter 12). They required the baseline to be left at the default and the threshold to be set by hand in the exponential region. Under the instructions, a target counted as detected below a Ct of 40.00.
Each setting changes what the instrument does or how its output is read. Running the same chemistry in fast mode would change the time the liquid spends at each temperature. Changing the threshold would change every Ct, and changing the cut-off would change the calls.
In a validated test, the instrument's run mode, reference-dye setting, analysis settings and software version are part of the method whose performance was established. Any change to them, however reasonable it looks, means that the established performance no longer applies until it has been shown again.
The test system
Regulators draw the consequence explicitly. What they evaluate is a test system: the assay, the instrument and the software together. The FDA finalized its guidance on assay migration in April 2013. The guidance covers moving an assay to a new instrument system whose performance the agency has not evaluated, and it defines the system as the assay, the instrument and the software. It is written for approved class III assays, licensed donor-screening tests and cleared assays whose migration raises particular concerns. Approval, licensing and clearance are separate FDA routes to market, and Chapter 25 explains the device classes and clearance.
For a quantitative test the guidance recommends, among other studies, testing precision at several sites, comparing results on at least 180 samples across the measuring range and showing that the limits of blank and detection are equivalent (Chapter 21). For a qualitative test it recommends panels of low-level samples and comparison panels of at least 100 positive and 100 negative samples. For every nucleic acid test it recommends studies of carry-over between samples.
The FDA finalized a later policy in August 2022, and that policy starts from the premise that an assay is cleared for use on specific instruments. It describes an instrument family: instruments from the same manufacturer with the same general architecture, design, tolerances and capabilities. It also describes the conditions under which a cleared assay may be extended to another member of the family without a new submission. The policy notes that a new submission is likely to be needed if the software must be modified to run the assay on the added instrument.
The CDC test is an example of a test system in practice. Its instructions named one instrument, with one software version. They named the master mixes that could be used: initially one, with three more added on 30 March 2020. They named the extraction methods: a list of kits and automated systems, extended in June and July 2020 as new platforms were shown to work. They also named a heat-treatment alternative, with a warning that it had been tested on few clinical specimens and might lose sensitivity. Each addition came with data and a revised authorization. A laboratory that ran the test on another instrument, or with another extraction kit, was no longer running the authorized test. It was running a modified test whose performance it had to establish itself.
What this means for developers
For a company developing an assay, the choice of instrument platform is therefore a long-term commitment: the assay's performance will be established on that platform, and moving the assay later is a project with its own studies. For a company developing an instrument, the assays that will run on it are part of its design inputs: the plates, volumes, dyes, profiles and analysis settings they need, and the stability of those parameters over the instrument's life. Software updates, a new lamp or LED supplier, a changed block design or a different plate supplier all affect the tests that run on the instrument, so each change must be controlled with those tests in mind.
Some instruments are closed: they run only the manufacturer's own assays, in its own consumables, with fixed settings, so that the test system is defined once. Others are open: they run any assay a laboratory develops. That openness offers flexibility, and it moves the burden of establishing the test system's performance onto the laboratory.
Once an assay's performance is established on an instrument, write down the complete test system: instrument model and software version, plate and seal part numbers, reaction volume, master mix, profile, run mode, dye and analysis settings, extraction method, controls and interpretation rules. Treat any change to any item as a change to the test. Assess it, and show with data that performance still holds before the changed test is used.
- The plate or tube sits on both the thermal and the optical interface, so changing it can change the temperature the liquid reaches and the light the optics see.
- Reagents are matched to the instrument's profile, speed and calibrated dyes, and run settings such as mode, reference dye, threshold and cut-off are part of the method.
- Regulators evaluate the assay, the instrument and the software together as a test system, assays are cleared for specific instruments, and moving an assay to a new instrument calls for migration studies or an instrument-family justification.
- The CDC test's instructions named one instrument, one software version, specific mixes and extraction methods, and each addition needed data and a revised authorization.